Everspring Z-Wave Temperature and Humidity Sensor EVR_ST814 Manual

June 8, 2024
EVERSPRING

Everspring

Z-Wave Temperature and Humidity Sensor

SKU: EVR_ST814

Quickstart

This is a ****

Measuring Sensor Device
for
Europe
.

To run this device please insert fresh *2 AA** batteries.

Please make sure the internal battery is fully charged.

Inclusion, Exclusion and wakeup are confirmed by three times quickly hitting the Linking button (second from left) on the front of the device.

Important safety information

Please read this manual carefully. Failure to follow the recommendations in this manual may be dangerous or may violate the law.
The manufacturer, importer, distributor and seller shall not be liable for any loss or damage resulting from failure to comply with the instructions in this manual or any other material.
Use this equipment only for its intended purpose. Follow the disposal instructions.

Do not dispose of electronic equipment or batteries in a fire or near open heat sources.

What is Z-Wave?

Z-Wave is the international wireless protocol for communication in the Smart Home. This
device is suited for use in the region mentioned in the Quickstart section.

Z-Wave ensures a reliable communication by reconfirming every message ( two- way
communication
) and every mains powered node can act as a repeater for other nodes
( meshed network ) in case the receiver is not in direct wireless range of the
transmitter.

This device and every other certified Z-Wave device can be used together with any other
certified Z-Wave device regardless of brand and origin
as long as both are suited for the
same frequency range.

If a device supports secure communication it will communicate with other devices
secure as long as this device provides the same or a higher level of security.
Otherwise it will automatically turn into a lower level of security to maintain
backward compatibility.

For more information about Z-Wave technology, devices, white papers etc. please refer
to www.z-wave.info.

Product Description

The ST812 is a combined temperature/humidity sensor which is designed for both – desktop use or wall mountable. The device has a local LCD display which either shows the temperature or the humidity. Four buttons allow to switch between minimim, maximum or actual values, humidity versus temperature and Fahrenheit versus Celsius. The device can report its values via Z-Wave and allows to set trigger level for direct control of other Z-Wave devices, when the measured value hits the trigger point. This makes the ST814 a very powerful and flexible tool to measure environmental values and control device within a Z-Wave network.

Prepare for Installation / Reset

Please read the user manual before installing the product.

In order to include (add) a Z-Wave device to a network it must be in factory default
state.
Please make sure to reset the device into factory default. You can do this by
performing an Exclusion operation as described below in the manual. Every Z-Wave
controller is able to perform this operation however it is recommended to use the primary
controller of the previous network to make sure the very device is excluded properly
from this network.

Reset to factory default

This device also allows to be reset without any involvement of a Z-Wave controller. This
procedure should only be used when the primary controller is inoperable.

Quickly clicking three times on the Linking button on the front of the device will enter inclusion mode. Within 1 second, press the Linking button again and hold it until the long signal tone is off. Now the device is set to factory settings.

Safety Warning for Batteries

The product contains batteries. Please remove the batteries when the device is not used.
Do not mix batteries of different charging level or different brands.

Installation

The Temp./Humid. Detector can either be mounted on a wall or can be freestanding on a table. Please consider a most suitable way before mounting/ placing it.

1. Release the Sensor from the mounting plate and the battery cover by removing the fixing screws.

2. Put the 3 * AA 1.5V batteries in the battery compartment.

3. For wall mounting use mount plate and the screws.

4. Refit the device to the back cover and secure with the fixing screw.

5. For table placing insert the stand into the hole on mounting bracket and turn 90 degrees lockwise.

6. Once snapped in place, the detector can be placed on a shelf, table or other surface where the temperature and humidity measurements are desired.

Inclusion/Exclusion

On factory default the device does not belong to any Z-Wave network. The device needs
to be added to an existing wireless network to communicate with the devices of this network.
This process is called Inclusion.

Devices can also be removed from a network. This process is called Exclusion.
Both processes are initiated by the primary controller of the Z-Wave network. This
controller is turned into exclusion respective inclusion mode. Inclusion and Exclusion is
then performed doing a special manual action right on the device.

Inclusion

Make sure that your Z-Wave Controller is in the Inclusion-/Exclusion-Mode. Quickly click three times on the Linking button on the front of the device to confirm the process.

Exclusion

Make sure that your Z-Wave Controller is in the Inclusion-/Exclusion-Mode. Quickly click three times on the Linking button on the front of the device to confirm the process.

Product Usage

There are 8 modes available for selection. Press the Select Mode button to select desired mode for different settings.

MODE 1 and MODE 2 are showing as main displays on the screen. Once mode setting is finished (MODE 3 to MODE 8), the screen will return to main display automatically after 12 seconds, or by pressing the Linking button to return to main display.

