Unit and LCD Display

MEDISANA-51430-Blood-Pressure-Measuring-Device-HGH-
Instruction-fig-4

IMPORTANT

  • Non-observance of these instructions can result in serious injury or damage to the device.

WARNING

  • These warning notes must be observed to prevent any injury to the user.

CAUTION

  • These notes must be observed to prevent any damage to the device.

NOTE

  • These notes give you useful additional information on the installation or operation.
  • Device classification: type BF
  • Lot number
  • Manufacture
  • Date of manufacture

Safety instructions

Read the instruction manual carefully before using this device, especially the safety instructions, and keep the instruction manual for future use. Should you give this device to another person, it is vital that you also pass on these instructions for use.

  • The unit is intended only for use in the home. Consult your doctor prior to using the blood pressure monitor if you are concerned about health matters.
  • Only use the unit for its intended purpose in accordance with the instruction manual. Warranty claims become void if the unit is misused.
  • Please consult your doctor prior to using the unit if you suffer from illnesses such as arterial occlusive disease.
  • The unit may not be used to check the heart rate of a pacemaker.
  • Pregnant women should take the necessary precautions and pay attention to their individual circumstances. Please consult your doctor if necessary.
  • If you experience discomfort during a measurement, such as pain in the upper arm or other complaints, press the START button  to release the air immediately from the cuff. Loosen the cuff and remove it from your arm.
  • Do not take any therapeutic measures on the basis of a self measurement. Never alter the dose of a medicine prescribed by a doctor.
  • The unit is unsuitable for children.
  • Do not allow children to use the unit. Medical products are not toys!
  • The unit may not be operated in the vicinity of appliances that emit strong electric radiation such as radio transmitters, mobile phones or microwaves.
  • Please do not attempt to repair the unit yourself in the event of malfunctions, since this will void all warranty claims. Only have repairs carried out by authorised service centres.
  • Protect the unit against moisture. If liquid does manage to penetrate into the unit, remove the batteries immediately and do not continue to use the unit. In this case contact your specialist dealer or notify us directly.
  • If the unit is not going to be used for a long period, please remove the batteries.

SAFETY NOTES FOR BATTERIES

Keep out of reach of children! Not rechargeable!

  • Do not throw into open fires!
  • Do not short-circuit! There is a danger of explosion!
  • Do not throw spent batteries into the domestic waste.
  • Dispose of them as special waste or put them in a battery collection container at a specialist dealer’s!

Unit and LCD display

  1. Battery Compartment
  2. Blood Pressure Indicator
  3. (green – yellow – orange – red)
  4. LCD screen (display)
  5. M button (Memory Recall)
  6. START button
  7. Wrist cuff
  8. Save symbol Display of Date/Time
  9. Display of Systolic Pressure
  10. Display of Diastolic
  11. Pressure Deflate to Measure symbol
  12. Arrhythmia display
  13. Blood Pressure Indicator (Bar display)
  14. Change Battery symbol

Scope of supply

Please check first of all that the device is complete. The following parts are included:

  • 1 MEDISANA blood pressure measuring device HGH
  • 2 batteries (type AAA, LR03) 1.5V
  • 1 storage box
  • 1 Instruction manual

Please contact your dealer if you notice any transport damage on unpacking the unit.

Starting up

Insert / replace batteries

You must insert the batteries provided before you can use your unit. The lid of the battery compartment is located on the left hand side of the unit. Open it, remove it and insert the 2 AAA LR03 type 1.5 V batteries supplied. Ensure correct polarity when inserting the batteries (as marked inside the battery compartment). Close the battery compartment. Replace the batteries if the change battery symbol  appears on the display  or if nothing appears on the display after the unit has been switched on.

Setting the date and time

  1. With the unit switched off, simultaneously press the START button  and the M button  to set the date and time. First of all the entry point for the month flashes (1, 2, 3… 12).
  2. Press the M button repeatedly until the desired number for the month appears.
  3. Press the START button  again to accept the setting. Now the entry point for the date flashes.
  4. Proceed in exactly the same manner as described in point 2 when entering the date, hours and minutes.
  5. In each case, press the M button  to select and the START button  to accept the entry.
  6. The date and time are displayed continuously. If the batteries are changed the information is lost and the date and time must be set again.

