Oscilloscope UT-P33 Differential Probe Active Probe Instruction Manual

June 5, 2024
Oscilloscope

Oscilloscope UT-P33 Differential Probe Active Probe

Features

  • The UT-P33 differential probe provides a safety means for measuring differential voltage to all models of oscilloscopes.It can convert the high differential voltage(≤14KVp-p) into a low voltage(≤7V)and display on the oscilloscope.Its bandwidth is up to 120MHz,which is ideal for big power testing,development and maintain.
  • The UT-P33 is designed to operate with the 1MΩ impedance oscilloscopes. When combine with the 50Ω load , the attenuation will be 2 times.
  • UT-P33 is recommend to use with our own manufactured PL-10 to expand the measuring with the electricity meter to observe more accurate measurement.The accuracy of oscilloscope is 1%and the DMM is less than 1%)

Specifications

  1. Bandwidth:DC-120MHz

  2. Attenuation: X100,X1000

  3.  Accuracy:+/-1%

  4. Input voltage range(DC+AC PEAK TO PEAK )

    • ≤1.4KV for x100,(about 490V RMS)
    • ≤14KV for x1000,(about 4900V RMS )
  5. Permitted max input voltage:

    • Max differential voltage: 14KV(DC+AC PEAK to PEAK)
    • Max voltage between each input terminal and ground: 5KV RMS
  6. Input Impedance:
    Differential: 20MΩ /1pF
    Between terminal and ground:10MΩ /2pF

  7. Output voltage: ≤7V

  8. Output impedance: 50Ω

  9. Rise time: 3ns

  10. Rejection rate on common mode:
    60Hz:>80dB;100Hz:>60dB;1MHz:>50dB

  11. Power Supply:Only External 6V DC power supply.

  12. Consumption: 150mA max (0.9 Watt)

Operatingenvironmental conditions

  Reference Use Storage
Temperature +20℃…+30℃ 0℃….+50℃ -30℃….+70℃
Relative Humidity ≤70%RH 10%…85%RH 10%…90%RH
  1. Dimensions and weight:245X84X36mm; 500g
  2. Electrical safety to IEC 1010-1
    • Dual insulation
    • Installation category Ⅲ
    • Degree of Pollution 2
    • Related voltage or max line-earth:5KV RMS CE:EN50081-1 and 50082-1

Operating procedure

  • Connect the probe to the oscilloscope with the insulated BNC/BNC lead.
  • Adjust the vertical zero adjustment of the oscilloscope if necessary.
  • Select the attenuation ratio* and the vertical deviation of the oscilloscope in accordance with the conversion table below.

NB: The POWER light must come on.

Attenuation ratio X1000 X100

Voltage Input Range

(DC+AC Peak to PEAK)

| 14KV| 1.4KV

[N.B]
The real vertical deviation in V/div is equal to the attenuation factor multiplied by the range of vertical deviation selected on the oscilloscope. It will be doubled in the case of use of a 50 Ω load.

 Maintenance

For maintenance,only use specified spare parts.The manufacturer can not be held responsible for any accident arising following a repair made other than its after sales service or approved repairs.

Cleaning

This probe does not require any particular cleaning.If necessary,clean the case with a cloth slightly moistened with the soapy water.

Warranty

  • Unless notified to the country,our instruments are guaranteed against any manufacturing defect or material defect. They do not bear the specification known as the safety specification.
  • Our guarantee,which may not under any circumstances exceed the amount of the invoiced price,goes on further than the repair of our faulty equipment,carriage paid to our workshops.

Repair

Maintenance,repairs under or out of guarantee.Please return the product to the manufacturer.