GENEQ CC4648-000 Vaccine Fridge Thermometer Temperature Monitoring System Instruction Manual

June 6, 2024
GENEQ

GENEQ CC4648-000 Vaccine Fridge Thermometer Temperature Monitoring System

GENEQ CC4648-000 Vaccine Fridge Thermometer Temperature Monitoring System
IMAGE

FEATURES

  • Triple display for current, minimum, and maximum temperature
  • Programmable low and high alarm limit
  •  Outdoor temperature display
  •  °C / °F exchange
  •  Low battery indication
  •  Foldaway stand
  •  Mounting keyhole

SPECIFICATIONS

Measuring range| 0 ~ 50°C / 32 ~ 122°F – internal probe (ambient)

– 50 ~ 70°C / -58 ~ 158°F – external probe (bottle)

---|---
Display accuracy| ****

-20 ~ 50 ± 0.3°C

Display resolution| 0.1°
Display reading update| 10 seconds
Battery| 1.5 volt, type AAA x2 pieces
Accessories| 1)  Outdoor sensor with 3 meters long and glycol bottle

2)  Battery 1.5 volt x2 pieces

INSTALLATION

  • Plug outdoor sensor into the unit.
  •  Install battery with correct polarity positioning.
  •  The unit will perform a self-test and then display readings normally after 3 seconds.
  •  Press (°C/°F) to select the temperature unit.
  •  Place the unit at the place away from direct sunlight, rain, or extreme heat.
  •  Click on OUT to display the refrigerator temperature.

OPERATION LOW AND HIGH ALARM LIMIT SETTING

  • Press (MODE) to enable alarm setting.
  • Press (LO ) to set a low alarm limit
  • Press (HI ) to set a high alarm limit
  •  Press (MODE) to return to the normal display.
  •  When the current reading is lower or higher than the alarm limit the alarm will sound.
  •  To stop the alarm sound press any button once. The LO or HI icon will still be flashing until the current reading is within the alarm limit.
  •  Press ( ) to disable or enable the alarm function.
    Holding down the button will increase the value automatically.

RESET MINIMUM (MIN) and MAXIMUM (MAX) READING MEMORY

  •  In MIN-MAX display mode, press (MIN-RESET) to reset minimum reading memory to the current reading.
  •  Press (MAX-RESET) to reset maximum reading memory to the current reading.
  •  Reset the memory once before taking a new minimum or maximum reading.

IN / OUTDOOR TEMPERATURE DISPLAY
Press (IN-OUT) to select outdoor or indoor temperature display.
NOTE:

  • When the low battery symbol( )appears, replace the battery with the same type of new battery.
  •  If the thermometer is not in use for a long period of time then remove the battery from battery compartment to avoid battery leakage.
    •   *  If there is an abnormal operation or display then press (RESET) once.

Read User Manual Online (PDF format)