BANTE ORPscan10/20 Pocket ORP Tester
Product Overview
Thank you for selecting the ORPscan series pocket ORP tester. This user manual provides a step-by-step guide to help you operatethe tester, please carefully read the following instructions before use.
Product Installing the Batteries
- Twist the electrode collar counter clockwise, pull theelectrode (or BNC connector) away from the tester.
- Insert two AAA alkaline batteries into the battery compartment, note polarity.
Product Prior to Use
ORPscan10:
- Remove the protective cap (and translucent cover) from the bottom of the tester.
- If some salt crystals deposited on the electrode, rinse with tap water to clean these deposits.
- If the platinum sensor has dried out, soak the electrode in 4MKCl solution for about 30 minutes.
- Push the electrode (or BNC connector) into the tester and twistthe electrode collar clockwise until tight.
ORPscan20:
- Take out the ORP electrode from carrying case.
- Insert the BNC connector into the connector socket on tester, rotate and push the connector clockwise until it locks.
Keypad
Key | Description |
---|---|
measurement | Switches the tester to measurement mode |
Cal | Enters the calibration mode or setup menu |
Switching the Tester On and Off If the battery voltage falls below the minimum powerrequirements, the icon automatically disappears. If you do not press any key within 8 minutes, the tester will switch off automatically to conserve energy.
Product Setup Menu
The ORPscan10/20 tester contains 3 menu items in the setup menu, the following table describes the functions of each option.
Menu Item | Option and Description |
---|---|
Auto-Hold | If enabled, the tester will automatically sense and lock the |
measurement endpoint.
Auto-Power Off| If enabled, the tester will automatically switch off if no key
is pressed within 8 minutes.
Factory Reset| If enabled, all of the calibration values and current settings
will be deleted or reset to the factory defaults.
Setting the Default Option
- In the measurement mode, press and hold the Cal key to enterthe setup menu.
- If necessary, press the Cal key again to select an option.
- Press the Enter key, the tester saves the current option andmoves to the next menu item.
- Repeat steps above until the tester returns to the measurement mode.
To exit the setup menu, press the Meas key.
Product Measurement
- Rinse the electrode with distilled water.
- Place the electrode into the sample solution and stirgently.
- Note, DO NOT completely immerse the tester in water.
- Wait for the measurement to stabilize and record the reading.
If the sample solution contains chromous, vanadous and titanous ions or other
ions that interfere with the ORP measurement range,remove the tester from the
sample immediately. If the Auto-Hold option is enabled in the setup menu, the
tester will automatically
lock the measurement endpoint and show HOLD icon. Press the Measkey to resume
measuring.
Relative mV Calibration The ORPscan10/20 tester allows 1 point calibration, butcalibration is not necessary unless exact readout agreement with a work standard and at a specific ORP value is needed.
- Rinse the electrode with distilled water and place into thestandard.
Overview
Thank you for selecting the ORPscan series pocket ORP tester. This user manual provides a step-by-step guide to help you operate the tester, please carefully read the following instructions before use.
Installing the Batteries
- Twist the electrode collar counter clockwise, pull the electrode (or BNC connector) away from the tester.
- Insert two AAA alkaline batteries into the battery compartment, note polarity.
- Push the electrode (or BNC connector) into the tester and twist the electrode collar clockwise until tight.
Keypad
Display
Prior to Use
ORPscan10
Remove the protective cap (and translucent cover) from the bottom of the
tester. If some salt crystals deposited on the electrode, rinse withtap water
to clean these deposits. If the platinum sensor has dried out, soak the
electrode in 4M KCl solution for about 30 minutes
ORPscan20
Take out the ORP electrode from carrying case. Insert the BNC connector into
the connector socket on tester, rotate and push the connector clockwise until
it locks.
Switching the Tester On and Off
- Press and hold the Meas key for about 5 seconds to switch on the tester.
- Press and hold the Meas key to switch off the tester.
- If you do not press any key within 8 minutes, the tester will switch off automatically to conserve energy.
Setup Menu
The ORPscan10/20 tester contains 3 menu items in the setup menu, the following table describes the functions of each option.
Setting the Default Option
- In the measurement mode, press and hold the Cal key to enter the setup menu.
- If necessary, press the Cal key again to select an option
- Press the Enter key, the tester saves the current option and moves to the next menu item.
- Repeat steps above until the tester returns to the measurement mode.
- To exit the setup menu, press the Meas key.
Measurement
Rinse the electrode with distilled water, place the electrode into the sample solution and stir gently. Note, DO NOT completely immerse the tester in water. Wait for the measurement to stabilize and record the reading.
- The platinum sensor may give unstable readings in solutions that contain chromous, vanadous and titanous ions or other ions that are stronger reducing agents than hydrogen or platinum.
- If the display shows —- indicating the measurement exceeds the range, remove the tester from the sample immediately.
- If the option is enabled in the setup menu, the tester will automatically lock the measurement endpoint and show HOLD icon. Press the Meas key to resume measuring.
Relative mV Calibration
The ORPscan10/20 tester allows 1 point calibration, but calibration is not necessary unless exact readout agreement with a work standard and at a specific ORP value is needed.
- Rinse the electrode with distilled water and place into the standardsolution. Stir the tester (or electrode) gently and wait until the measurement is stable.
- Press the Cal key, the display shows 000/CAL.
- Press the Cal key again to modify the calibration value, press the Enter key to confirm and move to the next digit. When the setting is completed, make sure that the displayed value matches your calibration standard.
- Press the Enter key, the tester begins the calibration. When the reading has stabilized, the display will show .
- After the calibration, the measurement unit “mV” will automatically convert to “R.mV” (relative millivolt). If you want to resume the mV measurement, reset the tester (refer to the Setup Menu section).
- To exit the calibration without saving calibrated value, press the Meas key.
Electrode Maintenance
- Rinse the electrode thoroughly with distilled water after use.
- In the corrosive chemicals, viscous solutions and solutions with heavy metals or proteins, take readings quickly and rinse electrode immediately.
- If you do not use the tester for a period longer than 1 month, store the electrode in 4M KCl solution or electrode storage solution.
Appendix
Preparation of ORP Standard Solutions
Quinhydrone solution A:
Dissolve 3 grams of quinhydrone reagent in 500 ml of the pH 4.01 buffer
solution, stir the solution for about 10 minutes. Undissolved quinhydrone
reagent must be present. If necessary, add the reagent.
Temperature | Quinhydrone in pH 7.00 (±10 mV) |
---|---|
20°C (68°F) | 94 mV |
25°C (77°F) | 87 mV |
30°C (86°F) | 80 mV |
Quinhydrone solution B:
Dissolve 3 grams of quinhydrone reagent in 500 ml of the pH 7.00 buffer
solution, stir the solution for about 10 minutes. Undissolved quinhydrone
reagent must be present. If necessary, add the reagent.
Temperature | Quinhydrone in pH 4.01 (±10 mV) |
---|---|
20°C (68°F) | 268 mV |
25°C (77°F) | 263 mV |
30°C (86°F) | 260 mV |
- Due to the quinhydrone solution is susceptible to air oxidation in storage, make sure to prepare the fresh solution before use.
Preparation of Electrode Storage Solution
Dissolve 29.82 grams of analytical grade KCl reagent in 100 ml distilled
water. Add pH 4.01 standard buffer and adjust solution to pH 4.
Optional Accessories
Solutions
Order Code | Description |
---|---|
PHCS-ES | Electrode storage solution, 480 ml |
ORP Electrodes
Tester Specifications
Model | ORPscan10 |
---|---|
mV Range | ±999 mV |
Relative mV Range | ±999 mV |
Resolution | 1 mV |
Accuracy | ±2 mV |
Calibration Point | 1 point |
Operating Temperature | 0 to 50°C (32 to 122°F) |
Storage Temperature | 0 to 60°C (32 to 140°F) |
Relative Humidity | < 80% (non-condensing) |
IP Rating | IP54 |
Connector | — |
Display | LCD, 21 × 21 mm (0.82 × 0.82 in.) |
Power Requirements | 2 × 1.5V AAA alkaline batteries |
Auto-Off | 8 minutes after last key pressed |
Dimensions | 185 (L) × 40 (Ø) mm (7.28 × 1.57 in.) |
Weight | 100g (3.5 oz.) |
Model | ORPscan20 |
--- | --- |
mV Range | ±999 mV |
Relative mV Range | ±999 mV |
Resolution | 1 mV |
Accuracy | ±2 mV |
Calibration Point | 1 point |
Operating Temperature | 0 to 50°C (32 to 122°F) |
Storage Temperature | 0 to 60°C (32 to 140°F) |
Relative Humidity | < 80% (non-condensing) |
IP Rating | IP54 |
Connector | BNC |
Display | LCD, 21 × 21 mm (0.82 × 0.82 in.) |
Power Requirements | 2 × 1.5V AAA alkaline batteries |
Auto-Off | 8 minutes after last key pressed |
Dimensions | 175 (L) × 40 (Ø) mm (6.89 × 1.57 in.) |
Weight | 100g (3.5 oz.) |
Disposal
This product is required to comply with the European Union’s Waste Electrical
and Electronic Equipment (WEEE) Directive 2002/96/EC and may not be disposed
of in domestic waste. Please dispose of product in accordance with local
regulations at the collecting point specified for electrical and electronic
equipment.
Warranty
The warranty period for tester is one year from the date of shipment. Above
warranty does not cover the electrode and standard solution. Out of warranty
products will be repaired on a charged basis.
The warranty on your tester shall not apply to defects resulting from:
- Improper or inadequate maintenance by customer
- Unauthorized modification or misuse
- Operation outside of the environment specifications of the products.
For more information, please contact the supplier.
Contact
- Office: 4715 Castlewood St., Sugar land, TX 77479, USA
- Tel: (+1) 346-762-7358
- E-mail: banteinstruments@yahoo.com
- Factory: F3, Building 2, No.2185, Laifang Rd., Shanghai 201615, China