1. Temperature

1.1 Display current Temperature

Press Select Mode button and select MODE 1 to display current temperature, and toggle Select Temperature Unit button to select the unit of temperature (°C /°F).

The temperature ranges from -20°C to 50°C . To show the last record of max/min temperature, press Max. or Min. button. To clear the record, press both Max. and Min. button at the same time.

Threshold Limit Warning:

If the temperature is reaching the limit, the icon of MAX or MIN will be displayed on the screen.

Ice Warning:

If the temperature falls to 0°C , temperature display will illuminate with LCD back light and beep tone will sound continuously for 1 second. Press any key to stop the beep tone.

1.2 Temperature Trigger-ON

Press Select Mode button and select MODE 3 to enter setting of temperature trigger-ON. Icon flashes and the screen shows the recorded trigger-ON temperature. If no value is preset, the screen will display ” – – -°C “.

To adjust trigger-ON value, press and hold Select Mode button for 5 seconds until a long beep is sounded. The “- – -” starts flashing. Use Min. and Max. button to adjust the degree of temperature, and hold Min. or Max. button to scan through the temperature reading from -20°C to 50°C. Once the value is selected, press Select Mode button to confirm setting or press Linking button to cancel.

To clear the trigger-ON record, press both Min. and Max. button at the same time. The record is cleared after a long beep is sounded.

If the temperature reaches the preset trigger-ON value, Temp./Humid. Detector will send a Z-Wave signal. The screen of detector returns to MODE 1 and the icon “ON” is flashing with backlight illuminate and siganl tone sounds for 1 second. Press any key to stop the signal tone.

1.3 Temperature Trigger-OFF

Press Select Mode button and select MODE 4 to enter setting of temperature trigger-OFF. The “OFF” icon flashes and the screen shows the recorded trigger- OFF temperature. If no value is preset, the screen will display “- – -°C”.

To adjust trigger-OFF value, press and hold Select Mode button for 5 seconds until a long beep is sounded. The “- – -” starts flashing. Use Min. and Max. button to adjust the degree of temperature, and hold Min. or Max. button to scan through the temperature reading from -20°C to 50°C. Once the value is selected, press Select Mode button to confirm setting or press Linking button to cancel.

To clear the trigger-OFF record, press both Min. and Max. button at the same time. The record is cleared after a long beep is sounded.

If the temperature reaches the preset trigger-OFF value, Temp./Humid. Detector will send a Z-Wave signal. The screen of detector returns to MODE 1 and the icon is flashing with backlight illumination and siganl tone sounds for 1 second. Press any key to stop the siganl tone.

Note:

The temperature trigger-ON and trigger-OFF cannot be set equal; there MUST be at least 2°C difference in between. For example, if now the trigger-OFF temperature is already set to be 20°C, so trigger-ON temperature can only be lower 18°C or higher 22°C (values between 18°C and 22°C cannot be set).

Once the detector has been triggered, the temperature must increase or cool down at least 2°C from the preset value before it can be triggered again. For example, if the detector is triggered on at 20°C, then the temperature must be higher 22°C or lower 18°C before it can be re-triggered.

2. Humidity

2.1 Display of Current Humidity

Press Select Mode button and select MODE 2 to display current humidity.

The humidity ranges from 20% RH to 90% RH. To show the last record of max/min humidity, press Min. or Max. button. To clear the record, press both Min. and Max. button at the same time.

Threshold Limit Warning:

If the humidity is reaching the limit, the icon of MAX or MIN will be displayed on the screen.

2.2 Humidity Trigger-ON

Press Select Mode button and select MODE 5 to enter setting of humidity trigger-ON. The “ON” icon flashes and the screen shows the recorded trigger-ON humidity. If no value is preset, the screen will display “- – -%”.

To adjust trigger-ON value, press Select Mode button and hold for 5 seconds until a long beep is sounded. The “- – -” starts flashing. Use Min. and Max. button to adjust the percentage of humidity, and hold Min. or Max. button to scan through the humidity reading from 20%RH to 90%RH. Once the value is selected, press Select Mode button to confirm setting or press Linking button to cancel.

To clear the trigger-ON record, press both Min. and Max. button at the same time. The record is cleared after a long beep is sounded.

If the humidity reaches the preset trigger-ON value, Temp./Humid. Detector will send a Z-Wave signal. The screen of detector returns to MODE 2 and the “ON” icon is flashing with backlight illuminate and signal tone sounds for 1 second. Press any key to stop the signal tone.

2.3 Humidity Trigger-OFF

Press Select Mode button and select MODE 6 to enter setting of humidity trigger-OFF. The “OFF” icon flashes and the screen shows the recorded trigger- OFF humidity. If no value is preset, the screen will read “- – -%”.

To adjust trigger-OFF value, press and hold Select Mode button for 5 seconds until a long beep is sounded. The “- – -” starts flashing. Use Min. or Max. button to scan through the humidity reading from 20%RH to 90%RH. Once the value is selected, press Select Mode button to confirm setting or press Linking button to cancel.

To clear the trigger-OFF record, press both Min. and Max. button at the same time. The record is cleared after a long beep is sounded.

If the humidity reaches the preset trigger-OFF value, Temp./Humid. Detector will send a Z-Wave signal. The screen of detector returns to MODE 2 and the “OFF” icon is flashing with backlight illuminates and signal tone sounds for 1 second. Press any key to stop the signal tone.

Note:

The humidity of trigger-ON and trigger-OFF cannot be set equal; there MUST be at least 5% difference in between. For example, if now the trigger-ON humidity is already set to be 50%, so trigger-OFF humidity can only be lower 45% or higher 55%. (Values between 45% and 55% cannot be set.)

Once the detector has been triggered, the humidity must raise up or drop down at least 5% from the preset value before it can be triggered again. For example, if the detector has been triggered on at 50%, then the temperature must be higher 55% or lower 45% before it can be re-triggered.

3. Z-Wave Function

This function is designed to enable or disable the sending Z-Wave commands to the associated nodes, as Temp./Humid. Detector triggered on/off.

Press Select Mode button and select MODE 7, the icon for Radio Frequenzy Status should flash. Press Max. button to turn On (enable) the function or Min. to turn OFF (disable) the function.

3.1 Programming Z-Wave

Z-Wave”s Groups (Association Command Class Version 2)

The Temp./Humid. Detector can be set to send reports to or control associated Z-Wave devices. It supports two association groups with one node support for Grouping 1 and three nodes support for Grouping 2.

Grouping 1 includes POWER_APPLIED, SENSOR_MULTILEVEL _REPORT, ALARM_REPORT and BATTERY_REPORT_COMMAND. Grouping 2 includes BASIC_SET.

Grouping 1

POWER_APPLIED command

Whenever power is applied, it will send ALARM_REPORT command to the nodes of Grouping 1 to inform the devices that the detector is powered up.

MULTILEVEL_SENSOR_REPORT

The detector will emit SENSOR_MULTILEVEL_REPORT to inform the nodes of Grouping 1 automatically its current temperature and humidity. Refer to the section of Z-Wave”s Configuration for settings of auto report configuration.

Low Battery Report

When the battery level of the detector drops to an unacceptable level, the Low Battery icon will appear on the LCD and the detector will emit ALARM_REPORT command to the nodes of Grouping 1.

Grouping 2 (Max. Node = 3)

When the detector is triggered, it will emit BASIC_SET_COMMAND to the nodes of Grouping 2.

Please refer to the table below for the setting of basic set command.

Note:

If the Z-Wave mode is OFF, no command will be sent even the Temp./Humid. Detector has been triggered. If Z-Wave mode is ON and the detector has been triggered, the device will send commands to nodes of Grouping 2.

If the Z-Wave mode reading is 00, it implies no node ID has been allocated by Z-Wave Controller. Please execute inclusion mode as described in the manual.

4. Beep Tone

To set the beep tone, press Select Mode button and select MODE 8. The Beep Tone icon flashes. Press Max. for ON and Min. for OFF. If it is ON, a beep tone will be sounded whenever a button is pressed; 4 continuous beep tones will be sounded for 1 second if the detector has been triggered.

Node Information Frame

The Node Information Frame (NIF) is the business card of a Z-Wave device. It contains
information about the device type and the technical capabilities. The inclusion and
exclusion of the device is confirmed by sending out a Node Information Frame.
Beside this it may be needed for certain network operations to send out a Node
Information Frame. To issue a NIF execute the following action:

Quickly clicking three times on the Linking button on the front of the device will issue a Node Information Frame and keep the device awake to receive configuration commands.

Communication to a Sleeping device (Wakeup)

This device is battery operated and turned into deep sleep state most of the time
to save battery life time. Communication with the device is limited. In order to
communicate with the device, a static controller C is needed in the network.
This controller will maintain a mailbox for the battery operated devices and store
commands that can not be received during deep sleep state. Without such a controller,
communication may become impossible and/or the battery life time is significantly
decreased.

This device will wakeup regularly and announce the wakeup
state by sending out a so called Wakeup Notification. The controller can then
empty the mailbox. Therefore, the device needs to be configured with the desired
wakeup interval and the node ID of the controller. If the device was included by
a static controller this controller will usually perform all necessary
configurations. The wakeup interval is a tradeoff between maximal battery
life time and the desired responses of the device. To wakeup the device please perform
the following action:

Quickly clicking three times on the Linking button on the front of the device will issue a Node Information Frame and keep the device awake to receive configuration commands.

Quick trouble shooting

Here are a few hints for network installation if things dont work as expected.

  1. Make sure a device is in factory reset state before including. In doubt exclude before include.
  2. If inclusion still fails, check if both devices use the same frequency.
  3. Remove all dead devices from associations. Otherwise you will see severe delays.
  4. Never use sleeping battery devices without a central controller.
  5. Dont poll FLIRS devices.
  6. Make sure to have enough mains powered device to benefit from the meshing

Association – one device controls an other device

Z-Wave devices control other Z-Wave devices. The relationship between one device
controlling another device is called association. In order to control a different
device, the controlling device needs to maintain a list of devices that will receive
controlling commands. These lists are called association groups and they are always
related to certain events (e.g. button pressed, sensor triggers, …). In case
the event happens all devices stored in the respective association group will
receive the same wireless command wireless command, typically a ‘Basic Set’ Command.

Association Groups:

Group NumberMaximum NodesDescription

1 1 Battery Status and Sensor Report
2 3 Sensor Triggered

Configuration Parameters

Z-Wave products are supposed to work out of the box after inclusion, however
certain configuration can adapt the function better to user needs or unlock further
enhanced features.

IMPORTANT: Controllers may only allow configuring
signed values. In order to set values in the range 128 … 255 the value sent in
the application shall be the desired value minus 256. For example: To set a
parameter to 200 it may be needed to set a value of 200 minus 256 = minus 56.
In case of a two byte value the same logic applies: Values greater than 32768 may
needed to be given as negative values too.

Parameter 1: Basic Level Set

Set basic set value to be on or off
Size: 1 Byte, Default Value: 99

SettingDescription

0 Disable
1 – 99 Use value

Parameter 2: Temperature Trigger ON value

Temperature level when a ON command is sent out
Size: 1 Byte, Default Value: 99

SettingDescription

0 – 50 Trigger On on
99 Disable
236 – 255 Trigger On on

Parameter 3: Temperature Trigger OFF value

Trigger Temperature level when a OFF command is sent out
Size: 1 Byte, Default Value: 99

SettingDescription

0 – 50 Trigger OFF on
99 Disable
236 – 255 Trigger OFF on

Parameter 4: Humidity Trigger ON value

Humidity level when a ON command is sent out
Size: 1 Byte, Default Value: 99

SettingDescription

20 – 90 Trigger On on
99 Disable

Parameter 5: Humidity Trigger OFF value

Humidity level when a OFF command is sent out
Size: 1 Byte, Default Value: 99

SettingDescription

20 – 46 Tigger OFF on
99 Disable

Parameter 6: Auto Report Time

Sets the time interval when a sensor report is sent
Size: 2 Byte, Default Value: 0

SettingDescription

0 Disable
1 – 1439 Auto report periodically

Parameter 7: Auto Report Temperature

Sets the temperature change causing a sensor report
Size: 1 Byte, Default Value: 0

SettingDescription

0 Disabled
1 – 70 Auto report on temp. change

Parameter 8: Auto Report Humidity

Sets the humidity change causing a sensor report
Size: 1 Byte, Default Value: 0

SettingDescription

0 Disable
5 – 70 Auto report on humidity change

Technical Data

Dimensions 0.0850000×0.0850000×0.0390000 mm
Weight 125.62 gr
Hardware Platform ZM3102
EAN 0696859123733
Battery Type 2 * AA
Device Type Routing Multilevel Sensor
Generic Device Class Multilevel Sensor
Specific Device Class Routing Multilevel Sensor
Firmware Version 01.08
Z-Wave Version 00.00
Certification ID ZC08-13100006
Z-Wave Product Id 0060.0006.0001
Frequency Europe – 868,4 Mhz
Maximum transmission power 5 mW

Supported Command Classes

  • Basic
  • Multi Channel
  • Wake Up
  • Association
  • Version
  • Battery
  • Configuration
  • Sensor Multilevel
  • Manufacturer Specific

Explanation of Z-Wave specific terms

  • Controller — is a Z-Wave device with capabilities to manage the network.
    Controllers are typically Gateways,Remote Controls or battery operated wall controllers.

  • Slave — is a Z-Wave device without capabilities to manage the network.
    Slaves can be sensors, actuators and even remote controls.

  • Primary Controller — is the central organizer of the network. It must be
    a controller. There can be only one primary controller in a Z-Wave network.

  • Inclusion — is the process of adding new Z-Wave devices into a network.

  • Exclusion — is the process of removing Z-Wave devices from the network.

  • Association — is a control relationship between a controlling device and
    a controlled device.

  • Wakeup Notification — is a special wireless message issued by a Z-Wave
    device to announces that is able to communicate.

  • Node Information Frame — is a special wireless message issued by a
    Z-Wave device to announce its capabilities and functions.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals