PROPulse Irrigator Instruction Manual
- June 6, 2024
- PROPulse
Table of Contents
- INTENDED PURPOSE
- TECHNICAL DATA
- GUIDE TO SYMBOLS
- OPERATING INSTRUCTIONS
- DETAILED INFORMATION
- Cleaning Instructions
- Cleaning
- Power Adaptor
- Contraindication to irrigation
- Maintenance & Safety Inspections
- Warranty
- Environmental Protection
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
**PROPulse Irrigator Instruction Manual
**
INTENDED PURPOSE
The Propulse Ear Irrigator is intended to:
-
Facilitate the removal of cerumen and foreign bodies that are not hygroscopic from the meatus.
-
Remove discharge, keratin or debris from the external auditory meatus by irrigation with warm water.
Reasons for using this procedure are to: -
Correctly treat otitis externa where the meatus is obscured by debris.
-
Improve conduction of sound into the ear, where impacted wax is believed to be the cause of a hearing defect.
-
Examine the external auditory meatus and the tympanic membrane.
-
Remove a cause of discomfort.
This procedure should ONLY be carried out by a suitably qualified healthcare professional.
WARNINGS AND CAUTIONS
- This manual must be read and understood before the Propulse Ear Irrigator is used.
- Only suitably trained staff should use the device. Mirage can advise on the availability of training courses offered by relevant organisations.
- The Propulse QrX™ Tip is “Single Use” and should be disposed of in accordance with local . authority guidelines after use.
- Propulse Ear Irrigator must not be immersed in water.
- Only clean the device as specified in this manual (See page 10).
- If any changes in performance occur, turn off the Propulse Ear Irrigator disconnect from the mains electricity supply and DO NOT use (See page 11).
- The device has no user serviceable parts (See page 13).
- Use recommended Propulse accessories only.
- Do not use Propulse accessories with other device
- If the device is to be used for domestic visits, it is strongly recommended that a Propulse Carry Case is used to prevent damage and contamination.
- The Propulse Ear Irrigator is not user repairable and should be returned
to your Propulse supplier or Mirage Health Group (UK customers only) for service and/or
repair. It is recommended that the Propulse Ear Irrigator is serviced annually
Please note: Damage caused to your Propulse Ear Irrigator by the use of accessories, consumables or service agents not recommended by Mirage Health Group, will invalidate your warranty.
DO NOT IRRIGATE the ears if:
- Previous complications occurred following this procedure.
- There is a history of a middle ear infection in the last six weeks.
- The patient has undergone ear surgery (apart from grommets that have extruded at least 18 months previously and the patient has been discharged from the ENT dept.)
- The patient has a perforation or there is a history of a mucous discharge in the last year.
- The patient has a cleft palate (repaired or not).
- In the presence of acute otitis externa; an oedematous ear canal combined with pain and tenderness of the pinna.
- If patient complains of pain.
STOP IMMEDIATELY
Please refer to Contraindications on page 12.
COMPONENT / PARTS INDENTIFICATION
- Reservoir
- Lid
- QrX™ tip
- Handle and Hose
- Handle holder
- Waterflow/Pressure control switch
- On/Off switch
- Footswitch
- Mains power adaptor
The Propulse Ear Irrigator consists of:
- The main unit and the following user controls:
- An On/Off switch
- A footswitch which (when pressed) starts the flow of water. The water stops when the footswitch is released.
- A mains power adaptor
- Water container/reservoir (1) is removable to facilitate filling and cleaning. Horizontal line indicates the correct level of water required for normal use, as well as the correct level of water required to dissolve cleaning tablet.
- Mushroom Valve – to retain the water in the reservoir when it is removed from the Propulse machine.
- Handle and non-detachable hose. The Handle accommodates the Propulse QrX™ Single Use Tips.
- The footswitch – is connected to the main body via a jack plug/socket connection. The device will only function if the footswitch is connected.
Please note: Residual water in the handle and hose will continue to flow if the handle is not held in the vertical position or, if the handle is held in a position that is lower than the machine. To prevent residual flow, it is recommended that the handle be returned to its holder on the machine.
