User Manual
EPC600
Environmental Particle Counter
Introduction
Thank you for selecting the Triplett Model EPC600 Particle Counter with Camera.
The EPC600 has a Color TFT LCD display, a 74MB internal data storage bank, a
micro-SD memory card slot for capturing images and video for viewing on a PC,
and a built-in 320×240 pixel camera for capturing stills/video of test
locations. The EPC600 offers quick and accurate readings for particle count,
air temperature, most surface temperatures, and relative humidity. The EPC600
also offers a
Dew Point and Wet Bulb temperature calculation geared for energy efficiency
and environmental protection applications. Supplied PC software allows the
user to generate reports based on records data, images, and video.
This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Please visit our website (www.Triplett.com) to check for the latest version of this User Guide, Product Updates, and Customer Support.
Features
2.8” TFT Color LCD display
Convenient keystroke menu structure for measurements, setup, and another
programming 320 x 240 pixel built-in camera for still and video imaging
JPEG images and 3GP video
Meter’s internal memory 74MB.
Meter can store 5000 records (date, time, counts, humidity, temperature,
sample volumes) and 20 minutes of video to internal memory.
Micro-SD memory card slot (8GB maximum); card not included.
Selectable Test sample time, count data, and programmable delay Simultaneously
measure and display 6 channels of particle sizes (down to 0.3μm) Air
temperature and relative humidity
Dew point and Wet bulb temperature calculations MIN/MAX/AVG/DIF record with
date/time stamp Auto power off
Language selection for English, French, German, and Spanish Meter is shipped
NIST calibrated with calibration certificate Tripod mounting option
Includes universal AC adaptor/charger with a variety of plug options, USB
cable, PC software, tripod, Zero count filter, 7.4V Lithium-ion polymer
rechargeable battery, and protective case
Meter Description
1. Isokinetic probe
2. Temperature and RH protective cover that slides
3. Color TFT LCD
4. Compartment for USB
AC Adaptor/Charger
5. Function F1, F2, F3| 6. Control buttons: Page Run/Stop, and ESC
7. ON/OFF button
8. Battery and Micro-SD card (SD card slot underneath
9. Tripod mount
10. Camera lens
---|---
Operation
Power ON-OFF
Press and hold the power button to switch the instrument ON or OFF.
Important Notes
Note: Operating the particle counter with the isokinetic probe cap in
place will damage the pump and other internal components.
Note: Electrical noise, sensor leakage, or other interference may cause the EPC600 to show incorrect particle count data.
Getting Started
When the meter powers ON, the Particle Counter icon is displayed at the center of the LCD and the date, time, and battery status can be observed on the top row of the LCD. Three bottom row options are also displayed and can be selected using the Function keys (F1, F2, and F3) as shown below. These are explained in greater detail later in this user guide:
| Memory (where data, images, and video are located)
---|---
| System Setup Parameters
| Help File
Symbol Glossary
| Cumulative mode| | Video Record start
---|---|---|---
| Concentration mode| | Video mode
| Test is paused| | Color-coded particle scale (green is low, yellow is
moderate, and red is high particle count)
| CAM (Camera) mock.| | Alarm ON
| Differential mode| | Batten/Status
| Test is running| | Particle Count Setup
Sequence of Operation
To ensure that you measure the number of particles accurately, perform this
sequence of operations.
Zero the particle count sensor.
Setup your counter for proper operation. Sample time, mode, cycle, and
interval.
Run your test cycles and evaluate the results.
Basic Operation
Press and hold the ON/OFF button for 3 to 5 seconds until the display illuminates to power on the meter.
Press the enter button to get to setup mode.
Press F3 to enter settings mode.
Set each parameter to your preferred setting by pressing ENTER to enter each
parameter.
Then press ENTER to enable editing of each parameter. Press ESC to exit
parameter setup.
Sample Time: Length of time each test runs (3 to 60 seconds)
Start Delay: Duration of time from RUN to the actual start of the test.
(3 to 100 seconds)
Channel Display: Particle sizes displayed during the test.
Ambient Temp/RH: Allows the display of the air temp and %RH values
Sample Cycle: Number of test cycles to run (1 to 100)
Sample Mode: Cumulative, differential, concentration
Interval: Test cycle delay measured in seconds.
Level Indication: Particle level selection (highlighted on display)
Press ESC to return to the testing screen.
Press RUN/STOP to start the automated test sequence.
Taking Measurements
After powering the meter, press the Enter button to access the Particle
Measurement screen. Then open the temperature/RH sensor by sliding its
protective cover downward.
When measuring particle counts, these icons correspond to the (F1, F2, F3). F1
is to enter Camera and video mode, F2 is to view your saved data, and F3 is
for Setup mode.
