INSTALLATION INSTRUCTIONS
STEP 1
Removing Cartridge.
- Turn off the X-PLO₂R® by pressing the power button for 2 seconds.
- Remove the X-PLO₂R from the carrying bag, if applicable.
- Remove the battery from the X-PLO₂R.
- Place the X-PLOâ‚‚R on its side to access the underside of the POC.
- Pull the cartridge out of the POC by pulling on the cartridge pull handle.
- Remove the cartridge completely from the X-PLO₂R and throw away.
STEP 2
Installing Cartridge.
- Remove the cartridge from the mylar bag. The new cartridge has a dust cap on top. Remove the dust cap. Make sure there is no dust or debris where the dust cap was located.
- Insert the new cartridge top first into the X-PLO₂R. Do not leave the cartridge ends exposed; the cartridge should be inserted into the X-PLO₂R as soon as the used cartridge is removed.
- Push the cartridge into the X-PLOR so it is fully inside.
- The cartridge pull handle should lay flat once the cartridge is secured.
- Connect charged battery to the X-PLO₂R.
- Turn on the X-PLO₂R by pressing the power button for 2 seconds. Set the flow setting to your prescribed setting and use it normally.