GRAM K3iX Stainless Steel Indicator User Guide

June 9, 2024
GRAM

GRAM K3iX Stainless Steel Indicator User Guide

GRAM K3iX Stainless Steel Indicator User Guide

PACKAGING

1 x 220 V/12 VDC 1A mains adapter.
1 x K3iX weight indicator
Accompanying documentation

LCD DISPLAY

GRAM K3iX Stainless Steel Indicator User Guide - LCD
DISPLAY

GRAM K3iX Stainless Steel Indicator User Guide - KEYBOARD &
DISPLAY GRAM K3iX Stainless Steel
Indicator User Guide - KEYBOARD & DISPLAY

SCALE SETTINGS ­ MENU OPTIONS

For access the settings menu press the Esc / M key. ¿ validate the current setting / moves to the next digit when editing a numeric value ¿¿ validate a manually input value.
ESC Returns to the menu’s previous level without making any change.
ß à change to the next / previous option. ¯ changes between the different values that can be assigned to a specific option.
= Factory setting

GRAM K3iX Stainless Steel Indicator User Guide - SCALE SETTINGS – MENU
OPTIONS GRAM K3iX
Stainless Steel Indicator User Guide - SCALE SETTINGS – MENU
OPTIONS

For access the settings menu press the Esc/M key. ¿ validate the current setting / moves to the next digit when editing a numeric value. ¿¿ validate a manually input value.
ESC Returns to the menu’s previous level without making any change.
ß à change to the next / previous option. ¯ changes between the different values that can be assigned to a specific option.
= Factory settings

SCALE CALIBRATION

At the SCAlE menu you will find the settings needed to define and adjust the measurement scale of the instrument: Measurement unit, maximal capacity, scale interval (division), decimal point, as well as different options related to the operation of the auto-zero device.
Access to these configuration options is reserved for technical personnel and is protected by a keyword to avoid accidental changes that would cause the instrument to malfunction.
It is possible to directly access the adjustment (calibration) function of the instrument when the indicator is turned on.

To do this, turn on the indicator, and while the LCD test appears with all segments on, press the and same time (one short press, not sustained).

keys at the

Once you have entered the access code to the scale settings menu, with the scale empty, select the option CALIB.

  1. The display will show that the acquisition of the initial zero value is in progress with the blinking message “CAL 0”.

  2. Once the zero value has been adjusted, place the adjustment weight (a standard weight) on the load receptor.

  3. Enter the weight value in the indicator, including the decimal positions. Use the cursor movement keys to move through the different positions on the display.

  4. Once you enter the weight value, double click the ¿ key to validate and move to next step. The display will show the blinking message “-CAL-” while acquiring the adjustment value.

  5. Lastly, it will show the message “GEO” for a few seconds, asking for the code of the geographical location where you did the adjustment.
    The geographical location code is a value from 0 to 31, which you must choose from the attached table. Use the and ¯ keys to change the value and validate by clicking on the ¿ key.

  6. Lastly, the message “SAVE” will briefly appear, indicating that the adjustment has been saved in the non-volatile memory. The indicator returns to normal use mode, displaying the weight on the load receptor.

If the automatic correction of the weight according to the geographical latitude and height (“G-COR option”) is set to ON, the next time you switch on the indicator after an adjustment, once the display test and initial welcome message is completed, the user will be asked to enter the value corresponding to the geographical area where the scale will be used.
Once the value has been entered for the geographical area where the scale is placed, it is recorded in the nonvolatile memory of the indicator, and the user will not be asked for it again.
The geographical area where the scale is used can be modified later whenever you wish by entering the menu with NEnU àSCALE à CAL à G-SET à GEO à G nn (being nn {0-31}).
The automatic correction of the setting according to geographical area can be disabled by entering the menu with NEnU àSCALE à CAL à G-SET à G-Cor à OFF

TABLE OF GEOGRAPHICAL ADJUSTMENT VALUES

GRAM K3iX Stainless Steel Indicator User Guide - TABLE OF GEOGRAPHICAL
ADJUSTMENT VALUES

CONNECTIONS

GRAM K3iX Stainless Steel Indicator User Guide -
CONNECTIONS

MORE INFORMATION

Download the full manual from the following link:

GRAM K3iX Stainless Steel Indicator User Guide - QR
Codehttps://gram-group.com/wp- content/uploads/2022/02/MU_K3X_V6_001_ENG.pdf

Gram Precision S.L. Travesía Industrial, 11 · 08907 Hospitalet de Llobregat · Barcelona (Spain)
Tel. +34 902 208 000 · +34 93 300 33 32
Fax +34 93 300 66 98
[email protected]
www.gram-group.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

GRAM User Manuals

Related Manuals