BANTE ISE-NH3 Ammonia Ion Selective Electrode User Guide
- June 6, 2024
- BANTE
Table of Contents
Ammonia
Ion Selective Electrode I User Guide
This ion-selective electrode is designed for the detection and analysis of dissolved ammonia in an aqueous solution and is suitable for laboratory applications.
Required Equipment and Solutions
-
An ion meter
-
Filling solution (order code: FS-NH3)
-
Ionic strength adjuster (order code: ISA-NH3)
-
Volumetric flasks and beakers
-
Distilled or deionized water:
To prepare the standard solutions or rinse the electrode between measurements. -
Ammonia standard solution 0.1M:
To prepare this standard solution, half fill a 1 liter volumetric flask with distilled water and add 5.35 grams of analytical grade ammonium chloride (NH₄Cl) reagent. Swirl the volumetric flask gently to dissolve the reagent and fill it to the mark with distilled water. Cap and upend the volumetric flask several times to mix the solution. -
Ammonia standard solution 1000 ppm:
To prepare this standard solution, half fill a 1 liter volumetric flask with distilled water and add 3.15 grams of analytical grade ammonium chloride (NH₄Cl) reagent. Swirl the volumetric flask gently to dissolve the reagent and fill to the mark with distilled water. Cap and upend the volumetric flask several times to mix the solution.
Ammonia ISE
Prior to Use
- Remove the membrane module and fill with filling solution.
- Tap the side of the cap to remove air bubbles and screw the module onto the body. Do not over-tighten.
Measurement and Calibration Hints
-
Do not use this electrode to measure the strongly acidic or alkaline samples, strong detergents, and organic solvents, these solutions will cause permanent damage to the electrode.
-
For better accuracy, we recommend adding the ionic strength adjuster (ISA) to all of the standards and samples.
A typical addition would be 2 ml ISA to 50 ml of standard and sample solutions. -
During the calibration and measurement, ensure that all standard and sample solutions are at the same temperature.
-
The calibration should be from the lowest concentration standard to avoid cross-contamination.
-
The sample solution must fall in the pH range of 11 to 13. If necessary, add the ionic strength adjuster.
-
Stir the standard and sample solutions at a uniform rate that will be promoting the accuracy of measurement.
1. Calibrate the meter according to the manufacturer’s instructions.
2. Rinse the electrode with distilled water and blot dry.
3. Place the electrode into the sample and record the stable reading.
Electrode Maintenance
- Rinse the electrode and membrane cap thoroughly with distilled water after use.
- Never touch or scratch the ion sensitive membrane on the bottom of the electrode.
- If the electrode response becomes sluggish, soak the electrode in 100 ppm standard solution for at least 1 hour
Specifications
Model | ISE-NH3 |
---|---|
Concentration Range | 0.01 to 17000 ppm |
Slope | 52 to 59 mV/decade |
pH Range | 11 to 13 |
Interferences | Hydrazine |
Operating Temperature | 5 to 50°C (41 to 122°F) |
Electrode Dimensions | 150 (L) × 12 (Dia.) mm (5.9 × 0.47″) |
Cable Length | 1 m (3.3 ft) |
Connector | BNC |
Body Type | Epoxy |
Optional Accessories
Order Code | Description |
---|---|
ION-NH3 | Ammonia standard solution 1000 ppm, 480 ml |
ISA-NH3 | Ionic strength adjuster, 480 ml |
FS-NH3 | Ammonia filling solution, 480 ml |