VOLTCRAFT ET-100 Insulation Tester Instruction Manual
- June 9, 2024
- VOLTCRAFT
Table of Contents
- Introduction
- Delivery content
- Up-to-date operating instructions
- Intended use
- Safety instructions
- Control elements
- Product description
- Displayed information and symbols
- Measuring
- Cleaning and maintenance
- Troubleshooting
- Disposal
- Specifications
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Operating Instructions
Insulation Tester ET-100
Item No. 2633261
Introduction
Dear customer,
Thank you for purchasing this product.
If there are any technical questions, please contact:
www.conrad.com/contact
Delivery content
- Insulation tester
- 2 safety measuring lines (red and black)
- 1 safety measuring line with alligator clip
- 6 AA batteries
- Bag with carrying strap
- Operating instructions
Up-to-date operating instructions
Download the latest operating instructions at www.conrad.com/downloads or scan the QR code shown. Follow the instructions on the website.
Intended use
– Measuring and displaying electric quantities in the range of measurement
category CAT III (up to 1000 V/DC or 750 V/AC to ground potential, in
accordance with EN 61010-1) and all lower categories.
– Measurement of direct and alternating voltages to max. 1000 V/DC or 750
V/AC.
– Measurement of insulation resistances of up to 4000 MΩ with an auxiliary
supply of 125 V to max. 1000 V/DC in accordance with EN 61557-2.
– Measurement of resistances of up to 400 Ω with a test current of approx. 200
mA
– Acoustic continuity check (< 35 ohms).
The measuring functions are selected with a rotary switch.
The low-resistance measuring range is secured against overload with a ceramic
quick-break fuse. The voltage in the measuring circuit must not exceed 1000
V/DC or 750 V/AC.
The measuring device may only be used by electrical specialists or technically
skilled persons familiar with the required standards and possible resulting
risks.
If you use the product for purposes other than those described, the product
may be damaged. Improper use can result in short circuits, fires, electric
shocks or other hazards.
The product complies with the statutory national and European requirements.
For safety and approval purposes, you must not rebuild and/or modify the
product.
Read the operating instructions carefully and store them in a safe place. Make
this product available to third parties only together with the operating
instructions.
All company names and product names are trademarks of their respective owners.
All rights reserved.
Safety instructions
Please read through the operating instructions before using the product for
the first time; they contain important information on proper operation. The
guarantee will be void if damage is incurred resulting from non-compliance
with these operat-ing instructions!
We assume no liability for any consequential damage! We do not assume any
liability for material and personal damage caused by improper use or non-
compliance with the safety instructions! In such cases the warranty/guarantee
is voided.
This device left the factory in perfect condition in terms of safety. To
maintain this status and to ensure risk-free opera-tion, the user must comply
with the safety instructions and warnings in these operating instructions.
The following symbols must be observed:
| An exclamation mark in a triangle indicates important information in these
operating instructions that has to be observed.
---|---
| A lightning symbol in a triangle warns of danger of an electric shock or of
the impairment of the electrical safety of the device.
| The arrow symbol informs you that there are special tips and hints
concerning the operation.
**| This product has been CE certified and meets the required European
guidelines.
| Protection class II (double or reinforced insulation)
CAT II| Measurement category II for measurements on electric and
electronic devices connected to the mains supply.
This category also covers all smaller categories (e.g. CAT I for measuring
signal and control voltages).
CAT III**| Measurement category III for measuring building wiring
installation (e.g. outlets or sub-distributions). This
category also covers all smaller categories (e.g. CAT II for measuring
electronic devices).
| Earth potential
-
For safety and licensing reasons, unauthorised conversion and/or modification of the device is not permitted.
-
Consult an expert when in doubt as to the operation, the safety or the connection of the device.
-
Measuring devices and accessories are no toys and do not belong in the hands of children!
-
In commercial facilities, the regulations for the prevention of accidents as laid down by the professional trade association for electrical equipment and devices must be observed.
-
In schools, training centres, hobby and self-help workshops, the operation of measuring devices must be supervised by trained personnel.
-
Before measuring voltages, always make sure that the measuring device is not set to resistance or insulation measuring range.
-
The voltage between the connection points of the measuring device and the earth potential must not exceed 1000 V/DC or 750 V/AC in CAT III.
-
When using the measuring lines without cover caps, measurements between the meter and the earth potential must not be performed above the measuring category CAT II.
-
When measuring in the measuring category CAT III, the cover caps must be pushed onto the measuring prods to avoid accidental short circuits during measurement.
-
Push the cover caps onto the measuring prods until they latch. To remove, pull the caps from the prods with a little force.
-
The test prods have to be removed from the test object before the measuring range is changed.
-
Pay particular attention when dealing with voltages exceeding 33 V AC or 70 V DC! If electrical conductors are touched, even these voltages involve the risk of a fatal electric shock.
-
Before measuring, always check your measuring device and its measuring lines for damage. Never carry out measurements when the protective insulation is damaged (ripped, torn off etc.).
-
The enclosed measuring cables have a wear indicator. When they are damaged, a second insulation layer in a different colour becomes visible. The measuring accessories must no longer be used and must be replaced.
-
To avoid an electric shock, do not to touch the connections/measuring points directly or indirectly during measurements. When measuring, only touch within the tangible grip markings on the test prods and the alligator clip.
-
Do not use the measuring device shortly before, during or directly after an electrical storm (lightning! /high-energy overvoltages!). Please make sure that your hands, your shoes, your clothing, the floor, circuits and circuit components are dry.
-
The measuring device must not be operated when it is open, i.e. with an open battery compartment or when the battery compartment lid is missing.
-
Measuring in potentially explosive areas (Ex) or damp rooms or under unfavourable ambient conditions is not permitted.
-
Adverse ambient conditions include:
– Dust or combustible gases, vapours or solvents
– Storms or stormy conditions, strong electrostatic fields, etc. -
Avoid operation near strong magnetic or electromagnetic fields, transmitter aerials or HF generators. This might lead to distorted measuring results.
-
If you have reason to believe that the device can no longer be operated safely, disconnect it immediately and make sure it is not unintentionally operated. It can be assumed that safe operation is no longer possible if:
– the device shows visible damage,
– the device does not function any longer or
– after it has been stored under unfavourable conditions over a period of time or
– after it has been exposed to heavy stress during transport. -
Do not switch on the measuring device immediately after it has been taken from a cold to a warm environment. The condensation generated can lead to faulty measurements or even destroy the device. Allow the device to reach room temperature before switching it on.
-
Do not leave packaging material lying around carelessly. It may become a dangerous toy for children.
-
For safety reasons, only use the provided measuring lines and accessories in correspondence with the specifications of the measuring device.
-
You should also observe the safety instructions in each chapter of these instructions.
(Rechargeable) batteries
- Correct polarity must be observed while inserting the (rechargeable) batteries.
- The (rechargeable) batteries should be removed from the device if it is not used for a long period of time to avoid damage through leaking. Leaking or damaged (rechargeable) batteries might cause acid burns when in contact with skin, therefore use suitable protective gloves to handle corrupted (rechargeable) batteries.
- (Rechargeable) batteries must be kept out of reach of children. Do not leave (rechargeable) batteries lying around, as there is risk, that children or pets swallow them.
- All (rechargeable) batteries should be replaced at the same time. Mixing old and new (rechargeable) batteries in the device can lead to (rechargeable) battery leakage and device damage.
- (Rechargeable) batteries must not be dismantled, short-circuited or thrown into fire. Never recharge nonrechargeable batteries. There is a risk of explosion!
Control elements
See fold-out page
- LCD display
- LOCK button for permanent insulation measurement
- HOLD button to “freeze” the current display and MAX/MIN button to display min. and max. values
- TEST button for insulation measurement
- Rotary switch
- LIGHT button to turn on the backlight and ZERO function
- Attachment point (for holding strap etc.)
- Fold-out tilt stand
- Battery compartment
- Red measuring socket “VΩ” (positive potential)
- Black “COM” measuring socket for earth potential or negative potential
Product description
The measuring device shows measured values and the corresponding units and
symbols on the digital display. The display for measured values comprises 4000
counts (count = smallest display value). A bargraph shows rapid changes in
value.
The measuring device shuts down automatically if it is not operated for about
10 minutes. This saves battery power and thus extends operating time. Battery
voltage is displayed in the “750 V~” range, which helps estimate the remaining
operating time.
The measuring device can be used for do-it-yourself or for professional
applications.
For better readability, the measuring device can be placed ideally using the
tilt stand on the back.
Rotary switch (5)
The individual measuring functions are selected via the rotary switch. The
automatic range selection “auto range” is active in the “400 Ω” measuring
function.
The appropriate measuring range is set automatically.
