APERA TDS20 Value Portable Pocket TDS Tester Meter Instruction Manual

June 9, 2024
APERA

TDS20 Pocket TDS Tester
Instruction Manual

APERA INSTRUMENTS (Europe) GmbH
www.aperainst.de

Battery Installation

Please install batteries according to the following steps. Please note polarity:
“+” (positive) is upward; “-” (negative) is downward

Keypad Functions

Short press——- < 2 seconds
Long press——– > 2 seconds

| 1. Short press to turn on, long press to turn off;
2. When turned off, long press to enter setup;
3. In mode setting, short press to change parameter;
---|---
| 1. When turned on, long press to enter calibration mode.
2. In calibration mode, short press to confirm calibration;
3. In mode setting, short press to confirm parameter selection.

Diagram – 1

Complete Kit

Calibration

  1. Rinse the probe in distilled water and dry it. Short press to turn on the tester.
    Diagram – 3

  2. Long press to enter calibration mode; Short press to exit.

  3. Dip the probe into 1413 µS calibration solution. Stir gently, leave it to stand, LCD displays (Diagram 3), short press to complete 1- point calibration.
    The tester returns to measuring mode, and calibration icon appears at the button left of LCD.

  4. Rinse probe in distilled water and dry it. Follow the steps in 4.2-4.3 to complete 2nd point calibration in 12.88 mS calibration solution, tester returns to measuring mode, calibration icons display on bottom left of LCD.

TDS Measurement

  1. Short press to turn on tester. Rinse probe in distilled water and dry it.
  2. Stir probe in the sample solution gently, leave it to stand. Get readings after the smile icon comes up and stays.

Notes

  1. The tester adopts 1413 µS and 12.88 mS standard calibration solutions. Users can use 1-2 point calibrations as needed. For most circumstances, calibrating in 1413 µS to complete 1st point calibration will meet testing requirements.
  2. The tester has self-diagnosis functions:
    Symbol | Self-Diagnosis information| How to fix
    ---|---|---
    | Wrong calibration solution, which exceeds the recognizable range of the meter.| 1. Check if calibration solution is correct
    2. Check if probe is damaged.
    | is pushed before measurement is stable ( comes up)| Wait for the smile icon to stay, and then short press
  3. The tester has already been calibrated after manufacture. Usually users can use the tester right away, or test it in the calibration solutions to test its accuracy. When error is large, calibrate it before using.
  4. We recommend replacing new calibration solutions after 5-10 times of calibrations to keep the solution’s accuracy.

Parameter Setting

  1. Parameter setting reference chart:
    Symbol| Parameter Setting Items| Code| Factory Default
    ---|---|---|---
    P1| TDS Factor| 0.40 to 1.00| 0.71
    P2| Select Temperature Unit| °C – °F| °C
    P3| Restore to factory default| No – Yes| No

  2. How to setup parameters:
    When turned off, long press to enter setup → short press to switch P1-P2-P3 Short press , parameter flickering → short press to choose, short press to confirm parameter selection → Long press to switch off.

  3. Parameter setting notes
    a) TDS Factor (P1):
    Users can adjust TDS Factor by experimental data or experience. The following chart lists some commonly used TDS Factors for reference.
    Conductivity and TDS Factor

Conductivity of the solution TDS Factor
0~100 μS/cm 0.6
100~1000 μS/cm 0.71
1~10 mS/cm 0.81
10~100 mS/cm 0.94

b) Restore to factory default: Select Yes to restore the calibration to the theoretical values and parameter setting to original values. When meter’s calibration or measurement performs abnormally, this function can be adopted so the meter goes back to factory default setting and then users can conduct calibration or take measurements again.

Technical Specifications

TDS| Range| 0 — 100.0 ppm, 0 — 1000 ppm, 0 —
10.00 ppt
---|---|---
Resolution| 0.1/1 ppm, 0.01 ppt
Accuracy| ± 1% F.S
Calibration points| 1-2 points
Automatic Temp.
Compensation| 0-50°C
Temp.| Range| 0-50°C
Resolution| 0.1°C
Accuracy| ± 0.5°C

Other Functions & Parameters

Indication of calibration points| | Auto Power-off| Power-off in 8 minutes if no operation
---|---|---|---
Indication of stable measurements| | Waterproof level| IP67, floats on water
Self-Diagnosis information| Erl, Er2| Power Supply| AAA batteries4
Low battery reminder| Flashes to remind to replace batteries| Battery Life| 1000 hours
Dimensions/Weight| Instrument: 40
31178mm/107g; Carrying case: 190165*140mm/438g

Warranty

We warrant this instrument to be free from defects in material and workmanship and agrees to repair or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a period of two years from the delivery (a six-month limited warranty applies to probes). This warranty does not apply to defects resulting from actions such as misuse (violation of the instructions in this manual or operations in the manner not specified in this manual), improper maintenance, and unauthorized repairs. Warranty period is the time limit to provide free service for the products purchased by customers, not the service life of the tester or probe.
Apera Instruments reserves the right to update the information in this manual without giving notice in advance.

APERA INSTRUMENTS (Europe) GmbH
Wilhelm-Muthmann-Straße 18
42329 Wuppertal, Germany
Contact: info@aperainst.de
Website: www.aperainst.de
Tel. +49 202 51988998

Documents / Resources

| APERA TDS20 Value Portable Pocket TDS Tester Meter [pdf] Instruction Manual
TDS20 Value Portable Pocket TDS Tester Meter, TDS20, Value Portable Pocket TDS Tester Meter, Pocket TDS Tester Meter, TDS Tester Meter, Tester Meter
---|---

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

APERA User Manuals

Related Manuals