blackline safety G7 DOCK Bump Test and Calibration Station
OVERVIEW
CONNECTED GAS CALIBRATION AND BUMP TESTING
What is G7 Dock?
G7 Dock is Blackline Safety’s solution to gas sensor calibrating, bump testing
and charging G7. It supports both G7c and G7x devices with single-gas, multi-
gas diffusion or multi-gas pumped cartridges. Simply insert G7, and G7 Dock
will do the rest. It automatically charges G7 and places G7 into a test-ready
mode to eliminate false alerts. G7 Dock helps you keep fully compliant
according to your business calibration and bump testing policies, so that it
always has your back.
BLACKLINE SAFETY NETWORK
How am I connected?
G7 Dock is not inherently connected to the Blackline Safety Network. A
connected G7 device placed in G7 Dock communicates to the network and to G7
Dock.
The Blackline Safety Network is the cloud-hosted system used to monitor your
safety. It includes 2G, 3G and 4G cellular networks, satellite networks, our
Blackline Live™ portal application, your monitoring account and your personal
safety monitoring device. Each device requires an active service plan to
connect to the Blackline Safety Network.
BLACKLINE LIVE WEB PORTAL
What is Blackline Live?
The cloud-hosted Blackline Live portal monitors employee safety and allows you
to manage your devices. Blackline Live keeps track of G7 and G7 Dock events,
including alert history, gas calibrations, and bump tests, and eliminates the
need to manually retrieve data logs from the field. All G7 data is
communicated in real-time. Blackline Live tailors user access depending on
employee roles. This ensures that everyone has access to the right tools to
accommodate their role in a comprehensive monitoring program.
WHAT’S IN THE BOX
Your G7 Dock comes with the following components:
- G7 Dock
- Hose
- Quick-connects (four male, one female)
- Power components:
- USB cable
- USB power adapter
- Dock filter
- Multi-lingual sticker sheet
- G7 Dock getting started guide
HARDWARE DETAILS
GAS INLET CONFIGURATION
The back of G7 Dock is equipped with a Gas Out exhaust and four gas inlets with quick-connects. The inlets are pre-programmed for you.
How are my inlets configured?
When looking at the back of G7 Dock, the Gas Out exhaust is the connector on
the far left side of the dock. It is used to transfer calibration gas to a
safe exhaust location.
NOTE: You should never plug gas into the Gas Out exhaust.
The next three inlets (to the right of the Gas Out exhaust) are configured to support your selected gases and gas mixtures. The table belows shows an example of a configuration with and without Chlorine (Cl2). Your configuration may be different depending on the gases configured in Blackline Live for your site. NOTE: For G7 Docks with serial numbers beginning with 8247 and higher, inlet 1 has been optimized for use with reactive gases such as Chlorine (Cl2). If you are configuring G7 Dock for use with Cl2, you must use inlet 1 for the Cl2.
Inlet 4 is configured as a purge inlet, supplying clean air to remove residual gas from G7 Dock. When using dock in dusty or humid environments, you can attach the optional dock filter subassembly to the purge inlet to ensure dock is being purged with clean air.
The dock filter is made up of four pieces:
- a disposable filter
- a luer fitting
- 2” hose
- one of the provided male quick connects
The filter piece may need to be replaced as it absorbs contaminants from the
air.
Can I customize my inlets?
All four inlets are customizable and a multi-lingual sticker sheet is provided
to label customized inlets. Changes to your inlet configuration can be made in
G7 Dock’s configuration profile, from the Docks page on Blackline Live. The
activation code on the back of Dock can be used to identify it from the list.
If you are having difficulties accessing this page or finding G7 Dock in
Blackline Live, please contact our Customer Care team.
Reactive gases
If you are using reactive gases (Cl2, HCN, NO2, NH3, SO2, or H2S) with G7
Dock, you must use either a Viton-lined (ACC-DFR-V) or stainless steel demand
flow regulator and a Teflon-lined hose (ACC-FEP-T2).
How do I connect G7 Dock to a gas cylinder?
G7 Dock provides quick-connects that attach to one end of the hose. Join the
hose to the configured inlet at the back of G7 Dock by connecting both quick-
connect ends and twisting clockwise until it clicks securely into place.
Ensure the other end of the hose is attached to a demand flow regulator on the
gas cylinder.
REQUIREMENTS
What do I need for G7 Dock to operate?
- G7c or G7x personal safety monitoring device (single-gas or multi-gas cartridge)
- Gas cylinder with specific single-gas or multi-gas mixture
- Teflon-lined hose (ACC-FEP-T2) is calibrating with Chlorine (Cl2)
- 120/240 VAC Power source and power components
NOTE: You must use the power components provided with your G7 Dock.
HOW IT WORKS
How do I power G7 Dock?
Insert the power cable into the port on the bottom of G7 Dock. Feed the power
cable through the tracks along the bottom to allow the dock to lie flat on a
table surface or mount on a wall. When plugged in, the green light at the
charge clip will turn on to indicate power.
How do I open G7 Dock?
Press inward on the two red release tabs until dock’s lid clicks
open.
How do I place my G7 device into G7 Dock?
- Slide G7 device into G7 Dock’s charge clip at an angle. Ensure the charge clips of both devices are connected.
- Keeping the charge clips connected, lie G7 down into G7 Dock and push down until it clicks in place.
- Close G7 Dock’s lid and press down firmly until the release tabs click into place on both sides.
FEATURES
CHARGING G7
How does G7 Dock charge my G7 personal safety monitor?
G7 devices will automatically start charging when placed into a powered G7
Dock. It will take a maximum of four hours to fully charge G7.
NOTE: If G7 is not charging, confirm that G7 Dock is powered, the charge
clip is clean of debris and G7 device is properly placed on the dock charge
clip.
BUMP TEST
What is a bump test?
It is safe practice to regularly test gas sensors by applying the target gas.
G7’s bump test also tests the operation of light, sound and vibration
indicators. The bump test schedule depends on your company’s safety policy. G7
automatically communicates bump test data to the Blackline Safety Network, and
will remind you when a bump test is overdue. The bump test schedule is
configurable.
NOTE: To meet CSA LEL performance standard, you are required to bump test
before each day’s use. Blackline recommends you do not exceed 30 days without
a bump test.
Overdue bump test
If your G7 device is overdue for a bump test according to its configuration,
it will automatically begin a bump test when placed into G7 Dock.
How do I perform a bump test that is not overdue?
- Power on your G7 device.
- Place G7 into G7 Dock.
- Close the lid until it clicks in place.
- Press G7’s arrow buttons to navigate to Bump Test on the LCD screen menu and press G7’s OK button to select. G7 Dock will perform the bump test.
- G7 will indicate whether the bump test has passed or failed and when it is ready to be removed from G7 Dock.
CALIBRATION
What is calibration?
Gas sensors periodically need to be calibrated by applying a known
concentration of gas for a set amount of time. This procedure ensures the gas
sensor can accurately detect gas levels throughout its operating life. The
calibration schedule depends on your company’s safety policy. Blackline
recommends not exceeding 180 days without a calibration.
Calibration settings
G7 devices have their own gas calibration settings, which can be modified in
Blackline Live. If the device’s configured gas settings do not match G7 Dock’s
settings, the calibration will continue using Dock’s settings as a default. As
long as G7 Dock is configured correctly to the gas concentrations being used,
the calibration will be successful.
Overdue calibration
If your G7 device is overdue for a gas calibration, it will automatically
begin calibrating when placed into G7 Dock.
How do I perform a calibration that is not overdue?
- Power on your G7 device.
- Place G7 into G7 Dock.
- Close the lid until it clicks in place.
- Press G7’s arrow buttons to navigate to Calibration on the LCD screen menu and press G7’s OK button to select. G7 Dock will perform the calibration.
- G7 will indicate whether the calibration has passed or failed and when it is ready to be removed from G7 Dock.\
G7 DOCK UPDATE
Updates to your Dock’s settings can be made in your G7 Dock’s configuration
profile, from the Docks page on Blackline Live. The activation code on the
back of Dock can be used to identify it from the list. If you are having
difficulties accessing this page or finding G7 Dock in Blackline Live, please
contact our Customer Care team.
NOTE: G7 Dock does not need to be connected to the Blackline Safety
Network to charge a G7 device or complete calibrations and bump tests, but the
G7 device does need to be connected to the network to update G7 Dock
configuration settings.
MULTI DOCK SET-UP
What is a multi dock set-up?
In situations where multiple safety monitoring devices require calibration and
bump testing, docks can be joined together to perform these functions across
all connected devices. Multiple G7 Docks can be connected to a single gas
cylinder by using tubes, check valves and T-connectors. One gas cylinder is
able to support five G7 Docks at one time. Blackline offers a multi-dock setup
kit to help with multi-dock setup. To order, please contact our Customer Care
team or your distributor.
WARNING: Do not use a multi-dock set-up if you are calibrating with
Chlorine (Cl2). Calibration with Cl2 requires that the hose length be kept as
short as possible, which is not supported by a multi-dock set-up.
What do I need for a multi dock set-up?
- G7c or G7x personal safety monitoring devices (single-gas or multi-gas cartridge)
- G7 Docks for each device
- Hose (1/8” inner diameter)
- Nylon or polyethylene T-connectors
- Nylon check valves
- Gas cylinder with specific single-gas or multi-gas mixture
- 120/240 VAC Power source for each dock
How do I calibrate and bump test multiple devices?
G7 Docks in a multi dock set-up act independently of each other – calibrations
and bump tests are carried out in the same manner that they would be if they
were set up separately. This means that G7 Docks connected to the same gas
cylinder can do calibrations and bump tests at the same time without
interfering with each other.
MOUNTING
G7 Dock can be placed flat on a table or mounted to a wall.
WALL-MOUNTING
How do I mount my G7 Dock on a wall?
When mounting G7 Dock to a wall, ensure the location is within reach of a
power source and gas cylinder.
To mount G7 Dock, remove the mounting guide located on page 23. Place the
mounting guide on wall to plan where to place the mounting screws. Attach
screws to the wall according to the ‘+’ marks.
Align G7 Dock wall-mounting holes (located on the back of the dock) with the
screws and slide down to lock in place.
MAINTENANCE
How do I care for my G7 Dock?
To make sure G7 Dock is in working condition, ensure the following:
- Gas valves inside the G7 Dock lid move easily and are free of debris
- Gas inlets and outlet are free of debris
- Charge clip is free of debris
To clean G7 Dock, disconnect from power and wipe down with a damp cloth. Do not use pressure spray or cleaning solvents. For best performance, G7 Dock should be operated in a dry and dust-free environment.
TROUBLESHOOTING
SETUP
Why is G7 telling me to close the lid when I have already closed it?
If your device is telling you to close the lid when you try to run a bump test
or calibration, it is likely that the lid has not been closed properly. Ensure
that you press down firmly on both sides of the lid so that both release tabs
click into place.
Why is my device unresponsive when I put it into G7 Dock?
If your device does not respond when placed into G7 Dock, it may not be
properly connected to the dock, or the dock may not be connected to a power
supply. Remove your device from G7 Dock and check to see if the green power
light is on.
If… Dock’s green light is on:
Your device may not be properly connected to G7 Dock. Ensure that:
- Charge clips on both the device and dock are connected
- Device clicks into place when set down into the dock
- Lid is properly closed. Ensure that you press down on both sides so that both release tabs click into place.
- Charge clips are clean and free of debris
Dock’s green light is off:
G7 Dock is likely not connected to a power supply. Ensure that:
- Power cable is properly plugged into dock and the power adaptor
- Power source being used is 120/240 VAC
CALIBRATION AND BUMP TESTS
Why are my calibrations and bump tests failing?
Calibrations and bump tests can fail if there is not enough gas reaching the
sensors. To ensure that gas is flowing from the gas cylinder to G7 Dock, check
that:
- There are no kinks or crimps in the hose
- The gauge on the gas cylinder indicates there is gas inside (i.e., check that the gas cylinder is not empty)
- The gas cylinder is using a demand-flow regulator
- The quick connects are properly connected to G7 Dock
- Gas concentrations in the gas cylinder match Dock configurations
If G7 Dock is still not operating correctly, please contact Customer Care.
