Respiree RS001 Cardio Respiratory Monitor Owner’s Manual
- June 5, 2024
- Respiree
Table of Contents
- Respiree RS001 Cardio Respiratory Monitor About this ManuaL
- Product Description and Indications
- Intended Use
- Safety Information, Warnings and Cautions
- Technology Overview
- General Description for Oxygen Saturation (Sp02)
- Operation
- Using RS001.2.C and RS001.2.S Cardio-respiratory Monitor
- PULSE OXIMETRY (PR, Sp02) Measurement (HR Mode)
- Connecting to a Smart Device via Bluetooth (Optional)
- Models : RS001.2.D and RS001.2.DS
- RS001.2.DS Cardio-respiratory Software Dashboard Features
- Using RS001.2.D Cardio-respiratory Monitor
- Cleaning, Disinfecting, Charging and Service
- Service
- Disposal
- Troubleshooting
- Product Support
- Limited Warranty
- Specifications
- Sp02 Performance Specifications
- Medical Conditions
- Battery Life
- Physical Characteristics
- Compliance
- Symbols
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Respiree RS001 Cardio Respiratory Monitor About this ManuaL
Do not operate the RS001 Respiree Cardio-respiratory Monitor without
completely reading and understanding the instructions.
Always use the RS001 Respiree Cardio-respiratory Monitor precisely in
accordance with the directions in this manual including site selection and
sensor placement. Failure to follow all of the directions in this manual could
lead to inaccurate measurements.
Read and follow any warnings, cautions and notes presented throughout this
manual. The following are explanations of warnings, cautions and notes.
A warning is given when actions may result in a serious outcome to the patient
or user (for example, injury, serious adverse effect or death).
WARNING: This is an example of a warning statement.
A caution is given when any special care is to be exercised by the patient or
user to avoid injury to the patient, damage to this instrument, or damage to
other property.
CAUTION: This is an example of a caution statement.
A note is given when additional general information is applications
Note: This is an example of a note.
Product Description and Indications
Product Description
The RS001 Respiree Cardio-respiratory Monitor is intended as a noninvasive device that measures and displays arterial oxygen saturation (Sp02), Pulse Rate (PR) and Respiratory Rate (RR). RS001 Respiree Cardio-respiratory Monitor is available in the following versions:
Product Models | Features |
---|---|
RS001.2.S | Intended to measure and display arterial oxygen saturation (Sp02), |
Pulse Rate (PR) and Respiratory Rate (RR).
RS001.2.C| Intended to measure and display arterial oxygen saturation (Sp02),
Pulse Rate (PR) and Respiratory Rate (RR).
Bluetooth LE radio is intended to transfer of parameter data to a compatible smart device.
RS001.2.D| Intended to measure arterial oxygen saturation (Sp02), Pulse Rate (PR) and Respiratory Rate (RR) and used with RS001.2.DS for display.
Bluetooth LE radio is intended to transfer of parameter data to a compatible smart device.
RS001.2.A| Android App used for the display of result data (arterial oxygen
saturation (Sp02), Pulse Rate (PR) and Respiratory Rate (RR) from RS001.2.C.
RS001.2.DS| Intended to use data from RS001.2.D and display arterial oxygen
saturation (Sp02), Pulse Rate (PR) and Respiratory Rate (RR).
Intended Use
RS001 breathing frequency monitor is intended for measurement of respiratory
rate, heart rate and SPO2 in a non-invasive manner. Equipment is not intended
for connection to separate power supply when in use. The monitor is indicated
for use as suitable for use on adult patients. The operation of the monitor
has been studied on adults (age >18 years old) during rest conditions.
Indications for Use
The RS001 Respiree pulse oximeter and respiratory monitor is indicated for the
noninvasive spot checking of arterial oxygen saturation (Sp02) and pulse rate
(PR) for adult patients during no motion conditions and for patients who are
well or poorly perfused.
The RS001 Respiree pulse oximeter and respiratory monitor is indicated for the
noninvasive spot checking of respiration rate (RR) for adult patients.
RS001.2.D and RS001.2.DS is indicated for use in a clinical setting under a
physician or by healthcare professionals under the supervision of a physician.
RS001.2.C, RS001.2.S and RS001.2.A is indicated for use in a home setting and
may be applied by the patient. Training by a healthcare professional before
use is required and will be arranged by Respiree.
