Tuttnauer T-Edge 10 Fully Automatic Tabletop Autoclave User Guide

June 4, 2024
Tuttnauer

Tuttnauer T-Edge 10 Fully Automatic Tabletop Autoclave User Guide

REQUIRED SAFETY CAUTIONS

IMPORTANT: Read the safety section in the T-Edge operation manual prior to operating T-Edge

CAUTION: Hot surface.
CAUTION: It is recommended that a ground fault circuit breaker (GFCI) be installed on the power outlet

NOTE: Refer to the T-Edge manual before operating the device/autoclave

T-EDGE OVERVIEW

T-Edge is a fully automatic tabletop autoclave designed for sterilization of equipment used in first aid rooms, hospitals, dental clinics and medical clinics. It provides the following benefits:

  • Strict compliance with international safety standards.
  • Clinically proven sterilization capabilities.
  • Optimal chamber capacity with convenient equipment loading.
  • High-resolution tactile touchpad with a user friendly interface.
    T-EDGE OVERVIEW
    T-EDGE OVERVIEW

Optional equipment is supplied based on specific customers order.

CONNECTING T-EDGE

CONNECTING T-EDGE

POWERING UP T-EDGE

POWERING UP T-EDGE
POWERING UP T-EDGE

²Use of these ports and connectors is optional

INITIAL DATE AND LANGUAGE SETUP

When T-Edge powers on for the first time, perform the following steps:
DATE AND LANGUAGE SETUP

USER INTERFACE

Top bar
USER INTERFACE
USER INTERFACE

RESETTING ATMOSPHERIC PRESSURE

Prior to running the first cycle, resetting the atmospheric pressure is required.

To reset the atmospheric pressure, you must first open the Autoclave door and verify that the chamber’s temperature is below 113°F.

ATMOSPHERIC PRESSURE
ATMOSPHERIC PRESSURE

FILLING T-EDGE WITH WATER

Fill the water reservoir with distilled water as described below:
FILLING T-EDGE
NOTE: The upper water insertion cover design may vary.
FILLING T-EDGE
NOTE: Use flexible soft bottles (do not use glass or rigid plastic bottles)

NOTE: You may only fill T-Edge manually if it is not connected to the water supply facilities.

LOADING EQUIPMENT INTO T-EDGE

Load equipment into T-Edge in one of the following ways:
LOADING EQUIPMENT
NOTE: Equipment must be thoroughly cleaned prior to loading.
LOADING EQUIPMENT

SELECTING STERILIZATION PROGRAM

STERILIZATION PROGRAM
NOTE: The door must be open in order to select a program.

LOCKING THE DOOR

LOCKING THE DOOR

RUNNING STERILIZATION PROGRAM

STERILIZATION PROGRAM
STERILIZATION PROGRAM

REMOVING STERILIZED EQUIPMENT

STERILIZED EQUIPMENT
STERILIZED EQUIPMENT

CAUTION!
Equipment rack is extremely hot watch for vapor release when opening the chamber

TURNING OFF T-EDGE

TURNING OFF T-EDGE

TROUBLESHOOTING

Issue/Message

| Description|

Required Action

---|---|---
| Full water level (waste water tank)| Empty waste water reservoir
| Low water level (waste water tank)| Informative symbol. No action required
| Empty water level (waste water tank)| Informative symbol. No action required
| Full water level (clean water tank)| Informative symbol. No action required
| Low water level (clean water tank)| Fill the water reservoir (clean water tank)
| Alert| Tap icon to read a description of the alert

Cycle Failed

| Sterilization cycle failed due to the cycle being stopped by the user or due to a run-time error| Perform a new sterilization cycle
Blank display| User interface display is blank when T-Edge is turned on| Make sure the power switch on the back is turned ON Make sure the power cord is properly connected to the device and to the wall outlet Verify proper electrical power supply
“Low Temp” “High Temp” “High Temp (Ending)” “Low Pressure” “High Pressure” “High Pressure (Ending)” “High Pressure (Exhaust)” “High Pressure (Dry)” “Door is open (During the cycle)” “Canceled By User” “Air Error” “Power Down”| Various reasons (please refer to the Operation and Maintenance manual for more detailed information)| Perform a new sterilization cycle
“Heat Time Error” “Pressure Time Error”| Various reasons (please refer to the Operation and Maintenance manual for more detailed information)| Verify that T-Edge is not overloaded

TECHNICAL SPECIFICATIONS – CLASS B PROGRAMS

Autoclave Model| Chamber ( Size ) Inches| Volume| Overall Dimensions D x W x H (Inches)| No. of Trays| Connections| Standard Cassette Capacity| Item Supplies| Voltage (V) Freq. (Hz)| Weight
---|---|---|---|---|---|---|---|---|---
T-Edge10 220V| 18” depth x 10” diameter| 6 gal/ 23 liters| 21.1”x19” x19.7”| 5| 1 ethernet port, 4 USB ports, WiFi, intergrated automatic water filling and drain connection| 3 Full + 3 Half| Drain tube, tray holder| 230V, Frq 50- 60Hz| 117 lbs.
T-Edge10 110V| 18” depth x 10” diameter| 6 gal/ 23 liters| 21.1”x19” x19.7”| 5| 1 ethernet port, 4 USB ports, WiFi, intergrated automatic water filling and drain connection| 3 Full + 3 Half| Drain tube, tray holder| 120V, Frq 50- 60Hz| 117 lbs.
T-Edge11 110V| 18” depth x 11” diameter| 7 gal/ 27 liters| 21.1”x20” x19.7”| 5| 1 ethernet port, 4 USB ports, WiFi, intergrated automatic water filling and drain connection| 4 Full + 4 Half| Drain tube, tray holder| 120V, Frq 50- 60Hz| 124 lbs.

STERILIZATION PROGRAMS

 |

Icon

| Cycle Program| Load Type| *T-Edge10 Cycle Times Including Drying (Minutes)|

*T-Edge11 Cycle Times Including Drying (Minutes)

---|---|---|---|---|---
CLASS S| | Unwrapped Instr. 2700 F| Unwrapped Instruments| 20 Minutes| 25 Minutes
| Wrapped Pouches 2700 F| Wrapped Pouches| 32 Minutes| 43 Minutes
| Unwrapped Delicate 2500 F| Unwrapped Delicate Instruments| 45 Minutes| 47 Minutes
| Handpieces 2700 F| Handpieces| 45 Minutes| 59 Minutes
 | Icon| Cycle Program| Load Type| T-Edge10 Cycle Times Including Drying (Minutes)| T-Edge11 Cycle Times Including Drying (Minutes)
---|---|---|---|---|---
CLASS B|


| Unwrapped Instr. 2730 F| Unwrapped Instruments| 20 Minutes| 26 Minutes
| Wrapped Pouches 2730 F| Wrapped Pouches| 39 Minutes| 49 Minutes
| Unwrapped Delicate 2500 F| Unwrapped Delicate Instruments| 35 Minutes| 35 Minutes
| Wrapped Delicate 2500 F| Wrapped Delicate Instruments| 58 Minutes| 75 Minutes

*Cycle times are approximate and may vary based on the instrument load, heating, sterilization and drying.

Support

International Sales and Marketing
E-mail: [email protected]
www.tuttnauerusa.com

Tuttnauer USA Co., Ltd.
25 Power Drive, Hauppauge, NY 11788
Tel: +800 624 5836, +631 737 4850
Fax: +631 737 0720
E-mail: [email protected]
www.tuttnauerUSA.com

Distributed by:
Copyright © 2022 Tuttnauer Ltd, all rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language or computer language, in any form or by any means, electronic, mechanical or otherwise without prior written permission of Tuttnauer Ltd.
Learn from our experts
Join our blog tuttnauerusa.com/blog

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Tuttnauer User Manuals

Related Manuals