SPECIFICATIONS
DETAILED SPECIFICATIONS
-
Size & Weight
Size: 215mm x 102mm x 95mm (8.46” x 3.94” x 3.74”) Weight: 202g (7.1oz)
Gas configuration 4 customizable gas inlets, 1 gas outlet -
Bump test
25 second bump test, less than 10 seconds of gas applied during a bump -
Calibration|
4 minute calibration, less than 2 minutes of gas applied during a calibration -
Gas Usage
Optimized gas delivery (less than 500mL/min) -
Power Supply
Blackline requires using the power supply provided with your G7 Dock. Input voltage: 5 VDC @ 1000 mA Power connector: Micro-USB 120/240 VAC power adaptor and USB cable included. -
Regulatory Compliance RCM
CAN ICES-3(B)/NMB-3(B) FCC CE
Features -
Over-the-air (OTA) configuration updates
-
Automatic Bump
-
Automatic Calibration
-
Dock usage reports
-
Gas usage reports
Warranty
Two-year warranty
SAFETY PRECAUTIONS
DANGER
Do not use G7 Dock in areas classified as hazardous locations, where there
is risk of explosion due to presence of gas, vapor, or dust. G7 Dock is not
certified as intrinsically safe. Do not place G7 Dock in or near open flame,
or submerge in liquid.
WARNING
Electrical equipment may be hazardous if misused. Seek advice from your
local electronics recycling authority regarding the disposal of your device.
Do not dispose Blackline products in your household trash.
LEGAL NOTICES
Information in this document is subject to change without notice. This
document is provided “as is” and Blackline Safety Corp. (“Blackline”) and its
affiliated companies and partners assume no responsibility for any
typographical, technical or other inaccuracies in this document. Blackline
reserves the right to periodically change information that is contained in
this document. However, Blackline makes no commitment to provide any such
changes, updates, enhancements or other additions to this document to you in a
timely manner or at all. Copyright © 2019 Blackline Safety Corp. All rights
reserved. Except as expressly provided herein, no part of this manual may be
reproduced, copied, transmitted, disseminated, downloaded, or stored in any
storage medium, for any purpose without the express prior written consent of
Blackline Safety Corp. (“Blackline”). Blackline hereby grants permission to
download a single copy of this manual onto some form of electronic storage
medium to be viewed and to print one copy of this manual or any revision
hereto, provided that such electronic or printed copy of this manual must
contain the complete text of this copyright notice. Further, any unauthorized
commercial distribution of this manual or any revision hereto is strictly
prohibited. The Blackline, Alert. Locate. Respond. families of related marks,
images and symbols, including Blackline, G7, G7c, G7x, G7 Bridge, G7 Dock,
LiveResponse, Loner, Loner IS, Loner IS+, Loner M6, Loner M6i, Loner Mobile,
Loner 900, and SureSafe are the exclusive properties and trademarks of
Blackline Safety Corp. All other brands, product names, company names,
trademarks and service marks are the properties of their respective owners.
FCC Compliance
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Note: the grantee is not responsible for any changes or modifications not expressly approved by the party responsible for compliance. Such modifications could void the user’s authority to operate the equipment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for further assistance.
Warranty
Your G7 Dock is warranted against defects in materials and workmanship for up
to two year from date of purchase. For further details regarding your
Blackline warranty, please refer to your Terms and Conditions of Service.
Visit http://www.blacklinesafety.com for
more information.
SUPPORT
LEARN MORE
Visit support.BlacklineSafety.com to find support and training materials for G7 Dock.
CUSTOMER CARE
For technical support, please contact our Customer Care team. North America (24 hours) Toll Free: 1-877-869-7212 | support@blacklinesafety.com United Kingdom (8am-5pm GMT) +44 1787 222684 | eusupport@blacklinesafety.com International (24 hours) +1-403-451-0327 | support@blacklinesafety.com