Fitting the cuff

  • Place the cuff on the left wrist with the palm of the hand facing upwards. (fig.1).
  • The distance between the cuff and the palm of the hand must be approx. 1 cm. (fig.2).
  • Securely fasten the monitor around your wrist using the Velcro fastener. (fig.3).

The correct position

  • Carry out the measurement whilst seated.
  • Relax your arm and place it comfortably, on a table for instance.
  • Raise the wrist so that the cuff of the monitor is at heart level (fig.4, position a = too high, b = correct, c = too low).
  • Keep still during the measurement: do not move and do not talk as this may alter the measurements.

Taking a blood pressure measurement

  1. Switch the unit on by pressing the START button 5.
  2. If the START button 5 is pressed, a long beep is heard and all display characters are shown (display test). This test can be used to check that the display is indicating properly and in full. The display shows the values from the previous measurement. If no measurement has been taken, zeroes will appear next to the date and time.
  3. The unit is ready for measurement, the number 0 is displayed. It automatically inflates the cuff slowly in order to measure your blood pressure. The pressure on the display changes continuously during the time it continues to inflate.
  4. The unit inflates the cuff until sufficient pressure has built up for a measurement. Then the unit slowly releases air from the cuff and carries out the measurement.
  5. When the measurement is finished, the cuff is deflated completely. The systolic and the diastolic blood pressure, as well as the pulse rate, appear in the display 3 in succession. The blood pressure indicator 2 flashes next to the relevant coloured bar depending on the WHO blood pressure classification. If the unit has de-tected an irregular heartbeat, the arrhythmia display 12 also flashes.
  6. The measured values are stored automatically. Up to 60 measured values with date and time can be stored in the memory.

WARNING

Do not take any therapeutic measures on the basis of a self measurement. Never alter the dose of a medicine prescribed by a doctor.

The unit switches off automatically after a minute if no buttons are pressed. The unit may also be switched off using the START button 5.

Displays of stored values
You can store up to 60 measurements (blood pressure and pulse with date and time) in the memory of your blood pressure monitor. The measured values are stored automatically Press the M button 4 to call up the measured values stored. The stored data appears on the display. First the number of occupied memory locations is displayed. Then the values from the previously stored measurement are displayed, together with the flashing blood pressure indicator. Repeatedly pressing the M button 4 displays the respective values measured previously. If you have reached the final entry and do not press a button, the unit automatically switches itself off after approximately 10 seconds in memory recall mode. Memory recall mode can be exited at any time by pressing the START button 5 , which will also switch the unit off. The oldest value is deleted if 60 measured values have been stored in the memory and a new value is saved.

Deleting the memory
Press and hold the M button 4 for approximately 3 seconds and all measured data stored will be deleted automatically after three beeps.

Blood pressure classification according to the WHO

These values have been specified by the World Health Organisation (WHO) without making any allowance for age.

Low blood pressure

  • systolic <100 diastolisch <60

Normal blood pressure

  • (green display area 2 ) systolic 100 – 139 diastolic 60 – 89

F orms of high blood pressure / hypertension

  • Mild hypertension (yellow display area 2) systolic 140 – 159 diastolic 90 – 99

Moderate hypertension (orange display area 2 )

  • systolic 160 – 179 diastolic 100 – 109

Severe hypertension (red display area 2 )

  • systolic >180 diastolic >110

WARNING

Blood pressure that is too low represents just as great a health risk as blood pressure that is too high! Fits of dizziness may lead to dangerous situations arising (e.g. on stairs or in traffic)

Error messages and error remedying

  • : Replace both batteries.
  • Er 0 (Pressure system instable at start of the measurement)
  • Er 1 (Error in measuring the systolic pressure)
  • Er 2 (Error in measuring the diastolic pressure)
  • Do not move.
  • Er 3 (Pressure system is blocked or sleeve is too tight during inflation)
  • Er 4 (Pressure system has a leak or sleeve is too loose during inflation)
  • Put the sleeve on correctly.
  • Er 5 (Sleeve pressure above 300 mmHg)
  • Er 6 (Longer than 3 min. with sleeve pressure over 15 mmHg)
  • Er 7 (EEPROM access error)
  • Er 8 (Error when testing the functions on the device)
  • Er A (Error in recording the pressure)

Perform measurement again after 5 min. If you are unable to rectify the error, please consult the service centre.

