This procedure describes how to set the time of the GMSplus realtime clock (RTC) manually.
2. Required Tools
GMSplus
Computer, connected to Recorder Serial Console with ucon
https://www.geosig.com/files/FAQ_GMSplus_How_to_Connect_Serial_Console_with_uCon.pdf
3. Set RTC Time
In the main menu, select [H] to set RTC time
-
Enter the time in format YYMMDDHHMMSS and confirm with [Enter] The sample below is for a time of 18:45:00 on January 19th of 2024
-
When entering the time, it may happen that log messages are written into your input.
If this happens, just continue your input and confirm with [Enter]
The input may be split over several lines as below -
GMSplus will adapt its time to the new setting, it will close the old ringbuffer recording (RBF) andopen a new one with the new time setting.
This process may take a few seconds.
4. Check RTC Time
In the main menu, select [G] to see the current RTC time
-
In the main menu, select [G] to view current RTC time.
-
Current RTC Local Time is displayed in format YYYY.MM.DD
HH:MM:SS.ms. -
If new time isn’t reflected, repeat from step 3. and check
backup battery voltage. Refer to FAQ’s Maintenance Guideline for more details.
Specifications
-
Product Name: GMSplus
-
Function: Realtime Clock (RTC)
-
Manufacturer: Geosig
-
-
FAQ – Frequently Asked Questions
GMSplus Set RTC Time from Recorder Serial Console
Q: How do I set the RTC time manually on GMSplus?
A: Follow steps outlined in the user manual under “Set RTC Time” section.
Q: What should I do if the new time is not reflected?
A: Check backup battery voltage and refer to the FAQ’s Maintenance Guideline for troubleshooting.