BANTE DO100 Dissolved Oxygen Electrode User Guide
This polarographic dissolved oxygen electrode is designed for laboratory
and field use.
Required Equipment and Reagents
- A dissolved oxygen meter
- Sodium sulfite reagent (Na2SO3)
- Cobalt (II) chloride hexahydrate (CoCl2 • 6H2O)
- Electrolyte solution (order code: DO-ES)
Required Solutions
-
Air-saturated water:
To prepare this solution, use an air-pump to blow air into clear water at least 1 hour, while stirring the solution. -
Zero oxygen solution:
To prepare this solution, dissolve 500 mg of the sodium sulfate reagent and a small amount of cobalt (II) chloride hexahydrate in the 250 ml distilled water, mix the solution until reagent is completely dissolved.
Prior to Use
-
Unscrew the membrane cap from the bottom of the electrode, rinse the inside and outside with distilled water and blot dry.
-
Fill the membrane cap halfway with electrolyte solution
-
Screw membrane cap back onto the electrode. Some electrolyte solution will overflow during this process.
-
Check the electrode, make sure that the cathode of electrode makes contact with membrane cap, no air bubbles are trapped in the electrolyte solution and membrane is not creased or damaged.
Measurement
- Connecting the electrode and switch on the meter, wait 10 minutes for the electrode to polarize.
- Calibrate the meter according to the manufacturer’s instructions.
- Place the electrode into the sample and record the stable reading
Electrode Maintenance
- Rinse the electrode thoroughly with distilled water after use.
- DO NOT touch the membrane and always keep it is clean and wet.
- If the electrode will not use for long periods, screw off the membrane cap, rinse the electrode anode, cathode, membrane cap with distilled water and blot dry. Install the electrode and store dry.
Specifications
Model | DO100 |
---|---|
Range | 0 to 20 mg/L |
Temperature Sensor | NTC 10 KΩ |
Response Time | 95% of final reading in 30 seconds, 98% in 45 seconds |
Minimum Sample Flow | 20 cm per second |
Operating Temperature | 0 to 50°C (32 to 122°F) |
Electrode Dimensions | 150 (L) × 12 (Dia.) mm (5.9 × 0.47″) |
Cable Length | 2 m (6.5 ft) |
Connector | 6-pin mini-DIN |
Body Type | Epoxy |