DeVilbiss 6910 Series 6910P-DR Portable Compressor Nebulizer System Instructions

June 6, 2024
DeVilbiss

6910 Series 6910P-DR Portable Compressor Nebulizer System

CAUTION: Federal (U.S.A.) law restricts this device to sale by, or on the order of a physician.

Not manufactured with natural rubber latex. Made in China

IMPORTANT PARTS

OPERATION

CLEANING

CLEANING

INTRODUCTION

Thank you for purchasing the DeVilbiss Traveler portable compressor/nebulizer for aerosol therapy. Your doctor has prescribed a liquid medication to treat your respiratory condition. The portable compressor/nebulizer works to convert the medication to a high-quality mist of fine particles that penetrates deep into the lungs.
Be sure that you read and understand the information in this instruction guide. By following these simple instructions and the advice of your physician, your compressor will become an effective addition to your therapeutic routine.

Statement of Intended Use
This compressor is intended to be used with a compatible pneumatic nebulizer to convert certain inhalable drugs into an aerosol form for inhalation by a patient for the treatment of asthma, COPD, and other respiratory ailments. The device is intended for the home care market. It is intended for use with a single adult, pediatric, or infant patient.

NOTE:   If using a nebulizer other than one distributed by DeVilbiss, verify that the pressure and/or flow characteristics of the nebulizer (while driven by the Model 6910) meet the nebulizer manufacturer’s recommendation.

IEC SYMBOLS

IMPORTANT SAFEGUARDS

When using electrical products, especially when children are present, basic safety precautions should always be followed. Read all instructions before using. Important information is highlighted by these terms:
DANGER– Urgent safety information for hazards that will cause serious injury or death.
WARNING– Important safety information for hazards that might cause serious injury.
CAUTION– Information for preventing damage to the product.
NOTE– Information to which you should pay special attention.

Read all Instructions before using.

DANGER

To reduce the risk of electrocution:

1. Always unplug this product immediately after using.
2. Do not use while bathing.
3. Do not place or store product where it can fall or be pulled into a tub or sink.
4. Do not place in or drop into water or other liquid.
5. Do not reach for a product that has fallen into water. Unplug immediately.

WARNING
To reduce the risk of burns, electrocution, fire or injury to persons:

1. A product should never be left unattended when plugged in.
2. Close supervision is necessary when this product is used by, on, or near children or physically incapacitated individuals.
3. Use this product only for its intended use as described in this guide. Do not use attachments not recommended by the manufacturer.
4. Never operate this product if:
a. It has a damaged power cord or plug.
b. It is not working properly.
c. It has been dropped or damaged.
d. It has been dropped into water.
Return the product to an authorized DeVilbiss provider for replacement.
5. Keep the power cord away from heated surfaces.
6. Never use while drowsy or asleep.
7. Never drop or insert any object into any opening or hose.
8. Do not use in an oxygen enriched environment.
9. Do not use this product with a DC to AC converter, or with an AC voltage and/or frequency other than specified.
10. Care should be taken when this product is used in the vicinity of other electronic equipment concerning the possible interference between them. They may need to be moved apart.

NOTE: If replacement of the cord or plug is necessary, contact your authorized DeVilbiss provider.
NOTE: The exterior surfaces of the unit and the air outlet connector may become slightly warm to the touch during use. This is normal.

GENERAL PRECAUTIONS

  1.  Before use, thoroughly read and understand the information contained in this instruction manual.
  2. Ensure the dryness of the inlet filter for proper operation. If the inlet filter becomes moist, replace it with a dry filter and allow the wet filter to dry.
  3. This device is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
  4. Unplug the device before filling. Do not overfill.
  5. Always unplug the device immediately after using it.

DRUG USE WARNING
Always follow the doctor’s instructions regarding the type of drug to be used, the dosage, the frequency, and duration of the treatment. This device has been designed for use with aqueous solutions of drugs; drugs that are too dense or oily will not be nebulized correctly.

SAVE THESE INSTRUCTIONS

IMPORTANT PARTS OF YOUR PORTABLE COMPRESSOR (PAGE 2, FIGURES A & B)

1. Power Switch
2. Filter Cover and Filter (non-washable)
3. Built-In Reusable Nebulizer Holder
4. Air-Outlet Connector
5. Add-On Nebulizer Holder Bracket
6. Rechargeable Battery (optional w/some models)
7. Power Indicator Light

8. Battery Status Indicator Light
9. Power Socket
10. Battery Release Buttons (one on each side)
11. AC Adapter (not shown)
12. DC Car Adapter (not shown)
13. Battery Cover (not shown)

IMPORTANT PARTS OF YOUR DISPOSABLE AND REUSABLE NEBULIZERS (PAGE 2, FIGURES C & D)

1. Mouthpiece
2. T-piece
3. Corrugated Tubing
4. Cap
5. Cap w/Baffle

6. Baffle
7. Medication Cup
8. Air-Inlet Connector
9. Tubing

Nebulizer received may look different than pictured nebulizer. Nebulizer parts are not sold separately. Additional nebulizers (part #NEB KIT 500 or 18254) are available through your DeVilbiss provider.

NOTE: Reusable Nebulizer may be optional with some models.

Accessories

Accessories

HOW TO OPERATE YOUR PORTABLE COMPRESSOR

NOTE : Before initial operation, your nebulizer should be cleaned by following the Cleaning Instructions, or as recommended by your physician or DeVilbiss provider.

The device can be powered in several ways:

With the AC Adapter:
This is designed for use at home or wherever normal-voltage power supply is available (different connections are available for different voltages).

  1. Insert the AC adapter into an appropriate wall outlet (specifications are indicated on the adapter).
  2. Insert the plug of the adapter into the power socket of the device (Fig A).
  3. The power indicator light will display green when the unit is turned on (Fig B).

With the DC Car Adapter:
It is very convenient to allow the patient to take treatments when traveling by car.

WARNING
Do not use while driving.

  1. Make sure the car cigarette lighter or DC source is 12 VDC. Insert the plug of the DC adapter into the cigarette lighter or 12 VDC source.
  2.  Insert the other end of the adapter into the power socket of the device (Fig A).
  3.  The power indicator light will display green when the unit is turned on (Fig B).

With the Rechargeable Battery:
The rechargeable battery can provide two treatments of about 20 minutes each.

  1. Remove the battery cover on the end of the device if attached.
  2.  Insert the battery by matching one of the connectors of the battery into the groove on the side of the device. Push in until the second connector clicks into its groove. Installation is done.
  3. Remove the battery by pushing in both side release buttons on the battery and pulling away from the device (Fig A).
  4. The power indicator light will display green when the unit is turned on (Fig B).
  5.  The battery status indicator light will illuminate red if battery voltage is low during operation with battery (Fig B).

DANGER
The portable compressor/nebulizer must be operated on the specified power source to avoid risk of electric shock and damage to the compressor.

  1. Wash hands.
  2. Connect one end of the nebulizer tubing to the compressor air-outlet connector (Fig. 1).

NOTE:   During high-humidity weather conditions, condensation (water build-up) can occur in the nebulizer tubing. If this occurs, after the treatment operate the unit with the tubing connected but without the nebulizer to dry the tubing.

3. Assemble clean nebulizer parts by placing baffle down in medication cup. Add prescribed medication into the medication cup (Fig. 2). Holding cup stationary, screw on nebulizer cap (Fig. 3).

4. Assemble mouthpiece and t-piece (if applicable) and insert into the top of the nebulizer cap. If using an aerosol mask, insert the bottom part of the mask directly into the top of the nebulizer cap.

5. Attach tubing to nebulizer air-inlet connector (Fig. 4).

6. Press the power switch “ON” to start the compressor.

7. Begin treatment by placing the mouthpiece between teeth. With mouth closed, inhale deeply and slowly through mouth as aerosol begins to flow then exhale slowly through the mouthpiece (Fig. 5). If treatment needs to be interrupted, simply press power switch and insert neb or tubing into the holder (Fig B).

NOTE: Some physicians may recommend a specific breathing pattern during treatment. If so, follow their recommendation.

8. If an aerosol mask is used, place the mask over mouth and nose (Fig. 6). As aerosol begins to flow, inhale deeply and slowly through mouth, then exhale slowly.

9. When treatment is complete, turn unit off by pressing the power switch. Unplug the unit from the wall power outlet if using the AC adapter.

CHARGING THE BATTERY

1. The battery indicator light will illuminate red if battery voltage is low during operation with battery.

2. Recharge the battery by inserting plug of the either the AC or DC adapter into the power socket of the device while battery is installed (Fig A). The charging time is 4-12 hours. The battery indicator light will illuminate orange during the charging operation and turn green when fully charged (Fig B). Normal battery life is approximately 250 full charge/discharge cycles under normal conditions and if properly maintained.

NOTE– For a new rechargeable battery, the first charge will take at least 24 hours.
NOTE– Battery will not charge while device is in use. Charging the battery continually between treatments is recommended.

CAUTION– Make sure battery poles do not come in contact with keys, coins or other metal objects that can short circuit battery poles.

Battery Status Indicator Light

  • Red – Low Battery. Recharge ASAP.
  • Green – Fully Charged Battery.
  • Orange – Battery Recharging.

Proper Care for the Rechargeable Battery

1. The battery should always be stored FULLY CHARGED in a cool dry place.

2. Charge the battery continually to maintain maximum battery capacity.

3. If unit is not in use for extended periods, the battery should be recharged a minimum of every 3 months. It may take up to 3 charge/discharge/recharge cycles to return the battery to maximum capacity.

NOTE– When unit is not in use, disconnect battery from main unit to preserve the life of the rechargeable battery. If the battery goes for an extended period of time without recharging, the rechargeable battery will slowly discharge.

NEBULIZER CLEANING

All parts of the nebulizer, except tubing, should be cleaned according to the following instructions. Your physician and/or DeVilbiss provider may specify a certain cleaning procedure. If so, follow their recommendations.

WARNING
To prevent possible risk of infection from contaminated medication, cleaning of the nebulizer is recommended after each aerosol treatment. Disinfecting is recommended once a day.

Clean After Every Use:

1. When the unit is turned off, unplug power cord from wall outlet.

2. Disconnect tubing from the air-inlet connector and set aside.

3. Disassemble mouthpiece or mask from cap. Open nebulizer by turning cap counterclockwise and removing baffle.

4. Wash all items, except tubing, in a hot water/dishwashing detergent solution. Rinse under hot tap water for 30 seconds to remove detergent residue. Allow to air dry.

CAUTION– The Reusable Nebulizer is dishwasher safe, but do not place any disposable nebulizer parts in an automatic dishwasher; doing so could cause damage.

Disinfect Daily:

  1. Using a clean container or bowl, soak items in three parts hot water (131ºF-149ºF/55ºC-65ºC) to one part white vinegar (>=5% acetic acid concentration) for 30 minutes (Fig. 7) OR use a medical bacterial-germicidal disinfectant available through your provider. Be sure to follow manufacturer’s instructions carefully.
  2.  With clean hands, remove items from disinfectant solution, rinse under hot tap water, and air dry on a clean paper towel. Store in a Zip-lock® bag.

NOTE– Do not towel dry nebulizer parts; this could cause contamination.

WARNING
To prevent possible risk of infection from contaminated cleaning solutions, always prepare fresh solution for each cleaning cycle and discard solution after each use.

3. For reusable nebulizer only, clean daily in dishwasher (if available) using top shelf.

4. Keep the outer surface of the tubing dust-free by wiping regularly. Nebulizer tubing does not have to be washed because only filtered air passes through it.

NOTE: The Disposable Nebulizer will last at least 15 days and possibly longer, depending on usage. Proper cleaning will also help extend the life of the nebulizer. Because the nebulizer is disposable, we recommend that an extra nebulizer be kept on hand at all times. The Reusable Nebulizer is dishwasher safe and may be cleaned and reused for up to 6 months.

COMPRESSOR CLEANING

1. With the unit off, unplug power cord from wall outlet.

2. Wipe the outside of the compressor cabinet with a clean damp cloth every few days to keep dust-free.

DANGER
Do not submerge in water; doing so will result in compressor damage.

FILTER CHANGE

1. Filter should be changed every six months or sooner if filter discolors.

2. Remove filter (Fig. A) by pulling back the filter cover tab and then pulling out the filter from the unit. Discard old filter. The filter cover tab should remain attached to the device.

3. Replace with a new filter. Additional filters should be purchased from your DeVilbiss provider.

4. Push new filter into position, and replace filter cover.

CAUTION– Do not reuse or wash filter or substitute any other material such as cotton for a DeVilbiss air-inlet filter. It will result in compressor damage.

MAINTENANCE

All maintenance must be performed by a authorized DeVilbiss provider.

DANGER
Electric shock hazard. Do not remove compressor cabinet. All disassembly and maintenance must be done by a authorized DeVilbiss provider.

SPECIFICATIONS

Equipment Classifications
Protection against electric shock ………………………………………………………… Class II Degree of protection against electric shock ………………………………………………… Type BF Equipment not suitable for use in the presence of flammable anaesthetic mixture with air or with oxygen or nitrous oxide.

THREE-YEAR LIMITED WARRANTY ON COMPRESSOR PARTS

DeVilbiss Healthcare warrants the DeVilbiss Traveler Portable Compressor to be free from defective workmanship and material for a period of three years from the date of purchase. The rechargeable battery is warranted for one year from date of purchase. Any defective part(s) will be replaced or refunded prorated over the term of the warranty at the sole discretion and determination of DeVilbiss if the unit has not been tampered with or used improperly during that period. Make certain that any malfunction is not due to inadequate cleaning or failure to follow the instructions. This warranty is limited to the original purchaser. Normal maintenance items and disposable products are not covered under warranty. If replacement is necessary, contact your authorized DeVilbiss provider or DeVilbiss Customer Service Department at 800-338-1988 or 814-443-4881 for instructions.

NOTE– This warranty does not cover providing a loaner compressor, compensating for costs incurred in compressor rental while said unit is out of service, or costs for labor incurred in replacing defective part(s). For non- warranty replacement, the customer will be charged the current rate at the time of the receipt of unit; all shipping costs are the responsibility of the purchaser. Any replacement parts or units are warranted for the remainder of the original warranty period.

THERE IS NO OTHER EXPRESS WARRANTY. IMPLIED WARRANTIES, INCLUDING THOSE OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED TO THE DURATION OF THE EXPRESS LIMITED WARRANTY AND TO THE EXTENT PERMITTED BY LAW ANY AND ALL IMPLIED WARRANTIES ARE EXCLUDED. THIS IS THE EXCLUSIVE REMEDY AND LIABILITY FOR CONSEQUENTIAL AND INCIDENTAL DAMAGES UNDER ANY AND ALL WARRANTIES ARE EXCLUDED TO THE EXTENT EXCLUSION IS PERMITTED BY LAW. SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, OR THE LIMITATION OR EXCLUSION OF CONSEQUENTIAL OR INCIDENTAL DAMAGES, SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU.

This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

DeVilbiss User Manuals

Related Manuals