Reference Guide to Automatic Processing of INDUSTREX Films

NOTE: For more information, see CHSP-8790.

Recommended Chemicals and Mixing Instructions

Manual Mixing:

  • Measure water for correct dilution.
  • Add replenisher concentrate and mix using a separate paddle for Developer and Fixer to avoid contamination.
  • Mix only enough replenisher to last a week to assure fresh chemicals.
    IMPORTANT: use the floating lid for the developer replenisher tank to prevent oxidation and maximize solution life.

| Dilution
Concentrate
+ I-120| Replenishment
Rate| pH re St Working gth| Specific Gravity
Working Strength| To Make Working Strength Solution
---|---|---|---|---|---
Developer| | | | |
Use INDUSTREX Single Part
Developer and Replenisher.
Material No.
5315288| 1 + 3| 665mUm2 (e.g., 100 mL per 14 x 17 inch sheet)| 10.40 +/- 0.10
(25 °C,77 °F)| 1.072 +/- 0.005
(20 °C,68 °F)|
Fixer| | | | |
Use INDUSTRIES
LO Fixer and
Replenisher.
Material No.
5159082| 1 + 3| 1200mUm2 (e.g., 180mL per 14 x 17 inch sheet)| 4.90 +/- 0.10
(25 °C,77 °F)| 1.084 +/- 0.005
(20 °C.68 °F)|

Chemical Mixers

  • Proper templates that match the bottles should be used.
  • Thorough mixing is necessary after each addition.
  • Ideally, the mixer should have a floating lid to protect the solutions from oxidation.
  • The concentrates should be examined to be certain there is no premature oxidation due to excessive handling.
  • The manufacturer’s recommended water-mixing temperature should be observed.
  • A new mix of developer and/or fixer should only be made when indicated by the solution level in the mixer.
  • To aid in verifying that the chemistry is acceptable, the pH and the specific gravity can be measured. It is important to make these measurements at the recommended