If the rotary switch is set to “OFF”, the measuring device is switched off.
Always turn off the device when it is not in use.
Displayed information and symbols
The following is a list of symbols and information the measuring device can display.
MIN | Min value display |
---|---|
MAX | Max. value display |
HOLD | Data hold function is active |
ZERO | Zero adjustment symbol in the low-resistance and voltage measuring range |
OL | Overload; the measuring range has been exceeded |
OFF | Off position; the measuring device is switched off |
Battery replacement symbol | |
Warning symbol for dangerous voltage (<30 V) at the test prods | |
Acoustic continuity check symbol | |
Backlight symbol | |
BATT | Display of battery voltage in the measuring device (battery check) |
**** V | Direct voltage |
**** V | Alternating voltage |
mV | Millivolt (exp.-3) |
V | Volt (unit of electric voltage) |
Ω | Ohm (unit of electric resistance) |
kΩ | Kiloohm (exp.3) |
MΩ | Megaohm (exp.6) |
LOCK | Symbol for permanent measurement (in the insulation measuring range) |
Measuring
Do not exceed the permitted max. input values. Never touch circuits or parts
of circuits with a pos-sible voltage of more than 33 V/ACrms or 70 V/DC!
Danger to life!
Before measuring, check the connected measuring lines for damage such as cuts,
cracks or pinch-es. Defective measuring lines must no longer be used! Danger
to life!
When measuring, only touch within the tangible grip markings on the test
prods.
Measuring is only permitted with the housing and the battery compartment
closed.
During insulation tests, dangerous voltages can be present at the test prods.
A lightning symbol on the display warns of such voltages. The insulation may
only be tested on currentless components or devices.
As soon as “OL” (overload) appears on the display, you have exceeded the
measuring range.
a) Switching on the measuring device
The measuring device is switched on using the rotary switch (5). Turn the
rotary switch to select the desired measuring function. Tum the rotary switch
to “OFF” to switch off the measuring device. Always switch off the measuring
device when it is not in use (“OFF” position).
Before the measuring device can be used, the enclosed batteries have to be
inserted. See chap- ter “Cleaning and maintenance” for information on
inserting and replacing the batteries.
b) Voltage measurement “V”
Proceed as follows to measure DC voltages “V/DC”:
– Switch on the measuring device and select measuring range “1000 V=”.
– Plug the red measuring line into the “V” measuring socket (10) and the black
measuring line into the COM measuring socket (11).
– Now connect the two test prods with the test object (battery, switch etc.).
The red test prod is the positive terminal, the black test prod is the
negative terminal.
– The polarity of the measured value is displayed with the measured value.
If a minus “-” appears for direct voltage in front of the measured value, the
measured voltage is negative (or the measuring lines have been mixed up).
– After measuring, remove the measuring lines from the test object and switch
off the measuring device.
Proceed as follows to measure AC voltages “V/AC”:
– Switch on the measuring device and select measuring range “750 V~’.
– Plug the red measuring line into the “V” measuring socket (10) and the black
measuring line into the COM measuring socket (11).
– Connect the two test prods with the test object (generator, switch etc.).
– The measured value is shown on the main display. The small display at the
bot-tom right shows the current battery voltage in the measuring device.
– After measuring, remove the measuring lines from the test object and switch
off the measuring device.
c) Resistance measurement “Ω”
Make sure that all circuit parts, circuits, components and any other test
objects are currentless and discharged.
The measuring device allows measuring resistances in the low-resistance
measuring range from 0.01 to 400 Ω. For accurate results, resistance is
measured with a test current of approx. 200 mA.
Proceed as follows to measure resistances:
– Switch on the measuring device and select measuring range “400 0”.
– Plug the red measuring line into the 0-measuring socket (10) and the black
meas-uring line into the COM measuring socket (11).
– Check the measuring lines for continuity by connecting both test prods with
one another. The resistance value must be approximately 0 – 0.5 ohms (inherent
re- sistance of the measuring lines) along with a beep.
– Connect the test prods with the test object. The measured value is displayed
as long as the test object is not a high-resistance object or interrupted.
Wait until the display has stabilised.
– IF “OL” (overload) is displayed, you have exceeded the measuring range or
the measuring circuit is broken.
– After measuring, remove the measuring lines from the test object and switch
off the measuring device.
If no measurement can be carried out or if the continuity check fails (zero
display), check the built-in fuse. Fuse replacement is described in chapter
“Cleaning and maintenance”.
