HF scientific MCX 28141S Peristaltic Pump Head Replacing Instructions

July 16, 2024
HF Scientific

HF scientific MCX 28141S Peristaltic Pump Head Replacing

Product Information

Specifications

  • Product Name : MCX
  • Model : Peristaltic Pump Head Replacement Kit #28141S
  • Compatibility : Compatible with MCX instrument

Product Usage Instructions

To replace the peristaltic pump on the MCX instrument

  1. Push tabs in on sides and pull up to remove the pump head from the base.
  2. Remove existing clamps and detach hoses. Take care to keep hose ends in good condition and note which hose is connected to the IN and which is connected to the OUT; these will need to be reattached to the new pump head in the same order.
  3. Based on the arrows between the pump head fittings, reattach the IN hose to the IN pump head fitting and the OUT hose to the OUT pump head fitting. Secure using the new clamps.
  4. Align the cutout of the pressure rotor with the cutout in the base.
  5. Lower the new pump head down to securely reattach it to the base.

USA
T : 239-337-2116
Toll-Free 888-203-7248
F : 239-332-7643
HFscientific.com
Latin America
T : (52) 55-4122-0138
HFscientific.com

FAQ’S

Q: How often should I replace the peristaltic pump head on the MCX instrument?
A: The frequency of replacing the peristaltic pump head may vary based on usage and maintenance. It is recommended to inspect the pump head regularly for any signs of wear or damage and replace it if necessary.

Q: Can I use a different pump head replacement kit with the MCX instrument?
A: It is recommended to use the specified Peristaltic Pump Head Replacement Kit #28141S for the MCX instrument to ensure compatibility and proper functioning.