Incubation
- Prepare the Incubation mix in a microcentrifuge tube according to the table below.
Incubation mix | per ½ 96-well plate (μL) |
---|---|
Olink® Target 48 Incubation Solution | 168 |
Olink® Target 48 Frw-probes | 21 |
Olink® Target 48 Rev-probes | 21 |
Total | 210 |
- Vortex and spin down the Incubation mix. Transfer 23 µL of the Incubation mix to each well of a new 8-well strip.
- Transfer 3 μL of Incubation mix to each well of the first 6 columns of a 96-well plate by reverse pipetting and name the plate Incubation Plate.
- Add 1 µL of each sample using a multi-channel pipette to the bottom of the well, 1 µl of Sample Control to the three top wells (yellow), 1 µL of Negative Control to two wells (red), and 1 µL of Calibrators to three wells (green), according to the plate layout.
- Seal the plate with an adhesive plastic film, spin at 400 – 1000 x g, 1 min at room temperature. Incubate overnight at +4 °C.
Extension
- Prepare an extension mix according to the table below.
- Bring the Incubation Plate to room temperature, spin at 400 – 1000 x g for 1 min. Preheat the PCR machine.
- Vortex the Extension mix and pour it into a multichannel pipette reservoir.
- Start a timer for 5 min and transfer 96 µL of Extension mix to the upper parts of the well walls of the Incubation Plate by using reverse pipetting.
- Seal the plate with a new adhesive plastic film, use the MixMate® to vortex the plate at 2500 rpm for 30 sec, ensuring that all wells are mixed, and spin down.
- Place the Incubation Plate in the thermal cycler and start the PEA program. (50 °C 20 min, 95 °C 5 min (95 °C 30 sec, 54 °C 1 min, 60 °C 1 min) x 17, 10 °C hold)
Detection
- Prepare and prime an Olink® 48.48 IFC for Protein Expression. Briely, inject one control line fluid syringe into each accumulator on the chip, and then prime the IFC on the Olink® Signature Q100 following the instructions on the instrument screen.
- Thaw the Primer Plate, vortex and spin briefly.
- Prepare a Detection mix according to the table below.
- Vortex the Detection mix and spin briefly and add 46 µL of the mix to each well of an 8-well strip.
- Transfer 7.2 μL of the Detection mix to each well of column 1-6 in a new 96-well plate by reverse pipetting, and name it Sample Plate.
- Remove the Incubation Plate from the thermal cycler, spin down the content and transfer 2.8 µL to the Sample Plate, using forward pipetting.
- Seal the plate with an adhesive film, vortex and spin both at 400 – 1000 x g, 1 min at room temperature.
- Transfer 5 μL from each well of the Prime Plate and 5 μL of the Sample Plate into the primed 48.48 IFC left and right inlets, respectively. Use reverse pipetting and change tips after each primer or sample. Do not leave any inlets empty.
- Remove bubbles and load the chip in the Olink Signature Q100 and follow the instructions on the instrument screen.
- Run the IFC on the Olink Signature Q100.
- Carefully remove the adhesive film from the Primer Plate to avoid contamination between wells.
- Transfer 5 µL of each primer using reverse pipetting from each well in position 1 A-H (green) to the inlets in the first column on the left side of the IFC (green). Change pipette tips after each column. When using an eight-channel pipette every other inlet will be filled according to the image.
-
For research use only. Not for use in diagnostic procedures.
-
This product includes a license for non-commercial use. Commercial users may require additional licenses. Please contact Olink Proteomics AB for details.
-
There are no warranties, expressed or implied, which extend beyond this description. Olink Proteomics AB is not liable for property damage, personal injury, or economic loss caused by this product.
-
Olink® is a registered trademark of Olink Proteomics AB.
-
© 2018–2022 Olink Proteomics AB. All third-party trademarks are the property of their respective owners.
-
Olink Proteomics, Dag Hammarskjölds väg 52B , SE-752 37 Uppsala, Sweden
-
1126, v1.1, 2022-05-05
Documents / Resources
|
Olink Target 48 High Multiplex Immunoassay
Panels
[pdf] Instructions
Target 48, High Multiplex Immunoassay Panels, Target 48 High Multiplex
Immunoassay Panels, Immunoassay Panels, Panels, Target 48 Immunoassay Panels,
Target 48 Panels
---|---
|
Olink Target 48 High Multiplex Immunoassay
Panels
[pdf] Instructions
Target 48, High Multiplex Immunoassay Panels, Target 48 High Multiplex
Immunoassay Panels, Immunoassay Panels, Panels, Target 48 Immunoassay Panels,
Target 48 Panels