If you carry out a resistance measurement, make sure that the measuring points
where you connect the test prods are free from dirt, oil, solderable lacquer
or the like. Faulty measurements may result under such circumstances.
For low-voltage measurement, the resistance of the measuring lines is included
in the measuring result. Measuring line resistance is approx. 0.5 Q. The
equivalent circuit diagram is illustrated on the right. Resistances “R E1″ and
“R E2″ represent the measuring lines. “R” represents the test object. All
three resistances contribute to the measured value. In order to exclude
measuring line resistance from the result, activate the ZERO function.
d) Continuity check
Make sure that all circuit parts, circuits, components and any other test
objects are currentless and discharged.
– Switch on the measuring device and select measuring range
– Carry out the measurement as described for resistance measurement. A
measured value of less than 35 ohms is
identified as continuity and accompanied by a beep. The continuity check
symbol is displayed.
– After measuring, remove the measuring lines from the test object and switch
off the measuring device.
e) Insulation test with auxiliary supply
Make sure that all circuit parts, circuits, components and any other test
objects are currentless and discharged. If the measuring device recognises a
voltage of 230 V/DC or 230 VIAC in the measur- ing circuit, no insulation test
is possible. Always measure the voltage (AC + DC) before testing the
insulation.
During insulation tests, dangerous voltages can be present at the test prods.
A lightning symbol on the display warns of such voltages. The voltage is also
present in all connected circuits and conductors. Do not touch any circuits or
components when measuring the insulation.
Only touch within the tangible grip markings on the test prods or the
insulated alligator clip.
The insulation test allows measuring the insulation resistance of safety-
related protective measures with an adjustable auxiliary supply. Faulty or
insufficient insulation can be detected. The required insulation resistances
can be gathered from the respective standards.
– Switch on the measuring device and select the “4000 MΩ” measuring range and
the corresponding auxiliary supply.
– Plug the red measuring line into the “V” measuring socket (10) and the black
measuring line into the COM measuring socket (11). You can use the provided
(rey alligator clip to carry out a one-hand measurement.
– The display shows the unit “MΩ”.
– Connect the two test prods (red = positive/black = negative) with the test
object. The measuring line with the alligator clip makes handling easier
during measure-ments.
When measuring, a beep along with a flashing lightning symbol warns of test
voltage present in the measuring lines.
If a voltage of more than 30 V (AC/DC) has been detected, no test is carried
out and “>30 V” is displayed along with the voltage type (AC or DC). This is
indicated by a warning sound. Stop meas- uring immediately and disconnect the
measuring circuit, any components and the device from the power supply.
– For a single measurement, hold down the TEST button (4) for the duration of
the measurement. After a short while, the insulation resistance is displayed
along with the current auxiliary test voltage. Wait until the main display has
stabilised. This may take several seconds. Let go of the TEST button to stop
measuring. The last measured result
is recorded (“HOLD” displayed) and the measuring circuit is discharged.
– For permanent measurement, press the LOCK button (2). The display shows
“LOCK” and the lock symbol. Press the TEST button (4) shortly to start the
insulation test. Permanent measuring mode is active for about 10 minutes.
Then the measuring device turns off automatically. If you want to end the
measurement before that time, press the TEST button. The last measured result
is recorded (“HOLD” displayed) and the measuring circuit is discharged.
– IF “OL” (overload) appears on the display, you have exceeded the measuring
range or the measuring circuit is highly resistive.
– After measuring, remove the measuring lines from the test object and switch
off the measuring device.
f) HOLD function
The HOLD function locks the displayed value to allow you to read it
comfortably or log it.
If you test live conductors, make sure that this function is deactivated when
you start measuring.
Otherwise, a false measuring result is shown!
To activate the HOLD function, press the “HOLD” button (3). An audio signal
confirms the activation and “HOLD” appears on the display.
To deactivate the HOLD function, press the “HOLD” button again or switch to
another measuring function.
g) ZERO function
The ZERO function is only active in the 400 O, 1000 V/DC and 750 V/AC
measuring ranges.
The ZERO function allows zero adjustment of the display or a fast relative
measurement, e.g. to display component tolerances. The reference value is
measured and saved in the internal memory and then automatically deducted from
subsequent measurements. This allows the easy display of deviations from the
reference values.
– Connect the test object as described for resistance measurement or voltage
measurement. Wait until the measured value has stabilised.
