Welch Allyn 60300 Standard laryngoscope handles Instruction Manual

June 6, 2024
Welch Allyn

Welch Allyn logo Standard laryngoscope handles
Directions for use

Intended use
The standard laryngoscope handle is an accessory used with a compatible rigid standard laryngoscope blade which is used to examine and visualize a patient’s airway and aid placement of a tracheal tube.
About this document
These directions for use apply to Welch Allyn reusable standard laryngoscope handles REF:
60200, 60300, 60400, 60305. Welch Allyn reusable standard laryngoscope handles may be used with Welch Allyn standard laryngoscope blades MacIntosh REF: 6904X, English
MacIntosh REF: 6924X, and Miller REF: 6804X and REF 68470.
Warnings and cautions
WARNING:
Welch Allyn’s reusable standard laryngoscope handles must be reprocessed after each use.
WARNING: The reprocessing procedure and the equipment and materials described must be followed and conducted by persons trained and familiar with medical device
reprocessing.
WARNING: Consult cleaning and disinfecting agent manufacturer instructions for their proper preparation and use.
WARNING: Repeated reprocessing may degrade elements of the handle. Follow inspection procedures to assure damage has not occurred to the handle.
WARNING: High-level disinfection and/or sterilization are not achieved by these methods.
WARNING: Discard any component that shows evidence of damage or deterioration.
WARNING: Do not modify this equipment. Any modification of this equipment may lead to patient injury. Any modification of this equipment voids the product warranty.
WARNING: Personnel shall follow their facility policies and procedures and wear appropriate personal protective equipment when handling potentially contaminated equipment.
WARNING: Laryngoscope equipment is not suitable for use in intense magnetic fields
Welch Allyn 60300 Standard laryngoscope handles -
icon1CAUTION: Failure to follow these instructions may cause damage to this handle.
Welch Allyn 60300 Standard laryngoscope handles -
icon1CAUTION: Do not immerse/soak handle, damage to handle may occur.
Welch Allyn 60300 Standard laryngoscope handles -
icon1CAUTION: If the device will be unused for several months or longer, remove the batteries prior to storing the device.

Reprocessing instructions

These reprocessing instructions refer to procedures for cleaning and intermediate-level disinfection. Standard laryngoscope handles must be reprocessed prior to first use and between each use using the following method as outlined in this document:

  • Cleaning and intermediate-level disinfection

Welch Allyn has validated the above instruction as being capable of preparing these laryngoscope handles for re-use. The user must ensure that the reprocessing as actually
performed by the user’s personnel, with the user’s equipment and materials, achieves the desired result. This may require validation and routine monitoring of the user’s actual process
NOTE: The main handle and bottom cap components of handles marked “AUTOCLAVE” are compatible with the autoclave methods identified which are provided for facilities who wish to autoclave after cleaning and intermediate- level disinfection.

Cleaning and intermediate-level disinfection instructions

Point of use:|

  1. Separate blade assembly from handle and place handle into suitable containment for subsequent reprocessing per figure 1. Do not place the handle with sharp devices.
  2. Prevent the handle from drying (i.e. wrap/cover in the moist germicidal wipe).

---|---
Preparation for decontamination:|

  1. Select appropriate quaternary ammonium isopropanol-based germicidal cleaner labeled suitable for use on healthcare equipment and capable intermediate-level disinfection. Reference EPA-registered disinfectants: http://www.epa.gov/oppad001/chemregindex.htm Outside of the U.S., please consult applicable regulatory body for equivalent quaternary ammonium isopropanol germicidal cleaner.
  2. Remove batteries per figure 3.

Welch Allyn 60300 Standard laryngoscope handles - fig

Cleaning and intermediate-level disinfection:|

  1. Follow the germicidal wipe manufacturer’s instructions to clean all exposed surfaces of the handle and end cap.
  2. If necessary, brush with a dry, soft-bristled brush and re-wipe to loosen/remove excessive visible soil.
  3. After all visible soil is removed, re-wipe to wet all surfaces and allow adequate contact time for disinfection as directed by the germicidal wipe manufacturer.

Welch Allyn 60300 Standard laryngoscope handles -
icon1 CAUTION: Only use quaternary ammonium isopropanol-based germicidal wipes.

---|---
Drying:| Allow components to air dry.
Maintenance, Inspection, and Testing|

  1. Inspect each component area (per figure 3) for damage or deterioration.
    WARNING: Discard any component that shows evidence of damage or deterioration.
    Contact Welch Allyn for component replacement.

  2. Reassemble batteries into handle per figure 4 with new or batteries in known good condition.

  3. Attach handle to a clean and disinfected test blade assembly in known working condition. Verify that:
    • Blade assembly engages and locks onto the handle.
    • Blade assembly deploys into its locked position on handle AND lamp illuminates.
    • Light output is satisfactory.
    If the lamp fails to light or output is low, check or replace the batteries.

Storage:| Store handle per facility practice to allow the device to remain clean, dry, and ready for service.

End of reprocessing instructions for intermediate-level disinfection.

Welch Allyn 60300 Standard laryngoscope handles - fig1

Autoclave instructions

NOTE: The main handle and bottom cap components of handles marked “AUTOCLAVE” are
compatible with the autoclave methods identified which are provided for facilities who wish to
autoclave after cleaning and intermediate-level disinfection.

Disassembly: Remove batteries per figure 2 and set aside.

After battery removal, select ONE of the following autoclave methods below for the main handle and bottom cap (only):
Gravity autoclave: Follow equipment manufacturer and facility procedures in the set-up and operation of autoclave equipment. Gravity autoclave settings are as follows:

  • Temperature: 132 C (270 F)
  • Exposure time: 3 minutes (unwrapped)
  • Minimum dry time: 1 minute

Pre-vacuum autoclave: Follow equipment manufacturer and facility procedures in the set-up and operation of autoclave equipment. Pre-vacuum autoclave settings are as follows:

  • Temperature: 132 C (270 F)
  • Exposure time: 3 minutes (unwrapped)
  • Minimum dry time: 1 minute

Maintenance, Inspection, and Testing

|

  1. Inspect each component area per figure 3 for damage or deterioration.
    WARNING: Discard any component that shows evidence of damage or deterioration.
    Contact Welch Allyn for component replacement.
    2. Reassemble batteries into handle per figure 4.
    3. Attach handle to a clean and disinfected test blade assembly in known working condition. Verify that:
    • Blade assembly engages and locks onto the handle.
    • Blade assembly deploys into its locked position on handle AND lamp illuminates.
    • Light output is satisfactory.
    If the lamp fails to light or output is low, check or replace the batteries

Storage: Store|

Maintenance instructions

Replace the batteries

  1. Unscrew bottom cap of the handle per figure 5 and remove batteries.

  2. Alkaline batteries are supplied with your handle for maximum performance and are recommended as replacements; however carbon-zinc batteries may also be used.
    • Large handle, REF 60200 uses two “D” size
    • Medium handle, REF 60300 uses two “C” size
    • Penlight handle, REF 60400 uses two “AA” size
    • Stubby handle, REF 60305 uses two “AA” size

  3. Insert batteries and reinstall bottom cap per figure 5.

  4. Reprocess repaired assembly as appropriate per these instructions

Welch Allyn 60300 Standard laryngoscope handles - fig3

Specifications

Electrical:
For information about electromagnetic compatibility (EMC) see Welch Allyn website: http://www.welchallyn.com

Operating:

|

Storage/Transport:

---|---
Welch Allyn 60300 Standard laryngoscope handles - fig5| Welch Allyn 60300 Standard laryngoscope handles -
fig7

Approvals:
Conforms to ASTM F 965 and ISO-7376-1, IEC/EN 60601-1, IEC/EN 60601-1-2
The CE mark on this product indicates that it has been tested to and conforms with the provisions noted within the 93/42/EEC Medical Device Directive.
Complies with EMC Framework of Australia
Warranty:
One year
Service Information:
For Technical Support or to obtain information about any Welch Allyn product, contact
Welch Allyn Technical Support: www.welchallyn.com/support.

Welch Allyn logo©2014 Welch Allyn, Inc.
MM 721357 Ver. C
Welch Allyn, Inc.
4341 State Street Road
Skaneateles Falls, NY 13153-0220 the USA
www.welchallyn.com
Authorized European Representative Address:
Regulatory Affairs Representative
Welch Allyn, Limited
Navan Business Park
Dublin Road
Navan, County Meath, Republic of Ireland

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Welch Allyn User Manuals

Related Manuals