TENMARS TM-197 AC-DC Magnetic Field Meter User Manual
- June 3, 2024
- TENMARS
Table of Contents
- Accessories
- Safety Precaution
- Instrument Description
- Operation
- Zeroing
- Peak Measurement
- Relative Value Measurement
- Turn on backlight
- Manual Record for One Log
- Auto Record
- Read the record data from RAM
- Data Hold
- The Max./Min. Values Hold
- Disable Auto Power Off Function
- Stopping buzzer
- Manual range mode/Automatic range mode
- AC/DC Switching magnetic field
- Setup: Step 1~ Step 8
- Software Installation
- General Specifications
- Electrical Specification
- Maintenance or Repair
- Battery Replacement
- PRECAUTIONS
- End of Life Disposal
- Documents / Resources
TENMARS TM-197 AC-DC Magnetic Field Meter
Introduction
- Measurement for the magnetic field generated from Direct Current (DC, with display by N/S Magnetic polarity) or Alternating Current (AC in 40-500Hz).
- Measurement for the residual magnetic after processing of mechanical parts
- Measurement for the magnetic field strength of magnetic applications
- Measurement for the residual magnetic generated from the stress after processing of stainless material
- Measurement for the magnetic force of magnetic materials
- Measurement for the various steel materials with natural magnetism
- Measurement for the magnetic field strength of motors in various electrical appliances
- Measurement for the magnetic field strength of permanent magnets
- Detection for the magnetic field of leakage generated from superconductive magnets
Accessories
- 1 Meter
- 1 AC/DC magnetic probe
- 1 User’s Manual
- 6 Battery 1.5V(UM4/AAA)
- 1 Carrying case
- 1 Power adapter DC 9V
- 1 USB cable and Installation disk
Safety Precaution
| Caution! Please refer to this manual. Improper use may damage the meter
and its components.
---|---
| Complies with European Directive.
-
Do not operate in environments with flammable gas or humid environments.
-
Operating altitude: up to 2000M.
-
Operating environment: Indoor use; Pollution degree 2.
-
Clean with soft cloth when dirty, such as glasses cloth.
Do not clean with chemicals and other solvents. -
EMC: EN61326-1:CISPR 11:Group 1, Class B
-
Class B – Equipment for use in all establishments other than domestic.
-
Group 1 – RF energy generated is needed for internal functioning.
Instrument Description
- Data hold
- Power
- Backlight button
- SET/Peak value hold
- Read the record data from RAM (R-M)/Left
- Set to Zero (Zero)/ Relative Value (REL)
- AC/DC switch/Up
- Manual range mode/Automatic range mode
- Manual record (REC)/ Auto record (LOG)
- Buzzer/GO & NO-GO
- The max. & min. value hold/Down
- Probe connector
- External power DC 9V input
- USB data output jack
- Battery cover and tilt stand
- LCD display
Operation
-
Press to turn on or turn off.
-
Remove the cap of probe.
-
Move the HALL ELEMENT which is located on the tip of probe close to the magnet under test and start measuring. Please refer to Fig. 1.
-
Read the measured value from the LCD display.
-
Cover the cap of probe back.
| If the probe not inserted into the host or not inserted completely, will
appear. Please turn power off and ensure to insert the probe into the host;
again, turn power on.
---|---
| If the probe is disconnected with the host while working, will appear for
warning.
Zeroing
Move the probe away from the magnetic field and press , ZERO will appear on the LCD. After zeroing completed, the ZERO disappears.
| If more than 3G or 0.3mT, will appear to prevent from erroneous zeroing the
value in measuring. (from PC)
---|---
| If more than 3G or 0.3mT, will appear to prevent from erroneous zeroing the
value in measuring. (from HOST)
| The zeroing function is unavailable for temperature.
Peak Measurement
Move the probe away from the magnetic field, then press and hold for ≥2 seconds to start Peak Measurement. The upper digit row presents the instant measuring; the lower digit row presents the peak value x.x|
Again, press and hold to quit this function.
| Zeroing will run automatically in prior to Peak Measurement. If more than 3G
or 0.3mT, will appear to prevent from erroneous zeroing the value in
measuring.
---|---
Relative Value Measurement
Press and hold for ≥2 seconds to start Relative Value Measurement.
The difference between the two magnetic fields can be compared here. For
instance, the 1st magnetic field is measured to be 100. At the time, press
“REL” button, and the LCD displays 0; again, measure the 2nd magnetic field.
If the measured value of the 2nd magnetic field is 120, the LCD will display
20 (120-100=20).
Again, press and hold to quit this function.
Turn on backlight
Press to turn the backlight on or off. If connected via an external power
supply, the backlight will lit automatically and stay on.
*The backlight will automatically turn off after being lit for 30 seconds.
Manual Record for One Log
Press to save one log of data, the LCD will show “REC” and the number of logs. For instance, if the current number is 10, it will be increased by 1 at each pressing. The number of logs is limited to be 200; if exceeded, the LCD will show “Full”.
Auto Record
Press and hold , LCD displays , and the auto-recording function is enabled. Again, press to quit this function
| The record is setup on PC via USB connection.
---|---
| If the power is insufficient when recording in logs, will appears.
Read the record data from RAM
Press for the reading mode, and appears on the LCD. Press or to select the log number for reading. Press and hold to quit this mode.
Data Hold
Press to enable or disable the data hold function.
The Max./Min. Values Hold
Press to enable MIN/MAX function; again, press to show the MAX and MIN values by turns. Press and hold to quit. The measured data can be held and updated as the measured the MAX and MIN values.
Disable Auto Power Off Function
With power off mode, keep pressing and press to turn on power, LCD will show “n-SL” that means to disable auto power off function. Re-starting will reset to the auto power off function. The function will be disabled when an external power is detected.
Stopping buzzer
Press to stop the buzzer’s beep
Manual range mode/Automatic range mode
Press to operate in the manual range mode with the LCD showing . This can change into two different range modes. Press and hold for ≥ 2 seconds to return to automatic range mode with LCD showing .
AC/DC Switching magnetic field
Press to switch the measurement between AC magnetic field and DC magnetic field.
Setup: Step 1~ Step 8
Press to enter the setup procedure from step 1 to step 8, while SET flashes on the LCD.
Step 1. Setup unit G or mT.
- Press or to select the unit G or mT.
- Again, press to go to Step 2.
Step 2. Setup unit °C or °F
- Press or to select the unit ºC or ºF.
- Again, press to go to Step 3.
Step 3. Set up the upper limit (Max.) for alarm GO/NO-GO
- The default value is 20000G (2000.0mT). Press or to modify and setup the default value.
- Again, press to go to Step 4.
Step 4. Set up the lower limit (Max.) for alarm GO/NO-GO
- The default value is 0G (0.0mT). Press or to modify and setup the default value.
- Again, press to go to Step 5.
| If the min. value is more than the max. value, appears.
---|---
Step 5. Setup the measurement interval of alarm GO/NO-GO
- The default value is 2 seconds. Press or to modify and setup the default value.
- Again, press to go to Step 6.
Step 6. Setup year/month/date
- Go to the mode to setup the year/month/date.
- Press or to select the item to modify, the digits of the selected item will flash.
- Then, press or to correct the value.
- Again, press to go to Step 7.
Step 7. Setup hour/minute:
- Go to the mode to setup the hour/minute.
- Press or to select the item to modify, the digits of the selected item will flash.
- Then, press or to correct the value.
- Again, press to go to Step 8.
Step 8. Clear the manual record:
- Go to the mode to clear the manual record LCD shows .
- Press or to select or where the default value is
- If the record is not cleared, press again to return to the measurement mode when the LCD shows .
- If is selected, press to return to the measurement mode and clear the record while LCD shows “CLEAR”.
Software Installation
Windows 7 / Windows 10 operating systems
-
Place the CD included with this meter into the CD/DVD-ROM drive of the PC to connect to and install the desktop program:
-
Run the desktop application:
-
Connect the USB cable included with this meter to the PC, as shown in the figure below.
-
Execute the PC desktop software program:
Double-click the left mouse button on the desktop program (Thermometer Meter) to execute the desktop program.
General Specifications
-
Liquid crystal display (LCD) of 5 digits with the displaying value up to 30000.
-
Backlight display. The backlight is manually turned on (it will turn off automatically after turned on for 30 seconds) and turned off. If connected via an external power supply, the backlight will lit and stay on; it will turn off until the external power supply disconnected.
-
Enable and disable auto power off function (default 15 minutes).
-
Measurement range: 0~3000mT (milli Tesla) /0~30000G (Gauss); Auto Range function (with Automatic Measurement).
-
Conversion of units: 1mT (milli Tesla) = 10G (Gauss)
-
Relative value/Peak hold/Real time Zeroing function.
-
Enable and disable auto power off function (default power off is 15 minutes from the last operation).
-
DATA HOLD(Data hold)
-
MAX/MIN (Maximum hold/ Minimum hold)
-
The GO and NO-GO alarm function and its settings that can be applied to detect good and defective products on the production line.
-
Unit switch for Gauss/Tesla.
-
The maximum, minimum, and average values: auto export the top 10 max./top 10 min./ average values based on the data which is recorded and downloaded to the computer.
-
As connected to a PC, press REC to record the real-time data in the PC.
-
Overload display: “OL”.
-
Manual recording: up to 200 logs.
-
Automatic recording: up to 7000 logs where the
record interval shall be setup with PC. -
detection for battery low
-
Battery life: Approximately 100 hours.
-
Operation temperature and humidity: 5°C to 40°C (41°F
to 104°F), < 80%RH。 -
Storage temperature and humidity:0°C to 50°C, relative
humidity under 70%. -
Weight: Approximately 320 grams (with battery)
-
Dimensions: 156 (L) x 73 (W) x 35 (H) mm
-
AC to DC Adaptor
External AC 100~240V to DC 9V/0.5A power supply
Plug: The pin in the center connects to the positive electrode and the outer case is negative electrode
Diameter: 5.5mm; internal diameter: 2.1mm
Electrical Specification
Accuracy is indicated as [% rdg + dgt] at 23°C ± 5°C with RH < 80%.
Specifications | Range | Resolution | Accuracy |
---|
DC milli Tesla
| 300.00 mT| 0.01 mT| ±(4% + 10dgt)
2000.0 mT| 0.1 mT
2000.1 ~
3000.0 mT
| 0.1 mT| N/A
DC Gauss
| 3000.0G| 0.1G| ±(4% + 10dgt)
20,000G| 1G
20,001 ~
30,000 G
| 1G| N/A
AC milli Tesla| 150.00mT| 0.01 mT|
±(5% + 20dgt)
1500.0mT| 0.1 mT
AC Gauss| 1500.0G| 0.1G
15,000G| 1G
Temperatures:
Range | -20~50°C / -4~122°F |
---|---|
Resolution | ±0.1°C / ±0.1°F |
Tolerance | ±1.0°C / ±1.8°F |
Maintenance or Repair
- When the When symbol is displayed on the LCD, it means that there is insufficient power; please change the battery immediately in order to ensure its accuracy.
- Do not place the meter in locations that have high temperature, humidity or that are exposed to direct sunlight.
- Remember to turn off the power after usage; remove the battery if not used for a long period of time in order to prevent battery leakage and causing damages to internal components.
- When the instrument failure, only by the authorized service provider or return the original repair.
Battery Replacement
- Turn off the power.
- Open the frame and battery cover at the back of the meter, remove the batteries.
- Insert 6 pcs of new AAA 1.5V batteries according to the battery polarity.
- Put the battery cover and frame back in place.
PRECAUTIONS
Removing Probe (With LOCK):
End of Life Disposal
| Note : This symbol indicates that the meter and its accessories must be
separated and processed properly.
---|---
Professional Electrical and Environment Test & Measurement Instruments:
Battery Capacity / Impedance Tester
LED light meter, Temperature & Humidity meter,
Infrared Thermometer, Sound level meter
Light meter, EMF meter, UV Light meter,
RF meter, Hot wire Anemometer, CO meter
Anemometer, Lan cable tester, CO2 meter,
Solar power meter, Radiation meter,
Clamp meter, Multimeter, Phase Rotation test,
Digital Insulation tester
Our products of high quality are selling well all over the world
TENMARS ELECTRONICS CO., LTD.
6F, NO.586 Ruiguang Rd, Neihu Dist.
Taipei City, Taiwan
E-mail:service@tenmars.com
http://www.tenmars.com