Syphilis DAFntc2XodQ Fingerstick Whole Blood Procedure User Guide
- June 11, 2024
- Syphilis
Table of Contents
Syphilis DAFntc2XodQ Fingerstick Whole Blood Procedure
Specifications
- Procedure: Fingerstick Whole Blood Procedure for Rapid Syphilis Testing
- Materials Provided: Buffer/diluent in dropper bottle, Disposable fixed volume pipettes, Test cassettes
- Order Information: Box of 20 tests
- Reimbursement Information: CPT Code 86780QW
Product Usage Instructions
Step 1: Collect Specimen
- A. Wipe the finger with an alcohol pad.
- B. Prick the finger using a sterile lancet.
- C. Remove the first drop of blood.
- D. Collect blood using the pipette. Do not squeeze the bulb on the pipette. Allow the blood to flow into the pipette on its own. Milk the finger.
Step 2: Run the Test
- A. Add blood to the sample well by squeezing the bulb on the pipette.
- B. Slowly add four (4) drops of diluent to the sample well.
- C. Set a timer for 10 minutes.
- D. Read results no sooner than 10 minutes and no later than 15 minutes. Any line in the ‘T’ zone, even if faint, accompanied by a line in the ‘C’ zone should be interpreted as a positive result.
Note: Well will turn red. Allow blood to be absorbed.
Provided Materials
This document is to be used as a Quick Reference Guide only.
- For complete instructions, refer to the Package Insert.
- Individuals should receive proper training before using Syphilis Health Check™. Training is available at DiagnosticsDirect2U.com.
- Failure to follow the instructions may result in inaccurate test results.
USING INSTRUCTION
Step 1: Collect the specimen
- A. Wipe your finger with an alcohol pad.
- B. Prick finger using a sterile lancet.
- C. Remove the first drop of blood.
- D. Collect blood using a pipette.
DO NOT SQUEEZE THE BULB ON THE PIPETTE!
- Allow the blood to flow into the pipette on its own.
- “Milk” finger.
- Hold pipette horizontally, touch tip to sample.
Step 2: Run the test
- A. Add blood to the sample well by squeezing the bulb on the pipette.
- B. Slowly add four (4) drops of diluent to the sample well.
- C. Set a timer for 10 minutes.
- D. Read results no sooner than 10 minutes, and no later than 15 minutes.
Any line in the ‘T’ zone, even if faint, accompanied by a line in the ‘C’ zone should be interpreted as a positive result.
MORE INFORMATION
- Order Information Description: Box of 20 tests
- Reimbursement Information CPT Code: 86780QW
- Diagnostics Direct, LLC DiagnosticsDirect2U.com
- 359 96th Street Stone Harbor, NJ
- 866-358-9282
FAQ
- Q: What should I do if I see a faint line in the ‘T’ zone?
- A: Any line, even if faint, in the ‘T’ zone along with a line in the ‘C’ zone should be considered a positive result for the test.
- Q: How should I properly dispose of used materials after testing?
- `A: Please ensure proper disposal of all used materials according to biohazard waste disposal guidelines to prevent contamination and ensure safety.