The display will show the particle count channels with readings, air temperature (AT), Dew point temperature (DP), wet bulb temperature (WB), and Relative Humidity (%RH) as shown in the accompanying diagram.
Press the RUN/STOP button to start the air pump and the particle count
measurement process. When the programmed sampling time has elapsed, the
particle measurement cycle will automatically end and the measurement data
will be automatically stored. To set the sampling time, refer to the Particle
Count SETUP MODE section. If desired, press the RUN/STOP button and stop the
session manually.
When measuring particle counts, these icons correspond to the function keys
(F1, F2, F3). F1 is to enter Camera and video mode, F2 is to view your saved
data, and F3 is for Setup mode.
Capturing Images
While measuring particle counts, press F1 to enter the camera mode. Press F1 to capture a Still image. Then press F1 to Save the image or F3 to delete it. Press ESC to get back to exit this mode.
Capturing Video
While measuring particle counts, press F1 to enter the camera mode. Press F3
to enter the video mode. The LCD will show . Now press F2 to capture video and
press F2 again to stop the video capture.
The video file will be automatically stored. The System Setup mode section
allows the user to store video to the meter’s internal memory or to an
installed Micro-SD card. Press ESC to exit this mode.
Particle Counter Setup Mode
When measuring, these display icons are visible. Press F3 to enter the Setup mode. Use the up and down arrow keys to select an option to edit and press ENTER ( ) to access the selection.
SAMPLE TIME (Test Cycle Time)
Select Sample Time. Press the Enter button to enter the option and press Enter
again to activate the setting for editing. Use the arrow buttons to set the
sampling time/gas volume (3 to 60 seconds).
Use the Esc button to save and return to the menu.
START DELAY
The test start delay time is adjustable (3 to 100 seconds). From the Particle setup menu, use the Up and Down arrows to select Start Delay. Press the Enter button to enter the option and press Enter again to activate the setting for editing. Use the arrow keys to select the desired delay time. The test will not start until the programmed delay time has elapsed. Use the Esc button to save the setting and to return to the menu.
CHANNEL DISPLAY
Select channels that are to be displayed. From the Particle setup menu, use the Up and Down arrows to select Channel Display. Press the Enter button to enter the option. Use the arrow keys to select the channel and press the Enter key to select or de-select the channel. Use the Esc key to save the setting and to return to the menu.
AMBIENT TEMPERAURE / %RH
Enable or Disable the ambient temperature and relative humidity displays. From the Particle setup menu, use the Up and Down arrows to select Ambient Temp/RH. Press the Enter button to enter the option. Use the arrow keys to select Enabled or Disable and use the ESC button to save and return to the menu.
SAMPLE CYCLE
Set the desired number of test cycles to run. From the Particle setup menu, use the Up and Down arrows to select Sample Cycle. Press the Enter button to enter the option and press Enter again to activate the setting for editing. Use the arrow keys to select the number of cycles to run the test (1 to 100). Press ESC to save the setting and to return to the menu.
SAMPLE MODE
Select desired mode ( Cumulative, Differential, or Concentration). From the
Particle setup menu, use the Up and Down arrows to select Sample Mode. Press
the Enter ( ) button to enter the option. Use the arrow buttons to select the
desired mode and then use ESC to save and to return to the menu.
Cumulative Mode – measures all particles that are greater than or equal to the
particle size selected in the Sample Volume field.
Differential Mode – measures all particles that are greater than or equal to the particle size selected in the Sample Volume field, but less than the next largest particle size.
Concentration Mode – used to take a quick sample of airborne particulates, performed especially in areas where the particulate levels are unknown and may exceed the operating levels of this counter.
INTERVAL (TIME BETWEEN REPEATED TEST CYCLES)
Set the time between tests for tests with more than one cycle (1 to 100
seconds). From the Particle setup menu, use the Up and Down arrows to select
Interval Time. Press the Enter button to enter the option and press Enter
again to activate the setting for editing. Use the arrow buttons to set the
desired interval time. Use the ESC button to save settings and to return to
the menu.
Note: The INTERVAL setting is only for programming the time between test
cycles, not between individual measurements.
LEVEL (ALARM LIMIT) INDICATION
Select the Alarm Limit (Level) of the corresponding particle size. When the
selected particle size is exceeded, the instrument alerts the user. From the
Particle setup menu, use the Up and Down arrows to select Level Indication.
Press the Enter button to enter the option. Use the arrow buttons to select
the desired alarm limit and then press ESC to save the setting and return to
the menu
Memory Storage Browser
When the meter is switched ON, these LCD icons are visible. Press F1 to enter
the data memory section. There are three options available in the memory mode,
Picture, Video, and Particle Logs.
