Thermo WX 80 Ultraspeed Refrigerated Centrifuge Access Control System User Manual
- June 4, 2024
- Thermo
Table of Contents
Thermo WX 80 Ultraspeed Refrigerated Centrifuge Access Control System
Summary
This access control is used to restrict the access to a certain number of
users of the centrifuge. The users are required to log on to the centrifuge by
inputting the ID code and the password on the initial screen to operate the
centrifuge.
Before using the access control, carefully read through both: this instruction
manual and the instruction manual of the centrifuge. This system distinguishes
between the system administrator and the registered user. Up to 3 system
administrators and up to 37 users can be registered.
The system administrator is permitted to:
- Register, change, and delete the registered users
- Refer to the ID codes and passwords of the system administrators and the registered users on the user list screen, and
- Utilize the printout function by means of the optional printer.
The registered users are:
- Permitted to log on to the centrifuge by inputting an ID code and a password in the specified boxes on the initial screen to operate the centrifuge, but
- Not permitted to refer to the user list screen nor to utilize the printout function.
Both: the system administrators and the registered users can turn off the centrifuge power or log off the centrifuge while the vacuum pump is turned off or the rotor is stopping after operation. The display turns to the initial screen of the access control when the centrifuge is logged off.
How to operate (for system administrator) ?
CAUTION: Keep your password secret from others.
Using the access control for the first time
The system administrator should log on to the access control according to the following procedure. When installing the optionally available access control to your centrifuge, the user lists stored in your centrifuge must be deleted. In such cases, delete the user lists according to “2.4 Initialization (deletion) of user lists” then perform the following operation.
| No.| Operation| Key strokes and screen display
---|---|---|---
| 1| Turn on the main power switch.
Then the access control starts and the initial screen is displayed.|
| 2| Input ”1111” as an ID code.| 1 1 1 1 ENTER
| 3| Input ”1111” as a password.
Then the RUN screen is displayed as shown in the right figure.|
4| Open the MENU screen and select “User List”. “AAA” is temporarily
registered as a user name.
=NOTE=
Numbers 1 through 3 are assigned to the system administrators, and numbers 4
and later are assigned to the registered users.|
|
5| Put the cursor on the No. 1 and press the ENTER key.| |
6| Change the user name, ID code and password to those of the system
administrator to be registered according to the prompts displayed on the
screen. If necessary, register another system administrator or user.
Refer to the instruction manual of the centrifuge “2-8-2 User List” for the
registration procedure.
The right figure shows an example.
=NOTE=
Do not use the same ID code for registration of different users.
| |
| | | | | |
7| Press the ESC key. The display returns to the RUN screen.|
---|---|---
8| Put the cursor on the “ACCESS CONTROL” in the FUNCTION box and press the
ENTER key. Then the registered name is displayed on the USER NAME box.|
9| Log off the access control.
Turn off the centrifuge power or put the cursor on the “LOG OFF” and press the ENTER key.
|
Registration of new users by the system administrator (including change
or deletion of registered users)
Log on to the access control with the system administrator’s own ID code and password. Perform registration, change or deletion of the users on the User List screen according to the procedure specified in section 2.1.
Using the printout function
The printout function is available by means of the optional printer in a like manner as the standard centrifuge. However, use of the printout function is limited to the system administrator by the access control. Refer to the instruction manual of the centrifuge “2.9 Print Utilities” for details. Connect the printer to the centrifuge referring to the instruction manual of the printer.
Initialization (deletion) of user lists
CAUTION: Once the user lists are deleted, they cannot be recovered.
No. | Operation | Key strokes and screen display |
---|---|---|
1 | Turn on the main power switch. Press the keys as follows when the |
initial screen of the access control is displayed.
MENU 1 2 3 4
HOLD
The screen shown in the right figure is then displayed.
|
2| Select “1: Yes”. Then the screen shown in the right figure is displayed.
If you do not want to delete, select “2: No”.|
3| By selecting “1: Yes”, the screen shown in the right figure is displayed and the following data is deleted.
- Machine log
- Rotor log
- User log
- Machine management
- Print program
- Centrifuge scheduler
- Run result
If you do not want to delete, select “2: No”.
|
4| After initialization, following data is recorded at the No. 1 of the user
list.
ID code: 1111
Password: 1111 User name: AAA
Refer to “2-1. When using the access control for the first time” for the next
time use.|
How to operate (for registered user) ?
CAUTION: Keep your password secret from others.
Using the centrifuge
No. | Operation | Key strokes and screen display |
---|---|---|
1 | Input the registered user’s own ID code and password. |
(Example)
ID code: 9999
Password: 9999|
2| The screen shown in the right figure is then displayed.
The centrifuge becomes operable when this screen is displayed.
|
3| Put the cursor on the “ACCESS CONTROL” in the FUNCTION box and press the
ENTER key. Then the name of the user who is logging, on is displayed.
Without doing this operation, the centrifuge is operable when the screen shown in the above step 2 is displayed.
|
Programming the operation
The centrifuge equipped with the access control system differs from the standard centrifuge in the following points.
- The user name used for logging on is automatically registered.
- The run conditions stored (registered) in the centrifuge cannot be changed or deleted unless the logged on user is the one who stored the run conditions.
- You can refer to the stored (registered) run conditions even if the user name deviates. In addition, you can operate the centrifuge using the run conditions stored by other users.
CHINA
Tel.: +86 21 68 65 45 88
+86 10 58 50 35 88
E-mail Technical Service Representative for SORVALL® brand products at
techsupport.led.asheville@thermo.com
JAPAN
Tel.: +81 454 53 92 20
INDIA
Tel.: +91 22 55 42 94 94
OTHER ASIA PACIFIC COUNTRIES
Tel.: +852 28 85 46 13
CANADA or LATIN AMERICA
Tel.: +1 800 522 77 46
+1 866 984 37 66
In Other Countries: Contact your local SORVALL® products distributor or agent.
AUSTRIA
Tel.: +43 (1) 80 14 00
FRANCE
Tel.: +33 1 69 18 77 77
+33 2 28 03 20 00
NETHERLANDS
Tel.: +31 7 65 71 44 40
UNITED KINGDOM or IRELAND
Tel.: +44 87 06 09 92 03
BELGIUM
Tel.: +32 24 82 30 30
GERMANY
Tel.: +49 61 84 90 69 40
+49 61 03 40 81 012
SPAIN or PORTUGAL
Tel.: +34 9 32 23 31 54
Other EUROPE,
MIDDLE EAST, or AFRICA
Tel.: +49 61 84 90 69 40
+33 2 28 03 20 00
FINLAND
Tel.: +35 89 32 91 00
ITALY
Tel.: +39 02 95 05 91
SWITZERLAND
Tel.: +41 (1) 4 54 12 12
UNITED STATES of AMERICA
Tel.: +1 800 522 77 46
+1 866 984 37 66
Visit our web site at http://www.sorvall.com or http://www.thermo.com
RUSSIA/CIS
Tel.: +7 09 52 25 11 15
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>