PureHaven ND YAG 3 in 1 Multi Functional Laser Instruction Manual
- July 26, 2024
- PureHaven
Table of Contents
ND YAG 3 in 1 Multi Functional Laser
“`html
Specifications
- Product Name: ND YAG Laser Machine
- Main Components: ND YAG Laser Handle, ESHR Insert Handle, Ice
Wave Lift Handle, Laser Goggles, Photon Goggles, Hand Pendant,
Instructions, Power Cord, Safety Tube
Product Usage Instructions
1. Product Appearance and Composition
1.1 Description of Product Appearance and Components
The machine features a sleek design and includes various
components for different treatments. The appearance is designed
with handwork aesthetics.
Note: The image in the user manual is for
reference only. Please refer to the actual product for accurate
details.
1.2 Product Packing List
- Mainframe x 1
- ND YAG Laser Handle x 1
- ESHR Insert Handle x 1
- Ice Wave Lift Handle x 1
- ESHR Insertion x 6
- Eye Mask x 1
- Funnel x 1
- Laser Goggles x 1
- Photon Goggles x 1
- Hand Pendant x 3
- Instructions x 1
- Power Cord x 1
- Safety Tube x 2
2. Product Safety
This section outlines important safety precautions for using the
machine to ensure the well-being of operators and clients.
2.1 Optical Safety
-
Avoid wearing jewelry or reflective items to prevent light
reflection. -
Use protective goggles to shield eyes from strong pulse
light. -
Avoid looking directly at the light source or its
reflections. -
Maintain distance from the light outlet during operation.
-
Ensure light emission is within safe limits.
-
Prevent excessive light energy exposure to skin.
-
Store the operating handle safely when not in use.
FAQs (Frequently Asked Questions)
Q: How do I clean the machine?
A: Use a soft, dry cloth to clean the exterior of the machine.
Do not use liquid cleaners directly on the machine.
Q: Can I use the machine on sensitive skin?
A: It is recommended to perform a patch test on a small area
before using the machine on sensitive skin.
“`
Catalogue
1. Appearance and composition of products………………………………………………………………………………………….
3 1.1. Description of appearance and components of the
product……………………………………………………. 3 1.2. Packing
list……………………………………………………………………………………………………………………………….3
2. Product safety…………………………………………………………………………………………………………………………………….4 2.1
Optical safety…………………………………………………………………………………………………………………………….4 2.2 Electrical
safety…………………………………………………………………………………………………………………………4 2.3 Fire
prevention…………………………………………………………………………………………………………………………. 5 2.4 Instrument
Safety Devices……………………………………………………………………………………………………….. 5 2.5 Safety
Identification Notes……………………………………………………………………………………………………….. 5
3. Technical parameters of products……………………………………………………………………………………………………… 5
3.1. General technical parameters…………………………………………………………………………………………………. 5 3.2.
Environmental requirements …………………………………………………………………………………………………. 6 3.3. Type
of handle ………………………………………………………………………………………………………………………..6
4. Product installation……………………………………………………………………………………………………………………………. 6
4.1. Operation of handle support hanger mounting………………………………………………………………………..
6 4.2. Operation handle installation and
replacement………………………………………………………………………..7 4.3. Water injection and replacement
of cooling water……………………………………………………………………7
5. Guidelines for product operations……………………………………………………………………………………………………… 7
5.1. Boot ………………………………………………………………………………………………………………………………………..7 5.2.
Instrument initialization…………………………………………………………………………………………………………….8 5.3.
Choice of work modalities Interface………………………………………………………………………………………… 8 5.4.
ND: YAG Laser Interface Operations……………………………………………………………………………………….8 5.5. SHR
interface Operations……………………………………………………………………………………………………….. 9 5.6.E-light
interface Operations……………………………………………………………………………………………………… 9 5.7. RF Ice Wave
Laper Interface Operations…………………………………………………………………………………9 5.8. Single Count
Zero ……………………………………………………………………………………………………………….. 10 5.9. Shut down
…………………………………………………………………………………………………………………………….10 5.10. Alarm function
…………………………………………………………………………………………………………………….10
6. Daily maintenance of products…………………………………………………………………………………………………………10
6.1. Preventive maintenance……………………………………………………………………………………………………….. 10 6.2.
Mobile and Transport Machines……………………………………………………………………………………………. 12 6.3.
