Impex Produkter NANOCOLOR Ammonium 50 Instruction Manual
- June 3, 2024
- Impex Produkter
Table of Contents
Impex Produkter NANOCOLOR Ammonium 50
Overview
The test is suitable for the photometric determination of NH4+. The test is in
accordance with APHA 4500-NH3 F, ISO 7150-1, DIN 38406-E5 and EPA 350.1.
The test is suitable for surface water, groundwater and drinking water and
slightly contaminated wastewater.
Results are highly reproducible in water with low levels of pollutants.
- Measuring range:
- 1.0–40.0 mg/L NH4-N (method 0051)
- 1.0–50.0 mg/L NH4+ / NH3 (method 0052 / 0053)
- Number of tests: 20
- Wavelength for photometric determination: 660 / 690 nm
- Shelf life: 12 months
- Reaction time: 15 minutes
- Storage temperature: 15–25 °C
- Storage conditions: upright
Method
Photometric determination of a blue indophenol dye based on a reaction with hypochlorite and salicylate at a pH of 12.6 in the presence of sodium nitroprusside.
Interferences
Heavy pollution results in errors and requires prior distillation.
The method can be applied for analyzing seawater.
Turbidities cause higher measurement values.
Reagents and accessories
Contents of reagents set:
- 20 test tubes R0
- 1 NANOFIX R2
Required devices:
- MACHEREY-NAGEL photometer
- Digital piston pipette 100–1000 µL (REF 91677) with pipette tips (REF 91667)
- Digital piston pipette 50–200 µL (REF 916914) with pipette tips (REF 916915)
- Tweezers for sampling NANOFIX capsules (REF 916114)
Standards
- NANOCONTROL Multistandard Sewage inflow (REF 925012)
Sampling and preparation
See DIN EN ISO 5667-3-A 21.
Adjust to pH 1–10 prior to analysis.
Quality control
The measurement of a blank value and a standard is recommended before every
measuring series as quality control measure.
www.mn-net.com
Quality data:
The following data were determined during production according to ISO 8466-1
and DIN 38402-A51:
- Number of LOTs: 60
- Standard deviation of the method: Validade: 0,36
- Coefficient of variation of the process: ± 1.87 %
- Confidence interval: ± 0.86 mg/L NH4-N
Specified data for procedure:
- Sensitivity (absorbance of 0.010 A corresponds to): 0.26 mg/L NH4-N
- Accuracy of a measurement value: ± 0.91 mg/L NH4-N LOT-specific certificates are available at www.mn-net.com.
Procedure
- Open test tube
- Pipette 0.2 mL of sample into test tube
- Add 1 NANOFIX R2
- Seal test tube and shake vigorously
- Wait 15 min
- Clean outside of test tube
- Measure
Notes
When using other photometers, make sure measurements are possible in test
tubes (16 mm OD) and calibrate the method.
Correction value e. g. for colored or turbid samples possible (see photometer
manual).
Information regarding safety can be found on the box’ label and in the safety
data sheet. You can download the SDS from www.mn-net.com/SDS.
08/2020
DE / International:
Tel.: +49 24 21 969-0
Fax: +49 24 21 969-199
E-mail: info@mn-net.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>