LUDLUM 26S Integrated Scintillator Frisker User Manual
- June 9, 2024
- LUDLUM
Table of Contents
LUDLUM MODEL 26S INTEGRATED SCINTILLATOR
USER’S MANUAL
January 2023
LUDLUM MEASUREMENTS, INC
901 OAK STREET, P.O. BOX 810
SWEETWATER, TEXAS 79556
325-235-5494, FAX:
325-235-4672
26S Integrated Scintillator Frisker
STATEMENT OF WARRANTY
Ludlum Measurements, Inc. warrants the products covered in this manual to
be free of defects due to workmanship, material, and design for a period of
twelve months from the date of delivery. The calibration of a product is
warranted to be within its specified accuracy limits at the time of shipment.
In the event of instrument failure, notify Ludlum Measurements to determine if
repair, recalibration, or replacement is required.
This warranty excludes the replacement of photomultiplier tubes, G-M and
proportional tubes, and scintillation crystals which are broken due to
excessive physical abuse or used for purposes other than intended.
There are no warranties, express or implied, including without limitation any
implied warranty of merchantability or fitness, which extend beyond the
description of the face there of. If the product does not perform as warranted
herein, purchaser’s sole remedy shall be repair or replacement, at the option
of Ludlum Measurements. In no event will Ludlum Measurements be liable for
damages, lost revenue, lost wages, or any other incidental or consequential
damages, arising from the purchase, use, or inability to use product.
RETURN OF GOODS TO MANUFACTURER
If equipment needs to be returned to Ludlum Measurements, Inc. for repair
or calibration, please send to the address below. All shipments should
include documentation containing return shipping address, customer name,
telephone number, description of service requested, and all other necessary
information. Your cooperation will expedite the return of your equipment.
LUDLUM MEASUREMENTS, INC.
ATTN: REPAIR DEPARTMENT 501 OAK STREET
SWEETWATER, TX 79556
800-622-0828
325-235-5494
FAX325-235-4672
Introduction
The Model 26S is an ergonomic, lightweight instrument, which can be used for
locating gamma radiation or for measuring gamma exposure or dose. It features
the ability to measure radiation in count rate, exposure rate/dose, activity
rate, integrated exposure/dose, time-averaged rates, and scaler counts. The
Model 26S utilizes a CsI(Tl) crystal detector,a rugged, improved version of
the classic 1×1 NaI scintillator for detection and measurement of gamma
radiation. The scintillator is energy-dependent, and will over-respond to
energies in the 100 keV region.
Three modes of operation are available for the Model 26S – RATE, MAX, and
COUNT. RATE mode operation will display the current count, exposure, or
activity rate. MAX mode is used to capture the highest count, activity, or
exposure rate detected – useful for finding a peak rate, or frisking when the
display is not visible. Primary and Secondary Units for RATE and MAX modes can
be chosen from cps, cpm, Bq, dpm, mR/h, or µSv/h. COUNT mode allows the user
to perform a count for a predetermined time. Depending on the Count Units
chosen, the result can be a scaler count (in counts or disintegrations), a
time-averaged rate (cpm, dpm, Bq, cps), a timeaveraged exposure or dose (mR/h,
µSv/h), or an integrated exposure or dose (mR or µSv).
The instrument features a large backlit LCD (liquid crystal display), a
piercing audio warning, and easy, intuitive use in a single-handed platform. A
comfortable wrist strap and lanyard are also included for keeping the
instrument close and secure. The unit body is made of lightweight but durable
plastic. It is intended for outdoor use and can resist splashing water.
There is also a headphone option available (not included). The instrument can
be modified to include a standard 1/8 inch headphone jack (part number
4498-538).
The display will be automatically backlit if light levels are low. The display
backlight can also be configured for “Continuous On” operation. RATE and MAX
modes can be silent or utilize a “click” or a sigma audio; the audio is always
silent during COUNT mode. The “click” audio produces a short sound for each
gamma ray detected. The default sigma audio measures the radiation level on
power-up and, when a sudden increase is detected, emits a short beep every
quarter second.
Setup of the instrument is accomplished through the front-panel buttons.
(Setup can be disabled via the internal switch on the Model 26S in order to
protect settings.)
