CORPAK The FARRELL Valve System Instructions
- June 7, 2024
- CORPAK
Table of Contents
REORDER#: 20-4100
Total FARRELL Line Volume | 9.6 ml |
---|---|
Priming Volume* | 3.2 ml |
*Volume from in-line Y to step connector
The FARRELL® Valve System with EQUILIBRIUM TECHNOLOGY™
- For use with enteral feeding devices
- FOR ENTERAL USE ONLY, do not administer IV medications through this enteral device
- Single patient use only
- The FARRELL bag is vented to allow the escape of gas, while also providing a container to retain excess formula in a closed system
- FARRELL tubing is marked in 1/2 ml increments; accuracy: ± 1/2 ml
- Change every 24 hours
-
Hang FARRELL bag at the same height as the feeding bag/container.
-
Close WHITE clamp above FARRELL “Y” port.
-
a. Attach enteral administration set connector to “Y” port on FARRELL tubing.
b. Prime FARRELL tubing below the “Y” (3 ml).
c. Close BLUE clamp.
-
Attach FARRELL set connector to feeding tube.
-
Important: Position the FARRELL “Y” port at or below the patient’s stomach.
-
Open the BLUE clamp to establish flow, then open the WHITE clamp.
NOTE: Normal height of formula in the FARRELL tubing will be slightly above patient’s stomach level. Formula may continuously move up and down in the FARRELL tubing.
Refer to Tip #4 below for additional information on establishing flow.
WARNING: When FARRELL tubing line is open (Open WHITE clamp), the pump’s “occlude” alarm will not function, as formula will continue to flow into FARRELL bag.
MEDICATION ADMINISTRATION
- When administering medication, use the access port on the feeding tube if possible.
- Close the BLUE clamp before opening the feeding tube access port.
- Administer medication.
- Wait 510 minutes before re-opening the BLUE clamp.
ADDITIONAL TIPS AND INFORMATION
- In order for the FARRELL Valve System to work properly, the FARRELL “Y” port must be AT or BELOW the patient’s stomach.
- If a patient has a distended stomach, manually decompress the patient’s stomach with a syringe prior to the initial use of the FARRELL Valve System.
- It may be helpful to prime the FARRELL tubing slightly above the in-line “Y” port to the expected level of the fluid in the patient’s stomach.
- Patients on very low volume feeds (neonates) may require substantially longer time (6090 minutes) to establish flow before opening the WHITE clamp above the “Y” port.
- When the FARRELL Valve is working properly, formula may continuously move up and down in the FARRELL tubing.
- The FARRELL Valve continuously decompresses the patient’s stomach, but the FARRELL bag will not usually inflate with gas because it is vented.
- When administering medication, use the access port on the feeding tube if possible. Close the BLUE clamp before opening the feeding tube access port. Administer medication and wait 510 minutes before re-opening the BLUE clamp.
- If formula begins to back up into the FARRELL bag and a feeding tube occlusion is suspected, close the BLUE clamp and disconnect the FARRELL Valve System. Check for occlusions in the feeding tube. Once the occlusion is resolved reconnect the FARRELL Valve System.
Important Note:
If a large volume of fluid is in the FARRELL tubing and bag after resolving an
occlusion, adjust the head height of the FARRELL bag to prevent a free-flow
bolus of fluid into the patient.
CAUTION: While the FARRELL Valve is in operation, flow to the patient is essentially controlled by gravity. With normal (Non-FARRELL) operation, pump pressure (1218 psi) may overcome and clear occlusions or kinks in the feeding tube. With the FARRELL Valve in operation, these occlusions may not be overcome and formula will back up into the FARRELL bag. If formula begins to back up into the FARRELL bag, close the WHITE clamp located on the line above the FARRELL “Y” port and wait several minutes to determine if the pump can then overcome the occlusion. Make sure the WHITE clamp above the FARRELL “Y” port is CLOSED and the BLUE clamp below the FARRELL “Y” port is OPEN.
5769A 09-14
www.corpakU.com
www.farrellvalve.com
CORPAK, CORPAK logo, FARRELL, FARRELL logo, EQUILIBRIUM TECHNOLOGY, CORPAK
University and CORPAK U logo are either registered trademarks or trademarks of
CORPAK MedSystems, Inc. in the United States and/or other countries.
CORPAK MedSystems, Inc.
1001 Asbury Dr., Buffalo Grove, IL 60089
Phone: 1-847-403-3400
U.S. Toll Free: 1-800-323-6305
www.corpakmedsystems.com