Use the arrow buttons to select one and then press ENTER to access the
selection.
When browsing images, data, and video use the arrow buttons to navigate, use
the Enter button to select, and use the ESC or F3 button to return to the
menu. Press ESC again to exit to the power-on screen.
System Setup Parameters
When the meter is switched ON, these LCD icons are visible . Click on the F2
key and the menu will appear. Detailed explanations are provided below for
each parameter.
Use the Arrow keys to choose the desired parameter.
DATE/TIME
Set the Date and Time. Press the Enter button to open the selection for
editing. Use the arrow keys to change a value, and use the Enter button to go
to the next item. Press the ESC button to save the setting return to the Setup
menu.
FONT COLOR
Select the color for the Display font. Press the Enter button to open the selection for editing. Use the arrow keys to select a color and use the ESC button to save the setting and return to the Setup me
LANGUAGE
Select the desired language for menu text. Press the Enter button to open the selection for editing. Use the arrow keys to select a language and use the ESC button to save the setting and return to the Setup menu.
BRIGHTNESS
level. Press the Enter
Adjust the display to the desired brightness button to open the selection for editing. Use the arrow keys to select a level and use the ESC button to save the setting and return to the Setup
AUTO POWER OFF
OFF. Press the
Select elapsed time value before meter switches
Enter button to open the selection for editing. Use the arrow keys to select
the setting and use the ESC button to save the setting and return to the Setup
menu
DISPLAY TIME-OUT
Select display auto-off delay. Press the Enter button to open the
selection for editing. Use the arrow keys to select a setting and use the ESC
button to save the setting and return to the Setup menu.
MEMORY STATUS Display the memory and the micro-SD card
status. Press the Enter
button to open the selection for editing. Use the arrow keys to select ‘Device’ or ‘SD Card’ and then view the status for the selected item. Use the ESC button to return to the Setup menu
FACTORY DEFAULT
Restore factory default settings. Press the Enter button to open the selection for editing. Use the arrow keys to select YES or NO and then use the ESC button to save the setting and return to the Setup menu.
UNITS
Select the desired unit of measure for temperature C or F. Press the Enter
button to open the selection for editing. Use the arrow keys to select the
Unit and use the ESC button to save the setting and return to the Setup menu.
Zero the Particle count Sensor
Particles must be purged (removed) from the sensor before and after each use in a high sample count environment.
Zero the sensor:
-
Unscrew and remove the Isokinetic probe and attach the Zero count filter.
-
Turn on the meter and in the Setup mode,
Set the Sample mode to Cumulative.
Set the Channel Display to ensure all channels are selected.
Set Sample time to 60 seconds
Set Sample Cycle to 10 -
Start the meter
-
Allow the meter to run until all of the particle counts read zero. You may need to run the meter multiple times to ensure that all channels read zero.
-
Turn off the meter.
-
Remove the Zero count filter and attach the Isokinetic probe.
Particle Count Table
Particle count readings, grouped by color, Green (good), Yellow (caution), and
Red (hazard), are shown for each channel.
Channel
| Green| Yellow|
Red
---|---|---|---
0.3um| 0 to 100,000| 100,001 to
250,000| 250,001 to 500,000
0.5um| 0 to 35,200| 35,201 to 87,500| 87,501 to 175,000
1.0um| 0 to 8,320| 8,321 to 20,800| 20,801 to 41,600
2.5um| 0 to 545| 546 to 1,362| 1,363 to 2,724
5.0um| 0 to 193| 194 to 483| 484 to 966
10.0um| 0 to 68| 69 to 170| 171 to 340
Note: This table is intended for reference use only
Particle Counter Considerations
Particle pollution is a mixture of various components including smoke, dust, chemicals, mold, and gases. Particle concentrations can affect our health and comfort levels. Evaluating the quality of the air by determining the concentration and size of these particles can help to determine if a problem exists. Locating the source of these various particles can help to determine effective methods of reducing these and improving Indoor Air Quality (IAQ).
One way to determine the source of air particulates in a room is to test the air outside of the room under investigation. Determine the concentration and size of particulates at the air intake to that room. Purge or Zero the particle counter and then measure the concentration and size of particles in multiple areas of the room to be tested, to determine the effectiveness of the incoming air filtration. The air temperature, relative humidity, and CO2 levels are also important considerations for IAQ as high humidity and lower temperatures can bring about mold creation.
If a room is classified as an ISO cleanroom or clean zone, the counts of
particle sizes for that class cannot be exceeded when tested.
Refer to the ISO Clean Room Class chart (ISO 14644-1:1999) for values.
The EPC600 is equipped with a still/video camera to directly see the areas under test. Data, images, and video can be stored on up to an 8GB micro-SD card (not included) or in the meter’s internal 74MB memory.
The life of the meter can be maximized by testing only when necessary and securely storing the instrument when it is not in use. Continuous use is not recommended and can shorten the life of the instrument, especially in dusty environments. Setting up a common-sense maintenance schedule and testing to the schedule will go a long way in lengthening the life of the instrument sensor.
PC Interface and Supplied PC software
This meter has the capability to connect to and communicate with a PC. A USB
cable is supplied and connects to the jack on the meter’s left side
compartment.
To install and use the software, please refer to the instructions provided on
the supplied CD-ROM and/or the instructions provided in the HELP Utility
within the software program.
Check the software download page of the website
www.triplett.com for the latest version of the PC
software and its operating system compatibility.
Battery Charging and Replacement
When the battery icon appears drained on the LCD or if the meter will not
switch ON, the battery must be recharged using the supplied AC
adaptor/charger. The charger plug connects to the jack located on the meter’s
left side compartment.
The battery compartment is located on the rear of the instrument.
The battery is a 7.4V Lithium-ion polymer rechargeable battery.
To access the meter’s battery for replacement:
- On the rear of the meter, remove the Phillips head screw that secures the battery compartment.
- Open the compartment by carefully lifting the compartment cover.
- Replace the battery with one of the same type (note that the Micro-SD card slot is located underneath the battery as explained below).
- Close the compartment and secure the compartment cover with the Phillips head screw
Battery Safety Reminders
- Please dispose of batteries responsibly; observe local, state, and national regulations.
• Never dispose of batteries in fire; batteries may explode or leak.
• Always install a new battery of the same type.
Micro-SD Card Slot (optional)
The micro-SD card slot is located inside the battery compartment underneath
the battery.
Micro-SD cards up to 8GB can be inserted in the slot.
To access the SD memory card slot:
- Turn off power to the meter and remove the AC adapter cable if connected.
- Open the battery compartment as described in the battery replacement section.
- Remove the battery.
- Insert a micro-SD card (8GB max.) in the card slot underneath the battery by lifting up on the metal cardholder and placing the Micro SD card in the connector. The pins face towards the top of the meter. Close the metal holder.
- Insert the battery and secure the compartment cover before powering on the meter.
Specifications
Particle Counter
Channels
Six channels: 0.3, 0.5, 1.0, 2.5, 5.0, 10μm
Flow Rate
0.1ft³/min (2.83L/min)
Counting Efficiency
50% at 0.3μm; 100% for particles >0.45μm
Coincidence Limit (Coincidence Loss) 5% at 2,000,000 particles per ft³
Particle Count Modes
Cumulative, Differential, and Concentration modes
Air Temperature and Relative Humidity Measurements
Air Temperature Range 32 to 122°F (0 to 50°C)
Dew Point/Wet Bulb Temperature Range 32 to 122°F (0 to 50°C)
Relative Humidity Range 0 to 100%RH
Air Temperature
Accuracy
±0.9°F (0.5°C) 50 to 104°F (10°C to 40°C)
±1.8°F (1.0°C) all other ranges
Dew Point/Wet Bulb Tempera Accuracy
±0.9°F (0.5°C) 50 to 104°F (10°C to 40°C) 1.8°F (1.0°C) all other ranges
Relative Humidity Accuracy
±3%RH from 40 to 60%RH
±3.5%RH from 20 to 40%RH and from 60 to 80%RH
±5.0%RH from 0 to 20%RH and from 80 to 100%RH
Display | 2.8” 320 x 240 pixel TFT Color LCD |
---|---|
Low battery indication | The battery symbol appears on the LCD |
Micro-SD Card Slot | Accommodates up an 8GB card |
Meter internal memory | 5000 sample records; 74MB |
Power supply | Rechargeable 7.4V Lithium-ion |
polymer battery
Battery life| Approx. 4 hours of continuous use
Battery charge time| 2 hours approx. with AC
Operating Temperature| 32 to 122 ºF (0 to 50º C)
Operating/Storage Humidity| 10 to 90% RH non-condensing
Storage Temperature| 14 to 140º F (-10 to 60º C)
Dimensions| 9.4 x 3.0 x 2.2” (240 x 75 x 57mm)
Weight| 1.26 lbs. (570g)
Triplett / Jewell Instruments extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of (1) one year from the date of purchase. This warranty does not apply to any of our products that have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as, in our sole judgment, to injure their stability or reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers altered, defaced, or removed. Accessories, including batteries, are not covered by this warranty
Copyright © 2021 Triplett
www.triplett.com