Troubleshooting …………………………………………………………………………………………………………………… 12
7. Guidelines for product
troubleshooting…………………………………………………………………………………………….13 7.1. System Prompt
Language Control Table………………………………………………………………………………. 13
8. Guidelines for product application……………………………………………………………………………………………………
14 8.1. Mechanism of ESHR action ………………………………………………………………………………………………… 14 8.2.
RF Ice Wave Lap Pre-treatment preparation………………………………………………………………………… 15 8.3.
ND-YAG Laser Recommended Parameter Table…………………………………………………………………..17 8.4.
Operational matters………………………………………………………………………………………………………………. 18
1
Company Profile
Beijing Nubway S&T Co.,Ltd. was founded in 2003, headquartered in Beijing. The
company is engaged in medical ND-YAG laser, CO2 laser,808 laser, IPL, RF, E
light, frozen weight loss, and ultrasonic equipment production and sales as
one of the high-tech enterprises, the company has products covering electro-
optic Q-switched laser series, passive laser series, ultrasonic knife series,
Youli plastic weight loss series, semiconductor 808NM laser hair removal laser
series, super pulse CO2 laser series, infrared 1550 nm laser series,2940 nm
pixel laser series, pulse 1064 nm &532 nm laser series, CA ultrasonic series,
IPL (strong pulse light series), RF series, RFP series, E-light series, water
treatment series, T-finisher series, OOP series, high-frequency laser series,
no-light series, FIL series, FIL series, FIL series, FIL series, FIL series,
FIL series, At the same time, the company and the National Institute of
Optoelectronic Technology, Shandong University of Technology, the United
States Coherent and other scientific research institutes in the field of new
technology grafting cooperation, through the unremitting efforts of the people
to improve, product research and development and technology to tackle key
problems, R & D production of medical beauty products with international
first-class level, the products are widely praised and paid attention to at
home and abroad.
The company is based on the concept of mutually beneficial win-win, pragmatic
and innovative development, with reliable product quality and high-quality
after-sales service system as the backing to establish a good relationship of
mutual trust and cooperation with many enterprises and users at home and
abroad.
The company adheres to the pursuit of excellence to create the future of the
enterprise is willing to join hands with you to create a better tomorrow!
2
1. Product Appearance and Composition
1.1. Description of Product Appearance and Components
Machine appearance
Handwork Appearance
Note: The picture in the use manual is for reference only, if it does not
conform to the actual product, please refer to the actual product.
1.2. Product Packing List
Mainframe x 1
ND: YAG Laser Handle X1 ESHR insert handle x 1 Ice wave lift handle x
3
ESHR insertion x 6 Eye Mask x 1 Funnel x 1
ice electropola skin
therapy head ×3
ND: YAG Laser Handle X3
Laser goggles x 1
Photon goggles x 1
Hand pendant x 3
Instructions x 1
1 Power cord x 1 Safety tube x 2
2. Product safety
This chapter describes the safety precautions to be taken when using this
machine. Operators and technicians who maintain the machine must undergo
professional and good safety operation training, be familiar with the safety
requirements and operation steps of the machine.
2.1 Optical safety
1 Do not wear jewelry, watches or glasses and other reflectors to prevent
reflection of strong pulse light.
2 Strong pulsed light may damage the eyes and cause skin burns. The necessary
protective measures should be taken in use: all personnel should wear goggles
when using this machine, and customers should wear goggles that can completely
block the opaque material of light. Do not look directly at the high-intensity
pulsed light or its reflected light from the operating handle, even when
wearing goggles.
3 any part of the body should be away from the outlet of the operating handle
when booting. 4 The operating handle shall not be used outside the limits of
this manual, nor shall strong
pulse light be emitted beyond the operating area. 5 the closed light guide
crystal conducts the strong pulse light to the skin, and the strong
pulse light can only be emitted from the front end face of the light guide
crystal. 6 Emission of excessive light energy to the operating site may cause
skin damage. 7 The operating handle shall be placed on the hanger when not in
use, and the operating handle
shall point to the place to be operated when in use. 8 Filter crystals should
always remain clean and do not allow cold gels to penetrate into
the inside of the operating handle.
