berlinger Fridge-tag 3 Temperature Monitor User Guide
Important symbols
OK symbol:
Alarm symbol:
Warning symbol:
Indication of an OK display
In measurement mode, when no alarm has been triggered, the following information is indicated:
- The OK symbol
- The current date and time: 16.02.20 18/13:47
- The current temperature: 5.6 °C
Indication of an Alarm display
In measurement mode, when an alarm has been triggered, the following information is indicated:
- The alarm symbol
- An additional warning symbol
- The corresponding arrow (alarm indicator), upper/lower alarm
- The current date and time: 16.02.20 18/15:42
- The current temperature: 8.8 °C
Alarm display and confirmation options (factory preset)
Option 1- Alarm indication “all alarms”
With this option the alarms will be visible on the display with an for 30 days. By pressing the READ button, the will be disabled for the corresponding alarms. The cannot be canceled nor reset.
- In this mode only one upper and one lower alarm will be triggered per day.
- The will be present on the display for 30 days.
- The can be deactivated by confirming all existing alarms in the readout mode.
- The alarm buzzer stops when the alarm is confirmed within the set alarm limits. Otherwise the buzzer pauses for approx. 1 hour and starts again for up to 168 hours (7 days).
Option 2 – Alarm indication “unconfirmed alarms”
The alarms are shown with the until all alarms (in the 30-day history) have been confirmed as solved by pressing the READ button.
Afterwards the display will show the until a new alarm is triggered.
Confirmation: Temperature is within the set alarm limits.
Press the READ button and the and the will immediately disappear and the optional buzzer stops. A new alarm will be triggered as soon as the set alarm limits are exceeded again.
Confirmation: Temperature is outside the set alarm limits.
If the READ button is pressed still during a temperature violation the buzzer will be muted for approx. 1 hour. The and the will stay on the display for the corresponding alarm. If the temperature still exceeds the limit after 1 hour, the buzzer will restart beeping.
Confirming an alarm
An alarm must be confirmed on the device directly by the user. Confirming an alarm can only be made if the device gets back within the set temperature limits.
The alarm cannot be confirmed with a SMS command, only direct on the device.
Not confirmed alarms
If a an alarm event is not confirmed on the device and the device gets back within the set temperature limits, an alarm will be retriggered when the limit is violated again.
This ensures that a forgotten confirmation of an alarm event on the device itself, does not result in a missed following alarm.
Remote Alarm Notifications
In case an upper or lower alarm is triggered, the predefined recipient(s) will be notified as follows:
- Standalone Version: SMS Notification
- Cloud Version: Make sure that the system is set up to your needs (e.g. in Smart View SMS and Email Notifications are available).
Audio alarm (factory preset)
In case an upper or lower alarm is triggered, 3 audible alarm signals are emitted immediately. Thereafter:
- Every minute 1 alarm signal for maximally 168 hours (7 days).
- After 168 hours (7 days) the buzzer will stop.
- If an alarm event is confirmed (READ is pressed) while the limits are still exceeded the buzzer pauses for approx. 1 hour and then restarts beeping every 3 minutes.
- Confirmation within the alarm limits will stop the buzzer.
Reading the history Option 1: Read out day-per-day directly on the device
(30 day history)
Example of an OK display
The following information is indicated on the display:
- The OK symbol
- The corresponding flashing arrow (example: high arrow “today”)
- The highest recorded temperature of that day: 10.5 °C
- The time duration out of the preset temperature upper limit: 00:32 (hrs:min)
Example of an alarm display
The following information is indicated on the screen:
- The alarm and the warning symbol
- The corresponding flashing arrow (alarm indicator) (example: Temperature out of lower alarm limit)
- The date of alarm: 16.02.20 18
- The time of alarm: 18:21
The following additional information is indicated:
- Lowest recorded temperature of that day: -1.1 ° C
- The duration or the exceedance of the preset lower temperature limit: 01:35 (hrs:min)
Note: Continue repetitively pressing the READ button to read out the details of the past 30 days.
Option 2: Read only alarms directly Alarm-Super-Jump function (30-day
history)
Note: Press and hold the READ button again for at least 3 seconds and the next alarm event will appear on the screen.
Option 3: Read out data with a computer
Connect the device via USB with your computer.
Note: Disconnect the external sensor from the device first.
The Fridge-tag will now generate a PDF and ASCII report of max. 28 or 56 days (factory setting). This process may take up to 2 minutes.
The continuously appearing arrows in the upper display area indicate that the
device is processing.
This process must not be interrupted until the OK symbol appears on the display. This indicates that the creation of the ASCII and PDF files has been successfully completed. Open the disk drive named “FT3” and copy/open the files.
Option 4: SMS temperature reading
Status temperature query: send an SMS with the content „status“ to mobile number of the device. The device returns a SMS with the current STATUS of the device.
Status query OK
Fridge-tag 3 is in the OK state, the current temperature is 6.5°C and within the set alarm limits.
Status query alarm
Fridge-tag 3 is in the alarm state, the current temperature is 1.5°C and below the set alarm limits..
Note: Please confirm the alarm directly on the device. SMS confirmation is not possible.
Cloud Version
Please log in to your Data Management Software (Berlinger Smart View, etc.) to manage the data.
External sensor: Please connect the device with the external sensor again. After 10 minutes (factory standard) without a connection between the device and the external sensor the following display appears and:
- The buzzer will beep twice at intervals of three minutes for a maximum of 168 hours (7 days).
- The whole display starts blinking.
- Any confirmation will stop the display from blinking.
- The buzzer only stops if the connection error is corrected.
If the error still exists, the buzzer continuously beeps at a three-minute interval for 168 hours (7 days).
For more information download the whole user manual from www.berlinger.com
/user-manuals .