VISIONTECHSHOP VLP-100B Series Label Printing Scale User Manual
- September 13, 2024
- VISIONTechShop
Table of Contents
- VLP-100B Series Label Printing Scale
- Product Information
- Specifications:
- Normal Calibration in kg mode:
- Normal Calibration in lb mode:
- Q: How do I access the calibration mode on the VLP-100B Series
- Q: What should I do if the calibration fails?
- Q: What is the recommended weight to use for calibration?
VLP-100B Series Label Printing Scale
“`html
Product Information
Specifications:
- Model: VLP-100B Series Label Printing Scale
- Calibration Modes: Normal, Linear, Zero
- Units: kg, lb
- Capacity: Minimum 1/3 of the scale capacity
Product Usage Instructions
Normal Calibration in kg mode:
-
Press the calibration switch located at the back of the
indicator in approval version. -
Press the key to enter the setup menu.
-
Enter calibration using numeric keys 5351.
-
Follow the steps provided in the manual for calibration,
including emptying the scale, placing the test weight, and
confirming calibration success.
Normal Calibration in lb mode:
-
Select lb unit by pressing the calibration switch at the back
of the indicator. -
Enter normal calibration using numeric keys 51.
-
Follow the steps provided in the manual for calibration,
including emptying the scale, placing the test weight, and
confirming calibration success.
FAQ
Q: How do I access the calibration mode on the VLP-100B Series
Label Printing Scale?
A: To access calibration mode, you need to press the calibration
switch located at the back of the indicator in approval
version.
Q: What should I do if the calibration fails?
A: If calibration fails, ensure you followed all steps correctly
as per the manual. If issues persist, contact customer support for
assistance.
Q: What is the recommended weight to use for calibration?
A: It is recommended to use a test weight that is a minimum of
1/3 of the scale’s capacity for accurate calibration.
“`
VLP-100B Series Label Printing Scale Operation Manual
9. CALIBRATION (Dealer only) It allows selecting Normal calibrationLinear
calibrationZero calibration to calibrate the scale. 9.1 Normal Calibration in
kg mode
5.service> 3.calibration parameters > 5.Calibration> 1.Normal Calibration
REV:04
Before access to the parameter,it need to press calibration switch,locating back side of indicator in approval version.
Press
key to enter setup menu.
Using numeric keys 5351 to enter calibration.
Steps
1 Press
Key to enter.
The display will be show “Please empty the scale”
2 Remove goods from the platter.
Remark
– 42 –
VLP-100B Series Label Printing Scale Operation Manual
3 Press
key to confirm.
The display will be show “Enter calibration weight by using numeric keys “.
4 Press
key to confirm.
5 Place the test weight on the platter. Better to use minimum 1/3 of the scale capacity)
6 Press
key to confirm.
The display will be show “Please empty the scale”.
7 Remove the test weight from the scale.
8 Press
key to confirm
The display will be show “Calibration success ”
( Calibration Completed and back to the normal weighing mode)
– 43 –
REV:04
VLP-100B Series Label Printing Scale Operation Manual
9.2 Normal Calibration in lb mode
REV:04
Press
key to enter setup menu.
Using numeric keys 633 to select calibration unit.
6. Service> 3.calibration parameters > 3. calibration unit
Before access to the parameter,it need to press calibration switch,locating back side of indicator in approval version.
Press
key to select lb unit , press
Key to confirm
Using numeric keys 51 to enter normal calibration
5.Calibration> 1.Normal Calibration
Steps
Remark
1 Press
Key to enter.
The display will be show “Please empty the scale”
2 Remove goods from the platter.
3 Press
key to confirm.
The display will be show “Enter calibration weight by using numeric keys “.
– 44 –
VLP-100B Series Label Printing Scale Operation Manual
4 Press
key to confirm.
5 Place the test weight on the platter. Better to use minimum 1/3 of the scale capacity)
6 Press
key to confirm.
The display will be show “Please empty the scale”.
7 Remove the test weight from the scale.
8 Press
key to confirm
The display will be show “Calibration success ”
( Calibration Completed and back to the normal weighing mode)
– 45 –
REV:04