2.2 Electrical safety
1 confirm that the area power supply grid is in accordance with the local
input power supply. 2 Shut down and pull out the power plug for 10 minutes
before maintenance of the machine,
do not make the instrument in the cover plate open and unattended state,
otherwise there may be a risk of personal injury or damage to the instrument.
3 This machine has dangerous high pressure except Beijing Set sail Other than
technical personnel designated by Technology Limited, they are not detachable
Any parts. 4 Do not turn on the machine when the operation handle is leaking;
if it has been turned on, it must be shut down immediately. 5 This machine is
grounded through the ground wire in the three-core power line, and good
grounding is essential to the operation safety.
4
6 Make sure that all staff in the operation room are familiar with the control
of the machine and know how to shut it down quickly.
2.3 Fire prevention
- Please do not use this machine in an environment with explosive anesthetics or other combustible materials.
- Take precautions to avoid strong pulse light ignition operation area or
nearby flammables. 3) Use medical alcohol to clean and disinfect the parts of
the machine, must be dry before boot
use. 4) The instrument must not be left unattended. 5) The power cord should be unplugged for a long time without using the instrument.
2.4 Instrument safety devices - Leakage switch: when overcurrent or leakage, leakage switch will automatically cut off the power supply, to protect the instrument and personal safety; when the machine is not in use for a long time, please disconnect the leakage switch; the following figure “off” is disconnected state;” on “is on state;
- Key switch: instrument power switch control device, key switch and button
switch are connected on the fire line, convenient switch;
Leakage protector
Key
Button
Switch
Switch
2.5 Safety Identification Notes
- “AC220V” or “AC110V” means input power specification identification;
- “AC230V/10A” or “AC110V/15A” means fuse; 3) “Footpedal” means a foot mark;
- “WaterFall/WaterInlet” means the note/drain mark; 5) “Vents” means vent marking;
3. Technical parameters of products
3.1. General technical parameters
- Display: 2) Working mode: ND: YAG Laser, SHR, E-light, RF ice wave pimping
- Operation items: hair removal, tender skin, freckle, lifting, wrinkle
removal, whitening,
tattoo removal, 4) Type of Handle:3
5
-
Handle connection: Quick insertion
-
Light source and output: laser, strong pulse light, ice electro-bola power output
-
Filter Crystal Material: Quartz
-
Spectral range:430 nm-1200 nm
-
Spot size:50mm ×15mm
-
Light (IPL) energy density:5-12
Handle Type
Spot size
Spectral range
Fast mode Light energy
density (J/cm)2)
50 mm ×15 mm
430 nm-1200nm
5-12
-
Number of IPL pulses:1 to 10
-
Shock Protection: Type I BF
-
Input power: AC220V ±10%,50Hz ±1Hz/AC110V ±10%,60Hz ±1Hz
-
Input Power:1600W
-
Net Heavy:
-
Packing Weight:
-
Instrument Size:
-
Packing Sh ape Size:
3.2. Environmental requirements
- Working environment temperature:10oC~30oC 2) Storage ambient
temperature:0oC ~55oC 3) Working relative humidity:70% 4) Storage relative
humidity:93% 5) Atmospheric pressure:860hPa ~1060hPa 6) Environmental
requirements: Avoid direct glare 7) Cooling water: distilled or deionized
water
Note: If the indoor temperature is too low, or the instrument is not used for several days in a row, the ambient temperature must be restored to above 10oC, and then
3.3. Handle type
4. Product installation
4.1. Operation handle support hanger mounting
6
4.2. Operation Handle Installation and Replacement
Handle installation
- Hang the operation handle on the hanger; 2) Clear ND: YAG Laser handle
inserted in left fast slot, ESHR handle inserted in In the middle
slot, the RF ice wave lift handle is inserted in the right fast slot (from the front of the host); 3) Insert the handle quickly Align the interface of the machine and rotate the thread knob of the handle; 4) Tighten handle nut clockwise; 5) Operation handle installation completed.
