DOSTMANN electronic 5040-0301 PH Check PH Instrument
INTRODUCTION
Congratulations on your purchase of this pen type long probe pH meter. Please read the manual completely before using this meter. Filing and keeping the manual for future reference. Recommended to soak the electrode for at least 30 minutes before using to clear up the lazy effect or to make the electrode wet if the pH electrode dries out.
Features
- IP65 Waterproof housing.
- Dual display with ATC.
- Data hold to freeze readings.
- Compact size, easy to fit in pocket.
- Low battery indicator.
- Auto power off.
- °C/°F unit switchable.
- Power by 2pcs CR2032 batteries.
- Multi points calibration.
- One touch only for calibration.
Kindly note / Safety Instructions
- Check if the contents of the package are undamaged and complete.
- Remove the protection foil above the display.
- For cleaning the instrument please do not use an abrasive cleaner only a dry or moist piece of soft cloth. Do not allow any liquid into the interior of the device.
- Please store the measuring instrument in a dry and clean place.
- Avoid any force like shocks or pressure to the instrument.
- No responsibility is taken for irregular or incomplete measuring values and their results, the liability for subsequent damages is excluded!
- Do not use the device in explosive areas. Danger of death!
- Do not use the device in an environment hotter than 85°C! The lithium battery may explode!
- Do not expose the units to microwave radiantion. The lithium battery may explode!
- Keep these devices and the batteries out of reach of children.
- Batteries contain harmful acids and may be hazardous if swallowed. If a battery is swallowed, this can lead to serious internal burns and death within two hours. If you suspect a battery could have been swallowed or otherwise caught in the body, seek medical help immediately.
- Batteries must not be thrown into a fire, short-circuited, taken apart or recharged. Risk of explosion!
- Low batteries should be changed as soon as possible to prevent damage caused by leaking. Never use a combination of old and new batteries together, nor batteries ofdifferent types.
- Wear chemical-resistant protective gloves and safety glasses when handling l e a k i n g batteries.
MATERIAL SUPPLIED
This package contains:
1 x PH CHECK, 2 X CR2032 button battery,
1 x Operation manual and
1 x Container with storage solution
POWER SUPPLY
direction by coin, remove the battery stripe (DON‘T discard the washer!) Put
back the battery cover and turn it in counter-clockwise direction to lock.
The meter is powered by 2 pcs CR2032 batteries. To check the battery when:
- The battery symbol appears on LCD
- The meter can not power on
To install the batteries
- Turn off the meter.
- Loose the battery cover in clockwise direction by coin (DON‘T discard the washer!)
- Replace the old batteries with two new button cells CR2032.
- Make sure the batteries are in place and the polarity is correct.
- .Put back the battery cover and turn it in counter-clockwise direction to lock
NOTE:
- Please re-calibrate the meters after changing batteries.
- Remove battery from instruments that you do not plan to use for a month or more. Do not leave battery in instrument.
LCD DISPLAY
– The 1st display shows the measured pH reading.
– The 2nd display shows the temp.
– Cal = Calibration mode
– ATC = Auto Temp. Compensation
– Data hold
KEYPAD
OPERATION
-
Remove the probe storage bottle from meter to expose the electrode out. It is normal if you find white crystals are present on the cap or electrode assembly.
WARNING: For pH meters, please always make the bottle is filled with storage buffer to keep the electrode wet and in a good storage condition.
-
Dip the electrode into the test solution. Press „ „ to power on the meter and tir it to get a stable reading.
-
A small dot „ . „ is flashing while the meter is in measurement mode. The LCD not only shows the measured pH value but also displays the temperature.
-
Press „ „ to freeze current readings. The „ „ icon will appear on the LCD and the small dot will not flash. Press „ „ again to release the hold mode.
-
Turn off the meter by pressing „ „ button.
-
Once finishing the measurement, clean the electrode, put back the bottle, store the meter under the 0~50°C.
-
Please see page 18 for the calibration procedures
AUTO POWER OFF (SLEEP FUNCTION)
This meter will shut off automatically 20 min.of inactivity. For operating
longer time, you can disable the sleep mode.
To disable the auto power off: Before power on, pressing „ „ + „ „ keys
simultaneously until a „n“ appeared on the screen and then release keys to
return to normal mode.
Note:
The disable sleep mode will be invalid after every power off.
AUTOMATIC TEMPERATURE COMPENSATION (ATC) & TEMP. UNIT SETTING
pH CHECK is capable of measuring with Automatic Temperature Compensation. ATC will show in the middle top of the LCD.
Temp. Unit Setting
To select the temp. unit (°C or °F), turn off the meter first. When the meter is off, press „ „ and „ „ at the same time until the °C or °F appears on the LCD. Press „ „ to select the preferred unit and then press „ „ to save. „SA „ will appear on the LCD for one sec. and then go back to normal.
CALIBRATION MODE (CAL)
Calibration is necessary and should be done regularly, recommended everyday if the meter is used often. The unique calibration design of the meter features automatic buffer recognition to avoid errors.
Calibration
-
Power on the meter.
-
Place the electrode into a buffer solution (4,7 or 10), pH7 should be calibrated first and then 4 or 10 pH for better accuracy
-
Press „ „ to enter pH calibration mode. The icon „CAL” will appear on the LCD for one second and then pH value (4,7 or 10) will be displayed on the LCD.