The advanced user or administrator can set:
- Calibration Constant
- Dead Time Correction
- Efficiency
- Primary and Secondary Units
- Primary and Secondary Unit Alarm Levels
- Primary and Secondary Count Units
- Primary and Secondary Count Alarm Levels
- Response Time
- Auto-Response Rate (Fast or Slow)
- Available Operational Modes
- Count Time
- Click or Sigma Audio
- Overload Setpoint (not used)
The unit is operated with two alkaline AA batteries for operation from -20 to 50 °C (-4 to 122 °F). Battery life is approximately 200 hours under normal usage. A low-battery indicator on the LCD warns when less than 16 hours of battery life remain.
Getting Started
Unpacking and Repacking
Remove the calibration certificate and place it in a secure location. Remove
the instrument and ensure that all of the items listed on the packing list
are in the carton. Check individual item serial numbers and ensure calibration
certificates match. The Model 26S serial number is located on a label on the
front of the unit.
To return an instrument for repair or calibration, provide sufficient packing
material to prevent damage during shipment.
Every returned instrument must be accompanied by an Instrument Return Form,
which can be downloaded from the Ludlum website at
www.ludlums.com. Find the form by clicking the
“Support” tab and selecting “Service Department” from the drop-down menu. The
webpage has a link to download the correct return form depending on the
location of the service department that you wish to use.
Battery Installation
A low-battery indicator appears at the bottom of the LCD when less than 16
hours of battery life remain.
When this indicator is present, follow these steps to replace the two standard
AA batteries:
- Grab the ring on the screw.
- Turn the ring one quarter turn counter-clockwise.
- Release and remove the battery cover.
- Replace two each AA batteries, being careful to orient them correctly: the positive terminals should point towards the display.
- Firmly insert the barb of the battery cover completely into the body of the Model 26S.
- Replace the cover and turn ring one quarter of a turn clockwise to secure.
Warning: If this procedure is not followed correctly, and the barb is not inserted into the body of the instrument correctly, the barb may break off.
Instrument Operational Test
Turn the instrument ON by pressing the ON/ACK button for about a second, and
then releasing. The instrument should activate all the LCD segments and the
audio. Observe the device during this time. If any LCD segments are missing,
or audio fails to work, the device is in need of repair.
Please refer to Figure 1 below.
Figure 1: Startup display for Model 26S, with all LCD segments shown.
The instrument then displays the firmware version number and activates the
Alarm LED briefly. Should the Alarm LED fail to turn on, the device is in
need of repair. Please refer to Figure 2 below.
Figure 2: Firmware version display and Alarm LED check shown.
If sigma audio mode is active, then the unit will beep for 8 seconds while it
measures the current radiation level. The audio will then be silent and the
instrument will move to normal operation, displaying the current rate for the
Primary units (default: mR/hr).
The user may select the Secondary units (default: cpm) by pressing the Units
button.
Ensure that the low-battery indicator is not present. If the low-battery
indicator is present, replace the batteries as soon as possible. Should the
instrument detect a battery voltage that is high enough to power on, but too
low to safely operate, the display will blank and the low-battery icon will
flash. Normal operation will not be available until the batteries have been
replaced. Under extreme low-battery conditions, be aware that the unit may
not even turn on or may turn itself off abruptly.
A reference reading with a check source, 1 µCi (37 kBq) of 137 Cs for example,
should be obtained at the time the instrument is received in the field. Small
check sources of radiation are available from Ludlum Measurements. While
exempt from many regulations because of their small size, these sources are
large enough to produce a response on this instrument. The detector’s center
position is indicated by the intersection of virtual lines drawn between
small yellow dots on the back of the Model 26S, around the display window: the
edge of the circular detector housing indicates the approximate center of the
detector. If this procedure is done routinely with the same radiation source,
instrument malfunction may be detected when anomalous readings are observed.
If at any time the instrument fails to read within 20% of the reference
reading when using the same check source, it should be sent to a calibration
facility for recalibration and/or repair.
Example log reading:
Check Source # | Rate | Units |
---|
Once this procedure has been completed, the instrument is ready for use.
Detector Failure Diagnostic
Note that the Model 26S has its own diagnostic tests to ensure that the
detector is functioning correctly. The Model 26S can detect when the radiation
detector is malfunctioning and will flash the display to indicate a fault. If
the detector completely stops detecting radiation for 60 seconds, the Model
26S will flash a zero reading for the currently selected units. If this
indication is observed, remove the unit from service and have it evaluated by
a qualified repair and calibration technician.
Figure 3: Detector Failure display (shown for cpm); will also flash.