4.3. Water injection and replacement of cooling water
Note: Use Pure water, deionized water or distilled water are strictly prohibited
Water injection operation method:
1 Confirm vent Open, if Not open, please Unscrew the cap on the vent; 2 Insert the water injection joint of the funnel into the water injection port of the machine; 3 Raise the funnel, let the funnel rise above the water line, and the water pipe bends
naturally, Can be water injection (the first water injection of 2 liters is appropriate); 4 The outlet overflows the water, indicating that the water has been filled; 5 Unplug the funnel.
How to Replace Cooling Water:
1 Confirmation Vents open if Not open, please Unscrew the cap above the vent; 2 Insert the water injection joint of the funnel into the water injection port of the machine; 3 Place the funnel in a prepared wide-mouth container of more than 2 liters. The container
should be lower than the drain outlet. Unscrew the drain Cap, drainable; 4 After the drainage is completed according to the water injection operation method can be
water injection.
5. Product operations Guide
5.1. Boot
Connect the power cord, turn on the leakage protection switch, after 30S, insert the key into the key switch, twist 90 degrees clockwise, power the instrument, the power indicator light is on, meanwhile, the instrument
7
starts to work and enters the initialization.
5.2. Instrument initialization
Turn on the instrument, the system automatically enters the initialization,
the initialization interface displays the picture version number and the
program version number, the interface progress bar is filled after the
initialization is completed, the system automatically enters Choice of work
mode Interface.
5.3. Choice of work mode Interface
Choice of work mode Interface Display Selectable Three modes of operation, system support ND: YAG Laser, SHR, E-light, RF ice wave pimping A total of four operation modes.
5.4. ND: YAG Laser Interface Operations
Energy regulation
Frequency
Water flow Temperature
Single count Multiple counts
Work
8
5.5. SHR interface Operations
Pulse Width
Energy regulation Water temperature Refrigeration level
Selection of skin
Hair
5.6. E-light interface Operations
Thickness
Frequency
Capacitor Charge Xenon lamp lighting
5.7. RF Ice Wave Laper Interface Operations
9
5.8. Single Count Zero
5.9. Shutdown
After the operation is completed, exit the main operation interface, rotate
the key 90 degrees counterclockwise, turn off the key switch, after the
instrument is closed, turn off the leakage switch and pull out the power cord.
5.10. Alarm function
ND:YAG Laser
Error 1: The water is not running (check the tank for water and the handle is
plugged in) Error 2: The temperature is too high
ESHR
Error 1: The water is not running (check the tank for water and the handle is
plugged in) Error 2: Pre-combustion failure (replace the handle to see if it
can be pre-burned, if not, please contact the manufacturer after sale) Error
3: The charge is not finished (replace the handle to see if it can be charged,
if not, please contact the manufacturer after sale) Error 4: Too high water
temperature is not suitable for treatment (stop working, turn off electricity,
cool down the machine, or change water) Error 5: The temperature sensor is not
installed or damaged. (Replug the temperature sensor terminal to check if the
line is broken)
6. Daily product maintenance
Warning: Turn off and pull out the power plug 10 minutes after the maintenance
of the machine, do not make the instrument in the cover plate open and
unattended state, otherwise there may be personal injury or damage to the
instrument risk.
6.1. Preventive maintenance
(1) Preventive maintenance by Beijing Set sail Technology Co., Ltd. Authorized
technical personnel to execute. (2) When submitting a service application,
provide the serial number or photo on the nameplate on the back of the
instrument. (3) If there is a problem, contact the representative authorized
by our company.
10
6.1.1. Routine maintenance
Warning
This machine produces high voltage and pulse light energy radiation; the inside of the instrument can only be repaired by the authorized personnel of our company.
6.1.2. Visual inspection
Check the outside of the system once a week to ensure that there are no loose cables, the system is not damaged, and the LCD screen is in good condition. If problem is found, please contact our company.
6.1.3. Instrument cleaning
(1) Clean the outside of the system upon receipt of the instrument. The surface of the crystal should be cleaned with lint-free cloth dipped in a 70% concentration of medical alcohol or medical disinfectant solution.
6.1.4. Handle cleaning and disinfection
Warning! Do not use unpasteurized handles. The use of unpasteurized handles can cause a potential risk of infection, leading to significant medical complications.
6.1.5. Crystal cleaning process
after use, clean before the crystal surface stain dry.
