microlife BP W3 Comfort Wrist Blood Pressure Monitor Instruction Manual

June 6, 2024
microlife

microlife - Logo

BP W3 Comfort
Wrist Blood Pressure Monitor

microlife BP W3 Comfort Wrist Blood Pressure Monitor -
Cover

IB BP W3 Comfort E-V11 1521
Revision Date: 2021-03-24

1. ON/OFF button
2. Display
3. M-button (memory)
4. Time Button
5. Battery Compartment
6. Cuff
Display
7. Systolic Value
8. Diastolic Value
9. Pulse Rate
10. Date/Time| 11. Stored Value
12. Traffic Light Indicator
13. Irregular Heartbeat (IHB) Symbol
14. Cuff Signal Indicator «ERR 1»
15. Cuff Fit Check
-A: Suboptimal Cuff Fit
-B: Arm Movement Indicator «ERR 2»
-C: Cuff Pressure Check «ERR 3»
16. Pulse Indicator
17. Battery Display
---|---
| Read the instructions carefully before using this device.
---|---
| Type BF applied part
| Keep dry

Intended use:
This oscillometric blood pressure monitor is intended for measuring non- invasive blood pressure in people aged 12 years or older. Dear Customer, This device was developed in collaboration with physicians and clinical tests carried out prove its measurement accuracy to be of a very high standard.* If you have any questions, problems or want to order spare parts please contact your local Microlife-Customer Service. Your dealer or pharmacy will be able to give you the address of the Microlife dealer in your country. Alternatively, visit the internet at www.microlife.com where you will find a wealth of invaluable information on our products. Stay healthy ­ Microlife AG!

  • This device is tested according to the ESH protocol and ISO81060-2:2013.

Using the Device for the First Time

Inserting the batteries
After you have unpacked your device, first insert the batteries. The battery compartment 5 is on the bottom of the device. Insert the batteries (2 x 1.5 V, size AAA), thereby observing the indicated polarity.

Setting the date and time

  1. After the new batteries are fitted, the year number flashes in the display. You can set the year by pressing the M-button 3. To confirm and then set the month, press the time button 4.
  2. Press the M-button to set the month. Press the time button to confirm and then set the day.
  3. Follow the instructions above to set the day, hour and minutes.
  4. Once you have set the minutes and pressed the time button, the date and time are set and the time is displayed.
  5. If you want to change the date and time, press and hold the time button down for approx. 3 seconds until the year number starts to flash. Now you can enter the new values as described above.

Checklist for Taking a Reliable Measurement

  1. Avoid activity, eating or smoking immediately before the measurement.
  2. Sit down on a back-supported chair and relax for 5 minutes. Keep the feet flat on the floor and do not cross your legs.
  3. Always measure in a sitting position and on the same wrist. Use the wrist which usually shows higher blood pressure values.
  4. Remove any items of clothing and your watch, for example, so that your wrist is free.
  5. Always ensure that the cuff is positioned correctly, as shown in the pictures illustrated on the short instruction card.
  6. Fit the cuff comfortably but not too tight. The cuff will cover a wrist circumference of 13.5 – 23 cm (5.25 – 9.05 inches).
  7. Support your arm in a relaxed position and ensure that the device is at the same height as your heart.

Taking a Blood Pressure Measurement

  1. Press the ON/OFF button 1 to start the measurement.
  2. The cuff fit check 15 on the display indicates that the cuff is perfectly placed. If the icon 15 -A appears, the cuff is fitted suboptimally, but it is still ok to measure.
  3. The cuff will now pump up automatically. Relax, do not move, and do not tense your arm muscles until the measurement result is displayed. Breathe normally and do not talk.
  4. The measurement is performed during the inflation. The inflation speed may vary, this is a normal occurrence.
  5. During the measurement, the pulse indicator 16 flashes in the display.
  6. The result, comprising the systolic 7 and the diastolic 8 blood pressure and the pulse rate 9 is displayed. Note also the explanations on further display symbols in this booklet.
  7. Remove and switch off the monitor and enter the result in the enclosed blood pressure pass. (The monitor does switch off automatically after approx. 1 min.).

You can stop the measurement at any time by pressing the ON/OFF button (e.g. if you feel uneasy or an unpleasant pressure sensation).

How not to store a reading
As soon as the reading is displayed press and hold the ON/OFF button 1 until «M» 11 is flashing. Confirm to delete the reading by pressing the time button 4.
«CL» is displayed when the reading is deleted from the memory successfully.

How do I evaluate my blood pressure?
The triangle on the left-hand edge of the display AL points at the range within which the measured blood pressure value lies. The value is either within the optimum (white), elevated (hatched gray) or high (black) range. The classification corresponds to the following ranges defined by international guidelines (ESH, ESC, JSH). Data in mmHg.

Range Systolic Diastolic Recommendation
1. blood pressure too high ?135 ?85
2. blood pressure elevated 130 – 134 80 – 84
3. blood pressure normal <130 <80

The higher value is the one that determines the evaluation. Example: a blood pressure value of 140/80 mmHg or a value of 130/90 mmHg indicates «blood pressure too high».

Appearance of the Irregular Heartbeat (IHB) Symbol
This symbol 13 indicates that an irregular heartbeat was detected. In this case, the measured blood pressure may deviate from your actual blood pressure values. It is recommended to repeat the measurement.

Information for the doctor in case of repeated appearance of the IHB symbol:
This device is an oscillometric blood pressure monitor that also measures the pulse during blood pressure measurement and indicates when the heart rate is irregular.

Data Memory

This device automatically stores the last 30 measurement values.

Viewing the stored values
Press the M-button 3 briefly, when the device is switched off. The display first shows «M» AK and «A», which stands for the average of all stored values.
Pressing the M-button again displays the previous value. Pressing the M-button repeatedly enables you to move from one stored value to another.

Blood pressure readings with suboptimal cuff fit AO-A are not considered in the average value.
Memory full
Pay attention that the maximum memory capacity of 30 memories is not exceeded. When the 30 memory is full, the oldest value is automatically overwritten with the 31st value. Values should be evaluated by a doctor before the memory capacity is reached ­ otherwise data will be lost.

Clearing all values
If you are sure that you want to permanently remove all stored values, hold down the M-button (the device must have been switched off beforehand) until «CL ALL» appears and then release the button. To permanently clear the memory, press the time button while «CL
ALL» is flashing. Individual values cannot be cleared.

Cancel deletion: press ON/OFF button 1 while «CL ALL» is flashing.

Battery Indicator and Battery change

Low battery
When the batteries are approximately ¾ empty the battery symbol AQ will flash as soon as the device is switched on (partly filled battery displayed). Although the device will continue to measure reliably, you should obtain replacement batteries.

Flat battery ­ replacement
When the batteries are flat, the battery symbol 17 will flash as soon as the device is switched on (flat battery displayed). You cannot take any further measurements and must replace the batteries.

  1. Open the battery compartment 5 by pulling off the cap.
  2. Replace the batteries ­ ensure correct polarity as shown by the symbols in the compartment.
  3. To set date and time, follow the procedure described in «Section 1.».

The memory retains all values although date and time must be reset ­ the year number therefore flashes automatically after the batteries are replaced.

Which batteries and which procedure?
Use 2 new, long-life 1.5 V, size AAA alkaline batteries.
Do not use batteries beyond their date of expiry.
  Remove batteries if the device is not going to be used for a prolonged period.

Using rechargeable batteries
You can also operate this device using rechargeable batteries.

Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat battery symbol appears. They should not remain inside the device as they may become damaged (total discharge as a result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor. Recharge batteries in an external charger and observe the information regarding charging, care and durability.

Error Messages

If an error occurs during the measurement, the measurement is interrupted and an error message, e.g. «ERR 3», is displayed.

Error Description Potential cause and remedy
«ERR 1» 14 Signal too weak The pulse signals on the cuff are too weak.

Re-
position the cuff and repeat the measurement.
«ERR 2» 15-B| Error signal| During the measurement, error signals were detected by the cuff, caused for instance by movement or muscle tension. Repeat the measurement, keeping your arm still.
«ERR 3» 15-C| No pressure in the cuff| An adequate pressure cannot be generated in the cuff. A leak may have occurred. Check that the cuff is correctly connected and is not too loose. Replace the batteries if necessary. Repeat the measurement.
«ERR 5»| Abnormal result| The measuring signals are inaccurate and no result can therefore be displayed. Read through the checklist for performing reliable measurements and then repeat the measurement.

« HI»| Pulse or cuff pressure too high| The pressure in the cuff is too high (over 299 mmHg) OR the pulse is too high (over 200 beats per minute). Relax for 5 minutes and repeat the measurement.
«LO»| Pulse too low| The pulse is too low (less than 40 beats per minute). Repeat the measurement.

  • Please immediately consult your doctor, if this or any other problem occurs repeatedly.
    If you think the results are unusual, please read through the information in «Section 1.» carefully.

Safety, Care, Accuracy Test and Disposal

Safety and protection

  • Follow instructions for use. This document provides important product operation and safety information regarding this device. Please read this document thoroughly before using the device and keep for future reference.
  • This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by incorrect application.
  • This device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in the «Technical Specifications» section.
  • Protect it from:
    • water and moisture
    • extreme temperatures
    • impact and dropping
    • contamination and dust
    • direct sunlight – heat and cold
  • The cuffs are sensitive and must be handled with care.
  • Only pump up the cuff once fitted.
  • Do not use this device close to strong electromagnetic fields such as mobile telephones or radio installations. Keep a minimum distance of 3.3 m from such devices when using this device.
  • Do not use this device if you think it is damaged or notice anything unusual.
  • Never open this device.
  • If the device is not going to be used for a prolonged period the batteries should be removed.
  • Read the additional safety information provided within the individual sections of this instruction manual.
  • If you suffer from a cardiac arrhythmia consult with your doctor before using the device. See also chapter «Appearance of the Irregular Heartbeat (IHB) Symbol» of this user manual.
  • The measurement results given by this device is not a diagnosis. It is not replacing the need for the consultation of a physician, especially if not matching the patient’s symptoms. Do not rely on the measurement result only, always consider other potentially occurring symptoms and the patient’s feedback. Calling a doctor or an ambulance is advised if needed.

Ensure that children do not use this device unsupervised; some parts are small enough to be swallowed. Be aware of the risk of strangulation in case this device is supplied with cables or tubes.

Device care
Clean the device only with a soft, dry cloth.

Cleaning the cuff
Carefully remove any marks on the cuff with a damp cloth and mild detergent.

Accuracy test
We recommend this device is tested for accuracy every 2 years or after mechanical impact (e.g. being dropped). Please contact your local Microlife- Service to arrange the test (see foreword).

Disposal
Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations, not with domestic waste.

Guarantee

This device is covered by a 5 year guarantee from the date of purchase. During this guarantee period, at our discretion, Microlife will repair or replace the defective product free of charge. Opening or altering the device invalidates the guarantee. The following items are excluded from the guarantee:

  • Transport costs and risks of transport.
  • Damage caused by incorrect application or non-compliance with the instructions for use.
  • Damage caused by leaking batteries.
  • Damage caused by accident or misuse.
  • Packaging/storage material and instructions for use.
  • Regular checks and maintenance (calibration).
  • Accessories and wearing parts: Batteries.

The cuff is covered by a functional guarantee (bladder tightness) for 2 years. Should guarantee service be required, please contact the dealer from where the product was purchased, or your local Microlife service. You may contact your local Microlife service through our website: www.microlife.com/support Compensation is limited to the value of the product. The guarantee will be granted if the complete product is returned with the original invoice. Repair or replacement within guarantee does not prolong or renew the guarantee period. The legal claims and rights of consumers are not limited by this guarantee.

Technical Specifications

Operating conditions: 10 – 40 °C / 50 – 104 °F 15 – 95 % relative maximum humidity
Storage conditions: -20 – +55 °C / -4 – +131 °F 15 – 95 % relative maximum humidity
Weight: 112 g (including batteries)
Dimensions: 81 x 65 x 21 mm
Cuff size: 13.5 – 23 cm (5.25 – 9.05 inches)
Measuring procedure: oscillometric, corresponding to Korotkoff method: Phase I systolic, Phase V diastolic
Measurement range: 20 – 280 mmHg ­ blood pressure 40 – 200 beats per minute ­ pulse
Cuff pressure display range: 0 – 299 mmHg
Resolution: 1 mmHg
Static accuracy: pressure within ± 3 mmHg
Pulse accuracy: ± 5 % of the readout value
Voltage source: 2 x 1.5 V alkaline batteries; size AAA
Battery lifetime: approx. 320 measurements (using new batteries)
IP Class: IP20
Reference to standards: EN 1060-1 /-3 /-4; IEC 60601-1; IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements
Accessories: 2 years

This device complies with the requirements of the Medical Device Directive 93/42/EEC. Technical alterations reserved.

Microlife AG
Espenstrasse 139
9443 Widnau / Switzerland
www.microlife.com|
Microlife UAB
P. Lukšio g. 32,
08222 Vilnius
Lithuania
---|---

Guarantee Card
Name of Purchaser __
Serial Number __
Date of Purchase ____
Specialist Dealer ____

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals