Moisture Meter Balance PCE-MA Series
Thank you for purchasing a moisture balance from PCE Instruments.
Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non- observance of the manual are excluded from our liability and not covered by our warranty.
- The device must only be used as describes in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
- The instrument may only be used if the environmental conditions (temperature, relative humidity,…) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
- The case should only be opened by qualified PCE Instruments personnel.
- Never use the instrument when your hands are wet.
- You must not make any technical changes to the device.
- The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
- The device must only be used with accessories from PCE Instruments or equivalent.
- Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
- Do not use the instrument in explosive atmospheres.
- The limit values for the measuring variables as stated in the specifications must under no circumstances be exceeded.
- If possible, the device should be transported horizontally, bottom down and well-cushioned. Remove the loose attachment parts from the measuring chamber to avoid damage. We recommend to use the original packaging for transport.
- Only use IEC power connectors with a PE (ground) contact.
- Do not touch the halogen lamps or the cover during the measurement as this can cause severe burns.
- After the measurement, let the sample cool down before touching it.
- Never operate the device when the cover is open.
This instruction manual is published by PCE Instruments without any guarantee. We expressly point to our general guarantee terms which can be found in our general terms of business. If you have any questions please contact PCE Instruments.
Safety symbols
Safety-related instructions the non-observance of which can cause damage to
the device or personal injury carry a safety symbol.
Symbol | Designation / description |
---|---|
General safety symbol |
Non-observance can cause personal injury and/or damage to the device.
| Warning of hot surface
Non-observance can cause burns.
Specifications
Technical specifications
Model | PCE-MA 100 | PCE-MA 110 | PCE-MA 200 | PCE-MA 202 |
---|---|---|---|---|
Weighing range | 110 g | 110 g | 200 g | 200 g |
Readability | 1 mg / 0.001 g | 10 mg / 0.01 g | 1 mg / 0.001 g |
10 mg / 0.01 g
Heating element| Halogen lamp / round / approx. Ø 90 mm
Drying temperature adjustable| +40…+199 °C
Drying time (time mode)| 1…99 min adjustable
Measurement range moisture| 0…100 %
Readability moisture| 0.01 %| 0.1 %| 0.01 %| 0.1 %
Memory| 16 drying programmes (parameters)
Interface| RS-232
Display| LCD, 17 mm digit height
Power supply| 220 V / 50 Hz or 110 V / 60 Hz
Dimensions| 200 x 180 x 380 mm
Weight| Approx. 4.4 kg
Parcel details| 500 x 350 x 360 mm / approx. 7 kg
Delivery content
- 1 x moisture balance PCE-MA series
- 1 x wind shield
- 10 x sample pan
- 1 x mains cable
- 1 x sample pan holder / handle
- 1 x 100 g weight in plastic box
- 1 x instruction manual
Optional Accessories
– PCE-BP 1 thermal printer
System description
Device
- Heat protection plate
- Wind shield
- Insertion aid
- Pan retainer
- sample pan
Assemble the parts as illustrated above.
- Start with the heat protection plate (1).
- The notch in the side wall of the windshield (2) must be placed in a way that it faces the user.
- The handle of the insertion aid (3) is placed to rest in this notch.
- Place the sample pan retainer (4) in the middle of the measuring chamber. Make sure that the cone of the sample pan retainer rests in the guide properly.
- The sample pan (5) is the last part you need to install.
Display
- Heat-up mode (Standard / Quick / Slow)
- Measuring / stop mode (constant reading / user / time lapse)
- Drying temperature (40°C bis Zu 199°C)
- Drying time / Duration of the drying process
- Percentage moisture
- Percentage of dry weight / dry matter %
- Grams
Function keys
Key
| ****
Function
---|---
|
Power on / off
|
Start / stop measurement
|
Tare / Leave settings
|
Open settings
|
Switch between absolute and relative measuring result
|
Confirm current settings / go to next setting
| Return to weighing mode (short keystroke) / Calibration (long
keystroke)
| ****
Change settings
RS-232 interface / printer
The RS-232 interface is used for connecting the optional PCE-BP 1 printer to
the scales. To do so, you have to set the following on the printer:
The printer comes with an RS-232 cable which is used for connecting both devices:
Preparation
Preparation of sample
You should always prepare only one sample for measurement to avoid that it
loses moisture to the environment or absorbs air humidity. In case several
samples have to be taken at the same time, they should be put into airtight
containers to make sure their characteristics do not change when stored.
Spread the sample on the sample pan evenly and in a thin layer to get
reproducible results. Uneven application of the sample causes inhomogeneous
heat distribution in the sample to be dried. This means that the sample is
dried incompletely or that the drying time is increased. Due to piling
material, the upper layers will heat up more strongly, which will cause
combustion or incrustation. An excessively high thickness of the layer or
possible incrustations prevents the moisture from leaving the sample. The
residual moisture makes the measurement results non-verifiable and non-
reproducible.
Tools for sample preparation
The tools and instruments used for sample preparation have a huge effect on
the accuracy and reliability of the measurement. You should avoid tools with
heat-conducting properties, i. e. which are able to transfer heat to the
sample. Improper handling and preparation of the sample will falsify the end
result of the measurement.
Disposable aluminium sample panb
To be able to measure the moisture content of the sample, the sample should
be spread on the sample pan evenly and then be placed in the drying chamber of
the moisture balance. Re-using a sample pan can falsify the end result of the
measurement because of the adhering residues.
Spreading of the sample
ATTENTION:
The samples should not exceed a thickness of 8 mm and a diameter of 90 mm. If
the max. height of the sample is exceeded, this can lead to combustion or
incrustation of the sample. The sample should be spread on the sample pan
evenly, ideally in layers of 2 … 5 mm. Make sure that the temperature probe,
which is located in the cover of the moisture balance, does not touch the
sample as this will falsify the result.
Solids:
- Spread powdery and grained samples evenly on the sample pan.
- Crush coarse-grained samples using a mortar or similar device. Avoid any heat supply when crushing the sample as this will cause moisture loss.
Liquids:
For liquids, pastes or melting samples, it is recommended to use the glass
fibre filter.
Glass fibre filter circle
Glass fibre filter circles ensure that liquid samples are evenly spread on the
sample pan and prevent the combustion of solids. The glass fibre filter circle
has the following advantages:
- even spreading due to capillary action
- no formation of drops
- quick evaporation due to larger surface
Practical advice
Before starting the measurement, place the sample pan and, if appropriate, the
glass fibre filter on the pan retainer and tare it to make sure that only the
weight of your sample is assessed.
If you carry out several consecutive measurements, keep in mind that the
temperature from the previous measurement still obtains in the measuring
chamber and that evaporation already takes place when the cover is closed.
This might affect the measuring results. Therefore, you must either allow some
time for the drying chamber to cool down or accept the deviations and start
the next measurement as soon as possible.
Operation
Measurement
The moisture balance saves the drying parameters last used after switching off
and back on. These are shown in the display
Prepare the sample and place it in the drying chamber. You can now start the
measurement by pressing the key.
After the measurement, the measured values will be available until they are
deleted. You can use the key to switch between the weight indication / dry
matter indication in % and the indication of the moisture content in % during
and after the measurement. After the measurement, the measurement values last
determined must be deleted from the memory. To do so, press the key.
ATTENTION: Before placing the next sample on the sample pan, make sure
that the display shows 0.0 g respectively 0.00 g. If this is not the case,
press the key. If, however, different drying parameters are necessary due to
the different samples, you can use up to 16 memory locations. You can save the
selected heat-up mode/termination criterion/drying temperature, as well as the
drying time if necessary, to each of the memory locations.
ATTENTION: Do not touch the halogen lamps or the cover during the
measurement as this can cause severe burns. After the measurement, let the
sample cool down before touching it.
Drying modes
Press the key to get to the drying mode selection screen. Here you can select
different drying modes and set up new ones. A number that indicates the
memory location flashes on the left-hand side of the display. The user can
retrieve up to 16 different memory locations, using the arrow keys and . The
saved drying
parameters are displayed along with the selected memory location. If you wish
to use the drying parameters of a certain memory location, press the key. The
moisture the balance will then take over the drying parameters and you can
start the measurement. If you want to save some new drying parameters to a
memory location, select the memory location and confirm with the key.
Specify heat-up mode
After the selection of the memory location, the drop icon will flash in the
upper side of the display. The user can select one out of three heat-up
speeds.
Standard heat-up mode
This is the default mode which is suitable for most sample types. In this
heat-up mode, 120 °C are reached after approx. 4 minutes.
Quick heat-up mode
This mode is suitable for samples with a high moisture content. In this heat-
up mode, 120 °C are reached after approx. 1 minute.
ATTENTION: When using this heat-up mode, there might be increased
temperatures in the drying chamber for a short period as the sensor and the
control need some time to regulate the temperature.
Slow heat-up mode
This mode is suitable for samples with a low moisture content.
In this heat-up mode, 120 °C are reached after approx. 8 minutes.
ATTENTION: When using this heat-up mode, it is particularly important
that the drying time of the sample is adapted to the drying mode as otherwise
the drying process could be terminated (time expired) although there is still
some residual moisture in the sample.
You can select the desired heat-up mode using the arrow keys and .
To confirm the selected heat-up mode, press the key.
Specify measurement / stop mode
After the selection of the heat-up mode, the bar icon will flash in the upper
side of the display.
The user can select between three measurement/stop modes.
Constant measurement value
In this measurement/stop mode, the measurement is stopped automatically when
the measurement value is constant over a certain period of time.
The following table shows the stop conditions:
| Stop conditions
<period of time (weight change)>
---|---
| Fast
heat-up mode
| Standard heat-up mode| Slow
Heat-up mode
PCE-MA 110| 36 s (<20 mg)| 45 s (<20 mg)| 60 s (<20 mg)
PCE-MA 202| 36 s (<20 mg)| 45 s (<20 mg)| 60 s (<20 mg)
PCE-MA 100| 36 s (<3 mg)| 45 s (<2 mg)| 60 s (<2 mg)
PCE-MA 200| 36 s (<3 mg)| 45 s (<2 mg)| 60 s (<2 mg)
ATTENTION: In this mode, it is not possible to set a measuring time.
Manual / user stop mode
In this measurement/stop mode, the measurement can be ended by the user. To do
so, press the key.
ATTENTION: Depending on the temperature, combustion of the sample can
easily occur in this measurement/stop, mode. This mode should only be used if
constant monitoring can be guaranteed. The advantage of this mode is that the
sample can be dried for a long time to allow moisture to escape which is
located very deep in the sample.
Time stop mode
In this measurement/stop mode, the measurement is stopped automatically when
the set measuring/drying time has passed.
ATTENTION: The maximum measuring/drying time you can set is 99 minutes.
The desired measurement/stop mode can be selected by means of the arrow keys
and . To confirm the selected measurement/stop mode, press the key.
Setting the drying temperature
After selecting the measurement/stop mode, the temperature indicator will
flash in the upper side of the display.
You can set the drying temperature using the arrow keys and . The first
setting can be made in steps of 10 %. After confirming with the key, the
setting can be carried on in increments of 1 °C.
ATTENTION: During temperature regulation, temperature fluctuations of
approx.. 4 °C can occur. Mind this when selecting the temperature to make sure
the sample does not scorch. Otherwise, the moisture reading will be falsified.
Setting the drying/measuring time (only in time stop mode)
Depending on the previously set measurement/stop mode, the time indication
flashes in the upper side of the display.
The drying / measuring time can be set using the arrow keys and . To confirm
the set drying / measuring time, press the key.
ATTENTION: The maximum drying/measuring time you can set is 99 minutes.
Adjustment / calibration
The moisture analyser consists of a weighing unit and a drying chamber. The moisture measurement is carried out on the basis of the weight loss. This can be easily reproduced by a rule of three.
ATTENTION: The moisture analysers of the PCE-MA series have a higher
weighing resolution than the one indicated. The moisture balance uses the
higher resolution which is invisible to the user for calculation. This can
result in deviations after the decimal point when trying to reproduce the
result. The weighing unit can be adjusted by means of a 100 g external test
weight (min. M1). Special care must be taken when positioning the scales (as
little vibration as possible and balanced by means of the level). The
adjustment of the weighing unit should only be carried out after a heat-up
time of approx. 30 minutes.
To do so, press and hold the key for approx.. 6 seconds. “-CAL-” will be
displayed.
After the “-CAL-” indication, “100.00 g” will flash.
Now place the 100 g calibration weight in the middle of the weighing platform. The indication will change from “100.00 g” to “=====”.
When the indication changes from “=====” back to “100.00 g”, remove the weight from the weighing platform. After this, the display will show “=====” again and change to “0.00 g” automatically
The adjustment of the balance is now finished.
Print out results (optional)
You can print out the measurement results by using the optional PCE-BP 1 printer. To do so, follow these steps:
-
Make sure that the printer is set up properly and both devices are connected to each other.
-
During operation, press and hold the key until “PrInT” appears in the display.
-
After this, the options “L-C” (Chinese print-out) and “L-E” (English print-out) will appear in the display. Use the arrow keys and to select the desired option and press the key to
confirm. -
After this, numbers from 1 to 5 will appear in the display which represent the last 5 measurements (which are stored in the device). Select a measurement by using the arrow keys and and press the key to confirm.
-
The data is now sent to the printer via the RS-232 interface.
Print-out:
Troubleshooting
Error message | Error cause | Possible solution |
---|---|---|
-Err 1- | Sample weight below 1 g | Put more than 1 g of the sample on the scales and start the measurement. |
-Err 2- | Temperature setting below +40 |
°C
| Restart the scales.
-Err 3-| Measuring time setting below 30 seconds.| Restart the scales.
-Err 4-| Halogen lamp problem| Please contact our technical support.
-Err 5-| Time setting for slow heat-up mode below 3 minutes| Restart the scales.
-Err 6-| Temperature sensor broken| Please contact our technical support.
Warranty
You can read our warranty terms in our General Business Terms which you can find here: https://www.pce-instruments.com/english/agb.
Disposal
For the disposal of batteries, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose. In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law. If you have any questions, please contact PCE Instruments.
Contact
If you have any questions about our range of products or measuring instruments
please contact PCE Instruments.
PCE Instruments UK
By post:
PCE Instruments UK Ltd.
Units 12/13 Southpoint Business Park Ensign Way, Southampton Hampshire United
Kingdom, SO31 4RF
By phone: 02380 987 035
PCE Americas
By post:
PCE Americas Inc. 711 Commerce Way Suite 8 Jupiter 33458 FL USA
By phone: 561 320 9162
DECLARATION OF CONFORMITY
PCE Deutschland GmbH
Im Langel 4
59872 Meschede
Tel: +49 (0) 2903 / 976 99 0
Fax: +49 (0) 2903 / 976 99 29
E-Mail: info@pce-instruments.com
Internet: https://www.pce-instruments.com/
Declaration of conformity
We hereby declare that the product to which this declaration refers conforms
with the following standards.
Moisture Analyzer: PCE-MA 110, PCE-MA 50, PCE-MA 105, PCE-MA 102, PCE-MA
100, PCE-MA 202, PCE-MA 50T, PCE-MA 100T, PCE-MA 200T, PCE-MA 202T