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
Safety Information, Warnings and Cautions
Safety Warnings and Cautions
- WARNING: Do not use RS001 Cardio-respiratory Monitor during magnetic resonance imaging (MRI) or in an MRI environment.
- WARNING: Do not place RS001 Cardio-respiratory Monitor or accessories in any position that might cause it to fall on the patient.
- WARNING: Do not use RS001 Cardio-respiratory Monitor during defibrillation.
- WARNING: Do not use RS001 Cardio-respiratory Monitor during electrosurgery.
- WARNING: Do no use RS001 Cardio-respiratory Monitor in the presence of flammable anesthetics or other flammable substances, oxygen-enriched environments or nitrous oxide to avoid the risk of explosion.
- WARNING: Only use the RS001 Cardio-respiratory Monitor in pulse oximetry mode to secure on the finger. Excessive pressure to a finger can cause skin damage.
- WARNING: Check the sensor site every hour to ensure adequate circulation, skin integrity, and sensor alignment. Skin damage, pressure necrosis or inaccurate readings may result.
- WARNING: Do not leave the RS001 Cardio-respiratory Monitor unattended around children. Small items such as the sensor may become choking hazards.
- WARNING: Turn off RS001 Cardio-respiratory Monitor when charging as use is not intended while charging.
- WARNING: Avoid charging for extended periods – recommended maximum of 120 minutes.
- WARNING: Do not open the RS001 Cardio-respiratory Monitor.
- WARNING: RS001 Cardio-respiratory Monitor is without any user-serviceable part inside and only qualified service personnel can perform maintenance service.
- WARNING: Do not expose the RS001 Cardio-respiratory Monitor to moisture (such as rain or washing under tap or shower) to ensure performance and device safety. Device is not waterproof and should not be immersed in fluids.
- WARNING: RS001 Cardio-respiratory Monitor can be affected by strong electromagnetic interferences during operation. When not in use, switch off the device or shift device to another location in case of strong interference.
- WARNING: RS001 Cardio-respiratory Monitor is only to be used for charging with the provided charging cable.
- WARNING: Use only Respiree RS001 approved patches and accessories with RS001. Any other 3rd party accessories may lead to degradation of signals.
- WARNING: Reposition RS001 Cardio-respiratory Monitor when in respiratory rate on chest once at least once every 24 hours to allow patient’s skin to breath.
- WARNING: No modification of this equipment is allowed. Do not modify equipment without authorization of manufacturer.
- WARNING: Reposition device if patient shows signs of dermal irritation.
- WARNING: Do not use the RS001 Respiree patch on open-wounds, injured skin or sensitive and fragile skin. If the patient has a history of known skin allergies, consult a doctor before using the patch.
Performance Warnings and Cautions
- WARNING: RS001 Cardio-respiratory Monitor is not an apnea monitor and should not be used for arrhythmia analysis.
- WARNING: RS001 is non-diagnostic and non-confirmatory for any disease. RS001 Cardio-respiratory Monitor should not be used as the sole basis for medical decisions. It must be used in conjunction with clinical signs and symptoms.
- WARNING: Do not self-diagnose or self-medicate on the basis of the measurements. Always consult your doctor if you are not feeling well.
- WARNING: Do not use RS001 Cardio-respiratory Monitor for continuous monitoring. It is intended for spot-check use only. No alarms are provided to support continuous monitoring.
- WARNING: RS001 Cardio-respiratory Monitor does not provide low SPO2 alarms.
- WARNING: Do not use RS001 Cardio-respiratory Monitor if it appears or is suspected to be damaged. Damage to internal parts can result in no or inaccurate readings.
- WARNING: Do not repair, open or modify the RS001 Cardio-respiratory Monitor. Damage to internal parts can result in no or inaccurate readings.
- WARNING: Do not use the RS001 Cardio-respiratory Monitor if the internal parts have been exposed to liquids. Damage to the internal parts may result in no or inaccurate readings.
- WARNING: Optical, pleth based measurements (e.g. Sp02 and RR) can be affected by the following:
- Improper placement or alignment of RS001
- Intravascular dyes such as indocyanine green or methylene blue
- Externally applied coloring and texture such as nail polish, acrylic nails, glitter, etc.
- Blood pressure cuff applied to the same arm as the sensor site under pulse oximetry mode
- Placing the RS001 sensor on any extremity with an arterial catheter
- Elevated levels of bilirubin
- Venous congestion
- Abnormal venous pulsations
- Abnormal pulse rhythms due to physiological conditions or induced through external factors
- Physiological conditions that can significantly shift the oxygen disassociation curve
- A physiological condition that may affect vasomotor tone or changes in vasomotor tone
- WARNING: Inaccurate Sp02 readings may be caused by:
- Elevated levels of COHb and/or MetHb
- Severe anemia
- Extremely low arterial perfusion
- Excessive induced motion
- Hemoglobinopathies and Hemoglobin synthesis disorders
- WARNING: Inaccurate respiration rate (RR) readings may be caused by:
- Improper RS001 placement or alignment on the chest
- Excessive motion
- WARNING: Properly apply and avoid using the RS001 Cardio-respiratory Monitor under high ambient light sources, fluorescent lights, infrared heating lamps and direct sunlight to minimize interference that may result in no or inaccurate readings.
- WARNING: RS001 Respiree patch accessory to be used within 1 year from exposure to air.
- CAUTION: When using RS001 Cardio-respiratory Monitor with a smart device, keep both devices within the recommended range of each other (see Specifications for details): moving outside of this range may cause a loss in connection with the smart device.
- cAUTION: When using the RS001 Cardio-respiratory Monitor with a smart device, relocate the devices away from sources that may interfere with the Bluetooth connection. The presence of other devices that may create radio frequency interference (RFI) may result in loss of Quality of Service (see Specifications for details) of the Bluetooth connection. Devices that may cause RFI include but are not limited to the following: electrocautery equipment, diathermy equipment, other cellular telephones, wireless PC and tablets, pagers, RFID devices, MRI and electromagnetic security systems.
- CAUTION: Do not attempt to remanufacture, recondition or recycle RS001 Cardio-respiratory Monitor as these processes may damage the internal parts. Damage to internal parts can result in no or inaccurate readings.
- Note: The RS001 Cardio-respiratory Monitor may be difficult to view when exposed to direct sunlight or bright lights.
- Note: Do not assess the accuracy of the RS001 Cardio-respiratory Monitor using a functional tester.
- Cleaning, Disinfecting, Service Warnings and Cautions
- WARNING: RS001 Cardio-respiratory Monitor uses a rechargeable battery. Limit the charging time to maximum of 120 minutes.
- WARNING: Thoroughly clean and low level disinfect the RS001 Cardio-respiratory Monitor before applying it to on a new patient
- CAUTION: Do not clean RS001 Cardio-respiratory Monitor with any chemical other than those specified in Cleaning, Disinfecting and Service. These substances may affect the device’s materials and damage internal parts.
- CAUTION: Do not submerge RS001 Cardio-respiratory Monitor in any cleaning solution or attempt to sterilize by autoclave, irradiation, steam, gas, ethylene oxide or any other method. This will seriously damage the device.
- CAUTION: Do not use undiluted bleach (5% – 5.25% sodium hypochlorite) or any cleaning solution other than those recommended in Cleaning, Disinfecting and Service of this manual. Permanent damage to RS001 Cardio-respiratory Monitor may occur if other unspecified solutions are used.
- CAUTION: Never submerge RS001 Cardio-respiratory Monitor in water or any other liquid solution. This may cause permanent damage to the sensor.
- CAUTION: The RS001 Respiree patch is a single-use and strictly one-time use patch only. Dispose off the patch after a maximum recommended time of 24 hours upon use.
- Compliance Warnings and Cautions
- WARNING: For RS001.2.C, RS001.2.S and RS001.2.A, device must be operated by patients that have received appropriate training by registered nurses or physicians prior to usage.
- WARNING: For RS001.2.C, RS001.2.S and RS001.2.A, results presented are non-diagnostic and non-confirmatory for any disease. Patients that are trained users are encouraged to see their physician or registered nurses if they are unwell and not rely on their results for any form of self-diagnosis.
- WARNING: For RS001.2.D and RS001.2.DS, device must be operated by trained personnel under the supervision of a physician. trained personal under the supervision of physician.
- WARNING: Any changes or modifications not expressly approved by Respiree shall void the warranty for this equipment and could void the user’s authority to operate the equipment.
- CAUTION: Comply with local laws in the disposal of the RS001 Cardio-respiratory Monitor and Respiree patch.
- CAUTION: This device has not been evaluated for use in aircrafts.
- Note: When using the device with wireless features, consideration should be taken to local government frequency allocations and technical parameters to minimize the possibility of interference to/from other wireless devices.
- Note: In accordance with international telecommunication requirements, the frequency band of 2.4 GHz is only for indoor usage to reduce potential for harmful interference to co-channel mobile satellite systems.
- note: The device complies with the Part 18 of FCC 47 CFR rules.
Technology Overview
The following chapter describes the general principles of operation,
parameters and measurements used by the Respiree RS001 products.
The RS001 Respiree Cardio-respiratory Monitor uses an all optical approach for
direct measurement of pulse oximetry (similar as that of conventional pulse
oximeters) and direct measurement of respiratory rate from the chest. The Near
Infra-red sensor module is common to both measurements.
To measure pulse oximetry and that measures displays Pulse Rate (PR or HR) and
Arterial Oxygen Saturation (Sp02), RS001 Cardio-respiratory Monitor uses a
finger adapter and a switch on the RS001 monitor to switch to the pulse
oximetry mode.
To measure respiratory rate (RR), RS001 Cardio-respiratory Monitor uses a
disposable patch that allows the sensor to connect to the chest for
measurement of chest-thorax movements. A switch on the RS001 monitor allows
the sensor to switch the respiratory rate measurement mode.
To measure respiratory rate on the chest, the Respiree Cardio-respiratory
Monitor makes use of the direct-contract optical diffuse reflectance principle
of operation as shown in the figure below. The light absorption changes due to
stretching of the skin is measured from a photo-sensor. Adaptive filters to
remove noise and motion as well as to calculate respiratory rate accurately
are utilized.
This figure is for conceptual purposes only.
To measure pulse oximetry on the chest, the Respiree Cardio-respiratory Monitor makes us of the photo-photoplethysmography effect that measures the amount of light absorbed by the varying quantities of arterial blood changes.
General Description for Oxygen Saturation (Sp02)
Pulse oximetry is governed by the following principles:
- Oxyhemoglobin (oxygenated blood) and dexoyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry)
- The amount of arterial blood in tissue changes with your pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
General Description for Pulse Rate (PR or HR)
Pulse rate (PR or HR) is measured in beats per minute (BPM) and is based on
the optical detection of the peripheral flow pulse.
General Description for Respiratory Rate (RR)
Respiration rate can be determined by the direct absorption change due to skin
stretching. The method measures the respirations per minute (RPM).
Operation
Models: RS001.2.C and RS001.2.S
Note: This section refers to models RS001.2.C and RS001.2.S to be used by
users in a home setting and have been appropriately trained by Respiree.
RS001.2.C and RS001.2.S Cardio-respiratory Monitor Features
ID | Description | Function |
---|---|---|
1 | On-Off Button | Press once to turn on device. Press once again to turn off |
device.
2| Display Screen| Display for measurements and indicators.
3| Battery Status Indicator| Indicates the remaining relative life of the
battery.
4| Recharging Port| Interface to plug in micro-USB cable for recharging of
battery
5| Mode Switch| Switch downwards for Pulse Oximetry Mode and switch upwards
for Respiratory Rate Mode. Mode type will be displayed on screen.
6| Sp02/RR
Indicator
| Under Pulse Oximetry Mode, Sp02 is indicated on the top of “HR”, while PR is
indicated next to bpm. Under Respiratory Rate mode, Sp02 is not indicated and
“RR” is replaced with “HR”.
7| Warning Indicator| Indicates the type of error and warning flags. 4 types
of warning indicators are provided. “Ok” if no error or warnings. “MOT” if
excess motion. “SAT” if ambient light is too bright. “ERR” if sensor is
faulty.
---|---|---
Using RS001.2.C and RS001.2.S Cardio-respiratory Monitor
RESPIRATORY RATE (RR) Measurement (RR Mode)
To take readings for Respiratory Rate, follow the instructions below:
Note: Before use, ensure sensor is in good condition
- Take the sensor and piece of RS001.1.P patch for use with RS001 device. Ensure the “sensing-side” of the sensor is along the same direction as the “longest-side” of the patch window. Place the hook side of the patch with the fastener side of the RS001 device.
- Remove the non-adhesive part of the patch.
- Turn on the device by pushing the “on-off button” on the side of the sensor. The display will turn on with the initialization of “RR” as mode.
- Place the sensor 2 fingers from the collarbone on the left-side of the chest.
- Wait for 1 minute and respiratory rate (RR) will be displayed on the screen. RR is refreshed every 5 seconds.
Note: If readings are ok, display will show “ok” under Warning Indicators. If readings are perturbed by motion or ambient light, display will show “MOT” or “SAT”. If sensor is faulty, readings will display “ERR”. See Troubleshooting.
Note: In cases where the user has hyperopia or presbyopia, it is recommended that the patient attempt the Bluetooth pairing process first (Page
- before attaching the sensor to the patch. This will allow the readout to
be provided immediately on the Android app upon attachment.
Note: Once reading is complete and spot measurement is taken, remove the sensor, remove the patch, throw away the patch and turn-off the sensor. Proceed to clean down the sensor.
WARNING: Do not remove the sensor for 1 minute when the sensor is on the patient.
PULSE OXIMETRY (PR, Sp02) Measurement (HR Mode)
To take readings for Pulse Oximetry, follow the instructions below:
Note: Before use, ensure sensor is in good condition
- Take the sensor and the pulse oximetry adapter for use with RS001 device. Switch the mode of operation downwards, away from the on-off button, to HR mode.
- Ensure the sensor fits into the adapter provided and turn on the sensor. Use the holes on the adapter as guide for fitting the sensor.
- Place the finger along the sensing side of the RS001 monitor.
- Take the strap of the adapter and place into the slot and tighten accordingly. Ensure placement of the finger is according to the picture in 3 and 4.
- Wait for 45 seconds and Sp02 and PR will be displayed on the screen. Sp02 and PR are refreshed every 5 seconds.
Note: If readings are ok, display will show “ok” under Warning Indicators. If readings are perturbed by motion or ambient light, display will show “MOT” or “SAT”. If sensor is faulty, readings will display “ERR”. See Troubleshooting.
Note: Once reading is complete and spot measurement is taken, remove the sensor, remove the adapter and turn-off the sensor. Proceed to clean down the sensor and adapter.
WARNING: Do not remove the sensor for 45 seconds when the sensor is on the patient.
Connecting to a Smart Device via Bluetooth (Optional)
Bluetooth LE is an optional feature and comes only with the RS001.2.C Cardio-
respiratory model version.
The RS001.2.C version provides a Bluetooth LE wireless feature to allow
connection to a compatible smart device. The Bluetooth communication is only
available to smart devices using the Respiree Health App. RS001.2.C version
can only communicate to a single smart device at one time to minimize the risk
of unauthorized access.
Note: The RS001.2.C requires the use of the Respiree Health App to
communicate to a compatible smart device.
Pair RS001 Cardio-respiratory Monitor to Smart Device
- Ensure the Bluetooth is enabled on the smart device through the device settings.
- Ensure the Respiree Health App is installed on the smart device. This is only available on an Android version and provided by Respiree.
- Launch the Respiree Health App once installed.
- Turn on the RS001 Cardio-respiratory Monitor
- Follow the Respiree Health App instructions to pair the device.
- When the Respiree Health App identified the RS001 Cardio-respiratory Monitor, press/select it to pair.
- Once the RS001 sensor is connected, the Respiree Health App returns to the Main screen.
- Place the RS001 sensor on the patient’s finger, confirm that readings on the RS001 sensor and the Respiree Health App are the same without a delay greater than 10 seconds
Note: If the delay is greater than 10 seconds, move the RS001 closer to the smart device.
Verify Paired RS001 Cardio-respiratory Monitor
-
On the smart device, access the Register Device option
-
Locate Paired Device
Note: RS001.2.C will always be in Bluetooth mode -
Compare the Paired Device information to the Bluetooth Serial Number RS XXXX.
Disconnect Paired RS001 Cardio-respiratory Monitor
- On the smart device and access the Respiree Health App
- Press/select the Register Device option.
- Select the device and click on Unregister device. The RS001.2.C sensor will be disconnected from the smart device. The sensor will need to be paired if it is to be used again with this smart device.
Models : RS001.2.D and RS001.2.DS
Note: This section refers to models RS001.2.D and RS001.2.DS to be used
by a physician or healthcare professional under the supervision of a
physician.
RS001.2.D Cardio-respiratory Monitor Features
ID | Description | Function |
---|---|---|
1 | On-Off Button | Press once to turn on device. Press once again to turn off |
device.
2| Display Screen| Display for measurements and indicators.
3| Battery Status| Indicates the remaining relative life of the battery.
4| Recharging Port| Interface to plug in micro-USB cable for recharging of
battery
5| Mode Switch| Switch downwards for Pulse Oximetry Mode and switch upwards
for Respiratory Rate Mode. Mode type will be displayed on screen.
Note: RS001.2.D and RS001.2.DS are to be used under supervision of a physician. RS001.2.D is to be used in conjunction with RS001.2.DS. All other instructions and warnings are similar to RS001.2.S and RS001.2.C.
RS001.2.DS Cardio-respiratory Software Dashboard Features
The RS001.2.DS software dashboard can be accessed by any web browser. The
default username and password are “admin” and “password”.
Note: Before use, please consult the Respiree technical assistant for
installation of RS001.2.DS and setup.
RS001.2.DS main dashboard interface consists of the below:
- Patient ID
- Room Number
- Registration Time
- Respiratory Rate
- Heart Rate
- Oxygen Saturation
- RR Sensor Contact Quality
- HR Sensor Contract Quality
- Battery Level of RR Sensor (for sensor on chest)
- Battery Level of HR Sensor (for sensor on finger)
RS001.2.DS main dashboard interface appears as follows: RS001.2.DS dashboard permits 3 main use-cases as follows
- Patient registration
- Changing sensor
- Patient deregistration
- To register patient, Click on Add Patient first. Then Enter Patient ID. Subsequently, Select Room and Click Assign. Following this, Select the RS001.2.D sensor for Respiratory Rate measurement and the RS001.2.D sensor for Pulse Rate measurement. See below process for patient registration.
- To replace sensor, click on the Pencil Icon on Patient ID on main dashboard. Then Click on Remove Sensor under either the Respiratory Rate or Heart Rate. Click OK to confirm removal. Reassign a new RS001.2.D sensor. Following are processes for changing or replacing the sensor.
- To remove or deregister patient, simply click on the “trash” icon to delete patient on the Cmain dashboard page.
Using RS001.2.D Cardio-respiratory Monitor
RS001.2.D is to be used in conjunction with RS001.2.DS.
RESPIRATORY RATE (RR) Measurement (RR Mode)
To take readings for Respiratory Rate, follow the instructions under “Using
RS001.2.C and RS001.2.S Cardio-respiratory Monitor ” from points (1) to (4):
Note: Before use, ensure sensor is in good condition
Note: Ensure RS001.2.D is registered to the RS001.2.DS dashboard
PULSE OXIMETRY (PR, Sp02) Measurement (HR Mode)
To take readings for Pulse Oximetry, follow the instructions under “Using
RS001.2.C and RS001.2.S Cardio-respiratory Monitor ” from points (1) to (4):
Note: Before use, ensure sensor is in good condition
Note: Ensure RS001.2.D is registered to the RS001.2.DS dashboard
Cleaning, Disinfecting, Charging and Service
Cleaning and Disinfecting RS001 Cardio-respiratory Monitor
WARNING: Before cleaning, read Cleaning, Disinfecting, Service Warning
and Cautions in this manual.
WARNING: Before cleaning, make sure the device is off and is not applied
to a finger and chest.
CAUTION: Thoroughly clean and low level disinfect the RS001 Cardio-
respiratory Monitor before applying it to on a new patient.
CAUTION: Thoroughly clean and low level disinfect the Respiree RS001
finger adapter before applying it to on a new patient.
Note: Before cleaning, make sure the micro-USB charging port is closed
with the port cap.
To clean and low-level disinfect the RS001 Cardio-respiratory Monitor, follow
the instructions below:
- Damp a soft cloth with a commercial solution of 70% isopropyl alcohol in water.
- Lightly wipe the surfaces of the device.
- Lightly wipe the surfaces of the adapter.