6.1.6. Regular preventive maintenance
This equipment needs to be tested and maintained by specialized personnel to
ensure the best operating conditions.
Routine Inspection and Maintenance Recommendations
Testing / Services
Frequency
Operator
Crystal assembly cleaning
Used up every time
Operators
Routine external cleaning and disinfection
Used up every time
Operators
Check for serious damage to the surface of the instrument
Weekly
Operators
Press the Electric Leakage Switch Elliptical Orange button
Monthly
Operators
11
6.1.7. Replacement insurance Tube
- Turn off the instrument, disconnect the leakage switch, unplug the power
cord from the power outlet; 2) will Cross screw alignment, gently force the
top, turn to the left 90 degrees to remove
the fuse; 3) Replace the new fuse, align the fuse seat, gently push the top, move 90 degrees to the
right, the fuse is replaced successfully; Note:
To ensure that the specifications of the new safety pipe are consistent with those of the original safety pipe;
Replace both safety tubes at the same time.
6.2. Mobile and Transport Machines
6.2.1. Move This machine
When moving the machine, unlock the casters pedal, hold the top handle of the instrument, and push the instrument to move. Remove the foot and power cord before moving the machine. It is strictly forbidden to drag and drop the operation handle and power cord moving instrument.
6.2.2. Transport This machine
Preparation: Unplug the power outlet Unplug the power cord Unplug the foot Contact our company about transportation Attention! Be sure to use the original packing material of the equipment when shipping. If the original
packaging material of the equipment is not used, the damage caused by the equipment will destroy the warranty clause. if need packing material or need repacking instruction, please contact our company.
6.3. Troubleshooting
Reference 7.2 Troubleshooting Guide
12
7. Product Troubleshooting Guide
7.1. System Prompt Language Control Table
7.1.1. Troubleshooting guidelines
The following table lists some possible failures and measures that users can take:
Failure phenomenon
Possible reasons
Measures taken
No boot
Sound anomaly Touch screen No response
No precombust ion
Refrigerat ion anomalies
Display code, missing line
No light
The power cord is not connected Leakage Switch Unclosed Safety pipe burns Control board failure
Check that the power cord is reliably connected Close the leakage switch and reboot. If the leakage switch is disconnected again, contact authorized maintenance personnel Replacement of safety tubes Change the control board, please contact authorized maintenance personnel
Control system failure
Turn off the power switch, reboot, if there is any trouble, please contact authorized maintenance personnel
Operation malfunction
handle
Power failure
Control board failure
Refrigeration Power Failure
Switch
Damage to refrigerating sheet
Change the operation handle, try again,
if there is any trouble, please contact
the authorized maintenance personnel
Replace power supply, please contact
authorized maintenance personnel
Change the control board, please contact
authorized maintenance personnel
Please contact authorized maintenance
personnel to replace the refrigeration
switch power supply
Change the operation handle, please
contact
authorized
maintenance
personnel
Control board failure
Change the control panel, please contact authorized maintenance personnel
Handle button damaged Operating handle glows 100,000 times Control board failure
Please contact authorized maintenance personnel Replace operation handle, total number zero Change the control board, please contact authorized maintenance personnel
13
Power failure
Replace power supply, please contact authorized maintenance personnel
Low output light energy
Operating handle glows nearly 100,000 times
Poor water quality
Increase the energy level properly and
continue to use
Replace
quality
cooling
water
immediately
Water tank shortage
Fill the tank with water
Air in the water
Confirm that the vent plug is perforated
Water sensor damaged
Please contact authorized maintenance personnel
Water alarm
Pump not activated
Please contact authorized maintenance personnel
The waterway system has a Please contact authorized maintenance
bend
personnel
Water pump damaged
Please contact authorized maintenance personnel
- If the machine fails during start-up or normal operation, stop running immediately. Unless
the water tank is short of water, other malfunctions can only be removed by the maintenance
personnel authorized by Beijing sailing Junye Technology Co. Ltd.
- If the instrument fails, please record these malfunctions and contact Beijing Set sail
maintenance department contact of technology co., ltd.
If you encounter any other trouble, please contact Beijing Set sail maintenance department
contact of technology co., ltd.
8. Product Application Guide
The three-in-one machine shall be operated by a professionally trained
operator. Beijing Nubway co., ltd. requires operators to take part in the
operation training of this machine.
8.1. Mechanism of ESHR
- Biostimulation: The photochemical effect of strong pulsed light acting on
the skin causes chemical changes in the molecular structure within the
collagen and elastic fibers of the dermis, restoring the original elasticity.
in addition, the photothermal effect produced by it can enhance the vascular
function and improve the circulation, thus achieving the effect of reducing
pores. 2) Light Selective Pyrolysis principle: due to The pigment content in
the target tissue is far Higher than normal skin tissue To the light
Absorption Also higher than normal skin tissue, use This feature enables
target tissues (red blood cells/pigmentation/hair) to be placed in Normal Skin
Organization The effect of depilation / freckle / removal of red blood was
achieved without heat damage.
8.1.1. Applicable conditions
This machine can improve wrinkles, repair vascular lesions and pigment problems and the latest expansion of hair removal technology, provides a new comprehensive beauty concept, which
14
can effectively improve facial skin condition without requiring customers to stop work and rest. Can also be used in other parts of the body, such as the neck, chest and back of the hand.
- Scope of application:
a) hair removal: limb hair excess body hair, hairline, lip hair, armpit, bikini, etc. (b) Pigmentation changes in the skin: including pigmentation, freckles, skin heterochromism, senile plaques and other skin lesions caused by sunlight or photoaging. c) For normal skin aging, large pores. d) benign vascular changes in the skin: including capillary dilatation (red blood). e) acne. - No Scope of application:
(a) Skin malignancies or precancerous lesions. (b) Pregnancy. (c) People who have recently had a tan or sunburn or intend to spend their holidays in the sun. d) photosensitive diseases (sun rash, lupus erythematosus, epilepsy). (e) Acute skin infection (herpes, frostbite). (f) Recent (within one month) use of photosensitive drugs. (g) a thin, dry skin. h) Scar constitution. (i) Diabetes, heart disease and customers with unrealistic expectations. Customers with a heart starter. k) customers with metal in their bodies (including orthodontic appliances). l) where the face has a prosthesis. (m) Skin that is in the allergic phase. Customers who have used efficacy products must stop using them for at least a month, and are not allowed to use efficacy products during strong pulse lighting operations.
8.1.2. Recommended parameters for ESHR
8.2. RF Ice Wave Lap Pre-treatment preparation
A: Please have a skin test before treatment. Do not expose the area to strong light during the week before B: treatment. Remove cosmetic or oil stains from the area to be treated and keep dry before D: treatment. Remove the metal ornaments before E: treatment. Please remove the metal decration such as rings
or necklace before the therapy.
8.2.1. Recommended parameters
The following parameters are for reference only, and the actual treatment parameters should be set according to the actual feeling of the patient.
Men
15
Projects Location
Energy settings
Pulse mode
Continuous mode
Face
forehead Eyes
12-20 8-20
15-25 10-25
Nose & cheek
15-25
20-30
Body
Arms Abdomen
17-27 20-30
20-30 20-35
Legs
18-28
23-33
Projects Location
Ms.
Energy settings
Pulse mode
Continuous mode
Face
forehead Eyes
10-18 6-18
15-25 8-24
Nose & cheek
15-20
18-28
Body
Arms Abdomen
15-25 18-28
18-28 18-32
Legs
16-26
21-31
Table 1 Recommended parameters for ice wave lift
8.2.2. Note:
Please take moisturizing or sunscreen measures after treatment. Do not wash or
bathe with hot water (temperature above question) for 3 days after treatment.
Attention should be paid to the zygomatic area, dangerous triangle area and
mandibular area during face
treatment. Make sure you set the parameters according to the patient’s actual
conditions, or follow the doctor’s
instructions.
16
8.2.3. contraindications:
Scar constitution. If you have had plastic surgery, please inform your doctor.
Patients with high blood pressure, coronary heart disease or diabetes are not
suitable for such treatment.
