TRIPLETT MG420 Analog Megohmmeter/Insulation Tester User Manual
- June 6, 2024
- TRIPLETT
Table of Contents
TRIPLETT MG420 Analog Megohmmeter/Insulation Tester User Manual
Introduction
Congratulations on your purchase of the Triplett MG420 Analog Megohmmeter/Insulation Tester. The MG420 features three Insulation test ranges with a live circuit indicator, as well as low resistance and AC voltage measurements. The Power Lock On feature allows hands-free operation.
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present
Double insulation
Safety Notes
Do not exceed the maximum allowable input range of any function. Set the function switch OFF when the meter is not in use.
Remove the battery if meter is to be stored for longer than 60 days.
Circuits under test must be de-energized and isolated before connections are
made (except for voltage measurements).
Circuit connections must not be touched during a test. Use extreme caution
when working near bare conductors and bus bars.
Accidental contact with conductor could result in electrical shock.
After insulation tests, capacitors must be discharged.
Warnings
Set function switch to the appropriate position before measuring.
When changing ranges always disconnect the test leads from the circuit under
test.
Cautions
Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user manual before operating the meter.
Always remove the test leads before replacing the battery.
Inspect the condition of the test leads and the meter itself for any damage
before operating the meter. Repair or replace any damage before use.
Use great care when making measurements if the voltages are greater than 25VAC
rms or 35VDC. These voltages are considered a shock hazard.
Voltage checks on electrical outlets can be difficult and misleading because
of the uncertainty of connection to the recessed electrical contacts. Other
means should be used to ensure that the terminals are not “live”.
If the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired
Description
- Analog display – See Display Description below
- Batt Check key – Checks battery level (with TEST key)
- 0 ADJ key – Adjusts Zero point for Analog display
- Backlight key – Turns on Backlight
- TEST/LOCK key – Enables Battery Check, Resistance, and Megohmmeter functions (can be locked by rotating clockwise)
- Function Selector knob – Selects desired range and function
Top View
- V Test Lead Input
- COM Test Lead Input
Display Description
- Red: Megohmmeter Scale: Multiply reading by 0.5 (for 250V range), 1 (500V), 2 (1000V)
- Green: Low Resistance Scale (0 to 3 )
- Green: Resistance Scale (0 to 500
- Blue: AC Voltage Measurement Scale (0 to 600V)
- BATT GOOD Indicator
- Live Circuit Indicator LED
Operation
CAUTIONS
Observe all safety precautions when the FUNCTION switch is set to the 250V,
500V or 1000V positions.
Connect the meter test leads to the circuit under test before operating the
TEST key.
Do not touch the clip ends of the test leads when the TEST key is pressed.
Some electrical equipment, especially cables, may retain an electrical charge
when disconnected from the line. It is good practice to discharge such
equipment with grounding straps, or other suitable devices, before touching or
making connections. The meter automatically discharges the test circuits when
the spring loaded TEST key is released.
IMPORTANT NOTE
Remove all power to the circuit under test when making resistance measurements. If any voltage is present in the test circuit the red on the meter scale plate will light. Immediately disconnect test leads and turn off power to test circuit.
Live Circuit Indicator LED
If the Live Indicator status LED is lit, voltage is present on the device
under test. DO NOT proceed to test with voltage present. Shock to the user and
damage to the meter or device under test can result if testing is done with
voltage present. Proceed with testing only when the voltage is removed from
the device under test.
Connecting Test Leads
For all measurements, connect the red test lead to the V input terminal and
the black test lead to the COM input terminal.
Battery Check
- Set the range selector switch to the BATT.CHECK position
- Press and hold the TEST key
- Press the Batt Check key.
- Observe the meter pointer. If the pointer falls in the BATT GOOD range of the bottom scale, batteries are good. If the pointer falls outside of (to the left of) the BATT GOOD area, replace the batteries.
Test Lead Check
- Set the FUNCTION switch to the 3 range.
- Touch the test lead tips together and press the TEST key.
- Resistance should read less than 0.5 .
- With the leads not touching, the display should read infinity.
- Readings displayed other than the readings described above are indicative of a test lead problem. The test leads must be replaced before using the meter. Failure to do so could result in damage to equipment and electrical shock.
Megohm Measurements
- Disconnect all power from the circuit to be tested.
- Connect the red test lead to the V input terminal and the black test lead to the COM input terminal.
- Set the FUNCTION switch to the desired test range
- Connect the test leads to the circuit to be tested
CAUTION: If the LIVE CIRCUIT indicator lights up at this point, DO NOT operate the TEST key. Remove the test leads and disconnect all power from the circuit.
- Press and hold the TEST key to make the measurement. Turn the key clockwise to lock the test on.
- Read the value on the MΩ scale and apply the range multiplier to determine the Megohm resistance reading.
- Release or unlock the TEST key and allow the device to discharge before removing the test leads.
Range | Reading Multiplier |
---|---|
250V | 0.5 |
500V | 1 |
1000V | 2 |
Resistance Measurements
WARNING: Do not run this test unless the voltage on the device under test is zero. If the Live Circuit status LED lights when a test is initiated, immediately abort the test and check that the circuit under test is not powered.
- Set the Function select switch to the appropriate Resistance (3 or 500Ω) position.
- Connect the red test lead to the V input and the black lead to the COM input.
- Perform a zero adjustment using the 0 ADJ screw on the unit as described below.
- Connect the test lead tips to the circuit under test.
- Press the TEST key. Lock by rotating clockwise if desired.
- Read the resistance on the appropriate green display scale (3 or 500
- Release or unlock the Test key.
Zero Adjust
- Set the range selector switch to the 3Ω or 500Ω range.
- Short the test leads together.
- Press the Test key. Lock by rotating clockwise if desired.
- Use the 0 ADJ control to align the display pointer with the 0 point on the 3Ω or 500Ω scale.
- Release or unlock the Test key.
AC Voltage Measurements
- Set the Rotary switch to the OFF/VAC position.
- Connect the red test lead to the V input and the black lead to the COM input.
- Connect the other end of the test leads to the circuit under test.
- Read the voltage value on the meter.
Battery Replacement
- If the Battery Check indicates low battery power, the 6 AA batteries must be replaced.
- Remove the test leads and turn the meter over.
- Unsnap the tilt stand from the rear of the meter.
- Remove the 4 Phillips head screws that secure the battery cover.
- Remove the battery compartment cover
- Replace the batteries, observing correct polarity.
- Replace the battery compartment cover and secure the 4 screws.
- Re-attach the tilt stand.
You, as the end user, are legally bound (Battery ordinance) to return all used
batteries and accumulators; disposal in the household garbage is prohibited!
You can hand over your used batteries /accumulators at collection points in
your community or wherever batteries /accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the end of its lifecycle.
Specifications
General Specifications
Display
|
Multiband analog display with zero adjust
---|---
Live Circuit Indicator
| LED lights when live circuit is sensed
Power supply|
6 x 1.5V AA batteries
Power consumption
| 32mA
500mA
Fuse Protection|
600V
Operating Temperature
| 0 to 40oC (32 to 104oF)
Operating Humidity|
Below 80% RH
Altitude
| Up to 2000 meters
Storage Temperature|
-10 to 60oC (14 to 140oF) Below 70%
Storage Humidity
| RH
Dimensions|
200 x 92 x 50mm (7.9 x 3.6 x 2″)
Weight
| Approx. 700g. (24.7 oz) with batteries
CAT rating|
CATIII 1000V & CATIV 600V
RANGES
Range| Resolution| Accuracy| MIN. Open Circuit
Voltage| MIN Circuit Current
---|---|---|---|---
3Ω| 0.05Ω| +3%| 4.1V| 200mA
500Ω| 1Ω| 4.1V
AC Voltage
Range| Res.| Accuracy| Input impedance| Overload
Protect
---|---|---|---|---
600VAC| 20V| ±5%| 1.2M Ω| 1000Vrms
Megohmmeter Specifications
Range | Accuracy | Terminal voltage |
---|---|---|
100 Ω 250VDC | 0.2M Ω-5M Ω | ±5 % |
5M to 100 M Ω | ±10 % | |
200M 500VDC | 0.5M Ω-10M Ω | ±5 % |
10MM | ±10 % | 1000V + 10% ~ -0% |
Range | Test Current / Load | Short circuit Current |
--- | --- | --- |
100M 250VDC | 1m A | 250K |
200M 500VDC | 500K | |
400M 1000VDC | 1M |
Warranty
Triplett / Jewell Instruments extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of (1) one year from the date of purchase. This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as, in our sole judgment, to injure their stability or reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers altered, defaced, or removed. Accessories, including batteries are not covered by this warranty