KERN OPTICS ORM2CO Digital Refractometer Instruction Manual
- June 6, 2024
- KERN OPTICS
Table of Contents
- General Information
- Introduction
- Display & Operating Buttons
- Preparing Before Operating
- Turn On & Measure
- Calibration
- Changing Scale & Temperatures Unit
- Turning off
- Cleaning & maintenance
- Disposal
- Technical data
- Error codes
- Models and scales
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
KERN OPTICS ORM2CO Digital Refractometer
Carefully read through the operating manual even if you have prior experience with KERN refractometers.
General Information
Intended use
The refractometer is a measuring instrument for determining the refractive
index of transparent substances in liquid or in some cases also in the solid
state. It is used to observe the behaviour of light as it passes from a prism
with known properties to the substance being tested.
Use of the refractometer for other purposes is contrary to its intended use
and may be hazardous. The manufacturer shall not be liable for any damages
caused by improper use.
Warranty
The warranty shall be void in the event of
-
Failure to observe the instructions in the
operating manual -
Use for purposes other than those described
-
Modifications or opening the device housing
-
Mechanical damage and/or damage resulting from media, liquids, natural wear and tear
This digital refractometer cannot measure any liquid that is highly corrosive to metal or glass. When measuring liquids that are corrosive to plastics or react chemically with plastics, be careful not to drop the measured liquid onto the shell. Otherwise it will corrode the shell.
Introduction
Description
Scope of delivery
1x Storage box | 1x Digital refractometer |
1x Operating manual | 1x AAA Battery 1.5 V |
1x Pipette | 1x Screwdriver
Display & Operating Buttons
Description display & operating buttons
Note: Please replace the battery when the is displayed.
Preparing Before Operating
Turn On & Measure
Turn on
Note :
- When used outdoors, please avoid strong light so as not to affect the measurement accuracy.
- Please keep the instrument in a stable and still statement and position.
Measure
After turning on, clean the sample tank with distilled water and then dry it. Now fill the sample up to the mark, close the cover and press “READ”.
Average value measurement
Press “READ” for 2 seconds. The device starts an automatic measurement series
of 15 measurements and shows the average value. Afterwards, the device
automatically turns back to the normal measuring mode.
Calibration
The refractometer can only be calibrated with distilled water. To do this, fill the sample tank with distilled water up to the mark and close the cover.
While “CAL” is flashing in the display, press “CAL” again to start the calibration. When the calibration is finished, the display shows “End”. After approx. 10 seconds, the device automatically returns to normal mode.
If the calibration was not completed successfully, an error code appears in the display. Here, for example, A01.
Changing Scale & Temperatures Unit
Changing scale
Changing temperatures unit
If exceed the temperature limitations, the signs “HHH” or “LLL” would show.
Turning off
If without any operations for 1 minute, the instrument would be automatically turned off.
Cleaning & maintenance
- To avoid damages to the prism and the sample tank, clean them with distilled water after each use.
- Dry it with a soft cloth afterwards.
- Do not use hard or abrasive objects for cleaning.
- Do not leave any residue in the sample tank.
- If the refractometer is not going to be used for a longer time, remove the battery and store it at a cool and dry place.
Disposal
The packaging consists of environmentally friendly materials which can be
disposed of via local recycling facilities.
The device and storage box should be disposed of by the operator in accordance
with applicable national or regional regulations at the place of use.
NOTE: In accordance with the Battery Ordinance (Batt V), batteries must not be disposed of in household waste. The end user is legally obliged to return them.
Technical data
Scale + accuracy + resolution | Depends to the model |
---|---|
Temperature | 0,0 – 40,0 °C / 32,0 – 104,0 °F |
Automatic Temperature Compensation | Yes |
Minimum sample volume | 0.2 – 0.3 ml (Marking ring) |
AUTO-OFF | 60 seconds |
Averaging measurement | 15 measurements |
Battery | 1 × AAA 1.5 V |
Lifetime of the battery | Approx. 10.000 measurements |
Overall dimensions L×W×H | 125×65×30 mm |
Net weight | 140 g (without battery) |
Error codes
code | Instructions |
---|---|
A01 | Beyond the scope of calibration temperature.(0.0℃~40.0℃) |
A02 | During calibration, no solution or solution wrong. |
A03 | This instrument has a hardware failure. |
Models and scales
| Model| Scale| No.| Range| Unit| Resolution| Accuracy
---|---|---|---|---|---|---|---
| ORM 50BM| Brix| S01| 0.0~50.0| %| 0.1%| ±0.2%
Refractive Index| S02| 1.3330~1.4200| nD| 0.0001nD| ±0.0003nD
ORM 1RS| Brix| S01| 0.0~90.0| %| 0.1%| ±0.2%
Refractive Index| S02| 1.330~1.5177| nD| 0.0001nD| ±0.0003nD
Fructose| ORM 1SU| Fructose| S01| 0.0~68.9| %| 0.1%| ±0.2%
Glucose| S02| 0.0~59.9| %| 0.1%| ±0.2%
Brix| S03| 0.0~90.0| %| 0.1%| ±0.2%
Refractive Index| S04| 1.3330~1.5177| nD| 0.0001nD| ±0.0003nD
ORM 2SU| Lactose| S01| 0.0~16.5| %| 0.1%| ±0.2%
Maltose| S02| 0.0~15.6| %| 0.1%| ±0.2%
Dextran| S03| 0.0~10.6| %| 0.1%| ±0.2%
Brix| S04| 0.0~50.0| %| 0.1%| ±0.2%
Honey| ORM 1HO| Honey Water| S01| 5.0~38.0| %| 0.1%| ±0.2%
Honey Baume| S02| 33.0~48.0| °Bé| 0.1| ±0.2
Brix| S03| 0.0~90.0| %| 0.1%| ±0.2%
Refractive Index| S04| 1.3330~1.5177| nD| 0.0001nD| ±0.0003nD
Salinity| ORM 1NA| Salinity (NaCl) %| S01| 0.0~28.0| %| 0.1%| ±0.2%
Salinity (NaCl) ‰| S02| 0~280| ‰| 1‰| ±2‰
Specific Weight| S03| 1.000~1.220| –| 0.001| ±0.002
Brix| S04| 0.0~50.0| %| 0.1%| ±0.2%
Refractive Index| S05| 1.3330~1.4200| nD| 0.0001nD| ±0.0003nD
ORM 1SW| Salinity Seawater| S01| 0~100| ‰| 1‰| ±2‰
Chlorinity Seawater| S02| 0~57| ‰| 1‰| ±2‰
Specific Weight| S03| 1.000~1.070| –| 0.001| ±0.002
Brix| S04| 0.0~50.0| %| 0.1%| ±0.2%
Refractive Index| S05| 1.3330~1.4200| nD| 0.0001nD| ±0.0003nD
Alcohol| ORM 1AL| Alcohol Mass.| S01| 0~72| %| 1%| ±1%
Alcohol Vol.| S02| 0~80| %| 1%| ±1%
Brix| S03| 0.0~50.0| %| 0.1%| ±0.2%
Refractive Index| S04| 1.3330~1.4200| nD| 0.0001nD| ±0.0003nD
Beer| ORM 1BR| Plato| S01| 0.0~30.5| °P| 0.1| ±0.3
SG Wort| S02| 1.000~1.130| –| 0.001| ±0.002
Brix| S03| 0.0~50.0| %| 0.1%| ±0.2%
Refractive Index| S04| 1.3330~1.4200| nD| 0.0001nD| ±0.0003nD
Wine| ORM 1WN| Oechsle| S01| 0~150| °Oe| 1| ±2
Vol%| S02| 0.0~22.0| %| 0.1%| ±0.2%
KMW (Babo)| S03| 0.0~25.0| –| 0.1| ±0.2
Brix| S04| 0.0~50.0| %| 0.1%| ±0.2%
ORM 2WN| Oechsle France| S01| 0~230| °Oe| 1| ±2
Vol%| S02| 0.0~22.0| %| 0.1%| ±0.2%
KMW (Babo)| S03| 0.0~25.0| –| 0.1| ±0.2
Brix| S04| 0.0~50.0| %| 0.1%| ±0.2%
Coffee| ORM 1CO| Coffee TDS 1| S01| 0.0~25.0| –| 0.1| ±0.2
Brix| S02| 0.0~50.0| %| 0.1%| ±0.2%
Refractive Index| S03| 1.3330~1.4200| nD| 0.0001nD| ±0.0003nD
ORM 2CO| Coffee TDS 2| S01| 0.00~25.00| –| 0.01| ±0.20
Brix| S02| 0.00~30.00| %| 0.01%| ±0.20%
Refractive Index| S03| 1.3330~1.4200| nD| 0.0001nD| ±0.0003nD
Urine| ORM 1UN| Urine Human| S01| 1.000~1.050| –| 0.001| ±0.002
Serum Protein| S02| 0.0~12.0| g/100ml| 0.1| ±0.2
Brix| S03| 0.0~50.0| %| 0.1%| ±0.2%
Refractive Index| S04| 1.3330~1.4200| nD| 0.0001nD| ±0.0003nD
ORM 2UN| Urine Cat| S01| 1.000~1.060| –| 0.001| ±0.002
Urine Dog| S02| 1.000~1.060| –| 0.001| ±0.002
Brix| S03| 0.0~50.0| %| 0.1%| ±0.2%
Refractive Index| S04| 1.3330~1.4200| nD| 0.0001nD| ±0.0003nD
Car / Industry| ORM 1CA| Cleaner| S01| (-60.0)~0.0| °C| 0.1℃| ±0.5℃
AdBlue®| S02| 0.0~51.0| %| 0.1%| ±0.2%
Battery Fluid| S03| 1.000~1.500| –| 0.001| ±0.005
Brix| S04| 0.0~50.0| %| 0.1%| ±0.2%
Refractive Index| S05| 1.3330~1.4200| nD| 0.0001nD| ±0.0003nD
Ethylenglycol (%)| S01| 0.0~100.0| %| 0.1%| ±0.5%
Ethylene Glycol (°C)| S02| (-50.0)~0.0| ℃| 0.1℃| ±0.5℃
Propylene Glycol (%)| S03| 0.0~100.0| %| 0.1%| ±0.5%
ORM 2CA| Propylenglycol (°C)| S04| (-60.0)~0.0| °C| 0.1℃| ±0.5℃
Brix| S05| 0.0~90.0| %| 0.1%| ±0.2%
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>