Detector Over Range
If the unit is exposed to high levels of radiation or has an internal
malfunction that causes it to count high or excessively, the unit flashes the
maximum rate for the currently selected units as a warning. The user should
ensure whether this is being caused by a high radiation field or by internal
malfunction.
Figure 4: Detector Over Range (shown for µSv/h); will also flash.
Instrument Use and Controls
With three front-panel buttons, the Ludlum Model 26S is simple and easy to use
with minimal training required. Default operation is RATE mode, and the
display shows the current count rate using the Primary units. Pressing the
UNITS button will switch between Primary and Secondary units. Pressing the
MODE button will switch the instrument to MAX mode, which will display the
highest count rate detected. Pressing the MODE button again will switch it to
COUNT mode, which will display the COUNT timer. Note that, to provide the
simplest operation, either or both of the MAX and COUNT modes can be locked
out in the setup process.
See the Model 26S drawing at the beginning of this manual to reference the
following controls:
ON/ACK button: Used to power the Model 26S ON and OFF, silence click
audio, reset MAX mode, start/reset the COUNT Timer, and acknowledge audio
alarms.
- Power On: Press for approximately one second and release (all LCD segments will activate, and firmware version will be shown).
- Power Off: Press for approximately five seconds. The display will show a 3, 2, 1 countdown for the final three seconds of shutdown. Releasing the ON/ACK button during shutdown will return the device to the previous state of operation. At completion of the shutdown count, the LCD will go blank.
- Normal Operation: Will silence ‘click’ audio in RATE and MAX modes, reset MAX mode display, start/reset COUNT Timer in COUNT mode, and acknowledge/silence audio in all modes of operation. A long press of this button will also
MODE button: Used to advance between the three operating modes, RATE,
MAX, and COUNT. Note that MAX and/or COUNT mode may be disabled from use by
the administrator or calibrator.
UNITS button: Used to switch between Primary and Secondary units in RATE
and MAX modes. In COUNT mode, the UNITS button will switch between Primary
and Secondary units unless a countdown is active. The UNITS button is disabled
during an active countdown.
RATE Mode Operation
In RATE mode the current count rate will be displayed.
Pressing the UNITS button will switch the displayed value between the Primary
and Secondary Units.
Under a non-alarm condition, the Alarm Status LED will be off; pressing the
ON/ACK button will turn the audio on/off.
If an alarm condition is present, pressing the ON/ACK button will acknowledge
and turn off the continuous tone alarm audio. Under an alarm condition, the
ALARM display indicator will remain on, and the Alarm Status LED will be
flashing. Alarms are non-latching in RATE mode.
If other operational modes are available, pressing the MODE button will move
to the next available operational mode.
Figure 5: RATE mode display showing typical background radiation rate and the
low-battery icon.
MAX Mode Operation
While in MAX mode, the highest detected count rate (since the last reset) is
displayed. The word MAX will be displayed when in MAX mode. Pressing the UNITS
button will switch the displayed value between the Primary and Secondary
Units.
Under a non-alarm condition, the Alarm Status LED will be off; pressing the
ON/ACK button will turn the audio on/off. Pressing the ON/ACK button a second
time will reset the display and will enable the audio.
If an alarm condition is present, pressing the ON/ACK button once will
acknowledge and turn off the audio. Pressing the ON/ACK button a second time
will reset the display and clear the alarm condition. Under an alarm
condition, the ALARM display indicator will remain on, and the Alarm Status
LED will be flashing. Alarms in MAX mode latch with the display.
If other operational modes are available, pressing the MODE button will move
to the next available operational mode.
Figure 6: MAX mode operation display with ALARM indicator and alarm LED.
COUNT Mode Operation
When entering COUNT Mode from another operational mode, the currently selected
COUNT Units will be displayed for approximately one second. The purpose of
COUNT mode is to count for a predetermined amount of time, and to display the
results on the display. Note that the predetermined count time can be from 1
second to 20 minutes, or can be set to zero to enable continuous counting
until stopped by the user.
Count mode operation is very flexible, depending on the units chosen. A common
choice is for the count mode to just perform a scaler count for a specified
time, with a resulting answer in counts (equaling detected radiation events).
If a result in terms of activity is desired, the scaler count can also be in
units of “d” or disintegrations. But if the count units are chosen to be cpm
or cps, then the resulting answer is an averaged count rate over the time
interval. Similarly, if count units of Bq or dpm are chosen, the resulting
answer is an averaged disintegration rate.
NOTE: If the user desires the instrument to show results in terms of
disintegrations/area (eg. dpm/100cm² or Bq/cm²), then the appropriate factor
should be placed in the Efficiency parameter.
