LSI LASTEM ELR610M Heat Shield User Guide
- June 3, 2024
- LSI LASTEM
Table of Contents
LSI LASTEM ELR610M Heat Shield
INSTRUMENT DESCRIPTION
- LCD display 4×20 chrs.
- Keyboard
- Base mounting
- Tripod mounting (BVA325)
- Reset
- Power supply / battery charger
- Anemometer input
- RS-232 serial port
- Wet bulb temperature sensor
- Air temperature sensor
- Radiant temperature sensor
- Working status LEDs
- Protective shell
HOW TO USE HEAT SHIELD MASTER
Switch on/off
Press together +
After few seconds Heat Shield Shows its PN, version and serial number
Then it shows day/time. If day/time
is correct select ESC otherwise go to Change day/ time. During first switch-on Heat Shield asks to select the system language. Select the language using and confirm using
Change date/time
Move the cursor below the date to change using Press to confirm. Press to exit
to exit
Measurement display
Heat Shield MASTER shows measurements and indexes from its sensors and, when connected, from SATELLITE units. moves to different group of measurements/indexes
The proper way to select a combination of buttons is to keep pressure over the “Ctrl” button first, then push & release the second one, then release the “Ctrl” button
Data logging
To start data logging go to measurements screen Go to SURVEY for more
information.
Status bar
It contains some Heat Shield working information
- where
- storing in progress.
- nn: survey number.
- ss:mm:can have the following meanings:
- on programmed survey, it is the time to reach the starting time of the survey (s=”-“).
- on programmed survey, during survey, it is time to reach the end of the survey (s=”-“).
- During survey, it is the time from the beginning of the survey. It is available only if survey end has not been programmed (s=”+“).
- Eee: number of errors.ru: radio signal with battery level. “-“ radio on, “1” receiving data from SATELLITE unit 1, “2” receiving data from SATELLITE unit 2. It does not appear with mode set to WBGT 1 amb (no SATELLITE unit used) and in models without the in-ternal radio module.
- LI: shows alternatively the battery level (B) and Memory status (M). 100%: battery at full charge/full memory; 0%: low battery/memory empty; other levels indicated: 80%,60%,40%,20%.
SETUP
It is possible to change some factory setup parameters
Browsing
Push to go on the MAIN menu. Use to scroll each in formation, and to enter;to
return to previous selection, it goes to Measurements when pushed from MAIN
menu.
Change parameters
Depending by the situation it is possible to change the parameters from a
selectable list (exam-ple: language) or using the arrows (example: day/time).
SURVEY
Getting started
- Check battery status.
- Check memory status.
- Mount each unit. See examples at the end of this document.
- Top up the wet bulb temperature sensor (9) tank on the MASTER unit and any SATELLITE units, if used, with demineralized water. Unscrew the tank cap without removing the cotton sock. The water should not exceed the 4/5 of the tank capacity. Then, screw the cap back on. Remember to remove the water before putting away the instrumentation.
- First switch-on the MASTER unit and later the SATELLITE units, if used.
- Check measurements from MASTER unit and SATELLITE units, if used.
- Attention: radiant temperature sensor (11) would need up to 20 minutes to acclimatize
Survey setup
Go to Survey setup.
- Start/stop: select the way to start survey (manually/selected day-time) and stop survey (manually/selected day-time/after a period time). During survey all measurements are stored.
- Nr survey: Heat Shield assign a progressive number to each survey. It is possible to change (1 ÷ 99) the first assigned number during measurement.
- Mode: Heat Shield can perform measurement from the base unit only (WBGT1 amb), from MASTER unit plus one SATELLITE (WBGT 2 amb), from MASTER unit plus two SATEL-LITE units (WBGT 3 amb). Or perform one measurement which show average value of the measurement coming from base unit and two SATELLITE units (WBGT subject, WBGT+PMV/PPD) as per ISO7243-1989 requirements.
- Subject parameters: define the parameters of the subject according to the selected survey mode, in order to produce the calculation of indices whose result is applicable to the subjects considered.
- Anemometer: if not connected select Not used or select a default air speed value, otherwise specify the type: Hot wire or Rotor anemometer.
- Measure unit: select Celsius or Fahrenheit
Survey start/stop
Each survey starts and stops using the programmed mode. The start and stop
commands are enabled only in the measurements mask. If starting mode is
manual, push + to start. Push + to stop survey, this stops survey anyway even
if automatic stop mode is programmed. You can pause an active survey at any
time by pressing +. While paused, the instrument will update and show measures
on the display, but data logging is stopped. By pressing again the same key
combination, the data logging resumes normally. While paused, the instrument
will show a ‘P’ in the top-left of the display.
HS MANAGER
HS Manager is a PC tool used to download the stored data and perform data analysis.
Installation
The software can be found on the MW6501 USB pen drive and on the website www
.lsi-lastem.com. Run Setup and follow its instructions. During the first run,
the program asks to install a new Heat Shield unit.
- Connect Heat Shield MASTER to PC via RS-232 port (8) using its cable and USB adapter.
- Go to “Communication” and select the PC port where the USB adapter is connected.
- Select “Go” until the instrument is found. Select “end” to terminate this procedure
To detect the PC port where the USB adapter is connected, go to (Windows) Control Panel Device Manager Univer-sal Serial bus controller and look for the COM number of the USB se-rial port
Data download
While Heat Shield MASTER is connected to PC:
- Select “Communications” from Tools.
- Select the instrument serial number and press .
- Select the data limits and select
. - When finish press
Data display
- From the browser, expand “Database”, select instru-ment serial number.
- Select the survey of interest.
- Left key on the mouse to select “Show Data”.
- Select data limits and press
EXAMPLES
Fig. 1 – WBGT 1 environment measurement and index from MASTER unit only.
Fig. 2 – WBGT environments: separate measurements and indexes from MASTER unit and one SATELLITE unit.
Fig. 3 – Installation of the units in case of WBGT Subject.
Fig. 4 – WBGT subject: measurements and indexes averages from the MAS-TER and 2 SATELLITE units.
Fig. 5 – Usage of WBGT 1 env. with cup anemometer.
Fig. 6 – Use without data storing.
Fig. 7 – Usage of WBGT 1 env. with hot wire anemometer.
Fig. 8 – Exploded diagram
VIDEO TUTORIALS
The following video tutorials are available for Heat Shield:
NOTES
This manual is an introduction to the use of Heat Shield. For more information
refer to the IN-STUM_01517 User’s guide downloaded from the website www.lsi-
lastem.com.
Disposal
Heat Shield is a highly electronic scientific device. In accordance with the standards of environ-mental protection and collection, LSI LASTEM advises to handle Heat Shield as waste of elec-trical and electronic equipment (WEEE). It is therefore not to be collected with any other kind of waste. LSI LASTEM is liable for the compliance of the production, sales and disposal lines of Heat Shield, safeguarding the rights of the consumer. Unauthorized disposal will be punished by the law. Dispose of the dead batteries according to the regulations in force.
How to contact LSI LASTEM
If you need technical support, contact the LSI LASTEM at support@lsi-
lastem.com, or fill in the On-line technical support
request form accessible from the home page of the website www.lsi-
lastem.com. For further information:
- Telephone: +39 02 95.414.1 (switchboard operator)
- Address: Via ex S.P. 161 Dosso n. 9 – 20049 Settala (MI) – Italy
- Web site: www.lsi-lastem.com
- Sales: info@lsi-lastem.com
- After-sales service: support@lsi-lastem.com, Repairs: riparazioni@lsi-lastem.com