Bluelab METCOMPLUS Combo Meter User Guide
- June 5, 2024
- bluelab
Table of Contents
- Bluelab METCOMPLUS Combo Meter
- Features
- Bluelab Combo Meter
- Introduction
- Preparing for use
- Calibration
- Calibration cont.
- Changing nutrient and temperature display units
- Measuring hydroponic elements
- Battery replacement
- Cleaning the conductivity/temperature probe
- Cleaning the Bluelab pH Probe
- Hydrating the pH probe
- Storing the Bluelab Combo Meter
- Technical specifications
- Troubleshooting guide
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Bluelab METCOMPLUS Combo Meter
Features
Measures pH, conductivity/nutrient (EC, CF, ppm 500 and ppm 700) and
temperature (°C, °F)
Lightweight and portable| Replaceable double junction pH probe
Large easy to read display| Over range and under range indicators
Simple push button pH calibration| Low battery indicator
Successful pH calibration indicator| 2 x AAA alkaline batteries included
No calibration required for conductivity and temperature| Auto off function
Bluelab Combo Meter
Introduction
The Bluelab Combo Meter has three press buttons; ‘pH /
calibrate’, ‘nutrient / units’ and ‘temp / units’. The
buttons have a short press and long press function. A
short press means a button is released in about one
second. The long press is a button being held for at least
three seconds and released when the display starts flashing.
Turning the Combo Meter on
1 A short press of any button turns the Combo Meter on. The Combo Meter
automatically turns off after four minutes if no buttons are pressed. If the
Combo Meter turns off before the reading is taken, a short press of any button
will turn the Combo Meter on again.
Preparing for use
The following tasks must be performed before the Bluelab Combo Meter is used for
the first time.
1 Insert batteries
See section 6.0.
|
Inserting Twisting Attached
Attaching the
Bluelab pH Probe to the Meter
Removing pH probe storage cap
Ensure probe tip is covered by the KCl storage solution in cap
See section 3.0 for calibration steps
2 Connect pH probe
Connect the pH probe to the pH meter by lining up the lugs of the BNC fittings. Fasten securely by pushing the pH probe connector on and twisting one quarter turn.
3 Remove the storage cap
Remove the pH probe storage cap by gripping the top of the cap and gently twisting the base one rotation clockwise to loosen slightly.
Next slowly slide the cap off the pH probe.
DO NOT completely remove the base of the cap from the top of the cap.
CAUTION: When the pH probe is not in use, add enough Bluelab pH Probe KCl Storage Solution to the storage cap so the probe tip is covered. Then replace the cap and store in a secure place.
DO NOT use RO (Reverse Osmosis), Distilled or De-ionized water. Pure water changes the chemistry in the reference, causing the probe to die.
4 Calibrate the pH
Calibrate the combo meter by following the instructions in section 3.0 of this manual. This must be done before the
Combo Meter is used for the first time.
Calibration
pH calibration is required before first use to ensure that the first reading is accurate.
• The Bluelab Conductivity/Temperature Probe DOES NOT require calibration.
• The pH of the Bluelab Combo Meter DOES require calibration.
For accurate pH readings the pH probe is
cleaned and recalibrated when:
• The reading is different to what you were expecting.
• The batteries have been removed or changed.
• The pH probe is replaced with a new one or is disconnected from the Combo Meter.
• The pH calibration indicators have disappeared.
When calibrating the pH after first use the pH probe needs to be cleaned. See pH probe
cleaning in section 8.0.
For best pH calibration
pH reading accuracy is dependant on the
accuracy and age of the calibration solutions used,
and use and cleanliness of the pH probe tip.
• Ensure the pH probe has been cleaned and rinse the pH probe with clean water between calibration solutions to reduce contamination of the pH solutions.
• Only fresh uncontaminated solutions should be used.
• Calibrate the pH at the same temperature as the solution to be measured.
• ALWAYS calibrate the pH probe with pH 7.0 then pH 4.0 or pH 10.0.
NOTE: The conductivity/temperature probe does not need to be calibrated, but must be cleaned to remove any build up of nutrient salts. See Section 7.0.
The pH calibration involves cleaning the pH
probe tip and then calibrating in TWO SOLUTIONS.
If a reading below pH 7.0 is expected, use pH 7.0 and pH 4.0 calibration
solutions. If a reading above pH 7.0 is expected, use pH 7.0 and pH 10.0
calibration solutions. Follow the steps on the following page for Combo Meter
pH calibration.
Storage and use of calibration solutions
• Always place the lid back onto the bottle after use or evaporation will
occur rendering the solution useless.
• Store in a cool place.
• DO NOT measure directly into the bottle. Tip a small amount into a clean container and discard after use.