Cleaning and maintenance

Remove the batteries before cleaning the unit. Clean the unit using a soft cloth lightly moistened with a mild soapy solution. Never use abrasive cleaning agents or alcohol. Do not expose the unit to direct sunlight; protect it against dirt and moisture.

Disposal

This product must not be disposed of together with domestic waste. All users are obliged to hand in all electrical or electronic devices, regardless of whether or not they contain toxic substances, at a municipal or commercial collection point so that they can be disposed of in an environmentally acceptable manner. Please remove the batteries before disposing of the device / unit. Do not dispose of old batteries with your household waste, but at a battery collection station at a recycling site or in a shop. Consult your municipal authority or your dealer for information about disposal.

Guidelines / Standards

This blood pressure monitor meets the requirements of the EU standard for non- invasive blood pressure monitors. It is certified in accordance with EC Guidelines and carries the CE symbol (conformity symbol) “CE 0297”. The blood pressure monitor corresponds to European standards EN 60601-1, EN 60601-1-2, EN 80601-2-30, EN 14971, EN 15223-1. The specifications of EU Guideline “93/42/EEC of the Council Directive dated 14 June 1993 concerning medical devices” are met.

Electromagnetic compatibility:
The device complies with the EN 60601-1-2 standard for electromagnetic compatibility. Inquire at MEDISANA for details on this measurement data.

Technical specifications

  • Name and model : MEDISANA Blood pressure
  • measuring device HGH
  • Display system : Digital display
  • Memory slots : 60 with date and time
  • Measuring method : Oscillometric
  • Voltage supply : 3V=, 2×1.5V battery AAA LR03
  • Measuring range blood pressure : 0 – 300 mmHg
  • Measuring range – pulse : 40 – 180 beats/min.
  • Maximum measuring deviation of static pressure : ± 3 mmHg
  • Maximum measuring deviation of pulse values : ± 5 % of the value
  • Pressure generation : Automatic with pump
  • Air release : Automatic
  • Autom. switch-off : After 1 min.
  • Operating conditions : +5 °C to +40 °C, <80 % max. relative humidity
  • Storage conditions : -20 °C to +55 °C, <90 % max. relative humidity
  • Dimensions : 83 x 64 x 28 mm
  • Cuff : 300 x 70 mm, circumference 14 – 19,5 cm for adults
  • Weight (approx.) : 110 g without batteries
  • Item number : 51430
  • EAN number : 40 15588 51430 2

In accordance with our policy of continual product improvement, we reserve the right to make technical and optical changes without notice.

Warranty and repair terms

Please contact your dealer or the service centre in case of a claim under the warranty. If you have to send in the unit, please enclose a copy of your receipt and state what the defect is. The following warranty terms apply:

  1. The warranty period for MEDISANA products is three years from date of purchase. In case of a warranty claim, the date of purchase has to be proven by means of the sales receipt or invoice.
  2. Defects in material or workmanship will be removed free of charge within the warranty period.
  3. Repairs under warranty do not extend the warranty period either for the unit or for the replacement parts.
  4. The following is excluded under the warranty:
  5. All damage which has arisen due to improper treatment,
  6. g. nonobservance of the user instructions.
  7. All damage which is due to repairs or tampering by the customer or unauthorised third parties.
  8. Damage which has arisen during transport from the manufacturer to the consumer or during transport to the service centre.
  9. Accessories which are subject to normal wear and tear.
  10. Liability for direct or indirect consequential losses caused by the unit are excluded even if the damage to the unit is accepted as a warranty claim.