Other choices are to have count mode units of mR/h or µSv/h, in which case the
COUNT mode result is an averaged exposure or dose rate. But if count mode
units of mR or µSv are chosen, the result is shown in accumulated exposure or
accumulated dose over the chosen count time. The following tables lists the
possibilities:
UNITS | RESULT |
---|---|
c | counts per count time |
d | disintegrations per count time |
cpm, cps | count rate averaged over the count time |
dpm, Bq | disintegration rate, averaged over the count time |
mR/h, µSv/h | exposure or dose rate, averaged over the count time |
mR, µSv | integrated exposure or dose over the count time |
Audio is disabled in COUNT mode.
In COUNT mode, operation depends on the current state of the Count Timer.
When the Count Timer is Ready:
- The display will show the Count Time, and the Alarm Status LED will be off.
- Pressing the UNITS button will switch between the Primary and Secondary Count Units. The newly selected Count Units will be displayed for approximately one second, and the display will then return to the Count Timer.
- Pressing the ON/ACK button starts the Count Timer.
- If other operational modes are available, pressing the MODE button will move to the next available operational mode.
When the Count Timer is active:
- The display will show the Count Time remaining.
- Pressing the ON/ACK button will reset the Count Timer.
- The UNITS button is disabled.
- If an alarm condition occurs, the display will alternate between the Count Time remaining and the Count Rate. The ALARM display indicator and the Alarm Status LEDs will turn on. Alarms are latching in COUNT mode.
- If other operational modes are available, pressing the MODE button will cancel the current Count Timer and move to the next available operational mode.
When the Count Timer has finished:
- The display will show either the accumulated total for c, d, mR, and µSv, or the timed ratemeter average for cps, cpm, Bq, dpm, or the average exposure or average dose in mR/h and µSv/h.
- Pressing the UNITS button will switch between the Primary and Secondary Count Units. The newly selected Count Units will be displayed for approximately one second, and the display will then return to the accumulated total or timed ratemeter average, depending on the newly selected Count Units.
- Under a non-alarm condition, the Alarm Status LED will be off; pressing the ON/ACK button will reset the Count Timer.
- If an alarm condition occurred during the Timed Count, a continuous audio tone will sound, and the ALARM display indicator and the Alarm Status LED will turn on. Pressing the ON/ACK button once will acknowledge and turn off the continuous tone alarm audio. Pressing the ON/ACK button a second time will clear the alarm condition and reset the Count Timer. Alarms are latching in COUNT mode.
- If other operational modes are available, pressing the MODE button will move to the next available operational mode.
Figure 7: COUNT mode operation showing COUNT Timer of 10 minutes, 30 seconds.
Specifications
Detector: 2.5 x 1.9 cm (1 x 0.75 in.) (diameter x length) CsI(Tl)
scintillator (energy dependent)
Exposure Linearity: ±10% from background to 200 µSv/h (20 mR/hr)
Count Rate Linearity: ±10% from 8 cps to 8 kcps (200 cpm to 999 kcpm)
Energy Range: from 40 keV to 3 MeV
Operating Exposure Rate Range: from .05 to 200 µSv/h (.005 to 20 mR/hr)
Gamma Sensitivity: approximately 165 kcpm/mR/hr or 27.5cps/μSv/h (137 Cs)
Alarms: count rate and scaler alarm set points adjustable over the
display range
Loss of Count Protection: after 60 seconds of no pulses from detector,
unit will flash a zero reading and the alarm audio will be triggered
LCD Display: 3½ digit LCD with large 12.7 mm (0.5 in.) digits, (k)cps,
(k)cpm, (k)Bq, (k)dpm, mR(/h), µSv(/h), low-battery indicator, MAX, ALARM
Display Range:
- 0.00 cps to 19.9 kcps
- 0 cpm to 999 kcpm
- 0.00 Bq to 19.9 kBq
- 0 dpm to 999 kdpm
- 0.00 to 200 µSv/h
- .000 to 20 mR/h
- MAX RATE of 200 µSv/h or 20 mR/hr flashes and the alarm is triggered if the radiation field exceeds the display range.
Backlight: built-in ambient light sensor automatically activates low-
power LED backlight, or may be configured for “Continuous On” operation (will
reduce battery life)
User Controls:
- ON/OFF/QUIET – press to turn ON, tap to alternate between ‘click’ audio and QUIET, hold for OFF
- MODE – alternates between RATE (count rate), MAX ( captures peak rate), and COUNT ( preset count time) modes
- UNITS – alternates between Primary and Secondary units
Count Time Range: 1 second to 20 minutes, or “0” enables continuous
counting until stopped by user
Response Time: user-selectable from 1 to 60 seconds, or Auto-Response
Rate FAST or SLOW
Sigma Audio: greater than 60 dB at 0.6 m (2 ft), approximately 4.5 kHz
Power: two alkaline “AA” batteries
Battery Life: approximately 200 hours of operation
Construction: high-impact plastic with water-resistant rubber seals and
separate battery compartment
Environmental Rating: NEMA (National Electrical Manufacturers
Association) rating of 4 or IP (Ingress Protection) rating of 55
Temperature: -20 to 50 °C (04 to 122 °F)
Size: 4.6 x 6.9 x 27.2 cm (1.8 x 2.7 x 10.7 in.) (H x W x L)
Weight: 0.35 kg (0.75 lb)
Setup Mode
Warning! Only advanced users or administrators should consider changing any of the parameters in the following section. Incorrect settings could jeopardize the safety of users depending on this instrument.
Setup Overview
Your instrument has been shipped from Ludlum Measurements only after passing
electronic checkout, a 30-hour burn-in process, and a careful calibration
process. Calibration papers are supplied with each instrument shipped from
Ludlum Measurements.
Recalibration should be accomplished after maintenance or adjustments have
been performed on the instrument. Recalibration is not normally required
following instrument cleaning or battery replacement. Recalibration does not
require any special tools or software to perform.
Ludlum Measurements offers a full-service repair and calibration department.
Not only do we repair and calibrate our own instruments, we also service most
other manufacturers’ instruments. Calibration procedures are available upon
request for customers who choose to calibrate their own instruments.
Note: Ludlum Measurements, Inc. recommends recalibration at intervals no
greater than one year, assuming that regular operational checks are
performed. Check the appropriate local, state, and federal regulations to
determine required recalibration intervals.
Default Setup Values
Setup Parameter | Default Valu | Notes |
---|---|---|
Calibration Constant | 175 | kcpm per mR/h |
Dead Time Correction | 8 | Microseconds |
Efficiency | 15 | Efficiency % |
Primary Units | mR/hr | Non-SI exposure rate |
Primary Units RATE/ MAX
Mode Alarm Point| 0| Disabled
Primary Count Units| mR/h| Non-SI exposure
Primary Count Alarm Point| 0| Disabled
Secondary Units| cpm| Counts per minute
Secondary Units RATE/ MAX
Mode Alarm Point| 0| Disabled
Secondary Count Units| c| Counts per minute
Secondary Count Alarm Point| 0| Disabled
Response Time| 0| Enable Auto Response
Auto-Response Rate| S| Slow Auto Response
Operational Modes| 0| All Modes Available
Count Time| 1:00| One Minute
Low Light Threshold| 55|
Entering Setup Mode
To enter setup mode, power down the Model 26S, then turn the unit back ON.
Following the display of the Firmware version, when the instrument has begun
normal operation, press the MODE button three times (within four seconds) to
enter Setup mode. Entry to Setup mode can be confirmed when the numeric
portion of the display shows the Calibration Constant parameter, and no units
(cps, cpm, Bq, dpm, mR/h, or µSv/h) are displayed. If you simply wish to view
the parameters, you may do so by not pressing any other buttons. The
parameters will change every four seconds when no buttons are pressed, or the
UNITS button can be pressed to advance to the next parameter. The unit will
return automatically to normal operation after the last parameter is
presented.
SETUP PROTECT: The Model 26S parameters can be protected from
unauthorized changes via the internal switch located on the Model 26S circuit
board. To change the switch, open the battery compartment and remove the
batteries from the Model 26S. Next, loosen the four captive Phillips head
screws that fasten the detector cover.
Now turn the Model 26S over so that the detector is facing up. Gently remove
the plastic case that covers the back of the instrument, paying careful
attention not to lose the protective screen that covers the detector itself.
Remove the protective screen, and gently lift the GM pancake detector and set
it to the side. The DIP (dual in-line position) switch should now be visible
below the detector retainer.
To protect the Model 26S from changes in Setup mode, slide DIP Switch 2
(closest to the battery compartment) to the ON (right) position. If DIP
Switch 2 is in the OFF (left) position, changes are allowed in Setup mode.
Once the DIP Switch is set as desired, gently replace the detector in the
detector retainer. Place the detector screen over the detector, and fit the
plastic case on the back of the instrument. Turn the instrument over and
tighten the four Phillips head screws. Install the batteries, and replace the
battery cover. Note that with the DIP Switch 2 in the ON position, Setup mode
may be entered and parameters viewed, but changes cannot be made.
DISPLAY BACKLIGHT ‘Continuous On’: The Model 26S display backlight can be
set to remain on continuously during operation. Follow the steps above for
SETUP PROTECT, but use DIP Switch 1 for display backlight selection. Setting
DIP Switch 1 to the ON (right) position will configure the display backlight
to remain on during operation. Set DIP Switch 1 to the OFF (left) position,
and the display will be backlit only when light levels are low.
NOTE: Setting the display backlight for continuous-on operation can
result in reduced battery life.
Setup Mode Operation
Once the Model 26S is in Setup mode, the Calibration Constant will be
displayed on the LCD, and the hundredths digit will be blinking, indicating it
as the selected item. Use the MODE button to adjust the value for the selected
item. When the appropriate value is selected for that item, press the ON/ACK
button to move to the next item. When the desired value is displayed, either
wait for four seconds for the Model 26S to switch to the next parameter, or
press the UNITS button to advance to the next parameter. When the Model 26S is
in PROTECT mode (dipswitch setting), the Setup parameters will cycle through
to display the set values, but changes are not possible.
The order of Setup parameters for the Model 26S is as follows:
Calibration Constant (Default 175 kcpm per mR/h) – Use ON/ACK to select
the value to adjust, and MODE to adjust the value. Value is in kcpm per mR.
Available values are:
Ones Place (0-9)
Tens Place (0-9)
Hundreds Place (1-9)
Decimal Point
Dead Time Correction (Default 8 microseconds) – Use ON/ACK to select the
value to adjust, and MODE to adjust the value. Value is in microseconds.
Available values are:
Ones Place (0-9)
Tens Place (0-9)
Hundreds Place (0-9)
Efficiency (Default 15.0%) – Use ON/ACK to select the value to adjust,
and MODE to adjust the value. Available values are:
Ones Place (0-9)
Tens Place (0-9)
Hundreds Place (0-9)
Number of Decimal Places (0 or 1)
Normally the efficiency is used on a per detector basis, i.e. the efficiency
of the detector is calculated by dividing the count rate received from a
source by the total disintegration rate of the source.
Primary Units (Default mR/hr) – Use MODE to select. Available values are
cps, cpm, Bq, dpm, mR/h, and µSv/h.
Primary Units RATE/MAX Mode Alarm Point (Default 000) – Use ON/ACK to
select the value to adjust, and MODE to adjust the value. Units will be the
same as selected earlier with Primary Units. The ALARM LCD Segment will be on
to indicate an Alarm parameter. Set this Alarm Point to 000 to disable.
Available values are:
Ones Place (0-9)
Tens Place (0-9)
Hundreds Place (0-9)
Thousands Place (1 on or off)
Number of Decimal Places (0, 1, or 2– only available if k selected)
Range (k on or off – not available for mR/h or µSv/h)
Note: If the Primary Units has changed to a value other than that used to
previously set this Alarm Point, the Alarm Point will be reset to 000.
Primary Count Units (Default mR/hr) – Use MODE to select. Available
values are dependent on the selected Primary Units:
Primary Units | Primary Count Units Available |
---|---|
cps | cps |
c
cpm| cpm
c
Bq| Bq
d
dpm| dpm
d
mR/h| mR/h
mR
µSv/h| µSv/h
µSv
Primary Count Alarm Point (Default 000) – Use ON/ACK to select the value
to adjust, and MODE to adjust the value. Units will be the same as
selected earlier with Primary Count Units. Primary Count Units of c or d will
not be displayed, but mR or µSv will. The ALARM LCD Segment will be on to
indicate an Alarm parameter. Set this Alarm Point to 000 to disable. Available
values are:
Ones Place (0-9)
Tens Place (0-9)
Hundreds Place (0-9)
Thousands Place (1 on or off)
Number of Decimal Places (0, 1, or 2– only available if k selected)
Range (k on or off – not available for mR or µSv)
Note: If the Primary Count Units has changed to a value other than that
used to previously set this Alarm Point, the Alarm Point will be reset to 000.
