CAPP CRPI-412X Four Micro Plate Thermo Shaker User Manual
- June 3, 2024
- CAPP
Table of Contents
- BRIEF OVERVIEW
- INTENDED USE
- FEATURES
- TECHNICAL SPECIFICATIONS
- ACCESSORIES
- INSTALLATION
- INSTALLING MICRO-PLATES
- USER INTERFACE AND DISPLAY
- TROUBLESHOOTING
- TIPS FOR EFFICIENT OPERATION
- SAFETY PRECAUTION
- WARRANTY STATEMENT
- PRODUCT DISPOSAL
- Customer Support
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
CAPP CRPI-412X Four Micro Plate Thermo Shaker User Manual
BRIEF OVERVIEW
This four micro plate Thermos Shaker comes with long life maintenance free BLDC motor controlled by microprocessor, LED digital display and “PULSE” mode feature with an orbital diameter of 3 mm for proper mixing of samples and uniform temperature a cross the plate.
INTENDED USE
Thermos shaker is used for mixing biological samples and chemicals in micro-
plates. It is applicable in areas of Biochemistry, Biotechnology, Immunology,
Animal Biotechnology, Molecular Immunology, Enzymology, Protein Engineering,
Molecular Biology, Microbiology etc. laboratories.
NOTE: Before using the instrument, please read this user manual
carefully. This user manual is intended to assist with the operation and care
of the unit only and not its repair For repair please contact the supplier.
FEATURES
- Uniform heating and shaking across the plates
- Microprocessor controlled device and digital display
- Adjustable speed from 200 to 1200 RPM at every 10 RPM
- Adjustable time from 1min -99h 59min & infinite mode
- 3 mm orbital motion for thorough mixing in micro-plates
- Easy loading and unloading of micro-plates
- Brushless DC motor for quite operation and maintenance free life
- Spill safe operation even at higher speeds
- Pulse mode feature for clockwise and counter clockwise motion
- Temperature control from ambient to 70°C
TECHNICAL SPECIFICATIONS
Motor Type | BLOC |
---|---|
Type of Movement | Orbital |
Orbital Diameter | 3mm |
Plate Capacity | 4 Microplate |
Speed | 200 to 1200 RPM at every 10 RPM |
Timer | 1 to 99 h 59 min. & infinite |
Temperature Range | Ambient +3 to 70°C |
Temperature Stability | 10.5°C |
Temperature Accuracy | 10.5°C |
Pulse Mode | Yes (clockwise & counter clockwise) |
Display (Speed & Timer) | LED Digital |
TECHNICAL SPECIFICATIONS
Dimensions (W x D x H) mm | 267 x 361 x 155 |
---|---|
Permissible Ambient Temp. | 4 to 40°C |
Permissible Relative Humidity | 80% |
Weight | 8.45 Kg |
Operating Voltage | 230V, 50Hz |
Power Consumption | 264 W |
ACCESSORIES
- Replaceable 0-ring
- Product manual & warranty card
- Power Cable
INSTALLATION
Gently take the unit out of the box and place it on a flat, stable and non slippery surface. Keep 2-inches clearance from all 4 sides of the shaker so that it can be operated easily and safely. Do not lift the instrument by holding the platform. Check whether the rubber suction cup is attached to the table properly. Plug in the power adapter pin before connecting to the power source. The instruction manual and accessories should be kept near the device in a safe storage location for at least 2 years for warranty purposes.
INSTALLING MICRO-PLATES
The plates can be fixed on any side of the central pre-set 0-ring. Gently push
the 0- ring towards the center of the plate sets locked on the wall of the
platform. Ensure that the plate sits flat on the platform.
USER INTERFACE AND DISPLAY
CONTROL PANEL OPERATION
- SET and ACTUAL led will FLASH as per status of command.
- “+/-” button will increase and decrease the value of each parameter.
- Each LED light will flash as per status of parameter.
- “Start/Stop” button for operation start and stop.
- Speed, Time and Temp buttons to select it respective values.
- “Pulse” button to start pulse mode.
START / STOP BUTTON
To start the operation, press the “start” button after setting all the parameters(speed, timer & temperature) which will be indicated by LED. And to stop the operation at any time press “stop” button.
SPEED SETTING
To set speed in RPM, press “speed” button, which will blink on screen. It has
last run memory ie last set parameter will be on display upon turning ON [For
the first time “eeee” will appear – default factory setting], to set RPM
value, press “+ or -” key for increment or decrement of the values between 200
to 1200 RPM. The incremental and decremental values will be in 10 RPM upon
single press, for rapid increment & decrement of RPM keep “+/-” button long
pressed. The set value will be saved automatically when blinking of the value
stops.
At any point during the operation if speed needs to be changed then follow the
above instruction to set new desired values without turning off the
instrument.
For heating only (no shaking operation) set the speed values to “9999”
following above instruction &set the heating parameter.
TEMPERATURE SETTING
To set temperature press “TEMP” Key, the Actual and Set Led will flash alternatively, then select desired function, to set the temperature press the “+/-” key from ambient to 70 °C, and set the required value, the set value will be saved automatically when the blinking stops. For rapid increment and decrements long press the “+/-” key. The
USER INTERFACE AND DISPLAY
Value can be set at an increment or decrement of0.1 °C/°F. Ambient temperature
can be seen on the Home screen, it is automatically detected from the internal
unit sensor and will be displayed by default.
By pressing Temp, the set value blinks and when “+/-” key pressed “ambient
value” displayed this can be used for shaking when no heating is required.
Temperature can not be set below ambient temperature
At any point during the operation the temperature values can be changed
without turning off the instrument by following the above instruction.
Actual temperature will be seen throughout the heating operation.
NOTE: Temperature cannot be set below Ambient temperature.
Which is identified by internal temperature sensor
TIMER SETTING
To set time, press “Time” button, which will blink on screen. The time can be
always set during the blinking of the screen. It has last run memory i.e. last
set parameter will be on display upon turning ON [For the first time “JE” will
appear – default factory setting], to set fame value, press “+/-” key for
increment or decrement of the values between – 1min to 99hr 59min & JL. For
setting in hour or minute press “time” button until the desired segment in
screen blinks (Left values in time screen are for hours & right values for
minutes). The incremental and decremental values will be in 1min for minute &
1 hour for hour selection upon single press of“+/-” button. The set value will
be saved automatically when blinking of the value stops.
At any point during the operation if wants to set different AMER then follow
the above instruction to set new desired values without turning off the
instrument.
PULSE MODE
To activate pulse mode, press “Pulse” button. The activation of pulse mode
will be indicated by LED light. The pulse mode time can be set from 30-99
seconds (30 sec is by default factory setting). To program the pulse timer,
press “Pulse” button for longer To change the time from 30 sec – 99 sec by
pressing “+/-” for increment & decrement respectively. The set value will be
saved automatically when blinking of the value stops. The motion in pulse mode
would be CW/CCW. The pulse timer will set back itself to 30 sec by default
upon restart of the timer.
TEMPERATURE CALIBRATION MODE
Calibration of temperature is to reset the temperature accuracy. To enter into calibration mode press “Temp” button longer until “Temp” mode is activated. Enter the new temperature value from the reference unit. The new entered value will be saved automatically replacing older value. This is a single point calibration.
SPEED CALIBRATION MODE
By long pressing Speed button we can enter RPM calibration mode, any single
value of RPM can be calibrated. To enter into calibration mode long press
“Speed Time and Pulse” simultaneously at set rpm 1190 in running mode. (We can
change in + 50 rpm).
This is a single point calibration.
TROUBLESHOOTING
Error Code | Cause | Effect | Solution |
---|
E 01
| If safety circuit is fail, temp. will go beyond safe temp. 95 C| Heating off| Switch off device
- Warranty only to be carried out by authorized service personnel
E 02| Temperature sensor of shaking plate is fail| Over Heating
E 03| Temperature sensor of top lid is fail| Over Heating
E 04
| Heater connection is not proper/Heater fail (SHAKING PLATE)| —
E 05
| Heater connection is not proper/Heater fail (TOP HEATING PLATE)| —
TIPS FOR EFFICIENT OPERATION
Orbital shaking is a simple way for the mixing of assay components, but simply
putting the samples in a micro-plate does not guarantee reaching complete
blending after the mixing process.
The adjustment of the optimal mixing speed for micro-plates is dependent on
the size of the well and the filling volume. The below table is a short
summary for recommended mixing speeds for different micro-plates.
Mixing Speeds (RPM) for Different Micro-plates Depending on Filling Volume/Well (%)
Filling Volume in %| 96 Standard Well| 96 Deep Well| 384 Standard Well
50 %| 1000 – 1200 RPM| 800 – 1200 RPM| 1200 RPM
75 %| 800 – 1000 RPM| 600 – 1000 RPM| 900 – 1100 RPM
Larger volumes should be mixed with a higher orbit and low speed. Smaller
volumes requires a lower orbit, but a much higher speed.
NOTE: Overfilling of wells can lead to spillage.
SAFETY PRECAUTION
- Do not lift the shaker by holding the micro-plate or shaking platform.
- Unplug the power cord before cleaning.
- The whole unit should be cleaned regularly with a moist cloth only.
- Always operate the unit on a leveled and stable surface to avoid any kind of accident.
- Always keep the shaker clean. If any liquid spill occurs, it should be removed promptly. Do not immerse the unit in liquid for cleaning.
- The shaker should be used for specific applications only.
- Do not use shaker in the hazardous environment or with hazardous material for which the unit is not designed.
- Do not operate the unit if it shows signs of electrical or mechanical damage. If any such damage is found, contact the supplier.
- Care should be taken on the basis of sample used for mixing.
WARRANTY STATEMENT
This product is warranted to be free from defects in material and workmanship for a period of two (2) years from date of purchase. Your product will be duly repaired upon prompt notification in compliance with the following conditions:
This warranty is valid only if the product is used for its intended purpose
and within the guidelines specified in this instruction manual.
This warranty does not cover damage caused by accident, neglect, misuse,
improper service, natural forces or other causes not arising from defects in
original material or workmanship.
This warranty does not cover any incidental or consequential damages,
commercial loss or any other damages from the use of this product.
The warranty is invalidated by any non-factory modification, which will immediately terminate all liabilities on us for the products or damages caused by its use. The buyer and its customer shall be responsible for the product or use of products as well as any supervision required for Safety. If requested the products must be returned to the distributor in well packed and insured manner and all shipping charges must be paid.
Some states do not allow limitation on the length of implied warranties or the exclusion or limitation of incidental or consequential damages. This warranty gives you specific legal rights. This warranty is given expressly in lieu of all other warranties, expressed or implied.
Products received without proper authorization will not be entertained. All
items returned for service should be sent postage prepaid in the original
packaging or other suitable carton, padded to avoid damage. We will not be
responsible for damage incurred by improper packaging.
This warranty is valid only if the warranty is registered with the supplier
within 30 days from the date of purchase.
For you reference, make a Note of serial number, date of purchase and supplier here.
Serial No: Purchase:
Supplier:
PRODUCT DISPOSAL
In case the product is to be disposed off, the relevant legal regulations are
to be observed.
The disposal of electrical devices is regulated within the European Community
by national regulations based on EU Directive 2012/19/EU on waste electrical
and electronic equipment(WEE E).
According to these regulations, any devices supplied after 13.06.05 in the
business to business sphere, to which this product is assigned, may no longer
be disposed off in municipal or domestic waste. They are marked with the
following symbol to indicate this.
As disposal regulations within the EU may vary from country to country, please
contact your supplier if necessary.
Customer Support
AHN Biotechnologie GmbH
Uthleber Weg 14
D-99734 Nordhausen
Germany
Phone: +49(0)3631/65242-0
Fax: +49(0)3631/65242-90
E-Mail: info@cappahn.com
www.capp.dk