• Never add water to solutions.
pH reading accuracy is dependant on the accuracy and age of the calibration solutions used, and use and cleanliness of the pH probe tip.
Calibration cont.
To calibrate the pH
1 Clean pH probe tip.
See Section 8.0 (the pH probe does not require
cleaning before the first use).
|
clean the pH probe
pH 7.0
calibration solution
rinse in tap water
pH 4.0 or
pH 10.0
calibration solution
Successful pH 7 and pH 4 calibration
Ensure probe tip is covered by the KCl storage solution in cap
2 pH 7.0 calibration
a) Press any button to turn the pH meter ON.
b) Loosen and remove cap from probe.
c) Rinse pH probe tip in fresh water.
d) Press the CAL button until CAL appears on the screen and release.
e) A series of flashing ‘o’s will appear. Once calibration is complete, PH7 will be displayed.
f) If Err appears during the calibration process see Section 11.0.
g) The combo meter must be calibrated to two points. If after an hour the meter has not been calibrated with a second calibration point
the calibration indicators disappear and the combo meter reverts to an uncalibrated state. Calibration is required.
3 pH 4.0 / 10.0 calibration
a) Rinse pH probe tip in fresh water.
d) Press the CAL button until CAL appears on the screen and release.
e) A series of flashing ‘o’s will appear. Once calibration is complete, PH4 or PH10 will be displayed.
f) Rinse pH probe tip in fresh water.
c) The combo meter is now calibrated and ready for use.
d) The successful calibration indicator ‘ ’ or ‘ ’ will appear on screen.
d) Probes require monthly cleaning and calibration.
4 Store the pH probe
Add enough Bluelab pH Probe KCl Storage Solution into the probe storage cap to fully submerge the pH probe tip. Place storage cap on probe.
Changing nutrient and temperature display units
Nutrient and temperature can be displayed in different
units. The units available are shown in the tables below.
Changing nutrient units displayed
1 To select unit
a) Press and hold the ‘nutrient/units’ button until the display starts flashing.
Release, then short button press the same button to scroll through the available units.
Release when the required unit is displayed.
b) The display flashes four times after the last button press then returns back to a normal display, showing the selected unit.
|
Changing temperature units displayed
1 To select unit
a) Press and hold the ‘temp/units’ button until the display starts flashing.
Release, then short button press the same button to scroll through the available units.
Release when the required unit is displayed.
b) The display flashes four times after the last button press then returns back to a normal display, showing the selected unit.
|
Measuring hydroponic elements
The values measured in a hydroponic solution by the Bluelab Combo Meter
include nutrient (conductivity), temperature and pH levels.
Taking a conductivity/nutrient reading
1 Short press the ‘nutrient/units’ button to select conductivity.
2 Insert the conductivity/temperature probe into solution where there is
strong movement of the solution, or stir the solution with the
conductivity/temperature probe.
3 __ Wait 1-2 minutes or longer for the conductivity/temperature probe to
reach solution temperature. The conductivity reading will stabilize to a
constant value.
4 If the solution you are measuring is outside the measurement range of the
Combo
Meter, an ‘Or’ (over range) will be displayed.
NOTE: The conductivity/temperature probe tip should be cleaned at least once a
month to remove salt build-up ensuring accurate readings. If oily additives
are being used the conductivity/temperature probe tip should be cleaned after
each use. See Section 7.0.
Taking a temperature reading
1 Short press the ‘temp/units’ button to select temperature.
2 Insert conductivity/temperature probe into the solution.
3 Wait 1-2 minutes or longer for conductivity/temperature probe to reach
solution temperature. The temperature reading will stabilize to a constant
value.
4 If the solution you are measuring is outside the measurement range of the
Combo
Meter, a ‘Ur’ (under range) or ‘Or’ (over range) will be displayed
NOTE: For very cold or very hot temperatures it will take 4-5 minutes for the
conductivity/ temperature probe to reach solution temperature. To help
decrease time taken to reach the solution temperature, place the
conductivity/temperature probe in an area where there is strong movement of
solution, or stir solution with the conductivity/temperature probe.
Taking a pH reading
1 Short press the ‘pH/calibrate’ button to select pH.
2 Remove the storage cap from the pH probe and place the pH probe into the
solution.
3 Wait 1-2 minutes or longer for reading to stabilize to a constant value.
The pH reading is displayed.
4 If the solution you are measuring is outside the measurement range of the
Combo
Meter, a ‘Ur’ (under range) or ‘Or’ (over range) will be displayed.
NOTE: If taking readings of more than one solution, rinse the pH probe tip thoroughly in fresh water between solutions to avoid cross contamination.
Battery replacement
Batteries are replaced in the Bluelab Combo Meter when
the low battery indicator appears on screen. The low
battery indicator remains on and the Bluelab Combo Meter
continues to operate until the batteries die or are
replaced.
