GeoSIG VE-1x/2x Series Velocity Sensor User Manual
- June 9, 2024
- GeoSIG
Table of Contents
GeoSIG VE-1x/2x Series Velocity Sensor
Document Revision
Version| Date| Action| Prepared| Checked|
Released
---|---|---|---|---|---
1| 21.09.1995| First issue| | |
2| 21.01.1997| One document for all the GSV types| | |
3| 29.10.2001| Update technical issues| | |
4| 13.05.2003| Change model number to GSV types| | |
5| 24.09.2003| Update technical issues| | |
6| 17.02.2005| Updated naming for 1x and 2x only, TB| | |
7| 12.06.2014| New Logo and updated address| | |
8| 05.05.2015| Correction of description| | |
9| 14.09.2020| Update technical specification| | |
10| 24.02.2022| Update housing dimensions| KEC| ALB| KEC
11| 05.08.2022.| Updated chapter 2, electrical connection| ALM| ALB| KEC
12| 03.03.2022| Updated electrical connector information & photos| KEC| ALB|
VAG
Disclaimer
GeoSIG Ltd reserves the right to change the information contained in this
document without notice. While the information contained herein is assumed to
be accurate, GeoSIG Ltd assumes no responsibility for any errors or omissions.
Copyright Notice
No part of this document may be reproduced without the prior written consent
of GeoSIG Ltd. The software described in this document is furnished under
license and may only be used or copied in accordance with the terms of such a
license.
Trademark
All brand and product names mentioned are trademarks or registered trademarks
of their respective holders.
All rights reserved GeoSIG Ltd Switzerland
Warnings and Safety
The sensor housing provides no protection against explosive atmosphere. It must not be directly operated in area where explosive gases are present.
Introduction
The GeoSIG VE-1x/2x series velocity sensors consist of the following sensor types:
Frequency response 1 Hz to 315 Hz:
- VE-11 uniaxial
- VE-12 biaxial
- VE-13 triaxial
Frequency response 4.5 Hz to 315 Hz:
- VE-21 uniaxial
- VE-22 biaxial
- VE-23 triaxial
All sensor types are implemented in the same waterproof, 195 x 112 x 95 mm
cast aluminium housing. The modules inside the VE velocity sensors are 1 to 3
high-quality geophones, geophone signal amplifier, gain ranger, geophone
integrator (VE-1x only) and current loop interface for 0 to 20 mA output.
Since the modules can easily be added regarding specific user requirements,
the VE velocity sensors offer maximum versatility in obtaining the desired
performance.
Small size and single bolt fixation allow for both saving space and
installation time. Levelling is accomplished via three-point levelling screws.
Alternatively, the flanges that support levelling can be used for mounting if
desired.
Electrical Connection
The VE velocity sensors may be equipped with different connection versions:
- with 2 m cable and no connector
- with 2 m cable and a 12-pin male connector
- with a 12-pin male housing plug
- (open frame) with direct PCB connection
The connections and pin assignment are shown in Table 1 through Table 3. Based on the intended use, the 12-pin metallic-style connectors will be supplied in one of the following options: Binder Serie 623 or Binder Serie 423.
The following connectors are used in VE velocity sensors:
- Sensor housing plug, 12-pin male
- Sensor cable connector (to connect VE to GSR), 12-pin male
- Extension cable connector (to connect to VE cable connector), 12-pin female
Binder Serie 623
GeoSIG | P/N #J_CIR.012.002.F |
---|---|
Binder Serie 623 | P/N 99 4606 00 12 |
The cable gland nut is determined according to the cable’s external diameter and must be ordered separately. It must also provide the cable shield connection to connector case.
Binder Serie 423
GeoSIG | P/N #J_CIR.012.010.M |
---|---|
Binder Serie 423 | P/N 99 5629 00 12 |
The cable gland nut is determined according to the cable’s external diameter and must be ordered separately. It must also provide the cable shield connection to connector case.
Connector Pin Description
The connector pin assignment and cable colour code can be observed in the table below:
Binder Connector|
SIGNAL
|
Comment
|
Colour
---|---|---|---
Serie 623| Serie 423
Pinout| Pinout
1| A| OUTPUT X (+)| 0 V ± 10 V voltage output, 47 W output impedance| White|
2| B| OUTPUT X (-)| 0 V ± 10 V voltage output inverted, 47 W output impedance|
Brown|
3| C| OUTPUT Y (+)| 0 V ± 10 V voltage output, 47 W output impedance| Green|
4| D| OUTPUT Y (-)| 0 V ± 10 V voltage output inverted, 47 W output impedance|
Yellow|
5| E| OUTPUT Z (+)| 0 V ± 10 V voltage output, 47 W output impedance| Grey|
6| F| OUTPUT Z (-)| 0 V ± 10 V voltage output inverted, 47 W output impedance|
Pink|
7| G| TEST INPUT| Test input, output will result in a sensor step response|
Blue|
8| H| GND| Connected to Recorder’s GND| Red|
9| J| +12 VDC power| Power input, +9 to +15 VDC range| Black|
10| K| GROUND| Ground, not connected to mechanical ground| Violet|
11| L| AUX| Sensor Mode Signal| Grey/Pink| |
12| M| GROUND| Ground, not connected to mechanical ground| Red/Blue| |
The analog output voltages are referenced to 2.5 VDC. Analog Reference Voltage
available internally on the PCB represents the analog common.
Never connect the 2.5 VDC to power common (GND)
Internal PCB Pin Description
Sensor Housing Pin Assignment
Cable Configuration, Specification and Length
- The cable configuration, specification, length and quality of installation affect the quality of analog signal received, the cost of materials, and the long-term reliability of the system.
- When cabling is ordered as part of the system, GeoSIG engineers review the installation plan and the cable specifications as well as environmental conditions to assist you in achieving a reliable and cost effective installation. Following the guidelines outlined below will help further ensure your success.
- A cable must not only be able to transmit power and signals, but must also survive the environment in which it is placed. This includes chemical exposure, UV exposure, impact and cut protection, temperature extremes and any regulatory safety requirements. Because the permutations are so numerous, it is not practical to specify particular manufacturer’s cable part numbers here. However, GeoSIG engineers will work with you to help you specify an appropriate cable. The electrical parameters required for transmission of signals and power is discussed below.
- The cable construction must be an overall shielded twisted pair type for optimal protection from electromagnetic interference (EMI) sources along the path of transmission. Normally the shield can be a foil wrapper with a drain wire. However if the cable is to be installed in close proximity to high voltage power cables, an overall braided shield is additionally recommended.
- For optimum noise shield performance and maximum cable run lengths, the VE velocity sensor signals should be paired as shown on Table 4.
Pair| Wire Pair Function
---|---
1| X-Signal high and low
2| Y-Signal high and low
3| Z-Signal high and low
4| S_Test and AGND
5| V_EXT and AGND
6| S_MODE and AGND
Connect the cable shield to the local ground at the recorde
- Cables do not generate noise. However, longer cables increase the amount of the contributed noise from external sources. Cables should always be routed as far from power distribution and control wiring as possible. Again, if the cable needs to be installed close to power cables, an overall braided shield is additionally recommended.
- Cable resistance primarily determines the maximum cable length. This is not an issue related to analog signal degradation since both the signal currents and the transmission bandwidth are comparatively quite low. The main limitation is an outcome of voltage drop in the power supply due to cable resistance. The following table lists typical conductor resistance values for twisted pair shielded cables.
Diameter mm| Square mm2| Resistance
W /km
| AWG
---|---|---|---
0.25| 0.051| 371| 30
0.42| 0.14| 135| –
0.45| 0.159| 114| 25
0.51| 0.204| 93| 24
0.53| 0.22| 86| –
0.64| 0.321| 52| 22
0.80| 0.5| 39| –
0.81| 0.515| 34| 20
0.98| 0.75| 26| –
1.02| 0.817| 21| 18
1.13| 1.0| 19| –
- While selecting a cable, a maximum total resistance of 100 Ω should be taken into account and the values out of the table above must be doubled (forward and back path) to have the correct resistance value.
- Cable length may be extended or a thinner cable might be used via utilizing several conductors for the VA+ and GND signal.
Verify that voltage drop does not lower the voltage at the geophone to less than 10 VDC.
Current Draw
Sensor| Idle I [mA]| Excited I [mA]| Avrg I [mA]|
Voltage
---|---|---|---|---
VE – 11| 26.22| 27.70| 26.80| 15V
VE – 12| 26.23| 29.58| 27| 15V
VE – 13| 26.23| 31.10| 27.30| 15V
VE -21| 25.55| 65.84| 44| 15V
VE – 22| 25.55| 98.4| 50| 15V
VE – 23| 25.55| 115.2| 74| 15V
Geophones using current loop output will draw 75 mA, maximum.