TECHNICAL DATA
Performance:
Flow rate| Up to 300ml/minute
Water jet pulses| 1200 per minute (approximately)
Maximum operating time:| 10 minutes continuous use (with a recommended rest
time of 2 hours)
Storage temperature range:| -5°C to 65°C
Storage relative humidity:| up to 80%
Power adaptor:| Input 100-240v ~ 50/60Hz Max 0.45A Output 9v DC2A
Electrical Safety:| EN6061-1
EMC Compliance:| EN60601-1-2
GUIDE TO SYMBOLS
Attention – Consult Accompanying Documents
Type BF Electrical safety
Protected against water droplets
Single use item
Conforms to the Medical Device Directive 93/42/EEC
Must be disposed of in accordance with European waste electrical and electronic equipment directive 2002/96/EC
Manufactured by
Power On
Power Of
DC Current
Variable Flow Rate
Duty Cycle
Read Instruction Manual
Indoor use only
OPERATING INSTRUCTIONS
These instructions are for general use. When required, refer to the detailed information in the second half of this manual.
- Ensure that ONLY suitably trained clinicians operate the device.
- Ensure that the warnings and cautions are observed.
- Ensure that the patient exhibits no contraindications (refer to page 12).
- Ensure the unit has been cleaned prior to first use , and every day prior to use (refer to page 10 for detailed cleaning guidance).
- The device may only be operated whilst connected to the mains electrical supply using the Power Adaptor and footswitch provided.
- The reservoir should be removed prior to filling.
- The water reservoir must be filled to the horizontal line on the front of the reservoir.
- Monitor temperature regularly to ensure that patient comfort and safety is maintained. Refill as necessary.
- Fit a new Propulse QrX™ Single Use Tip to the Handle.
- Adjust the Waterflow Switch (6) to an appropriate level.
- Turn the On/Off Switch (7) to the position marked “I”.
- Direct the irrigator tip into the noots tank and switch on the machine for 10-20 seconds in order to circulate the water through the system and eliminate any trapped air or cold water.
- Ensure the water is warm before presentation to the patient.
- During treatment you can pause the flow by releasing the footswitch.
- After treatment empty the reservoir and operate the device to purge any residual water.
- Remove the Propulse QrX™ Tip and dispose of in accordance with local authority guidelines. ®
- Turn off the On/Off switch after use and disconnect from the power supply.
- Clean the Propulse Ear Irrigator unit ® every morning prior to use, with a Propulse cleaning tablet (See page 10).
- The Propulse Ear Irrigator should only be transported in a Propulse approved carry case toprevent damage or contamination.
- If any changes in performance occur, turn off the Propulse Ear Irrigator, disconnect Adjust the waterflow to an appropriate value that is compatable with the treatment requirements and patient comfort. from the mains electricity supply and DO NOT use. Please refer the machine to Mirage.
DETAILED INFORMATION
Fitting the Footswitch
The footswitch is connected to the main unit by a socket in the side of the device. The Propulse Propulse Ear Irrigator will NOT operate unless the footswitch is connected.
Filling the water reservoir
It is recommended that:
- The water reservoir is removed from the device for filling and that the lid is always in place when the water container is in-situ on the device.
- The water reservoir should be filled to the horizontal line on the front. This helps to eliminate the risk of spillage.
- Water at 40°C is recommended. Higher temperatures increase the risk of scalding and burns to the patient. Lower temperatures increase the risk of patient discomfort and dizziness.
Fitting the Propulse QrX™ Tip
The Propulse Ear Irrigator is designed to be used only with Propulse QrX™ Single Use Tips. Use one Propulse QrX™ Tip per treatment.
To fit a Propulse QrX™ Tip
- Remove Tip from packaging – Tips are non-sterile.
- Push Tip into Handle until a click can be heard.
To remove a Propulse QrX™ Tip
- Retract the QrX™ Locking Collar using a thumb.
- Grasp the used QrX™ Tip between forefinger and thumb and pull gently from the QrX™ Handle.
- Dispose of the used Tip in accordance with local authority guidelines. DO NOT REUSE TIPS.
Propulse QrX™ Tips are available to purchase in boxes of 100 individually wrapped (non-sterile) tips from your normal Propulse supplier or from Mirage directly (UK only). Propulse QrX™ Tips are clearly branded with the Propulse logo on the tip and its packaging. Only branded Propulse QrX™ Tips should be used with the Propulse Ear Irrigator.
Replacing the Mushroom Valve
The Mushroom Valve is specifically designed to prevent water flowing out of the reservoir whilst filling. Should a replacement Mushroom Valve be required, please follow the steps below .
Firstly ensure you are Fitting the correct mushroom valve for this model Propulse Ear Irrigator This will help prevent possible damage to the water inlet valve.
- Remove the reservoir from the Propulse Ear Irrigator.
- Remove the old Mushroom Valve from the reservoir and discard
- Insert a new Mushroom Valve without bending or exerting undue force on the legs of the Mushroom Valve, into the reservoir.
- Check the condition of the ‘O’ ring on the base of the reservoir and if worn replace with a new unit specific ‘O’ ring.
- Return the reservoir to the machine.
Guidance and Manufacturer’s Declaration – Electromagnetic emissions
The Propulse is intended for use in the electromagnetic environment specified below. The customer
or user of the Propulse should assure that it is used in such an environment.
Emission tests| Compliance| Electromagnetic environment –
guidance
RF emissions CISPR 11| Group 1| The Propulse uses only energy for its internal
function. Therefore, its RF emissions are not likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR 11| Class A| The Propulse is suitable for use in all
establishments other than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuation/flicker emissions IEC 61000-3-3| Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Propulse is intended for use in the electromagnetic environment specified below. The customer or the user of the Propulse should assure that it is used in such an environment
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic
environment guidance
Electrostatic discharge (ESD) IEC 61000-4-2| 6 kV contact 8 kV air| 6 kV
contact 8 kV air| Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity should be at least
30%.
Electrical fast transient/ burst IEC 61000-4-4| 2 kV for power supply lines 1
kV for input/output lines| 2 kV for power supply lines Not applicable| Mains
power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5| 1 kV differential mode 2 kV common mode| 1 kV
differential mode 2 kV common mode| Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input
lines IEC 61000-4-11| 5 % UT (> 95 %dip in UT) for 0.5 cycle 40 % UT (60%
dip in UT) for 5 cycles 70 % UT (30% dip in UT) for 25 cycles< 5 % UT (> 95
%dip in UT) for 5 s| < 5 % UT (> 95 %dip in UT) for 0.5 cycle 40 % UT (60% dip
in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles< 5 % UT (> 95 % dip
in UT) for 5 s| Mains power quality should be that of a typical commercial or
hospital environment. If the user of the Propulse requires continued operation
during power mains interruptions, it is recommended that the Propulse be
powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8| 3 A/m| Not
applicable| Power frequency magnetic fields should be at levels characteristic
of a typical location in a typical commercial or hospital environment.
NOTE UT is the A.C. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Propulse is intended for the use in the electromagnetic environment specified below.
The customer or the user of the Propulse should assure that it is used in such an environment
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic
environment – guidance
| | | Portable and mobile RF communications equipment should be used no closer
to any part of the Propulse, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter:
Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3| 3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to 2,5 GHz| 3 Vrms 3 V/m| Recommended protection distance:
d = 1.17 vP d = 1.17 vP for 80 MHz to 800 MHz
| | | d = 2,3 vP for 800 MHz to 2.5 GHz
| | |
where P is the maximum output power rating of the transmitter in watts (W) acc. To the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol:
| | |
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflections from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Propulse is used exceeds
the applicable RF compliance level above, the Propulse should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the Propulse. b.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
Recommended Separation Distance between portable and mobile RF
Communications Equipment and the Propulse
The Propulse is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
Propulse can help prevent electromagnetic interference by maintaining minimum
distance between portable and mobile RF communications equipment
(transmitters) and the Propulse as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output power of transmitter (W)| Separation distance according
to the transmit frequency (m)
150 kHz to 80 MHz d = 1.17 v P| 80 MHz to 800 MHz d = 1.17 v P|
800 MHz to 2,5 GHz d = 2.33 v P
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where p is the
maximum output power rating of the transmitter in watt (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Cleaning Instructions
The importance of using the correct strength cleaning solution cannot be overstated. A solution that is too strong will in time damage the Propulse Ear Irrigator. A solution that is too weak will fail to provide the correct level of cleaning and decontamination. Mirage Health Group recommends the use of Propulse CHLOR-CLEAN Tablets. They are easy and effective to use and provide a measured / specific strength of cleaning solution that is safe and kind to the Propulse internal components.
Ensure the unit has been cleaned prior to first use.
- Place warm tap water into the reservoir up to the horizontal line on the front.
- Place one Propulse CHLOR-CLEAN tablet into the reservoir and allow it to dissolve completely.
- Once dissolved run the machine until the cleaning solution leaves the handle. This ensures that the cleaning solution has reached all of the internal components.
- Leave the solution in place for 10 minutes.
- After 10 minutes remove the reservoir with the remaining cleaning solution and discard.
- Fill the reservoir with clean, well run, cold tap water and return to the Propulse.
- Run the Propulse ensuring that all remaining cleaning solution has been flushed through.
- Remove reservoir, discard water and dry reservoir thoroughly with a paper towel.
- Return reservoir to the Propulse – it is now ready for use.
Cleaning
Do not atiempt to clean the Propulse QrX™ Tip. Use one Propulse QrX™ Tip per patient treatment and discard to clinical waste after use as this reduces the risk of cross infection between patients.
External cleaning of the Propulse Ear Irrigator should be done by hand, wiping with a damp cloth only. Apply liquids to the cloth not the unit. Do not immerse the unit in water. Mild detergents and disinfectants may be used externally.
Power Adaptor
Connect the outlet lead of the Power Adaptor to the Power Adaptor socket marked on the end of the product and to the mains electrical supply. Make sure the cord and Power Adaptor are positioned so that they will not be subjected to damage or stress or present a trip hazard.
Only use a Propulse branded Power Adaptor.
To reduce the risk of electric shock, unplug the unit from the power source before attempting to clean it externally
The Power Adaptor must not be used outdoors or in damp areas.
The Propulse Power Adaptor will have been supplied with an appropriate plug for your region OR a selection of international plugs. Please fit the appropriate plug for your region. If there are any problems connecting to the mains electrical supply consult a qualified electrician.
Contraindication to irrigation
If the patient has experienced any complications from a previous episode of
irrigation with water.| If the patient did not tolerate a previous episode of
irrigation it would be unwise to repeat the procedure in case the symptoms are
exacerbated.
---|---
There has been evidence of a middle ear infection (otitis Media) in the last 2
months| The tympanic membrane may be vulnerable to damage due to the adverse
effect infected fluid may have on the ear drum
The patient has undergone any form of ear surgery apart from grommets, which
are documented to be extruded from the tympanic membrane for over 2 years and
the patient is discharged from the ENT department.| There will be a weakness
to the structure of the ear canal and tympanic membrane after surgery. This
does not include cosmetic surgery to the pinna (for example repair of bat
ears). If the tympanic membrane is intact 2 years post grommet extrusion,
there should not be an increased risk of damage to the tympanic membrane
There is a suspected or actual perforation present or there is a history of
mucous discharge from the ear in the last 2 years| A mucous discharge would
indicate a perforation and water entry under pressure could cause infection or
damage the delicate middle ear structures.
If the patient has a cleft palate (regardless of whether it has been repaired
or not).| A cleft palate indicates an underdeveloped facial skeleton and as
such the tympanic membrane and middle ear structures could be more vulnerable
to damage
In the presence of acute otitis externa (pain, swollen ear canal and
tenderness of the pinna).| Although it is essential to thoroughly clean the
infected ear canal, when it is swollen debris should be removed by
microsuction.
Profound hearing loss in one ear.| There is a risk associated with any
intervention and when a patient completely relies on one ear for hearing (as
the other ear has a profound hearing loss) any risk to this ear is
unacceptable
Caution when irrigating with water in the following groups of patients
Patient is taking anti-coagulants| The lining of the ear canal is delicate and
there is a higher risk of bleeding so ensure trauma to the ear canal is
avoided.
The patient is diabetic.| The pH of wax in patients who are diabetic is a
higher pH than average, increasing their vulnerability to infection.
Tinnitus.| Although wax impaction can cause tinnitus, trauma to the tympanic
membrane may exacerbate this.
Vertigo.| This is also a symptom of wax impaction but irrigation can trigger
an episode so ensure appropriate water temperature and patient safety.
Radiotherapy that has involved the ear canal.| A radiated ear canal can
develop bony necrosis so wax should be removed before it becomes hard and
trauma to the canal should be avoided.
Maintenance & Safety Inspections
To ensure optimum performance the Propulse Ear Irrigator should be serviced
every 12 months. Service or repairs conducted by unauthorised
agencies/organisations invalidate any or implied
warranties from Mirage.
The Propulse Ear Irrigator should under-go routine electrical safety testing to ensure that it remains safe to use, in accordance with EN ISO 62353:2014.
Users of the Propulse Ear Irrigator should carry out regular inspections to ensure that the handle and hose, power adaptor and cable, reservoir, footswitch and main body of the machine are free from damage prior to use. If any damage is evident, the Propulse Ear Irrigator should NOT be used until replacement parts have been fitted.
Only Propulse branded items should be used with the Propulse Ear Irrigator.
The Propulse Ear Irrigator is not user repairable and should be returned to your Propulse supplier or Mirage Health Group (UK customers only) for service and/or repair:
Mirage Health Group Service Centre
11 Tewin Court, Welwyn Garden City,
Hertiordshire
AL7 1AU
UK
Tel – +44 (0) 845 130 5445
The clinical procedures relating to the use of ear irrigators can be found on the following websites:
www.earcarecentre.com
www.entnursing.com/earcare.htm
Mirage is not responsible for the content or maintenance of third party internet sites. Mirage can also advise on the availability of training courses offered by the relevant organisations. Additional information on use of the Propulse can be found at: httip://www.youtube.com/user/MirageHealthGroup
Warranty
The Propulse Ear Irrigator carries a twelve month warranty (*subject to conditions) from the date of original purchase. Should any defect arise due to faulty material or workmanship, Mirage Health Group will, upon receipt of the faulty Propulse Ear Irrigator, proof of purchase, information relating to the nature of the fault and details of where the item was purchased, rectify the fault at no cost to you.
Should any of the “Accessory” items (listed below) prove to be faulty as a result of defective material or workmanship, Mirage Health Group will rectify the issue free of charge upon receipt of the faulty accessory (subject to conditions)
“Accessory” items are: Footswitch; Reservoir / Tank and Lid; Mushroom Valve
and Washer; QrX™
Tip; Power Supply Lead and Power Transformer.
Conditions of Warranty (applicable to Propulse Electronic Ear Irrigator and
“Accessory” items).
The warranty does not cover:
- Accidental damage or damage caused by misuse.
- Faults caused due to lack of maintenance.
- Damage caused by using the Propulse Ear Irrigator for any use other than its intended use.
- Damage caused as a result of repair by any unauthorised agents – ONLY Mirage Health
- Group should undertake repairs.
- Damage caused by the use of accessories / cleaning products that have not been recommended by Mirage Health Group as suitable for your model irrigator.
This warranty is in addition to, and does not diminish your statutory or legal rights.
Additional user manuals and other accessories are available from Mirage Health Group Ltd at:
Mirage Health Group
11 Tewin Court, Welwyn Garden City,
Hertfordshire AL7 1AU UK
Tel – +44 (0) 845 130 5440
Fax – +44 (0) 845 130 6440
uksales@miragehealthgroup.com
internationalsales@miragehealthgroup.com
Environmental Protection
This symbol on the products and/or accompanying documents means that used electrical and electronic products should not be mixed with general waste. Please return to Mirage Health Group or dispose of through locally approved disposal service for electronic equipment. Penalties may be applicable for incorrect disposal of this waste, in accordance with national legislation.
Disposing of used Propulse® QrX™ Tips
Disposal should be in accordance with local authority guidelines and regulations for the disposal of clinical waste. Propulse® QrX™ Tips should not be disposed of in municipal waste.
Transportation
Before the Propulse Ear Irrigator is transported, the reservoir must be emptied and the machine should be operated until the handle and hose are empty of liquid. The reservoir should then be dried using a paper towel.
For safe transportation of Propulse® Ear Irrigator, Mirage recommends that the Propulse Carry Case is used to prevent damage or contamination. For internal cleaning.