APERA LabSen 371 Plastic Flat pH Electrode User Manual
- June 8, 2024
- APERA
Table of Contents
LabSen 371 Plastic Flat pH Electrode
User Manual
LabSen 371 Plastic Flat pH Electrode
LabSen electrochemical sensors are premium pH electrode with manufacturing
technology and key components imported from Switzerland. LabSen 371 plastic
flat pH electrode adopts flat membrane and PTFE junction structure. It is
suitable for wet solid and semi-solid surface, such as paper, skin, textile
and leather.
This probe has following features:
- Anti-corrosion POM (Polyoxymethylene) shaft, have good mechanical performance.
- Blue gel inner solution, does not flow and will not cause bubble.
- Long life reference system, has better stability and service life.
Technical Data
Measuring Range | (0 ~ 14) pH | Electrolyte | Gel |
---|---|---|---|
Temperature Range | (0 ~ 80) °C | Soaking Solution | 3M KCl |
Shaft Material | POM | Electrode Dimension | (Ø12×105) mm |
Membrane Type | S | Membrane Resistance | <500MΩ |
Reference | Long Life | Connector | BNC |
Junction | PTFE | Cable | Ø3×1m |
Usage and Maintenance
2.1 Testing surface need to be clean and wet, such as wet the surface of skin
or textile by distilled water.
2.2 Connect BNC plug to the input on the pH meter.
2.3 When measuring, please unscrew the bottle cap, pull out the electrode and
rinse it with deionized water. After using, please put the electrode back into
the bottle and screw tight the cap.
2.4 The electrode’s measuring tip should be soaked in the soaking bottle
containing certain amount of storage solution to keep the membrane hydrated
and junction unblocked. Clean the bottle and replace the storage solution if
the storage solution gets turbid and mildewed. The electrode should never be
sinked in pure water or buffer solution for long.
2.5 The connector of the electrode should keep clean and dry. If being
contaminated, please clean it with medical cotton and absolute alcohol and
blow dry to prevent the short circuit of the electrode and slow reaction of
electrode.
2.6 After 1-year of use, we recommend replacing the electrode to ensure the
best accuracy.
Warranty
We warrant this electrode free from defects in material and workmanship and
agrees to repair or replace free of charge, at option of APERA INSTRUMENTS any
malfunctioned or damaged product attributable to responsibility of APERA
INSTRUMENTS for a period of six months. Warranty period is the time limit to
provide free service for the products purchased by customers, not the service
life of the tester or electrodes.
This limited warranty does not cover any damages due to:
transportation; storage; improper use; failure to follow the product
instructions or to perform any preventive maintenance; modifications;
combination or use with any products, materials, processes, systems or other
matter not provided or authorized in writing by us; unauthorized repair;
normal wear and tear; or external causes such as accidents, abuse, or other
actions or events beyond our reasonable control.
APERA INSTRUMENTS (Europe) GmbH
Address: Wilhelm-Muthmann-Str.18, 42329
Wuppertal, Germany
Phone: +49 202 51988998
Website: www.aperainst.de
Email: info@aperainst.de