AISON TECHNOLOGIES aiSon Focus Non-Sterile Adaptive Ultrasound Pad Instruction Manual
- June 3, 2024
- AISON TECHNOLOGIES
Table of Contents
AISON TECHNOLOGIES aiSon Focus Non-Sterile Adaptive Ultrasound Pad
Instruction Manual
Description and indications for use
The aiSonTM Focus is intended to be used by sonographers and can be used on any patient independent of age, weight, or other health conditions. It can be used on any ultrasound probe with a sensor window 7.6cm x 2.0cm or smaller.
It is intended to:
- avoid misdiagnoses from compression of underlying structures (e.g., all pathological, superficial fluid collections),
- avoid painful examinations for the patient (e.g., acute trauma with open or closed fractures, nerve lesions),
- facilitate ultrasound-guided interventions that do not require sterile conditions
- facilitate ultrasound-guided interventions using a sterile probe cover on top of the product when sterile conditions are required
- increase patient comfort (e.g., babies, or if a patient shows resistance towards the generous usage of ultrasound coupling agent),
- enable diagnoses in hard-to-reach areas of the patient’s body (e.g., Achilles tendon, anterior neck region, fingers, toes, pelvic floor, eyes, ears).
The aiSonTM Focus achieves its function by maximizing the contact area of the ultrasound probe with the patient’s body. In addition, it is an adaptive ultrasound standoff pad, which allows for a dynamic shift of the ultrasound image focus (point of highest image resolution). The shift can be achieved by increasing or decreasing the distance of the ultrasound probe to the body to be imaged by manually compressing or releasing the pad.
It is intended for single use on intact skin. It is intended to be used as an accessory to an ultrasound device to specifically and directly assist the medical functionality of the ultrasound device in terms of its intended purpose (please refer to the Instructions For Use provided by the manufacturer of your ultrasound system).
Instructions for Use
- Revision: 9
- Date: September 13th, 2022
Step-by-step setup guide:
Step 1:
Leave the ultrasound probe in its designated holder/storage on the ultrasound
machine.
Step 2:
Apply regular ultrasound gel (or any coupling agent allowed by the
manufacturer) to the sensor, at least enough to fully cover it.
Without applying ultrasound gel to the sensor before adhering the pad, image artifacts may appear.
Step 3:
Verify that the aiSonTM Focus package is completely sealed and unopened before
opening. Verify that the product has not expired.
Step 4:
Next, fully peel open the aiSonTM Focus packaging.
Step 5:
Pick up the aiSon™ Focus at the end with the integrated elastic opening.
Step 6:
Placing index and middle fingers on the inside of the integrated elastic
opening, stretch it to enlarge the aiSon™ Focus’ opening to approximately
match that of ultrasound probe’s widest point.
Step 7:
While keeping the integrated elastic opening stretched, gather the pad’s
material using the thumbs on the outside surface of the aiSon™ Focus, rolling
it up to form a ring-like shape with the liquid repository in the center.
The product is packaged such that it does not have to be turned inside out. If you turn the pad inside out, the product may not function as efficiently as expected.
Step 8:
Maintaining the aiSon™ Focus stretched and in a ring shape, turn it upside
down and place the liquid repository centered on the ultrasound probe’s
sensor. Once it is aligned with the probe’s sensor, hold the pad in place by
pressing it against the probe near the sensor with the thumbs while releasing
the integrated elastic opening by retracting index and middle fingers.
Step 9:
Release the thumbs, then use one hand to fix the pad against the probe on the
long sides of the probe while lifting it out of its holder/storage.
Step 10:
While holding the pad in place with one hand, pull down on the aiSon™ Focus’
bottom end with the integrated elastic opening with the other hand to fully
extend the pad and cover the largest possible area of the ultrasound probe
with its material.
Step 11:
Continue to hold the pad in place with one hand. With the other hand,
carefully pick up the loose elastic band provided in the package with the pad
and place it around the probe’s neck, or below its widest point.
Step 12:
Carefully twist the elastic band with the free hand and wrap it around the
probe as many times as necessary to tightly fasten the aiSon™ Focus on the
ultrasound probe.
Step 13:
If the aiSon™ Focus’ liquid repository is not centered on the ultrasound
probe’s sensor, pull on the pad’s material to adjust without loosening the
elastic band.
If the pad is fastened very tightly, image artifacts (e.g., reverberation artifacts) can appear in some ultrasound systems.
Step 14:
Remove any air pockets enclosed in the ultrasound gel between the pad and the
probe’s sensor by pressing down on the pad with a finger to avoid acoustic
shadows or artifacts. Then apply ultrasound gel, or another coupling agent
that is approved to be used on a patient’s skin, such as disinfectant or
water. Do not use disinfectants containing pyridine and pyridine derivatives
as coupling agent. The aiSon™ Focus is ready for use.
Without applying sufficient ultrasound gel, or another coupling agent, like
water or disinfectant, between the pad and the patient’s skin, there will be
limited image visibility and image artifacts (e.g., reverberation artifacts).
Step 15:
While scanning, dynamically adjust the pressure exerted through the probe as
to compress the pad and reduce the size of the standoff as required, or
release pressure to increase the size of the standoff as needed (see
horizontal arrow in the depiction below). The pad is designed to withstand
pressure and the liquid will move by bulging on the sides (see vertical arrows
in the depiction below), which is intentional, as shown in the depiction.
Removal and disposal
Step 1:
Unwind the elastic band around the ultrasound probe’s neck. If the elastic
band should be slippery due to excess ultrasound gel, first dry the elastic
band and then unwind it.
Step 2:
Pull the aiSon™ Focus upwards at the integrated elastic opening and pull it
off the ultrasound probe in one go. Dispose the aiSon™ Focus pad in a regular
waste bin.
Step 3:
Remove the ultrasound gel on the sensor and clean the ultrasound probe in
accordance with the manufacturer’s cleaning instructions.
Safety Measures
- The product does not contain any hazardous substances.
- The product does not contain latex.
- The product does not contain any endocrine-disrupting properties (e.g., bisphenol A or phthalates).
- Every aiSonTM Focus filling process has undergone quality checks.
- Does not contain animal or biological products.
Compatibility:
The aiSonTM Focus fits any ultrasound probe with a sensor area smaller or
equal to 8.0cm x 2.0cm.
Warnings and Precautions
-
Do not use if the package is damaged.
-
The product is provided as non-sterile.
-
Do not use the product after its expiration date.
-
Not made with natural rubber latex.
-
Do not reuse.
-
If the product dropped on the ground or was otherwise contaminated with dirt, we recommend its disposal.
-
Avoid applying excessive tension to the elastic band to prevent tearing. Tearing of elastic bands may result in injuries,
e.g., to the eyes, resulting from uncontrolled recoil of a tensioned elastic band upon tearing. -
Keep the elastic band in the package until the time of application to mitigate contamination. Do not use the elastic band if it falls on the floor or comes in contact with any other unclean surfaces.
-
Open the package only just before use.
-
Not meant for intracorporeal use.
-
Cannot be operated at temperatures below the freezing point.
-
Do not leave a contaminated product unattended.
-
Do not use disinfectants containing pyridine and pyridine derivatives as coupling medium.
-
The product is not optimized for linear “hockey stick” probes and may perform suboptimally with this probe design.
Contraindications:
No known contraindications.
Any serious incident that has occurred in relation to the product should be reported to AisonTM Technologies and the competent authorities.
Storage and Disposal
- The product can be used until the expiration date, which is indicated on the product label itself. The expiration date can be guaranteed if the product is stored at room temperature.
- After use, the aiSonTM Focus must be disposed of with the regular waste.
- Follow all local regulations for non-hazardous waste disposal.
Legal Notes:
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of AisonTM Technologies. AisonTM Technologies may revise this publication from time to time without written notice.
TRADEMARKS: All products and their name brands are trademarks of their
respective holders.
COPYRIGHTS: All Material Copyright© 2020-2022 by AisonTM Technologies AG.
All Rights Reserved
Support
AisonTM Technologies AG
Wiesenstrasse 10A
8952 Schlieren
Switzerland
Tel. no: +41 76 382 81 13
info@aisontechnologies.com
www.aisontechnologies.com
Article number: 001.002.001
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>