– Press the “ZERO” button (6) shortly while measuring. “ZERO” appears on the
display and the display is reset to zero. The reference value is saved.
– Continue with the differential measurements on the other test objects.
– Press the ZERO button again to deactivate the function.
After changing the measuring range or the measuring function, the ZERO memory
is deleted.
h) MAX/MIN function
This function shows the measured max. and min. values on the display. The
MAX/MIN function can only be used in the 400 Ω, 1000 V/DC and 750 V/AC
measuring ranges.
– Hold down the MAX/MIN button for about 1 second during the measurement. The
display shows “MAX” along with the highest value.
– Press the “MIN/MAX” button again to switch to “MIN”. The lowest value is
displayed.
– Pressing the button switches the display.
– To abort this function, hold down the “MAX/MIN” button for about 1s until
“MAX” or “MIN” is blanked accompanied by a beep.
i) Auto power-off function
The measuring device turns off automatically after 10 minutes if no button or
rotary switch is operated. This function saves battery power and extends the
service life.
To reactivate the measuring device after automatic power-off, use the rotary
switch or press any function key on the measuring device.
j) Switching on the backlight
The display can be illuminated in bad lighting conditions. To switch on the
backlight, hold down the light button (6) for about 2 seconds. The light will
remain on for about 220 seconds and switches off automatically to save the
batteries.
The lighting can be manually re-enabled by pressing the light button (6).
Press and hold the light button (6) for active illumination 2s, the display
light switches off manually.
Cleaning and maintenance
a) General
To ensure the accuracy of the measuring device over an extended period of
time, it should be calibrated once a year.
Apart from occasional cleaning and fuse replacements, the measuring device
requires no servicing.
See the next chapter for information on battery and fuse replacement.
Regularly check the technical safety of the device and measuring lines, e.g.
check for damage to the housing, the insulation or squeezing etc.
b) Cleaning
Always observe the following safety instructions before cleaning the device:
Live components may be exposed if covers are opened or parts are removed
(unless these parts
intended to be removed without tools).
Before cleaning or repairing the device, connected lines must be disconnected
from the measuring device and from any test objects. Switch off the measuring
device.
Do not use agents containing carbon, benzine, alcohol or similar substances to
clean the device. They corrode the surface of the measuring device. In
addition, the vapours are detrimental to health and explosive. Sharp-edged
tools, screwdrivers or metal brushes should not be used for cleaning purposes.
Use a clean, lint-free, antistatic and slightly moistened cloth to clean the
device, the display and the measuring lines. Allow the product to dry
completely before you use it again to carry out measurements.
c) Inserting and replacing the batteries
Operation of the measuring device requires six AA batteries. You need to
insert new, charged batteries before operat- ing the device or when the
battery replacement symbol appears on the display. Rechargeable batteries must
not be used!
Proceed as follows to insert or replace the batteries:
– Disconnect all measuring lines from the measuring device and switch it off.
– Fold out the tilt stand (8) and loosen the four screws of the battery
compartment(9).
– Remove the battery compartment lid from the measuring device. The batteries
are accessible now.
– Replace all batteries with new batteries of the same type. Observe the
polarity markings in the battery compartment when inserting the batteries.
– Close the battery compartment and screw on the lid in reverse order.
– The measuring device is ready for use again.
Never operate the measuring device when it is open.
RISK OF DEATH!
Rechargeable batteries must not be used!
Do not leave flat batteries in the device. Even batteries protected against
leaking can corrode and thus release chemicals which are detrimental to your
health or destroy the device.
Do not leave batteries lying around carelessly. They could be swallowed by
children or pets. If swal- lowed, consult a doctor immediately.
Remove the batteries if the device is not used for a longer period of time to
prevent leaking.
Leaky or damaged batteries may cause caustic bums if they come into contact
with skin. Therefore, you should use suitable protective gloves.
Make sure that batteries are not short-circuited. Do not throw batteries into
fire. Always replace all batteries at the same time. Mixing fully charged
batteries and half-full batteries reduces battery life. Always use batteries
of the same type and manufacturer in a set of batteries.
Batteries can be damaged by equalizing currents and may leak.
Batteries must not be recharged or dismantled. There is danger of explosion.
You can order suitable alkaline batteries stating the following order no.:
Item no. 652506 (please order six pcs).
Only use alkaline batteries. They are powerful and long-lasting.
d) Checking the batteries
The insulation tester ET-100 allows you to check the battery voltage anytime.
Switch the measuring device to the “600 V~” measuring range. After about 3
seconds, the small display at the bottom right shows the current battery
voltage.
At a voltage of <7.5 V the battery replacement symbol starts flashing
automatically.
e) Replacing the fuse
The low-voltage “400 Ω” measuring range is secured with a quick-break fuse. If
measuring in this range is no longer possible, you have to replace the fuse.
Proceed as follows to replace the fuse:
– Disconnect all measuring lines from the device and switch it off.
– Open the battery compartment as described in chapter “Inserting and
replacing the batteries”.
– The fuse is accessible.
– Replace the defective fuse with a new fuse of the same type and nominal
voltage. The fuse has the following characteristics: F 500 mA/1000 V, size 6.3
x 30 mm, ceramic.
– Properly close the housing again in reverse order.
Using repaired fuses or bridging the fuse holder is not admissible for safety
reasons. This can cause fire or electric arc explosions. Never operate the
measuring device when it is open.
Troubleshooting
By purchasing the measuring device you have acquired a state-of-the-art
product which operates reliably.
Nevertheless, problems or malfunctions may occur.
For this reason, the following is a description of how you can eliminate
possible malfunctions yourself:
Always adhere to the safety instructions!
Error | Possible cause | Remedy |
---|---|---|
The measuring device does not work. | Are the batteries exhausted? | Check the |
status. Replace the batteries
No change in measured| Is the wrong measuring function active (AC/DC)?| Check
the display (AC/DC) and values, switch the function,if necessary.
Is the fuse defective in the current range?| Check the fuse.
Is the HOLD function activated? (“HOLD” displayed)| Press the HOLD button to
deactivate this function.
Repairs other than those described should only be carried out by an authorised specialist.
Disposal
a) Product
**** This symbol must appear on any electrical and electronic equipment
placed on the EU market. This symbol indicates that this device should not be
disposed of as unsorted municipal waste at the end of its service life.
Owners of WEEE (Waste from Electrical and Electronic Equipment) shall dispose
of it separately from unsorted municipal waste. Spent batteries and
accumulators, which are not enclosed by the WEEE, as well as lamps that can be
removed from the WEEE in a non-destructive manner, must be removed by end
users from the WEEE in a non-destructive manner before it is handed over to a
collection point.
Distributors of electrical and electronic equipment are legally obliged to
provide free take-back of waste. Conrad provides the following return options
free of charge (more details on our website):
- in our Conrad offices
- at the Conrad collection points
- at the collection points of public waste management authorities or the collection points set up by manufacturers or distributors within the meaning of the ElektroG
End users are responsible for deleting personal data from the WEEE to be
disposed of.
It should be noted that different obligations about the return or recycling of
WEEE may apply in countries outside of Germany.
b) (Rechargeable) batteries
Remove batteries/rechargeable batteries, if any, and dispose of them
separately from the product. According to the Battery Directive, end users are
legally obliged to return all spent batteries/rechargeable batteries; they
must not be disposed of in the normal household waste.
**** Batteries/rechargeable batteries containing hazardous substances are
labelled with this symbol to indi- cate that disposal in household waste is
forbidden. The abbreviations for heavy metals in batteries are: Cd = Cadmium,
Hg = Mercury, Pb = Lead (name on (rechargeable) batteries, e.g. below the
trash icon on the left).
Used (rechargeable) batteries can be returned to collection points in your
municipality, our stores or wherever (rechargeable) batteries are sold. You
thus fulfil your statutory obligations and contribute to environmental
protection.
Batteries/rechargeable batteries that are disposed of should be protected
against short circuit and their exposed terminals should be covered completely
with insulating tape before disposal. Even empty batteries/rechargeable
batteries can contain residual energy that may cause them to swell, burst,
catch fire or explode in the event of a short circuit.
Specifications
Display …………………………..LCD, 4000 counts with bargraph
Measuring rate …………………………approx. 2.5 measuring operations/second
Measuring line length ………………….2 x approx. 120 cm (test prods), 1 x approx. 90
cm (alligator clip)
Measuring impedance ………………………………..>10 MΩ (V range)
Automatic power-off ……………………………………approx. 10 min
Power supply …………………………………………….6 x AA batteries
Operating conditions ………………………..0 to +40 °C, <80 % RH (non-condensing)
Operating altitude ……………………max. 2,000 m
Storage conditions. ……………….-10 °C to +60 °C, <70 % RH (non-condensing)
Weight ……………………………………………………..approx. 585 g
Dimensions (L x W x H) ………………………………200 x 92 x 50 mm
Measuring category ……………………………………CAT III 1000 V
Degree of pollution …………………………………….2
Measurement tolerances
Accuracy in ± (% of reading + display error in counts (= number of smallest
points)). The accuracy is valid for one year at a temperature of +23 °C (± 5
°C) with relative humidity less than 80 % RH, non-condensing.
Direct voltage
Range | Resolution | Accuracy |
---|---|---|
1000V | 1 V | ±(0.8% + 3) |
Overload protection: 1000 V; impedance approx.10 MΩ
Alternating voltage
Range | Resolution | Accuracy |
---|---|---|
750 V | 1 V | ±(1.2% + 10) |
Frequency range 40 – 400 Hz; effective average at sinusoidal voltage; overload protection 750 V; impedance approx. 10 MΩ
Resistance
Range | Resolution | Accuracy |
---|---|---|
40Ω | 0.01Ω | ±(1.2% + 3) |
400Ω | 0.1Ω |
Overload protection 1000 V, test voltage max. 5.8 V, test current approx. 200 mA; max. number of measurements with full batteries: x 500
Acoustic continuity check
Range | Resolution | Test voltage | Test current |
---|---|---|---|
<35 Ω permanent audio signal | 0.01 Ω | 5.8 V | 200 mA (typical) |
Overload protection 600 V
Insulation test
Nominal voltage at output
(0 to + 10 %)| Measuring range| Resolution| Accuracy| Nominal test current|
Short-circuit current
---|---|---|---|---|---
125 V| 0.125 – 4 MΩ| 0.001 MΩ| ±(2.0% + 10)| 1 mA
(at 125 kΩ)| 51 mA
4 – 40 MΩ| 0.01 MΩ| ±(2.0% + 10)
40 – 400 MΩ| 0.1 MΩ| ±(4.0% + 5)
400 – 4000 MΩ| 1 MΩ| ±(5.0% + 5)
250 V| 0.250 – 4 MΩ| 0.001 MΩ| ±(2.0% + 10)| 1 mA
(at 250 kΩ)| 51 mA
4 – 40 MΩ| 0.01 MΩ| ±(2.0% + 10)
40 – 400 MΩ| 0.1 MΩ| ±(3.0% + 5)
400 – 4000 MΩ| 1 MΩ| ±(4.0% + 5)
500 V| 0.500 – 4 MΩ| 0.001 MΩ| ±(2.0% + 10)| 1 mA
(at 500 kΩ)| 51 mA
4 – 40 MΩ| 0.01 MΩ| ±(2.0% + 10)
40 – 400 MΩ| 0.1 MΩ| ±(2.0% + 5)
400 – 4000 MΩ| 1 MΩ| ±(4.0% + 5)
1000 V| 1- 4 MΩ| 0.001 MΩ| ±(3.0% + 10)| 1 mA
(at 1 MΩ)| 51 mA
4 – 40 MΩ| 0.01 MΩ| ±(2.0% + 10)
40 – 400 MΩ| 0.1 MΩ| ±(2.0% + 5)
400 -4000 MΩ| 1 MΩ| ±(4.0% + 5)
Overload protection >/=30 V
Max. number of measurements with full batteries:
125V| x500| | | |
250V| x500| | | |
500V| x500| | | |
1000 V| x 500| | | |
Do not exceed the permitted max. input values. Never touch circuits or parts of circuits with a possible voltage of more than 33 V/ACrms or 70 V/DC! Danger to life!
This is a publication by Conrad Electronic SE, Klaus-Conrad-Str. 1, D-92240
Hirschau (www.conrad.com).
All rights including translation reserved. Reproduction by any method, e.g.
photocopy, microfilming, or the capture in electronic data processing systems
require the prior written approval by the editor. Reprinting, also in part, is
prohibited.
This publication represent the technical status at the time of printing.
Copyright 2022 by Conrad Electronic SE.
References
- France.fr : Explore France and its wonders - Explore France
- Conrad Electronic » Alle Teile des Erfolgs
- Conrad Electronic » All parts of success
- Conrad Electronic – Alle Teile des Erfolgs
- Conrad Electronic » All parts of success
- Contact
- Download center
- Conrad Electronic Online Shop » Alle Teile des Erfolgs
- Conrad Electronic » All parts of success
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>