Secondary Units (Default cpm) – Use MODE to select. Available values are
cps, cpm, Bq, dpm, mR/h, µSv/h, and off (no units displayed), excluding the
units selected as Primary Units.
Secondary Units RATE/MAX Mode Alarm Point (Default 000) – Use ON/ACK to
select the value to adjust, and MODE to adjust the value. If the Secondary
Units is off, this parameter will be skipped. Otherwise, units will be the
same as selected earlier with Secondary Units. The ALARM LCD Segment will be
on to indicate an Alarm parameter. Set this Alarm Point to 000 to disable.
Available values are:
Ones Place (0-9)
Tens Place (0-9)
Hundreds Place (0-9)
Thousands Place (1 on or off)
Number of Decimal Places (0, 1, or 2– only available if k selected)
Range (k on or off – not available for mR/h or µSv/h)
Note: If the Secondary Units has changed to a value other than that used
to previously set this Alarm Point, the Alarm Point will be reset to 000.
Secondary Count Units (Default c) – Use MODE to select. If the Secondary
Units is off, this parameter will be skipped. Otherwise, the available values
are dependent on the selected Secondary Units:
Secondary Units | Secondary Count Units Available |
---|---|
cps | cps |
c
off
cpm| cpm
c
off
Bq| Bq
d
off
dpm| dpm
d
off
mR/h| mR/h
mR
off
µSv/h| µSv/h
µSv
off
Secondary Count Alarm Point (Default 000) – Use ON/ACK to select the
value to adjust, and MODE to adjust the value. If the Secondary Units is off,
this parameter will be skipped. Otherwise, units will be the same as selected
earlier with Secondary Count Units. Secondary Count Units of c or d will not
be displayed, but mR or µSv will. The ALARM LCD Segment will be on to indicate
an Alarm parameter. Set this Alarm Point to 000 to disable. Available values
are:
Ones Place (0-9)
Tens Place (0-9)
Hundreds Place (0-9)
Thousands Place (1 on or off)
Number of Decimal Places (0, 1, or 2– only available if k selected)
Range (k on or off – not available for mR or µSv)
Note: If the Secondary Count Units has changed to a value other than that
used to previously set this Alarm Point, the Alarm Point will be reset to 000.
Response Time (Default 0=auto) – Use ON/ACK to select the value to be
adjusted and MODE to adjust the value. Setting the Response Time to a fixed
value is useful primarily when performing surveys to a fixed MDA (Minimum
Detectable Activity) level. Setting the Response Time to 0 will enable the
Auto Response mode for the Model 26S (see the next parameter). Available
values for the Response Time (in seconds) are:
Ones Place (0-9)
Tens Place (0-6, 6 forces max Response Time of 60)
Auto-Response Rate (Default S) – Use MODE to select Fast (F) or Slow (S).
When operating in Auto-Response mode, the Model 26S will vary the Response
Time based on the Auto-Response Rate selected (Fast or Slow) and the current
Count Rate. The following table shows the response time for different count
rates when these auto response modes are chosen:
Count Rate| Auto Response Time – Fast (Seconds)| Auto Response
Time – Slow (Seconds)
---|---|---
Less than 3 kcpm (50 cps)| 5| 10
Between 3 kcpm and 4 kcpm (67 cps)| 4| 8
Between 4 kcpm and 6 kcpm (100 cps)| 3| 6
Between 6 kcpm and 12 kcpm (200 cps)| 2| 4
More than 12 kcpm| 1| 2
The Model 26S also utilizes a Step function in Auto Response mode, which
enables faster response to a significant increase or decrease in Count Rate.
When the instrument detects a sudden change in count rate from the detector,
the response time is reduced to 1 second to quickly show the new value.
Operational Modes (Default 0=all modes available) – Use MODE to adjust
the value. Available values are:
0 – RATE, MAX, and COUNT Modes
1 – RATE and MAX Modes only
2 – RATE and COUNT Modes only
3 – RATE Mode only
Count Time (Default 1 minute) – Use ON/ACK to select the value to adjust,
and MODE to adjust the value. Setting Count Time to 0 enables continuous
count until reset. If 19 minutes are selected, then the maximum seconds value
is 60; otherwise, the maximum seconds value is 59. Available values are:
Ones Place (0-9)
Tens Place (0-6, 6 only available if minutes value is 19)
Hundreds Place (0-9)
Thousands Place (1 on or off)
Low Light Threshold (Default 55) – Use ON/ACK to select the value to
adjust, and MODE to adjust the value. Available values are:
Ones Place (0-9)
Tens Place (0-9)
Hundreds Place (0-9)
NOTE: If no buttons are pressed for four seconds, the Model 26S will
switch to the next parameter, and if on the last parameter, will save the
parameters and exit Setup mode, returning to RATE mode operation. The UNITS
button can be used to advance to the next parameter. To end Setup mode and
save the current setting, press and hold the UNITS button for approximately 5
seconds.
Safety Considerations
Environmental Conditions for Normal Use
Indoor or outdoor use
No maximum altitude
Temperature range of -20 to 50 °C (-4 to 122 °F)
Maximum relative humidity of less than 95% (non-condensing)
Pollution Degree 3 (as defined by IEC 664): (Occurs when conductive pollution
or dry nonconductive pollution becomes conductive due to condensation. This is
typical of industrial or construction sites.)
Not certified for use in an explosive atmosphere
Warning Markings and Symbols
Caution! The operator or responsible body is cautioned that the
protection provided by the equipment may be impaired if the equipment is used
in a manner not specified by Ludlum Measurements, Inc.
The Model 26S Integrated Scintillator is marked with the following
symbols:
CAUTION (per ISO 3864, No. B.3.1): designates hazardous live voltage and
risk of electric shock. During normal use, internal components are hazardous
live. This instrument must be isolated or disconnected from the hazardous live
voltage before accessing the internal components. This symbol appears on the
side panel. Be sure to take the precautions noted in the next section whenever
necessary.
The “crossed-out wheelie bin” symbol notifies the consumer that the product is
not to be mixed with unsorted municipal waste when discarding. Each material
must be separated. The symbol is placed on the label located on the side
panel. See section 8, “Recycling,” for further information.
The “CE” mark is used to identify this instrument as being acceptable for
use within the European Union.
Cleaning and Maintenance Precautions
The Model 26S may be cleaned externally with a damp cloth, using only water as the wetting agent. Observe the following precautions when cleaning or performing maintenance on the instrument:
- Turn the instrument OFF and remove the batteries.
- Allow the instrument to sit for one minute before cleaning the exterior or accessing any internal components for maintenance.
Revision History
NOTE: This section of the manual will be updated with each revision of
the Model 26S in order to document changes over time. Ludlum Measurements’
policy is to provide free software upgrades to instruments for the life of the
instrument.
January 2023: New manual.
Recycling
Ludlum Measurements, Inc. supports the recycling of the electronics products
it produces for the purpose of protecting the environment and to comply with
all regional, national, and international agencies that promote economically
and environmentally sustainable recycling systems. To this end, Ludlum
Measurements, Inc. strives to supply the consumer of its goods with
information regarding reuse and recycling of the many different types of
materials used in its products. With many different agencies – public and
private – involved in this pursuit, it becomes evident that a myriad of
methods can be used in the process of recycling. Therefore, Ludlum
Measurements, Inc. does not suggest one particular method over another, but
simply desires to inform its consumers of the range of recyclable materials
present in its products, so that the user will have flexibility in following
all local and federal laws.
The following types of recyclable materials are present in Ludlum
Measurements, Inc. electronics products, and should be recycled separately.
The list is not all-inclusive, nor does it suggest that all materials are
present in each piece of equipment:
Batteries | Glass | Aluminum and Stainless Steel |
---|---|---|
Circuit Boards | Plastics | Liquid Crystal Display (LCD) |
Ludlum Measurements, Inc. products that have been placed on the market after
August 13, 2005, have been labeled with a symbol recognized internationally as
the “crossed-out wheelie bin,” which notifies the consumer that the product is
not to be mixed with unsorted municipal waste when discarding. Each material
must be separated. On the Model 26S, the symbol will be placed on the serial
number label located on the side of the instrument.
The symbol appears as such:
Maintenance and Spare Parts
Maintenance
For external cleaning of the instrument, scrub with a dampened, soapy cloth. This is the only regular maintenance required.
Spare Parts
Following is a list of spare parts for a Model 26S:
Description | Part # |
---|---|
AA Batteries | 14-5240 |
M 26S Main Bd | 5520-217 |
CsI(Tl) detector with harness | 4498-1045 |
Battery Cover Hook | 7498-319 |
Options
Following is a list of optional items (not supplied) that could enhance the use of the Model 26S:
Description | Part # |
---|---|
Standard 1/8 in. jack | 4498-538 |
Canvas holster | 2312577 |
Model 180-28 sample holder | 47-3948 |
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>