1 Open battery compartment by sliding the back cover down and insert 2 x
AAA batteries as shown on the battery holder. Slide cover back on.
NOTE: Alkaline batteries are recommended.
|
Battery cover
2 IMPORTANT: Check the batteries at least once every six months for signs of deterioration, rusting or swelling.
If signs of deterioration are found, clean battery holder contacts and replace batteries.
Cleaning the conductivity/temperature probe
Cleaning the conductivity/temperature probe periodically ensures accurate readings. The conductivity/temperature probe is cleaned using the Bluelab Conductivity Probe Cleaner, or “Jif” a trade name for a liquid scourer cream used in home bathrooms and kitchens. Similar products are called “Liquid Vim”, “Soft Scrub”, “Cif cream”, or “Viss”.
Never use scented varieties as they contain oils that contaminate the conductivity/ temperature probe.
Follow the steps below to clean the conductivity/temperature probe.
1 Remove shroud.
Warm the shroud in your hand for a few seconds to help with removal. Hold the body and pull the shroud off.
|
1 2
3
3
clean, smooth uneven beads
film of water of water
5
2.77 EC Conductivity Standard Solution
2 Clean the conductivity probe face.
Place one or two drops of Bluelab Conductivity Probe Cleaner onto the probe face and rub with the Bluelab Chamois or your finger firmly and vigorously.
3 Rinse the conductivity probe face.
Rinse off all traces of cleaner under running tap water while scrubbing the probe face with the other side of the Bluelab Chamois or the same finger.
4 Check that the water forms a smooth film on the probe face. Ensure you have a clean, smooth film without any beads of water.
If you have beads of water, repeat steps 2 and 3.
5 Refit the shroud and test in 2.77 EC Conductivity Standard Solution to ensure adequate cleaning.
Place the probe tip into the solution, wait for the reading to stabilize to a constant value. This can take a few minutes while the probe adjusts to the temperature of the solution.
Repeat the cleaning process if the reading given is not within 0.1 EC, 1 CF, 50 ppm or 70 ppm of the values in the table below.
Testing the Bluelab Conductivity/Temperature Probe
The conductivity/temperature probe is tested in either
2.77EC/27.7CF/1385 ppm or 1940 ppm solution depending on the unit of
conductivity chosen.
Use the standard solutions in the table to the
right. Bluelab solutions are recommended.
NOTE: The shroud MUST be left on the probe when taking readings.
Cleaning the Bluelab pH Probe
To ensure accurate readings the pH probe tip needs to be rinsed in water after each use and cleaned prior to calibration using the following instructions.
The storage cap must always be put back on after cleaning. Always ensure it contains enough Bluelab pH Probe KCl Storage Solution to cover the probe tip.
1 Remove storage cap from pH probe. Hold the top of the storage cap, twist the cap to loosen then remove.
2 Rinse pH probe tip under fresh tap water.
Never use RO (Reverse Osmosis), Distilled or De-ionized water.
|
1 2
3 Fill a small plastic container with clean tap water.
Add a small amount of Bluelab pH Probe Cleaner or mild detergent (dishwashing liquid).
4 Gently stir the probe tip in the mixture. Ensure that you do not ‘knock’ the soil pH probe on the side of the container as this may cause damage to the probe.
Rinse well under fresh running water to remove all traces of the detergent mixture.
|
3 4
5 If the probe tip requires removal of heavy contamination:
Gently brush around the glassware with a few drops of Bluelab pH Probe Cleaner or mild detergent (dishwashing liquid) and a soft toothbrush.
|
5
6 Rinse well under fresh running tap water to remove all traces of the detergent mixture.
7 Calibrate pH probe after cleaning, see section 3.0
After calibration, store pH probe in the storage cap, ensuring there is enough KCl Storage Solution to cover the probe tip.
|
6 7
Hydrating the pH probe
Hydrate the pH probe in Bluelab pH Probe KCl Storage Solution when:
• the probe tip has not always been stored in KCl storage solution, to improve the reading response speed.
• the probe tip has been accidentally allowed to dry out.
Never use RO (Reverse Osmosis), De-ionized or Distilled water.
Pure water changes the chemistry in the reference, causing the probe to die.
1 Loosen, then remove the storage cap. Place the pH probe upright in a plastic container.|
Bluelab pH Probe KCl Storage Solution
2 Clean the pH probe tip.
Ensure the probe tip is cleaned before hydrating. See section 8.0 for instructions.
3 Add enough Bluelab pH Probe KCl Storage Solution to a plastic container
to submerge the pH probe tip.
4 Leave to soak for at least 24 hours.
After hydration, always calibrate the pH probe to ensure accuracy, see section 3.0.
Storing the Bluelab Combo Meter
1 Store the Combo Meter in a cool, dry and clean place when not in use.
2 Keep out of direct sunlight.
Keep Combo Meter out of direct sunlight to prevent irreparable damage to the LCD reading display.
3 __The Combo Meter is not waterproof but will withstand occasional water splashes.
If the Combo Meter is splashed, wipe dry as soon as possible.
4 __Remove batteries if the Combo Meter is
to be stored for a prolonged period.
5 Remove pH probe if storing the Combo Meter without use for longer than
two to three weeks and check regularly that the pH probe tip has not dried
out.
When storing the pH probe, the pH probe tip must be kept moist.
To prepare the pH probe for storage, add enough Bluelab pH Probe KCl Storage Solution to the storage cap so the probe tip is covered. Then replace the cap and store in a secure place. DO NOT use RO (Reverse Osmosis), Distilled or De- ionized water. Pure water changes the chemistry in the reference, causing the probe to die.
11.0 Error messages
An error message will only appear following pH calibration failure.
‘Err’ will be displayed for a few seconds then the display will show the previous reading. Successful pH calibration indicators will disappear. The Bluelab Combo Meter is in an uncalibrated state, therefore recalibration is required. See causes of Error messages below.
Possible causes for an ‘Err’ message: Calibration solutions contaminated Wrong solutions used
pH probe contaminated
pH probe not properly attached pH probe worn out or damaged
Calibrate to pH 7.0 FIRST then to pH 4.0/10.0
Technical specifications
| pH| Conductivity| Temperature
Measurement range
|
0.0 – 14.0 pH
| 0 – 9.9 EC
0 – 99 CF
0 – 4950 ppm 500 (TDS)
0 – 6930 ppm 700
|
0 – 50 °C
32 – 122 °F
Resolution
|
0.1 pH
| 0.1 EC 1 CF
10 ppm 500 (TDS)
10 ppm 700
|
1 °C
1 °F
Accuracy
(at 25 °C/
77 °F)
|
±0.1 pH
| ±0.1 EC
±1 CF
±50 ppm
±70 ppm
|
±1 °C
±2 °F
Calibration
| Two point pH 7.0 and pH 4.0
or pH 10.0
| Not required (factory calibrated)| Not required (factory calibrated)
Temperature compensation|
Not applicable
| Automatic temperature compensation|
Not applicable
Operating environment| 0 – 50 °C / 32 – 122 °F
Power source
|
2 x AAA alkaline batteries
Troubleshooting guide
Trouble| Reason| Correction
Nutrient reading low
| Contaminated conductivity/ temperature probe.| Clean
conductivity/temperature probe (see Section 7.0).
Solution temperature low/ high.| Wait 5 to 10 minutes for reading to stabilize
to a constant value.
Temperature reading inaccurate
| Temperature of conductivity/temperature probe different to solution temperature.| Wait 5 to 10 minutes for conductivity/ temperature probe to reach solution temperature.
pH reading inaccurate
| Contaminated pH probe / glassware not clean.| Clean pH probe (see Section
8.0); then calibrate (see Section 3.0).
Wick contaminated, blocked or dry.| Hydrate probe in KCl storage solution for
24 hours, see Section 9.0. Do not measure proteins or oils with this unit.
Replace unit.
Incorrect pH calibration.| Ensure calibration solutions are accurate. Replace
if in doubt.
Wait longer for readings to stabilize before calibrating to a constant value.
pH calibration unreliable.| Calibrate pH probe (see Section 3.0).
pH probe damaged or old.| Replace pH probe.
pH reading does not change from solution to solution| Broken glass bulb, tube
or connector.| Check pH probe for damage. Replace probe.
Displays low battery indicator| Insufficient power to take a
reliable reading.
| Replace the batteries.
DO NOT use rechargeable batteries.
No display
| Batteries dead or inserted incorrectly.| Check batteries are inserted correctly. Replace if necessary.
Display shows ‘Err’
| Problem with pH calibration.| See error message descriptions in Section 11.0
of this document.
Or Ur
While in pH mode
| Over range pH. Under range pH.| Solution > 14.0 pH. Solution < 0.0 pH. Check pH probe connection.
pH probe could be faulty.
Combo Meter could be wet inside.
Or Ur
While in temp mode
| Over range temperature. Under range temperature.| Solution >51 °C / 122 °F.
Solution <0 °C / 32 °F. Conductivity/temperature probe or Combo Meter faulty.
Or
While in conductivity/ nutrient mode
| Over range conductivity/nutrient.| Over range conductivity
9.9 EC, 99 CF, 4950 ppm 500, 6930
ppm 700. Conductivity/temperature probe or Combo Meter faulty.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>