Installation
The VE velocity sensors are fairly simple devices to use, but some care must
be taken in installation to be assured of proper performance. Because there
are many considerations, we recommend that before starting installation, you
review each section of this manual to ensure the best possible chance of a
simple installation that works right the first time. Prior to and after
installation we recommend that you verify functionality of the VE velocity
sensor and the cable assembly with testing the output signal and test pulse
response of the sensor. This may save time and trouble as well as give
confidence that connections are done correctly.
The location of the sensor, preferably as close as possible to the associated
recorder, should be as level and smooth as possible and the foundation should
be of concrete, rock or similar material which is perfectly bonded to the
ground or structure to be measured or monitored. Special installations such as
installing vertically on a reinforced concrete wall (i.e. vertical foundation)
are also possible, provided that the sensor is compatible with the required
orientation, the location is appropriately selected, and the sensor is
properly mounted.
Installing
The VE velocity sensor must be firmly mounted to the foundation and levelled
using the single centre pivot bolt and the three-point levelling screws as
shown in Figure 2. The “T” slot on the bottom of the sensor housing is made to
accept an M8 x 35 mm stainless steel bolt head. This centre pivot bolt is
first fastened to the foundation leaving approximately 18 – 20 mm of height
above the installation surface. The sensor, at its “T” slot, is then slipped
onto the bolt head and oriented in the proper direction. The three-point
levelling screws are then adjusted and tightened to both level the sensor and
securely fasten it to the surface.
In order to prevent any damage to the sensor housing or the fixation screw
and/or anchor, do not tighten the levelling screws using excessive force.
Use a bubble level and place it on the top of sensor surface; level first
along one axis, then the other as final levelling adjustments are made
.
Verifying Installation
Check to be sure that the sensor is aligned to produce the desired output signals as follows:
- For any axis or any combination of axes, a movement in the axis (or combination) direction indicated on the label of the sensor housing should produce a positive output signal, which can be observed on the output of the sensor (via the associated recorder).
Note that the standard orientations of the axes are in accordance with the right-hand rule:
- positive x-axis is in the direction of the thumb,
- positive y-axis is in the direction of the pointer finger
- positive z-axis in the direction of the middle finger of the right hand.
Make sure that your sensor is oriented properly for your requirements.
- Please contact GeoSIG if you wish to change any axis direction / orientation.
- You might also utilise your sensor by not fixing it to the foundation (in the case of horizontal foundations) but rather by placing it on the foundation, taking measurements and relocating for temporary or mobile measurements. However, the velocity sensors must be fixed to the ground as soon as the vibrations measured are higher than 1/3 of gravitational acceleration (> 3.27 g).
Connecting to a Recorder
The following points must be considered when connecting the VE sensor to a
recording system:
- In case of a voltage output the output range is 2.5 VDC to ± 2.5 V (i.e. 0 to 5 V range for peak to peak)
- In case of a current output the output range is 10 mA to ± 10 mA (i.e. 0 to 20 mA range for peak to peak).
Cable Configuration, Specification and Length
- The cable configuration, specification, length and quality of installation affect the quality of analog signal received, the cost of materials and the long-term reliability of the system. When cabling is ordered as part of the system, GeoSIG engineers review the installation plan and the cable specifications as well as environmental conditions to assist you in achieving a reliable and cost effective installation. Following the guidelines outlined below will help further ensure your success.
- A cable must not only be able to transmit power and signals, but must also survive the environment in which it is placed. This includes chemical exposure, UV exposure, impact and cut protection, temperature extremes and any regulatory safety requirements. Because the permutations are so numerous, it is not practical to specify particular manufacturer’s cable part numbers here. However, GeoSIG engineers will work with you to help you specify an appropriate cable. The electrical parameters required for transmission of signals and power is discussed below.
- The cable construction must be an overall shielded twisted pair type for optimal protection from electromagnetic interference (EMI) sources along the path of transmission. Normally the shield can be a foil wrapper with a drain wire. However, if the cable is to be installed in close proximity to high voltage power cables an overall braided shield is additionally recommended.
- For optimum noise shield performance and maximum cable run lengths, the VE velocity sensor signals should be paired as shown on Table 7.
