Carestream NDT Automatic Processing of Industrex Films User Guide

June 6, 2024
Carestream

Reference Guide to Automatic Processing of INDUSTREX Films

NOTE: For more information, see CHSP-8790.

Recommended Chemicals and Mixing Instructions

Manual Mixing:

  • Measure water for correct dilution.
  • Add replenisher concentrate and mix using a separate paddle for Developer and Fixer to avoid contamination.
  • Mix only enough replenisher to last a week to assure fresh chemicals.
    IMPORTANT: use the floating lid for the developer replenisher tank to prevent oxidation and maximize solution life.

| Dilution
Concentrate
+ I-120| Replenishment
Rate| pH re St Working gth| Specific Gravity
Working Strength| To Make Working Strength Solution
---|---|---|---|---|---
Developer| | | | |
Use INDUSTREX Single Part
Developer and Replenisher.
Material No.
5315288| 1 + 3| 665mUm2 (e.g., 100 mL per 14 x 17 inch sheet)| 10.40 +/- 0.10
(25 °C,77 °F)| 1.072 +/- 0.005
(20 °C,68 °F)|
Fixer| | | | |
Use INDUSTRIES
LO Fixer and
Replenisher.
Material No.
5159082| 1 + 3| 1200mUm2 (e.g., 180mL per 14 x 17 inch sheet)| 4.90 +/- 0.10
(25 °C,77 °F)| 1.084 +/- 0.005
(20 °C.68 °F)|

Chemical Mixers

  • Proper templates that match the bottles should be used.
  • Thorough mixing is necessary after each addition.
  • Ideally, the mixer should have a floating lid to protect the solutions from oxidation.
  • The concentrates should be examined to be certain there is no premature oxidation due to excessive handling.
  • The manufacturer’s recommended water-mixing temperature should be observed.
  • A new mix of developer and/or fixer should only be made when indicated by the solution level in the mixer.
  • To aid in verifying that the chemistry is acceptable, the pH and the specific gravity can be measured. It is important to make these measurements at the recommended

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals