DIGI-SENSE 20250-66 Digital Multimeter with Thermal Imager User Manual

June 5, 2024
DIGI-SENSE

20250-66 Digital Multimeter with Thermal Imager

User Manual
Digital Multimeter with Thermal Imager
with Bluetooth® Mobile App and NIST-Traceable Calibration
Model 20250-66

THE STANDARD IN PRECISION MEASUREMENT
1065DGMAN_20250-66_Rev1.indd 1

07/02/2019 11:03:39 AM

Introduction
The Digi-Sense Digital Multimeter with Thermal Imager and Bluetooth® Connectivity (model 20250-66) allows you to transmit data directly to your AndroidTM or iOS® device using our free D/S T-View app. Now you can view data at a safe distance from potentially hazardous parameters or even use your smart device as a real-time secondary display for checking measurements, eliminating the need to write down readings on paper. The results are automatically stored in the app with a date-and-time stamp and can be saved as a CSV file and emailed for future reference, manipulation, or analysis to help determine trends and conditions over a long period of time. You can also use the app to attach photos and notes to the records creating a clear reference point for your work–ideal for record keeping at large sites.
Setup is simple. Download the free D/S T-View app to your Android or iOS device. Place the meter in Bluetooth mode and open the app on your smart device. The meter will be sensed by your device and be listed as an available source that you can select. Once selected, the measured electrical data is displayed on your device and some of the instrument functions can be accessed. A full description of its operation is available for download in the app.
The Digital Multimeter with Thermal Imager provides fast A/D converting sampling time, high accuracy with data logging and trend capture features. This instrument is fully tested and calibrated to NIST-traceable standards for reliable measurements right out of the box, saving you time and money.

Key Features
· Bluetooth interface · User-friendly mobile app · 6000 count 2.8” TFT color LCD · Double-molded IP65-rated housing · Built-in Thermal imager with Max, Min
and Center crosshair targeting · 50 Hz fast thermal image frame rate · DC voltage · AC, AC+DC TRMS voltage

· DC current · AC, AC+DC TRMS current · Resistance and continuity test · Diode test · Capacity · Frequency · Duty cycle · Temperature with type K probe · Flexible coils current

Unpacking
Check individual parts against the list of items below. If anything is missing or damaged, please contact your instrument supplier immediately.

1. Meter 2. Test leads 3. One type K temperature probe 4. Two rechargeable Li-ion batteries 5. Battery charger cradle with
international adapters

5. Hard carrying case
6. User manual
7. NIST-traceable calibration report with data

2
1065DGMAN_20250-66_Rev1.indd 2

07/02/2019 11:03:39 AM

Safety
This symbol adjacent to another symbol, terminal or operating device indicates that the operator must refer to an explanation in the Operating Instructions to avoid personal injury or damage to the meter.
WARNING This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury.
This CAUTION symbol indicates a potentially hazardous situation, which if not
CAUTION
avoided, may result damage to the product.
MAX This symbol advises the user that the terminal(s) so marked must not connected to a circuit point at which the voltage with respect to earth ground exceeds (in this case) 1000 VAC or VDC.
This symbol adjacent to one or more terminals identifies them as being associated with ranges that may, in normal use, be subjected to particularly hazardous voltages. For maximum safety, the meter and its test leads should not be handled when these terminals are energized.
This symbol indicates that a device is protected throughout by double insulation or reinforced insulation.

Safety Instructions
This meter has been designed for safe use, but must be operated with caution. The rules listed below must be carefully followed for safe operation.
· NEVER apply voltage or current to the meter that exceeds the specified maximum:

Input Protection Limits Function V DC or V AC mA AC/DC A AC/DC Frequency, Resistance, Capacitance, Duty Cycle, Diode Test, Continuity Surge Protection: 8kV peak per IEC 61010

Maximum Input 1000VDC/AC RMS 800mA 1000V fast acting fuse 10A 1000V fast acting fuse 1000VDC/AC rms

· USE EXTREME CAUTION when working with high voltages.
· DO NOT measure voltage if the voltage on the “COM” input jack exceeds 1000V above earth ground.
· NEVER connect the meter leads across a voltage source while the function switch is in the current, resistance, or diode mode. Doing so can damage the meter.
· ALWAYS discharge filter capacitors in power supplies and disconnect the power when making resistance or diode tests.
· ALWAYS turn off the power and disconnect the test leads before opening the covers to replace the fuse or batteries.
· NEVER operate the meter unless the back cover and the battery and fuse covers are in place and fastened securely. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

3

1065DGMAN_20250-66_Rev1.indd 3

07/02/2019 11:03:39 AM

Description and reference guide

1

Front description

1. NCV detector area

2. LCD

2

3. Navigation/Menu buttons

4. MODE button

3

5. RANGE button

4

6. Rotary function switch

5

7. Positive (+) probe input jack for A (current)

8. Positive (+) probe input jack for mA (current)

6

9. COM(-) probe input jack

10. Positive (+) probe input jack for all inputs

7

except A and mA

8

11. Thermal mode/light button

12. Hold/capture button

13

14

Back description

15

13. No-slip slope

16

17

14. Thermal imager lens

15. Lens cover

16. Work light

17. Laser

18. Support plate

18

19. Battery cover lock

19

Understanding the Push Buttons
The 9 push buttons on the front of the Meter activate features that augment the function selected using the rotary switch, navigate menus or control power to Meter circuits.
· Cursor buttons:
Select an item in a menu, adjust display contrast, scroll through information, and perform data entry. PEAK Use Navigation UP button to select PEAK function.
REL Use Navigation Left button to select REL function. MAX Use Navigation Right button to select MAX function.

4

1065DGMAN_20250-66_Rev1.indd 4

12 11 10 9
07/02/2019 11:03:40 AM

· Physical buttons: HOLD Freezes the present reading in the display and allows display to be saved. ESC Also wake up for APO. MODE Press the MODE key to switch the functions.

RANGE
Press the RANGE key to manual range.

Enter function of the menu selects.

IR

Press the IR key to switch DMM MODE and IR+DMM MODE.

Navigation buttons.

Understanding the Display · Measurement on LCD 1. Indication of battery charge level 2. Indication of measuring result 3. Indication of Automatic/Manual mode 4. Analog bar graph 5. Indications associated with function keys 6. Indication of the system’s time 7. Indication of measuring unit 8. SD card 9. Temperature measuring result 10. Indication of Automatic/Manual mode 11. Temperature unit 12. IR camera 13. Indication of measuring unit 14. Indication of measuring result

1

6

7

2 3
4

5 DMM Mode

8

11

9

12

10

13

14

IR+ DMM Mode

· Icons on LCD
Voltage is over 30V (AC or DC)
Warning
Flexible coils
Traditional clamps

r Relative
High edge time
~ AC voltage or current
DC voltage or current

~ AC+DC voltage or Current ·))) Continuity function Diode function Ohms
5

1065DGMAN_20250-66_Rev1.indd 5

07/02/2019 11:03:41 AM

Understanding the Rotary Switch V~ V~ HZ % =CAP ·))) K Temp A mA
µA

AC voltage measurements DC and AC+DC voltage measurements Frequency and Duty measurements
Resistance, Diode test, Capacitance and Continuity measurements
Temperature measurements AC, DC and AC+DC amps measurements AC, DC and AC+DC milliamps measurements
AC, DC and AC+DC microampere measurements up to 6,000 µA
Flexible coils current

Select a primary measurement function by positioning the rotary switch to one of the icons around its perimeter. For each function, the meter presents a standard display for that function (range, measurement units, and modifiers). Button choices made in one function do not carry over into another function.

DMM Measurement and Setup
DC Voltage Measurements CAUTION: Do not measure DC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter. · Set the function switch to the VDC position. · Insert the black test lead banana plug into the negative COM jack. Insert the red test
lead banana plug into the positive V jack. · Read the voltage in the display.
AC+DC Voltage Measurements CAUTION: Do not measure DC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter. · Set the function switch to the VDC position. · Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack. · Press the MODE key to switch the V AC+DC Voltage functions. · Read the AC+DC voltage in the display.
AC Voltage Measurements WARNING: Risk of Electrocution. The probe tips may not be long enough to contact the live parts inside some 240V outlets for appliances because the contacts are recessed deep in the outlets. As a result, the reading may show 0 volts when the outlet actually has voltage on it. Make sure the probe tips are touching the metal contacts inside the outlet before assuming that no voltage is present. CAUTION: Do not measure AC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter.
6

1065DGMAN_20250-66_Rev1.indd 6

07/02/2019 11:03:42 AM

· Set the function switch to the VAC position. · Insert the black test lead banana plug into the negative COM jack. Insert red test lead
banana plug into the positive V jack. · Read the voltage in the main display.
Frequency Measurements · Set the function switch to the Hz% position. · Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack. · Read the Frequency in the display. · Press the MODE key to switch the Duty functions. · Read the Duty in the display.
Resistance Measurements To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements. Remove the batteries and unplug the line cords. · Set the function switch to the CAP ·))) position. · Insert the black test lead banana plug into the negative COM jack. Insert the red test
lead banana plug into the positive jack. · Read the resistance in the display.
Continuity Check WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements. Remove the batteries and unplug the line cords. · Set the function switch to the CAP ·))) position. · Insert the black test lead banana plug into the negative COM jack. Insert the red test lead
banana plug into the positive jack. · Press the MODE key to switch the continuity functions. · If the resistance is less than approximately 50 , the audible signal will sound. If the
circuit is open, the display will indicate “OL”.
4.7. Diode Test · Set the function switch to the CAP ·))) position. · Insert the black test lead banana plug into the negative COM jack and the red test lead
banana plug into the positive V jack. · Press the MODE key to switch the Diode functions. · Forward voltage will typically indicate 0.400 to 3.000V. Reverse voltage will indicate
“OL”. Shorted devices will indicate near 0V and an open device will indicate “OL” in both polarities.
7

1065DGMAN_20250-66_Rev1.indd 7

07/02/2019 11:03:42 AM

Capacitance Measurements WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any capacitance measurements. Remove the batteries and unplug the line cords. · Set the rotary function switch to the CAP ·))) position. · Insert the black test lead banana plug into the negative COM jack. Insert the red test lead
banana plug into the positive V jack. · Press the MODE key to switch the Capacitance functions. · Read the capacitance value in the display.
Temperature Measurements · Set the function switch to the TEMP (°C or °F) position. · Insert the temperature probe into the input jacks, making sure to observe the correct
polarity. · Read the temperature in the display. · Press the MODE key to switch the Unit (°C or °F).
Flexible Coil Current Measurements · Set the function switch to the Flexible coil position. · Insert the black test lead banana plug into the negative COM jack. Insert the red test lead
banana plug into the positive V jack. · Read the current in the display. · Press the MODE key to switch the AC, DC and AC+DC current. · Press the RANGE key to switch range. 1000mA, 10A, 30A, 40A, 100A, 300A, 400A, 1000A,
3000A.
DC Current Measurements · Insert the black test lead banana plug into the negative COM jack. · For current measurements up to 6000µA DC, set the function switch to the µA position
and insert the red test lead banana plug into the µA/mA jack. · For current measurements up to 600mA DC, set the function switch to the mA position
and insert the red test lead banana plug into the µA/mA jack. · For current measurements up to 10A DC, set the function switch to the 10A position and
insert the red test lead banana plug into the 10A jack.
· Press the MODE button to indicate “=” on the display.
· Read the current in the display.
AC Current Measurements CAUTION: Do not make 10A current measurements for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads. · Insert the black test lead banana plug into the negative COM jack.
8 ·)))

1065DGMAN_20250-66_Rev1.indd 8

07/02/2019 11:03:42 AM

· For current measurements up to 6000µA AC, set the function switch to the yellow µA position and insert the red test lead banana plug into the µA/mA jack.
· For current measurements up to 600mA AC, set the function switch to the yellow mA position and insert the red test lead banana plug into the µA/mA jack.
· For current measurements up to 10A AC, set the function switch to the yellow 10A position and insert the red test lead banana plug into the 10A jack.
· Press the MODE button to indicate “~” on the display.
· Read the current in the display.

AC+DC Current Measurements
CAUTION: Do not make 10A current measurements for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.
· Insert the black test lead banana plug into the negative COM jack.
· For current measurements up to 6000µA AC+DC, set the function switch to the yellow µA position and insert the red test lead banana plug into the µA/mA jack.
· For current measurements up to 600mA AC+DC, set the function switch to the yellow mA position and insert the red test lead banana plug into the µA/mA jack.
· For current measurements up to 10A AC+DC, set the function switch to the yellow 10A position and insert the red test lead banana plug into the 10A jack.
· Press the MODE button to indicate “~” on the display.
· Read the current in the display.

Using RANGE

Press the RANGE key to activate the manual mode and to disable the

Autorange function. The message “Manual Range” appears on the upper

left part of the display instead of “Auto Range”. In manual mode, press

the RANGE key to change measuring range: the relevant decimal point

w%ilTlecmhapn°gCe°iFts1p0oAs=~itio. Inn.

The RANGE key is not active in positions Autorange mode, the instrument selects the

·))) most

appropriate ratio for carrying out measurement. If a reading is higher

than the maximum measurable value, the indication “O.L” appears on

the display. Press and hold the RANGE key for more than 1 second to exit

the manual mode and restore the Autorange mode.

Hold and AutoHold Mode To freeze the display for any function, press key HOLD. And again press key HOLD to release freeze.

1065DGMAN_20250-66_Rev1.indd 9

9
07/02/2019 11:03:42 AM

Capturing Minimum and Maximum Values
The MAX MIN Record mode captures minimum and maximum input values. When the input goes below the recorded minimum value or above the recorded maximum value, the meter beeps and records the new value. This mode is for capturing intermittent readings, recording minimum and maximum readings unattended, or recording readings while equipment operation precludes watching the meter.
To activate the MAX MIN mode, press soft key labeled3. If the meter is already in MAX MIN function, pressing3causes the Meter to turn off
MAX MIN function.
Relative Values
To activate the Relative mode, press the soft key labeled s. If the meter is already in the relative function, pressing s causes the meter to turn off Relative mode.
Capturing Peak Values
To activate the Peak mode, press the soft key labeled4. If the meter is already in the peak function, pressing4causes the meter
to turn off Peak mode.
Non-Contact AC Voltage Detector (100 to 1000V AC)
WARNING: Risk of Electrocution. Before use, always test the Voltage Detector on a known live circuit to verify proper operation. WARNING: Insulation type and thickness, distance from the source, and other factors may effect operation. Always verify live voltage using other methods before working on electrical circuits. · The non-contact voltage detector operates when the meter is set to
any measuring function. The detector does not operate when Auto Power Off turns the meter off or when the rotary function switch is set to the Off position. · Slowly move the detector probe closer to the conductor being tested. · If AC voltage within the specified range is present, the indicator light will illuminate. NOTE: The detector is designed with high sensitivity. Static electricity and other sources of electrical energy may randomly activate the detector. This is normal operation. The detector only activates the indicator light when AC voltage is present. It does not indicate the voltage level on the LCD.

10
1065DGMAN_20250-66_Rev1.indd 10

07/02/2019 11:03:43 AM

Thermal imager and DMM Operation
Thermal imager basics
In the Thermal imaging and DMM mode user can measure a targeted surface’s temperature and use Multimeter at the same time. The measured result will display under the thermal image. · Press the red “IR” button to open the Thermal Imager. In the image shown the thermal
image is set to color palette IRON. Select other palettes in the Menu Settings. · Open the protective lens cover on the back of the meter.

1. Battery capacity indicator.

1 23 4 5 6

2. SD card icon. If this icon is displayed there is an

7

SD card inserted.

8

19

3. Bluetooth icon. If this icon is displayed the Bluetooth

9

is opened.

10

18

4. The currently selected emissivity value. Use the

Thermal Settings Menu to change the emissivity value.

17

11

5. Temperature unit icon. Use the Thermal Settings Menu

16

to select °C, °F, or K.

15

6. Current time display.
7. Center cross of the thermal imager temperature measurement represents the center spot temperature of the scene.

12 13 14

8. Highest temperature spot of the thermal imager temperature measurement, represents the highest spot temperature of the scene.

9. Minimum temperature spot of the thermal imager temperature measurement, represents the minimum spot temperature of the scene.

10. Current scene on the thermal image frame.

11. Range icon of the meter.

12. Max soft button.

13. REL soft button.

14. PEAK soft button.

15. DMM measurement, shown below thermal image.

16. Unit of the meter.

17. Lowest reading measured in the current frame.

18. Thermal scale shows the range color for thermal images. The lighter the color, the warmer the temperature; the darker the color, the cooler the temperature.

19. Highest reading measured in the current frame.

Using the thermal imager For basic operation follow these steps: 1. Set the function switch to any position. 2. Press the “IR” button to switch the thermal imager ON. Target the object by the thermal
imager lens.
11

1065DGMAN_20250-66_Rev1.indd 11

07/02/2019 11:03:43 AM

3. The display will show the temperature measurement in the upper left hand corner for the targeted area along with the currently selected emissivity value.

4. In the Thermal imaging mode, the laser pointer and display cross hairs can be used to assist in targeting. Tools can be switched ON or OFF in the Setting menu.

5. In the Thermal imaging mode, the highest temperature will be auto marked by a red cross, and the lowest temperature will be auto marked by a blue cross, the two spots can be switched ON or OFF in the Setting menu.

6. In the Thermal imaging mode, the meter continues to operate normally as a multimeter allowing any of the electrical functions to be used.

7. Press the HOLD button to hold the thermal image frame, then long press the HOLD button. You will capture the screen and save a bitmap with measure data into SD card. The saved bitmap later can be analyzed by PC software or smartphone apps.

8. The thermal imager FOV (Field of View) is 21 by 21°.

9. FOV is the largest area that your imager can see at a set distance.

10. This table lists the horizontal FOV, vertical FOV and IFOV for lens.

Focal Length 7.5mm

Horizontal FOV 21°

Vertical 21°

FOVIFOV 4.53mrad

IFOV (Instantaneous Field of View) is the smallest detail within the FOV that can be detected or seen at a set distance, the unit is rad. The formula is this:

IFOV = (Pixel Size)/(Lens focal length);
D:S theoretical ( =1/ IFOVtheoretical) is the calculated spot size based on the pixel size of the Thermal Imager detector array and lens focal length.
Example: if thermal imager uses 9mm lens, because the Pixel Size of detector is 34µm, horizontal FOV is 21°, vertical FOV is 21°, and the IFOV is 34µm/7.5mm = 4.53mrad;

80m 80m
220m

Spot Size=100.00cm*100.00cm (Based upon IFOVtheoretical)
21° 21°

D:Stheoretical ( = 1/ IFOVtheoretical) = 220:1

D:Smeasure ( = 1/ IFOVmeasure) is the spot size needed to provide an accurate temperature measure. Typically, D:Smeasure is 2 to 3 times smaller than D:Stheoretical, which means the temperature measurement area of the target needs to be 2 to 3 times larger than that determined by the calculated theoretical D:S

Using the Multimeter with the thermal imager on IR+DMM mode, MODE key, RANGE key, HOLD key and REL Function is same DMM mode.
· Capturing MAXMIN values on IR+DMM mode
1. To activate the MAXMIN mode, press the soft key labeled3,
and display max value.
2. If the Meter is already in the MAXMIN function, then press the
3key to display min value, then press the3key to display current measurement value. Next press the3key to display
max value again.
3. Press and hold the3key for more than 1 second to exit
MAXMIN mode.
12

1065DGMAN_20250-66_Rev1.indd 12

07/02/2019 11:03:44 AM

· Capturing Peak Values on IR+DMM mode
1. To activate the peak mode, press the soft key labeled4, and
display Peak max value. 2. If the Meter is already in the peak function, then press the
4key to display Peak min value, then Press the4key to display current measurement value. Next press4key again
to display peak max value.
3. Press and hold the4key for more than 1 second to turn off
peak.
Settings Menus
Using Settings Menus · Press MENU button to open the Settings Menus, as shown here: · Press UP/DOWN button to select menu item or
change the value of current focus item. · Press RIGHT/MENU button to enter the submenu
or set focus on the current selected item. Press LEFT button to return to the previous menu. · To exit settings menus, press MODE/RANGE/HOLD/IR button or press LEFT button in root menu.
Setting details
· Palette mode Thermal imager has five kinds of palettes, such as: Press RIGHT/MENU button to select one of the display color palettes.
Temp Unit Press RIGHT/MENU button to set focus on this option and the color of option value will change to black °C . In focus state, use the RIGHT/MENU button to toggle °C, °F, and K. Use LEFT/RIGHT/MENU button to exit focus state and the color of option value will change to white K .
Measure Press RIGHT/MENU button to enter measure menu. Two selections are available: HOT POINT and COLD POINT. Press RIGHT/MENU button to set your select item on or off. · Hot point: This option enables thermal imager to automatically
detect the highest temperature point. · Cold point: This option enables thermal imager to automatically
detect the lowest temperature point.
Emissivity · Press RIGHT/MENU button to set focus on this option. In focus state, use UP/DOWN button
to increase or decrease emissivity’s value, use LEFT/RIGHT/MENU button to exit focus state. The available range is 0.01 to 0.99 in 0.01 steps.
13

1065DGMAN_20250-66_Rev1.indd 13

07/02/2019 11:03:45 AM

Language · Press RIGHT/MENU button to enter language menu. Three options are available: Simplified Chinese, Traditional Chinese and English. Use UP/DOWN button to select language and use RIGHT/MENU button to set selected language to be valid.
Common Press RIGHT/MENU button to enter common menu. Five options are available: Beep, Bluetooth, Laser, Brightness and Auto Off. · Beep: Use RIGHT/MENU button to set beep on or off. · Bluetooth: Use RIGHT/MENU button to set bluetooth power on or off. · Laser: Use RIGHT/MENU button to set laser pointer on or off. · Brightness: Press RIGHT/MENU button to set focus on this option. In
focus state, use UP/DOWN button to change LCD’s brightness, use LEFT/RIGHT/MENU button to exit focus state. The available brightness range is 100% to 10% in 10% steps. · Auto Off: Press RIGHT/MENU button to set focus on this option. In focus state, use UP/DOWN button to choose the time period after which the meter enters the sleep mode.
Bluetooth® Connect 1. Turn on the Bluetooth function on the instrument. 2. Turn on the Bluetooth of smartphone, press
the icon Thermview+ and enter into the home interface. Press Connect Device icon on the Home interface, Bluetooth device name will appear.
3. Touch the device name listed in Bluetooth devices list to connect the device.
The detail information about Thermview+, please refer to Thermview+ APP help file. Thermview+ for Android: Please search in Google Play with keyword: “Thermview+”, download and run. Thermview+ for iOS: Please search in Apple store with keyword “Thermview+”, download and run.
Time/Date Press RIGHT/MENU button to enter time menu. In this menu, year, month, day, hour, minute and time format can be set. The changes take effect after exiting settings menus.

14
1065DGMAN_20250-66_Rev1.indd 14

07/02/2019 11:03:46 AM

Photo Press RIGHT/MENU button to enter photo menu. Two options are available: Photo Review and Delete Photo. · Photo Review: press RIGHT/MENU button to
enter image browser function, and exit settings menus immediately. · Delete Photo: After pressing RIGHT/MENU button dialog box will be displayed as shown at right. Warning: selecting ‘YES’ will delete all the photos in the memory card which has been captured by user.
Sys Info · Press RIGHT/MENU button to enter system information menu.
This menu contains software’s version, hardware’s version, and thermal imager’s version
Factory Set · Select Factory Set option after pressing RIGHT/MENU button.
The dialog box will be displayed as shown right. Select ‘YES’ button and system parameter will be reset.

Record measurements

· With a measurement on the display (Fig 1),

press Button key Menu to enter the instru-

ment’s general menu (Fig 2). The screen is
5 shown on the display. Press the Button or 6 key to select Record Item. Press the
Button and enter Record Menu (Fig 3).
4 5 6 · In Record Menu press the Button or key

Fig 1

5 6 4 to select Sample Interval Item or Duration Item
(Fig 3). Press the Button . Then Press the Button or key to adjust time. Setting of sampling

interval is from 1s to 59min:59s (Fig 4).

Setting of recording duration is from 1min to 9h:59min (Fig 5) .

Fig 2 Fig 4

Fig 3
Fig 5 15

1065DGMAN_20250-66_Rev1.indd 15

07/02/2019 11:03:48 AM

· In Record Menu. Press the Button 5 or 6 key to select

Start record Item. Press the button 4. Enter Save Record

Measurement (Fig 6).

In Save Record Measurement, press the Button 4 to stop

record. Press the Button 5 Save.
5 6 · In Record Menu. Press the Button or key to select 4 Review Item. Press the Button to view

record measurement (Fig 7).

Fig 6

Fig 7

3 4 · Press the Button MODE key to Trend record (Fig 137). Press the button or key to

select previous record measurement or next

record measurement. Press the button ESC key

to exit view record measurement.
3 · In Record View Display, press the Button or key to move the cursor on the graph. 4 5 Press the Button key to activate the zoom function of the graph (Fig 138), increasing

Fig 137

Fig 138

Fig 139

resolution (symbol “Zoom Xy” where y=max

zoom dimension appears at the top of the display on the right side). You can zoom

X1 for at least 10 measuring points, X2 for at least 20 measuring points, X3 for at

least 40 measuring points and so on for maximum of 6 zooming operations.
3 4 · In Record Menu. Press the Button or key to select Delete all Recordings Item 4 (Fig 139). Press the Button , enter Delete Box and select Yes or No.

Image Browser

· In Image Browser mode user can browse the pictures in the memory card. Press LEFT/RIGHT button to select previous or next picture. Press any other keys to exit Image Browser mode.

1. LEFT key instruction. 2. Current displayed picture’s filename. 3. RIGHT key instruction. 4. Picture display area.

4

2

1

3

· How to capture screen
When in DMM mode or Thermal imaging + DMM mode, use HOLD button to enter hold mode, as shown below. Press UP button to capture screen. After saving to TF card completely, multimeter will exit hold mode.

16
1065DGMAN_20250-66_Rev1.indd 16

07/02/2019 11:03:49 AM

Technical specifications

Technical characteristics

· Thermal imager

Field of view (FOV) / Minimum focus distance Spatial resolution (IFOV) IR resolution Thermal sensitivity / NETD Image frequency Focus mode Focal length Focal Plane Array (FPA) / Spectral range Object temperature range
Accuracy

21° x 21°/ 0.5 m 4.53 mrad 80 × 80 pixels < 0.1°C @ +30°C (+86°F) / 100 mK 50 Hz Focus free 7.5mm Uncooled microbolometer / 8­14 µm ­20°C to +260°C (­4°F to +500°F)
±3°C (±5.4°F) or ±3% of reading (Environment temperature 10 to 35°C, object temperature >0°C)

Accuracy calculated as [%reading + (num. digits*resolution)] at 18°C ÷ 28°C <75%RH.

· DC Voltage
Range
600.0 mV 6.000 V 60.00 V 600.0 V 1000 V

Resolution
0.1 mV 0.001 V 0.01 V
0.1 V 1 V

Accuracy
±(0.09% reading + 5 digits)
±(0.2%reading + 5 digits)

Input impedance >10 M

Protection against overcharge 1000 VDC/ACrms

· AC TRMS Voltage

Range
6.000 V 60.00 V 600.0 V 1000 V

Resolution
0.001 V 0.01 V 0.1 V
1 V

Accuracy(*)

(50 Hz÷60 Hz)

(61 Hz÷1k Hz)

±(0.8% reading + 5 digits)

± (2.4% reading + 5 digits)

Protection against overcharge 1000 VDC/ACrms

(*) Accuracy specified from 10% to 100% of the measuring range, sine wave. Input impedance: >9 M; Accuracy PEAK function: ±10% rdg, PEAK response time: 1 ms.

· AC+ DC TRMS Voltage

Range 6.000 V 60.00 V 600.0 V 1000 V

Resolution 0.001 V 0.01 V 0.1 V 1 V

Accuracy (50 Hz÷1k Hz)
±(2.4% reading +20 digits)

Input impedance >10M

Protection against overcharge 1000 VDC/ACrms

· DC Current
Range 600.0 µA 6000 µA 60.00 mA 600.0 mA 10.00 A

Resolution 0.1 µA 1 µA 0.01 mA 0.1 mA 0.01 A

Accuracy
±(0.9% reading + 5 digits)
±(0.9% reading + 8 digits) ±(1.5% reading + 8 digits)

Protection against overcharge Quick fuse 800 mA/1000 V Quick fuse 10 A/1000 V

1065DGMAN_20250-66_Rev1.indd 17

17
07/02/2019 11:03:49 AM

· AC TRMS Current

Range 600.0 µA 6000 µA 60.00 mA 600.0 mA 10.00 A

Resolution 0.1 µA 1 µA 0.01 mA 0.1 mA 0.01 A

Accuracy ±(1.2% reading + 5 digits) ±(1.5% reading + 8 digits)

Protection against overcharge Quick fuse 800 mA/1000V Quick fuse 10 A/1000V

(*) Accuracy specified from 5% to 100% of the measuring range, sine wave. Accuracy PEAK function: ±10%rdg , AC+DC TRMS Current: accuracy (50Hz÷1kHz): ±(3.0%reading + 20dgt).

· Flexible Coil Current

Range
30.00 A 300.0 A 3000 A

Resolution
0.01 A 0.1 A 1 A

· Diode test
Function

Test current <1.5 mA

Accuracy

(50 Hz÷60 Hz)

(61 Hz÷1k Hz)

±(0.8% reading + 5 digits)

±(2.4% reading + 5 digits)

Protection against overcharge 1000 VDC/ACrms

Max voltage with open circuit 3.3 VDC

· Resistance and Continuity test

Range 600.0 6.000 k 60.00 k 600.0 k 6.000 M 60.00 M

Resolution 0.1
0.001 k 0.01 k 0.1 k 0.001 M 0.01 M

Accuracy ±(0.5% reading + 10 digits)
±(0.5% reading + 5 digits)
±(2.5% reading + 10 digits)

Buzzer >50 M

Protection against overcharge 1000 VDC/ACrms

· Frequency (electronic circuits)

Range 40.00 Hz to 10 kHz

Resolution 0.01 Hz to 0.001 kHz

Sensitivity: 2Vrms

Accuracy ±(0.5% reading)

· Frequency (electronic circuits)

Range 60.00 Hz 600.0 Hz 6.000 kHz 60.00 kHz 600.0 kHz 6.000 MHz 10.00 MHz

Resolution 0.01 Hz 0.1 Hz
0.001 kHz 0.01 kHz 0.1 kHz 0.001 MHz 0.01 MHz

Accuracy
±(0.09% rdg + 5 digits)

Sensitivity: >2 Vrms (@ 20% 80% duty cycle) and f<100kHz; >5 Vrms (@ 20% 80% duty cycle) and f>100kHz

Protection against overcharge 1000 VDC/ACrms
Protection against overcharge 1000 VDC/ACrms

18
1065DGMAN_20250-66_Rev1.indd 18

07/02/2019 11:03:49 AM

· Duty Cycle

Range 5.0% to 95.0%

Resolution 0.1%

Accuracy ±(1.2% reading + 2 digits)

Pulse frequency range: 40 Hz to 10k Hz, Pulse amplitude: ±5V (100 µs to 100ms)

· Capacity
Range
60.00 nF 600.0 nF 6.000 µF 60.00 µF 600.0 µF 6000 µF

Resolution
0.01 nF 0.1 nF 0.001 µF 0.01 µF 0.1 µF 1 µF

Accuracy
±(1.5% reading + 20 digits) ±(1.2% reading + 8 digits) ±(1.5% reading + 8 digits) ±(1.2% reading + 8 digits) ±(1.5% reading + 8 digits) ±(2.5% reading + 20 digits)

Protection against overcharge 1000 VDC/ACrms

· Temperature with type K probe

Range

Resolution

Accuracy (*)

Protection against overcharge

­40.0 to 600.0°C 600 to 1000°C ­40.0 to 600.0°F 600 to 1800°F

0.1°C 1°C 0.1°F 1°F

±(1.5% reading + 3°C)
±(1.5% reading + 5.4°F)

1000 VDC/ACrms

(*) Instrument accuracy without probe; Specified accuracy with stable environmental temperature at ±1°C. For long-lasting measurements, reading increases by 2°C.

1065DGMAN_20250-66_Rev1.indd 19

19
07/02/2019 11:03:49 AM

· Reference standards Safety: EMC: Insulation: Pollution level: Overvoltage category: AT IV 600V, CAT III 1000V Max operating altitude:
· General characteristics Mechanical characteristics
Size (L x W x H): Weight (batteries included):
· Power supply Battery type: Battery charger power supply: Low battery indication: Auto Power Off: Fuses:
· Display Conversion: Characteristics: Sampling frequency:

IEC/EN61010-1 IEC/EN 61326-1 double insulation 2 C 6562 ft (2000 m)
7″ x 3″ x 2″ (175 x 85 x 55 cm) 19 oz (540 g)
1×7.4V rechargeable Li-ION battery, 2300 mAh 100/240 V AC, 50/60 Hz, 10 VDC, 1 A symbol ” ” on the display after 15 to 60 min minutes of nonuse (may be disabled) F10A/1000V, 10 x 38 mm (input 10A)
TRMS F800mA/1000V, 6 x 32 mm (input mA µA) color TFT, 6000 dots with bar graph 3 times/s

20
1065DGMAN_20250-66_Rev1.indd 20

07/02/2019 11:03:49 AM

Environmental conditions for use

Reference temperature: Operating temperature: Allowable relative humidity: Storage temperature: Storage humidity:

18 to 28°C (64 to 82°F) 5 to 40°C (41 to 104°F) <80% RH ­20 to 60°C (­4 to 140°F) <80% RH

1065DGMAN_20250-66_Rev1.indd 21

21
07/02/2019 11:03:49 AM

Maintenance, Recalibration, and Repair
· Clean the device with a soft, damp and lint-free cloth. Ensure that no moisture enters the housing. Do not use any sprays, solvents, alcohol-based cleaning agents or abrasive cleaners: only use clean water.
· The product must be repaired by the manufacturer.

22
1065DGMAN_20250-66_Rev1.indd 22

07/02/2019 11:03:49 AM

It is recommended that Digi-Sense products are calibrated annually to ensure proper function and accurate measurements; however, your quality system or regulatory body may require more frequent calibrations. To schedule your recalibration, please contact InnoCal, an ISO 17025 calibration laboratory accredited by A2LA.

Phone: 1-866-INNOCAL (1-866-466-6225) Fax: 1-847-327-2993 E-mail: sales@innocalsolutions.com Web: InnoCalSolutions.com
1065DGMAN_20250-66_Rev1.indd 23

23
07/02/2019 11:03:49 AM

For Product and Ordering Information, Contact:

Toll-Free: 1-800-323-4340 Phone: 1-847-549-7600 Fax: 1-847-247-2929 ColeParmer.com/Digi- Sense
1065DGMAN_20250-66_Rev1
24

Manual Part No. 00102-02

1065DGMAN_20250-66_Rev1.indd 24

07/02/2019 11:03:49 AM

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals