DIA GLOBAL DP 210 Duo Photometer Plus User Manual
Dear customer,
We are pleased that you have chosen the Duo Photometer plus from Diaglobal GmbH and thank you for the confidence you have placed in us.
The Duo Photometer plus belongs to a new generation of small mobile devices developed by Diaglobal GmbH and specially designed for on-site analysis.
With the software version V5.3 and higher, an automatic test of the device function has also been integrated. Therefore, the Duo Photometer plus complies with the requirements of the guidelines of the German Medical Association.
With the Duo Photometer plus, 9 clinical-chemical parameters can be
determined.
The device can be supplied with SI units of measurement on request (see
chapter 9, Technical Data, table Measuring Ranges).
The kits and accessories required for the test are also available from Diaglobal GmbH.
All the best for your work with the new Duo Photometer plus!
Yours
Diaglobal GmbH
General information on the Photometer
- Device name: Duo Photometer plus
- Model: DP 210
- Features: In-vitro diagnostics, measuring device for the determination of selected clinical-chemical parameters in blood and serum/plasma
The Duo Photometer plus fulfils the basic requirements of Appendix I of Directive 98/79/EC regarding in-vitro diagnostics.
The conformity of the device with Directive 98/79/EC is confirmed by the use of the CE marking.
Installation
For trouble-free operation of the device, the following environmental conditions must be met:
- Ambient temperature: 0 °C … 40 °C
- No direct exposure to sunlight or similar sources of radiant heat
- Free from excessive dust
- Free from vibrations
- Free from interference by electromagnetic waves
- Operation on a horizontal surface
Please observe the following instructions for use:
Insert a rechargeable battery or normal battery if the device is to be
operated
independently of a power supply or connect the photometer to a power supply
unit.
Press the < ON/ENTER> key (Fig. 1) to activate the internal device check which is automatically carried out by the device.
The device is then immediately ready for measurement.
Description of the device
Power supply
The Duo Photometer plus can be operated as desired using a power supply, a (9V block) battery or (model 6F22 or PP3) rechargeable battery.
Mains power operation
The Photometer is supplied with a power supply unit for operation on a mains
voltage in the range of 100 V … 240 V AC. The mains plug is marked with a
Diaglobal logo (sticker).
The connector plug of the power supply unit is connected to the power supply
socket on the back of the device.
Network-independent operation
To insert the rechargeable battery or the normal battery:
Unscrew the knurled screws on the bottom of the unit and remove the battery
compartment cover. Connect the battery to the push-button contact and insert
it into the device. Replace the battery compartment cover and screw in the
knurled screws.
Please note:
The Duo Photometer plus can be operated using a power supply without the need
to remove the rechargeable battery or the normal battery.
The rechargeable battery cannot be charged while it is installed. A separate
battery charger is required for this purpose.
Measuring system
The optical section is shown in Fig. 2.
The light emitted by an LED is first selected into its wavelength range (546 nm) by an interference filter IF (HBW ~ 5 nm) and then bundled and directed onto the cuvette in the shaft. After passing through the cuvette, a broadband photo sensor converts the light falling on its sensor surface into a current, proportional to the intensity.
Service
Adjustment and Calibration
The instrument is adjusted and calibrated at the factory on delivery,
adjustment by the customer is not necessary.
Adjustment is carried out via the interface socket on the rear panel. It can
only be carried out at the factory, adjustments by the customer are not
possible.
Information on calibrating the device can be found in chapter 6. Quality
control according to the Guideline of the German Medical Association.
Maintenance
The device is maintenance-free. Maintenance after the warranty period is
recommended, but not mandatory.
Due to the integrated test of the unit functions (chapter 8.5) and regular
tests with control material, maintenance is only recommended if one of these
two test functions indicates an error message.
Cleaning Instructions
Commercially available decontaminating solutions commonly used in clinical
chemistry laboratories, such as Mikrozid® AF Liquid, Bacillol® plus, 3 %
Kohrsolin® or similar, are recommended for cleaning the device and the
surface.
Before cleaning the unit with a soft cloth and the decontaminating solution,
it must be switched off and the electrical power supply must be disconnected.
Make sure that no liquids get into the device. There is no protection against penetrating liquids (Code IP X0)
The cuvette shaft must not be cleaned by the user of the device, as this may
damage the device. If cleaning is necessary, especially because of leaking
liquids or broken glass, please contact Diaglobal GmbH.
Malfunctions
If any malfunctions or problems occur, simply call us. Most questions can be
answered on the phone. Non-functional units should be sent to our Berlin
address.
We will provide a loan device for the duration of the repair.
Disposal
Diaglobal GmbH will take back and dispose of units that are no longer needed or cannot be repaired, free of charge.
Required reagents and laboratory accessories
Expiration date of consumables
It is important to ensure that all consumables may only be used within the
expiration date.
Reagents / parameter list
The following tests can be measured with the Duo Photometer plus:
Parameter | Sample material | Tests/pack | Art. no. |
---|---|---|---|
Blood | Serum | Plasma | |
Bilirubin | – | + | + |
Neonatale Bilirubin1) 2) | + | – | – |
Haemoglobin (SLS-method) | + | – | – |
Haemoglobin (HiCN-method) | + | – | – |
Erythrocytes | + | – | – |
Haematocrit | + | – | – |
- Mini centrifuge required (Art. no. DZ 002)
- From blood, with subsequent sample preparation (centrifugation with mini centrifuge)
Control materials
Art. no. | Description | Contents |
---|---|---|
HEM QS | Haemoglobin control |
Haemolysate for correctness and precision control of haemoglobin determination
in blood in the normal range| 5 x 1 mL
ERY QS| Erythrocytes- and Haematocrit control Control blood for accuracy and
precision control of haematocrit and erythrocytes determination in blood in
the normal range| 5 x 1 mL
BIL QS| Bilirubin control Lyophilisate for accuracy and precision control of
bilirubin determination| 20 caps
Laboratory aids and accessories
Art. no. | Description | Contents |
---|---|---|
LH 001 | Blood lancets | 500 |
LH 004 | Capillaries 10 µL, with ring mark | 250 |
LH 006 | Cuvette rack | 1 |
LH 007 | Micropipettor (pipetting aid) | 1 |
LH 009 | Cellulose swabs | 500 |
LH 010 | Cellulose swab box | 1 |
LH 011 | Alcohol pads, non-sterile | 100 |
LH 012 | Powder-free nitrile gloves size M | 200 |
LH 023 | Capillaries 100 µL, with ring mark | 250 |
LH 024 | Capillaries 20 µL, with ring mark | 250 |
LH 031 | Microvettes CB 300 | 100 |
LH 034 | Safety Lancets Neonatal, pink 1.2 mm | 200 |
LH 035 | Safety lancets extra, orange 1.8 mm | 200 |
DZ 002 | Mini centrifuge | 1 |
All reagent kits, control materials and other materials are supplied by Diaglobal GmbH and can be stored and transported together with the Duo Photometer plus in a practical case.
Quality control according to the Guideline of the German Medical
Association1)
The Duo Photometer plus has been specially developed for near-patient immediate diagnostics with unit-use reagents (German Medical Association, part B, chapter 2.1.5). According to the guideline of the German Medical Association, there is therefore no obligation to participate in surveys (German Medical Association, part B, chapter 2.2, paragraph (3) a). The user only has to carry out internal quality checks.
Internal quality assurance is carried out in the form of a weekly accuracy check (calibration) with subsequent documentation of the measured value. The corresponding protocol forms are available from Diaglobal free of charge.
We recommend using the Diaglobal control caps BIL QS to check the accuracy of bilirubin determinations.
We recommend using the blood control HEM QS and ERY QS with target values in the normal concentration area for checking the accuracy of determinations of haemoglobin, haematocrit and erythrocyte counts.
In agreement with the requirements of the German Medical Association, a test of the device function (see operating instructions, chapter 8.5) is integrated in the Duo Photometer plus, therefor a daily test by means of a standard manual test (German Medical Association, part B, chapter 2.1.5, paragraph (2) is not necessary.
Measuring process
Endpoint measurement
The absorbance is measured after reaching the endpoint.
It is measured against the reagent’s blank count.
Parameters: Haemoglobin SLS (HB SLS), Erythrocytes (ERY), Haematocrit (HCT), Haemoglobin (HiCN)
Calculation: Concentration = Absorbance x Factor
The erythrocyte and haematocrit counts are determined using stored reference curves.
Endpoint measurement with consideration of the sample blank value and
pre-programmed measuring time
After measuring the sample blank value, the colour reaction in the cuvette is
started and the endpoint absorbance is measured after a specified time has
elapsed.
Parameters: Bilirubin (BIL), Neonatal bilirubin (BIL N)
Calculation: Concentration = Absorbance x Factor
Measurement time: 2 Minutes
The samples are measured in succession:
- Sample 01: Measurement 1 (sample blank value)
- Sample 01: Measurement 2 (result)
- Sample 02: Measurement 1 (sample blank value)
- Sample 02: Measurement 2 (result)
etc.
Measurement
Switching the device on
- Press the <ON/ENTER> key
Self-test when switching on
When the device is switched on, a self-test of the digital and analogue circuitry is conducted. The operational device check proceeds automatically after it is switched on. It takes approx. 5 seconds, after which the unit is ready for measuring.
Note: If it becomes obvious during the test that one of the device functions does not correspond to the required settings, will appear in the display.
In this case, switch the device off.
Please call Diaglobal GmbH service (Tel. +49 (0) 30 6576 2597) or contact your specialist retailer.
Test selection
Press the <ON/ENTER> key.
The desired test is selected from the menu with the right or left arrow key:
HB – HB-SLS – ERY – HCT – BIL – BIL N – ABS546
Pressing the right arrow key activates the next test while pressing the left arrow key returns to the previous test. The selected test is shown in the upper right corner of the display.
Confirm test selection with the < ON/ENTER> key.
Switching the device off
To switch the device off, press both arrow keys simultaneously.
Integrated operational device checks
Self-test when switching on
Testing of the digital and analogue circuits of the device is automatically
performed by the device when it is switched on.
Please see chapter 8, point 8.2.
Differential measurements
All measurements are based on differential measurements. I. e. after selecting
the desired test, the device requests a zero measurement with a blank value
cuvette.
This creates a reference base to the measured value so that minor deviations
can be compensated.
Measurement range controls
The measurement ranges of all measurement results shown in the display are
verified by an integrated measurement range control. If the measurement range
is exceeded, an error is displayed.
The measurement ranges that are separately defined for each parameter are
documented on the respective package inserts as well as in this operating
manual, chapter 9, Technical Data.
Plausibility controls
For multi-point measurements, the absorbance measured first forms the reference basis. The programme verifies the plausibility of the individual measured values. If specific requirements (e.g. E2 > E1 for ascending reactions) are not met, an error message is displayed.
Notes on taking samples and carrying out measurements
Errors in taking samples will always lead to incorrect measurement results.
This chapter addresses the most common errors that can occur during taking
samples and measuring samples.
-
Before measuring, cuvettes stored in a refrigerator must be brought to room temperature. If the cuvettes are too cold, they will become misty with water on the outer wall due to the humidity, which will lead to incorrect measurement results.
-
Never touch the lower part of the cuvette (where the liquid is) with bare
hands. If this should happen accidentally, clean the vials with a fluff-free cloth before use. Cleaning with a fluff-free cloth is recommended in any case. Even if the package is still new and unopened. Fingerprints on the cuvette lead to incorrect measurement results. -
If the blood is taken from the heel using the microvette (neonatal bilirubin), make sure that there is enough blood (approx. 60 µL) in the microvette, as 20 µL serum/plasma is needed for the measurement. Close the microvette wellafter taking the blood sample and return it to the small sample tube for centrifugation.
-
After centrifugation of the microvette, please make sure that the centrifugate has separated completely and that the supernatant is clear and free of any solid particles. If not, repeat the centrifugation. If the supernatant is not free of suspended matter or if particles of the centrifugate accidentally enter the capillary, the measurement result will be incorrect.
-
If blood is taken from the fingertip or earlobe, note that the first drop that forms spontaneously must be wiped away with a cellulose swab. It contains a high proportion of tissue fluid, which will corrupt the measurement result.