-
If probe or buffer is in error: If the buffer is incorrectly inserted or the probe is damaged or the probe can not detect buffer in below voltage range, the meter will escape calibration mode automatically after 10 seconds. Text „End“ will appear on the LCD in one second and then will go back to normal status.
Acceptable voltage range of each point:
PH 4.00 : 97mV….. 250mV
PH 7.00 : – 60mV….. 60mV
PH10.00 : – 250mV….. -97mV -
If the probe recognizes the buffer: If the probe successfully recognizes the buffer, the buffer pH value (4 or 7 or 10) will appear on the display in 2 seconds.
If the calibration buffer is not 4,7,10 but another value, such as 7.01, just press „ „ to change the value.
-
The adjustable cal. point range for 4.0 pH is from 3.50 to 4.50. For 7.0 pH is from 6.50 to 7.50. For 10.0 pH is from 9.50 to 10.50
-
To save the calibration value: When the electrode reads a stable value and user stops pressing any keys, the meter will automatically save the value and then escape the calibration mode.
-
Rinse the probe with de-ionized water or a rinse solution (tap water…) after each measurement to last the meter‘s life.
-
Repeat above steps until the 3 point calibration are finished.
MAINTENANCE
Please always keep the pH glass bulb wet by using the storage bottle to
protect and store the electrode.
Always rinse the pH electrode in de-ionized water or rinse solution (tap
water……) before next use.
Never touch or rub glass bulb in order to last pH electrode life.
Make sure the electrode is clean. Between measurements, rinse the electrode
with deionised water. If the eletrode has been exposed to a solvent immiscible
with water, clean it with a solvent miscible with water e.g. ethanol or
acetone and rinse carefully with water. Store the electrode carefully
TROUBLESHOOTING Power on but no display
- Make sure the time of pressing power key is more than 100 mS.
- Check the battery are in place and make good contact and correct polarity.
- Replace with new batteries and try again.
- Move away the batteries for one minute and then put back again.
Slow response
Clean probe by immersing the electrode in tap water for 10-15 minutes, then rinse thoroughly with distilled water or use a general purpose electrode cleaner.
LCD display “ – – – “ Out of pH range, too acidic/or too alkaline.
H.“ or „L.“ Out of temp. range, too cold/or too hot.
pH value fluctuate quickly It is normal when the electrode is not immersed in
the water but exposed in air. Operating Temp.: 0°~50°C (32~122°F)
Operating Humidity: 0~80% RH pH calibration buffer suggested:
─ Standard USA buffers
─ Standard NIST buffers
─ Standard DIN buffers
Explanation of symbols
This sign certifies that the product meets the requirements of the EEC directive and has been tested according to the specified test methods.
Waste disposal
This product and its packaging have been manufactured using high-grade
materials and components which can be recycled and reused. This reduces waste
and protects the environment. Dispose of the packaging in an environmentally
friendly manner using the collection systems that have been set up.
Disposal of the electrical device: Remove non-permanently installed
batteries and rechargeable batteries from the device and dispose of them
separately. This product is labelled in accordance with the EU Waste
Electrical and Electronic Equipment Directive (WEEE). This product must not be
disposed of in ordinary household waste. As a consumer, you are required to
take end-of-life devices to a designated collection point for the disposal of
electrical and electronic equipment, in order to ensure environmentally-
compatible disposal. The return service is free of charge. Observe the current
regulations in place!
Disposal of the batteries: Batteries and rechargeable batteries must never be
disposed of with household waste. They contain pollutants such as heavy
metals, which can be harmful to the environment and human health if disposed
of improperly, and valuable raw materials such as iron, zinc, manganese or
nickel that can be recovered from waste. As a consumer, you are legally
obliged to hand in used batteries and rechargeable batteries for
environmentally friendly disposal at retailers or appropriate collection
points in accordance with national or local regulations. The return service is
free of charge. You can obtain addresses of suitable collection points from
your city council or local authority. The names for the heavy metals contained
are: Cd = cadmium, Hg = mercury, Pb = lead. Reduce the generation of waste
from batteries by using batteries with a longer lifespan or suitable
rechargeable batteries. Avoid littering the environment and do not leave
batteries or battery-containing electrical and electronic devices lying around
carelessly. The separate collection and recycling of batteries and
rechargeable batteries make an important contribution to relieving the impact
on the environment and avoiding health risks.
WARNING! Damage to the environment and health through incorrect disposal of the batteries!
WARNING! Batteries containing lithium can explode Batteries and rechargeable batteries containing lithium (Li=lithium) present a high risk of fire and explosion due to heat or mechanical damage with potentially serious consequences for people and the environment. Pay particular attention to correct disposal.
DOSTMANN electronic GmbH Mess- und Steuertechnik
Waldenbergweg 3b
D-97877 Wertheim-Reicholzheim Germany
Phone: +49 (0) 93 42 / 3 08 90
E-Mail: info@dostmann-electronic.de
Internet: www.dostmann-electronic.de
echnische Änderungen vorbehalten
Nachdruck auch auszugsweise untersagt
Stand06 2206CHB ● © DOSTMANN electronic GmbH
Technical changes, any errors and misprints reserved
Reproduction is prohibited in whole or part
Stand06 2206CHB ● © DOSTMANN electronic GmbH