Pair| Wire Pair Function
---|---
1| X-Signal high and low
2| Y-Signal high and low
3| Z-Signal high and low
4| S_Test and AGND
5| V_EXT and AGND
6| S_MODE and AGND
Connect the cable shield to the local ground at the recorder
- Cables do not generate noise. However, longer cables increase the amount of the contributed noise from external sources. Cables should always be routed as far from power distribution and control wiring as possible. Again, if the cable needs to be installed close to power cables, an overall braided shield is additionally recommended.
- Cable resistance primarily determines the maximum cable length. This is not an issue related to analog signal degradation since both the signal currents and the transmission bandwidth are comparatively quite low. The main limitation is an outcome of voltage drop in the power supply due to cable resistance. The following table lists typical conductor resistance values for twisted pair shielded cables.
Diameter mm| Square mm2| Resistance
W /km
| AWG
---|---|---|---
0.25| 0.051| 371| 30
0.42| 0.14| 135| –
0.45| 0.159| 114| 25
0.51| 0.204| 93| 24
0.53| 0.22| 86| –
0.64| 0.321| 52| 22
0.80| 0.5| 39| –
0.81| 0.515| 34| 20
0.98| 0.75| 26| –
1.02| 0.817| 21| 18
1.13| 1.0| 19| –
While selecting a cable, a maximum total resistance of 100 Ω should be taken
into account and the values out of the table above must be doubled (forward
and back path) to have the correct resistance value.
Cable length may be extended or a thinner cable might be used via utilizing
several conductors for the VA+ and GND signal. Geophones using current loop
output will draw 75 mA, maximum.
Verify that voltage drop does not lower the voltage at the geophone to less
than 10 VDC.
Operation and Configuration
Axis Orientation Configuration
The axis orientation configuration is printed on each sensor’s label that takes place on the housing. Geophones are sensitive to the gravity; therefore the VE velocity sensor has to be placed in accordance with the intended (and as purchased, i.e. horizontal or vertical) orientation.
- Do not exchange the geophone axis orientation(s) without contacting GeoSIG. Doing so will void the warranty of the instrument and might as well damage it completely.
- The polarity of the signal can be changed basically by exchanging the connections to the geophone. Also this action has to be coordinated with GeoSIG and must be executed by a skilled electrician.
Scale Factor / Gain
The standard scale factor is 100 mm/s. Other ranges are available. A gain
ranging option allows the user to select one of three scale factors by
connecting S_TEST and S_MODE to VA+ or GND. See the following table.
S-Mode | S-Test | Gain |
---|---|---|
0 | 0 | Mid Gain |
1 | 0 | High Gain |
0 | 1 | Mid Gain with sensor test |
1 | 1 | Low Gain |
Self Test
- When S_MODE is grounded and S_TEST is connected to VA+, the VE velocity sensor responds with a pulse. The pulse is positive and decays to zero after releasing the test input condition.
- This test is an excellent indication about the proper working of the sensor. Geophones might get damaged due to large shocks (e.g. by falling down to the ground). Such a damaged geophone would have developed internal contact between the moving part and have the coil fixed.
- The self-test wave perfectly shows such a defect because the response on the self-test input would be completely different than the standard.
Operation Confirmation
Confirmation of the operation is easily accomplished with a few simple
measurements, which involve checking of:
- the output voltage in normal operation,
- the power consumption,
- the wave-form of the test pulse.
Specifications
Please refer to the technical data sheet of your sensor.
Maintenance
The VE velocity sensors are sealed from the environment. As such, there is no
routine or additional maintenance required. For critical and long-term
applications, we recommend the periodic use of the self-test functionality
(see section 0, Table 9) to verify the integrity of the system and
installation.
In precision applications we recommend a calibration audit interval of 1 year.
GeoSIG can execute a calibration check, which is executed by using an
accelerometer with a hardware integrator as reference velocity sensor on a
shaker.
Theory of Operation
The VE-1x/2x contains the following electronic blocks:
- Geophone
- Standard Geophone 4.5 Hz
- Signal Conditioner
- Damping Resistor
- Amplifier
- Gain Selection (Optional)
- 3 Gains selectable 1:10:100 (standard)
- other ranges optional
- Geophone Integrator
- Makes the 1 Hz response (only VE-1x)
- Voltage to current converter (Optional-GS-320CL)
- 0 – 20 mA
Author: | Ralph Böniger |
---|---|
Checked: | Dr. Talhan Biro |
Approved: | Johannes Grob |
Distribution: | GeoSIG Ltd, Customer on request |