-
The second drop that forms is for blood sampling. To support blood collection, it may be pressed carefully (!). The emphasis on carefully, otherwise too much tissue fluid will get into the blood sample again.
-
Make sure that the blood drop that forms is large enough to fill the capillary with the required sample volume in one go. Repeated filling of the capillary leads to air bubbles that cannot be removed from the capillary. If air bubbles form, discard the capillary and start sampling again.
-
The capillary must be filled exactly up to the black ring mark.
Please note: A deviation of only 1 mm from the ring mark is sufficient to obtain a completely incorrect measurement result!
If the sample is above the black ring mark, this will lead to incorrect positive measurement results. A cellulose swab can be used to carefully soak up too much blood.
If the sample is below the black ring mark, this will lead to incorrect negative measurement results. In this case, correction is hardly possible due to the air bubble that will form when trying to collect more blood. -
Before the capillary is placed in the cuvette, the lower area must be carefully wiped on the outside with a cellulose swab to remove sample particles attached to the capillary. Otherwise, this would lead to incorrect positive measurement results.
-
With the help of the micropipetter, the sample is completely transferred into the cuvette. The complete transfer of the sample is done by ejecting it several times with the help of the push button on the micropipetter.
Please note: The micropipetter is only used when the capillary is filled with the sample. It is not needed for filling the capillary. The capillary is filled by the capillary action alone. -
When changing the cap with the starter cap, make sure that the substance in the starter cap has completely dissolved. Failure to do so will result in a nonlinear kinetic reaction process, which will lead to an error message in the display or unreliable measurement results.
Technical data
-
Storage temperature: -20 °C … 70 °C
-
Operating temperature: 0 °C … 40 °C
-
Dimensions: 200 x 100 x 50 mm
-
Weight: 450 g
-
Measuring principle: Absorption measurement with single beam
photometer (Fig. 2), chopped operation -
Projector: LED
-
Spectral apparatus: Interference filter
-
Measuring wavelengths: 546 nm
-
Spectral half-width value: ~ 5 nm
-
External light influence: Negligible
-
Interface: V24 (9600, 8, n, 2)
-
Power supply: 6 V … 12 V DC
-
Current consumption: max. 250 mA
-
Warm-up time: 0 min
-
Interference suppression: According to DIN VDE 0871 and DIN VDE 0875
-
Inaccuracy: < 0.5 % at A = 1.000
Relative photometric -
short-time standard deviation: < 0.1 %
Measuring ranges: | DP 210 | DP 210 SI |
---|---|---|
Haemoglobin (HiCN-method) | 0.0 – 50 g/dL | 0.0 – 31 mmol/L |
Haemoglobin (SLS-method) | 0.0 – 50 g/dL | 0.0 – 31 mmol/L |
Erythrocytes | 1.0 – 10 Mio/µL | 1.0 – 10 Mio/µL |
Haematocrit | 5 – 90 % | 5 – 90 % |
Bilirubin | 0.50 – 25 mg/dL | 8.50 – 428 µmol/L |
Neonatal Bilirubin | 2.30 – 50 mg/dL | 39.0 – 850 µmol/L |
ABS 546 nm | A = 2.500 | A = 2.500 |
General Guidelines and Notes
EC Directives
-
Directive 98/79/EC on in-vitro diagnostic devices
EN / ISO standards -
EN ISO 9001:1994, Quality Management Systems, Model for quality assurance in design, development, production, installation and customer service
-
EN ISO 13485, Medical devices, Requirements for regulatory purposes
(application of EN ISO 9001) -
EN ISO 14971, Medical devices – Application of risk management to
medical devices -
EN 61010 -1, Safety requirements for electrical equipment for
measurement, control and laboratory use – Part 1: General requirements -
EN 61010 -2-101, Safety requirements for electrical equipment for
measurement, control and laboratory use – Part 2-101: Particular requirements for in-vitro diagnostic (IVD) medical equipment -
EN 61326 -1, Electrical equipment for measurement, control and
laboratory use – EMC requirements – Part 1: General requirements -
EN 61326 -2-6, Electrical equipment for measurement, control and
laboratory use – EMC requirements – Part 2-6: Particular requirements – In- vitro diagnostic (IVD) medical equipment -
EN 592, Instructions for use for in-vitro diagnostic instruments for
professional use
National directives and recommendations (Germany) -
Guidelines for Quality Assurance of Laboratory Examinations of the
German Medical Association of 23.12.2019
Note on electromagnetic compatibility
-
The photometer meets the requirements for electromagnetic radiation
and interference immunity as described in the IEC 61326 series of standards. -
Do not use this device near sources of intense electromagnetic radiation
because they may interfere with correct functioning. A distance of at
least 1 m should be maintained between an operational (switched on)
mobile phone and the photometer during measurement.
Note on the unit’s internal quality control
The functionality of the device is checked when it is switched on. In addition, electronically controlled checks are carried out for individual tests during the measurement, which leads to an error message if specified requirements are not met.
Appendix: “Step-by-step measurement”
Please refer to the illustrations in the “Step by step” instruction manual.
Step by step instructions
Device manual
-
Switch on:
Press ON/ENTER key Wait for device check and confirm with ON/ENTER -
Select test:
Press arrow key until required test appears -
Confirm required test:
Press ON/ENTER -
Switch off:
Press both arrow keys at the same time
BIL / BIL N (BIL 142)
Additionally required: Mini centrifuge
-
Insert capillary with sample (serum/plasma) into cuvette
BIL: 100 µL (adults)
BIL N: 20 µL (newborns)
Sampling of BIL N see next page
-
Eject sample several times with micro pipetter into cuvette
-
Screw cap on Turn cuvette upside down several times
-
Switch photometer on with ON/ENTER key Wait for device check, confirm with ON/ENTER Select the required test, confirm with ON/ENTER
-
Zero point adjustment: Insert cuvette with sample(Fig. 3) into photometer, zero point is stored in memory Remove cuvette after signal tone
-
Replace screw cap with starter cap
-
Turn cuvette upside down several times
-
First press ON/ENTER Then insert cuvette into photometer
-
Time is displayed, wait for measured value
BIL / BIL N (BIL 142)
Sampling of BIL N
Additionally required: Mini centrifuge, Microvette
-
After pricking with the lancet, take about 60 µL of blood (approx. 1 drop) from the heel with the microvette
Note: Close the microvette carefully before placing it into the mini centrifuge
-
Insert microvette into mini centrifuge Centrifugate for 3-5 minutes
Note: Ensure an even loading inside the mini centrifuge
-
Take 20 µL plasma from the microvette
Continue with Fig. 1 on the previous page
HB 142 / HB 342 / ERY 142 / HCT 142
-
Insert capillary with 10 µL blood sample into cuvette
-
Eject blood several times with micro pipetter into cuvette
-
Screw cap on
Turn cuvette upside down several times Wait 3 minutes HB 342: wait 30 seconds
-
Switch photom
eter on with ON/ENTER key Wait for device check, confirm with ON/ENTER Select the required test, confirm with ON/ENTER
-
Zero point adjustment: Insert an unprocessed cuvette from package into photometer Zero point is stored in memory
-
Remove cuvette after signal tone
-
Insert cuvette with blood sample (Fig. 3) into photometer Read measured value
-
In regard to series measurement:
After zero point setting any number of additional samples can be measured
DIA QS / BIL QS
Quality assurance
Photometer testing with control caps
-
20 control caps with lyophilized control serum
DIA QS: Glucose, Lactate, Triglycerides, Cholesterol, HDLCholesterol, Uric acid, Bilirubin (adults) BIL QS: Bilirubin (adults) , Bilirubin N (newborns)
-
Screw the control cap onto a cuvette of the test you want to check
Mix well Now the cuvette contains a sample with a known concentration
-
Leave the cuvette for 1 minute
-
Switch photometer on with ON/ENTER key Wait for device check, confirm with ON/ENTER Select the required test, confirm with ON/ENTER
-
Zero point adjustment: Insert cuvette with sample (Fig. 3) into photometer, zero point is stored in memory Remove cuvette after signal tone
-
Replace control cap with starter cap
-
Turn cuvette upside down several times
-
First press ON/ENTER, then insert cuvette into photometer Wait for measured value Compare the measured value with the target value on the package insert