8.3. ND-YAG Laser Recommended Parameter Table
Ser Functi
ial
on
nu
mb
er
1
Acne
remova
l
Energy settings
57%–62%
2
Birthm
arks
55%–60%
3
Freckl
es
52%–57%
4
To get
a
tattoo
50%–55%
5
Eyebro
w
washin
g
6
Eye
wash
line
7
Lip
42%–47%
37%-42% 35%-42%
Frequency
1-3HZ
1-4HZ 2-4HZ 2-3HZ 1-3HZ 1-3HZ 1-3HZ
17
Course of treatment recommended
Flat mirror plus carbon powder
treatment, each treatment interval of 30-60 days, need
to do 2-6 times. Each treatment interval of 45-60 days, need to do 3-8 times,
spot repeat no more than 10%. Each treatment interval of 30-45 days, need to
do 2-4 times, spot repeat no more than 10%. Each treatment interval of 30-60
days, need to do 3-6 times, spot repeat no more than 10%. Each time interval
of 30-45 days, less pigment do 1-2 times, more pigment do 3-6 times.
wash
8
Attent The operation machine is not skilled, cannot wash the eye line, the eye
ion
line should have the effective protection measure.
1064nm to the blue and blue tattoo.
530nm go red, brown, brown tattoo.
Table 2 ND-YAG Laser Recommended Parameters
8.4. Operational matters
8.4.1. Communicate with customers
1 Ask the customer carefully what form of beauty he/she has received before to
determine his/her suitability for the machine.
2 Explain to the customer the discomfort and limitations that may arise in the
process of use, and dispel the unrealistic expectations of the customer.
3 The default value has been set for the convenience of the user. The default
value is a safety parameter, normally the operator does not need to adjust the
pulse parameter, can be adjusted according to the reference energy level range
(energy density) provided by our company.
8.4.2. Principle of energy level adjustment
1 Principle of one-by-one increase: prohibition of substantial adjustment; 2
Test spot principle (must close refrigeration): Before the official operation
should test
spot, and observe 5 minutes; 3 Use lower levels of complexion; 4 Lowering the
energy level to the ground; 5 proper use of cold gels.
8.4.3. Operation flow
1 Prepare cold gel, duck tongue, goggles, protective mirror, tissue paper,
record file; 2 Cleaning operation site; 3 Boot to enter the main operation of
the corresponding project interface; 4 Operating parameters are set according
to reference parameters (starting from low); 5 Operators and customers should
wear protective mirrors; 6 Apply the cold gel correctly at the operating
site;(thickness 1-2mm) 7 Test the spot (the outlet is perpendicular to the
operation site and gently pressure to
ensure the maximum area of contact), adjust the range of energy level
according to the effective characteristics and determine the parameters; 8
perform all-position operation, and lower some energy levels in the sensitive
part of the pain, the thinner part of the tissue and the darker part of the
skin, and observe the skin reaction in the whole process for timely
adjustment; 9 Clean up the operation area after operation; 10) Cold compress
for about 20 minutes;
18
- Replenish water and sunscreen. Note: hair removal operation should first
scrape off part of the hair, retain about 1.5 mm (light fine hair without
scraping), and then clean.
8.4.4. Operational Effective Features
To remove coarse hair: the skin has the process of heat-puncture-hot, the skin is slightly red; deep coarse hair capsule part has small bulge, small red papules, hair curl has paste flavor, individual hair can be easily pulled out; light fine hair capsule around slight red, part of hair curl has paste flavor. Tender skin: the skin has a warm-hot-spike-hot process, the skin is slightly red; the skin is bright and smooth, pores shrink; tight; lift. Acne blood: gray: blood coagulation.
perivascular microred: localized inflammation (2-3 days gone). Vasoconstriction: collapse of the vascular wall (the color gradually becomes lighter within 21 days).
8.4.5. Discomfort
1 There may be some pain in use. 2 Temporary redness/oedema may occur after use, usually within 3 to 7 days of detumescence. 3 There is a minor risk of producing side effects such as temporary pigmentation and changes
in skin texture. 4 may form scab shells or blisters that need to heal for 5-10 days. 5 The skin in or near the operating area may become temporarily vulnerable (do not wear makeup
or rub).
8.4.6. Points for attention after operation
1 If blister or ulceration occurs on the skin after using the machine, apply with antibiotic ointment or burn ointment.
2 Customers should use high-magnification sunscreen and hydration skin care products within one month after using this machine.
19