- Damp a moist cloth or cotton ball with a commercial solution of 70% isopropyl alcohol in water
- Clean the display
- Clean the photo-sensor parts gentle
- Avoid touching the cloth to the Velcro part
- Pay particular attention to cracks, crevices and hard to reach areas of the device
- Allow the RS001 Cardio-respiratory monitor and adapter to dry thoroughly before using them again The surfaces of the RS001 Cardio-respiratory Monitor have been treated to be chemically resistant to following solution:
- 70% Isopropyl Alcohol
Service
WARNING: Do not attempt to repair the RS001 Cardio-respiratory Monitor as
this may cause damage to the device and prevent it from operating properly.
If the device does not appear to be operating correctly, see Troubleshooting
in this manual.
Disposal
WARNING: Device to be cleaned and recycled as per local electronics
regulation. Patch is to be disposed as per normal hospital regulations.
Charging
WARNING: Do not overcharge the sensor. Maximum charge advisable is 120
minutes. WARNING: Turn off the RS001 Cardio-respiratory Monitor before
charging device.
Follow the following procedures to charge the sensor:
- Turn off the sensor before charging.
- Remove the USB charging port plastic cap.
- Insert provided micro USB cable into sensor.
- Connect USB cable to a 5V 1A maximum USB charger.
- Once fully charged (30 minutes), remove micro USB charger.
- Close the micro USB charger port with the plastic cap.
Troubleshooting
Error on Error Message | Possible Causes | Recommend Solutions |
---|---|---|
Device displays “ERR” | Faulty device | Contact Respiree Technical Services |
Device displays “MOT” | Significant amount of |
motion is being detected
| Check patient and request to keep stationary
Device displays “SAT”| Exposed to bright lights or sunlight| Relocate device
so that it is not directly under bright light
Device display does not switch to the correct mode| Mode switch is not in the
correct position
Faulty device
| Check that the mode switch is switched upwards (towards on-off button) for RR mode and downwards for HR mode
Contact Respiree Technical Services
Device does not turn on| Low Battery
Faulty device
| Charge monitor
Contact Respiree Technical Services
Product Support
For additional help, contact Respiree Technical Services at (+65) 90617570. Local contact information can be found at http://www.respiree.com
Limited Warranty
Respiree warrants to the original end-user purchaser the Respiree RS001
Cardio-respiratory Monitor and any software media contained in the general
packaging against defects in material and workmanship when used in accordance
with Respiree’s user manuals, technical specifications, and other Respiree
published guidelines for a period of 12 months from the original date the
Product was obtained by the end-user purchaser.
Respiree’s sole obligation under this warranty is the repair or replacement,
at its option, of any defective Product or software media that is covered
under the warranty.
To request a replacement under warranty, Purchaser must contact Respiree and
obtain a returned goods authorization number so that Respiree can track the
Product. If Respiree determines that a Product must be replaced under
warranty, it will be replaced and the cost of shipment covered. All other
shipping costs must be paid by purchaser.
The above warranty is in addition to any statutory rights provided to
Purchaser under applicable laws and regulations of the region in which the
product was sold to the extent that those rights cannot be disclaimed and are
superseded by the above described warranty to the extent permitted under
applicable laws and regulations of the region in which the product was sold.
All RS001 patch products sold separately by Respiree have a maximum 12 months
shelf life once the product is opened. This has been tested under accelerated
aging conditions.
Exclusions
The warranty does not cover the use or any defect in Respiree RS001 patch
products. The warranty does not apply to any non-Respiree branded product or
any software even if packaged with the Product or any Product that was (a) not
new or in its original packaging when supplied to Purchaser; (b) modified
without Respiree’s written permission; (c) supplies, devices or systems
external to the Product; (d) disassembled, reassembled or repaired by anyone
other than a person authorized by Respiree; (e) used with other products, like
new sensors, reprocessed sensors, or other accessories, not intended by
Respiree to be used with the Product; (f) not used or maintained as provided
in the operator’s manual or as otherwise provided in its labelling; (g)
reprocessed, reconditioned or recycled and (h) damaged by accident, abuse,
misuse, liquid contact, fire, earthquake or any other external cause.
No warranty applied to any Product provide to Purchaser for which Respiree, or
its authorized distributor, is not paid, and these Products are provide AS-IS
without warranty.
Limitation of Warranty
Except as otherwise required by law or altered by the purchase agreement, the
above warranty is the exclusive warranty that applies to the Product and
Software media, and Respiree does not make any other promises, conditions or
warranties regarding the Product. No other warranty applies, express or
implied, including without limitation, any implied warranty of
merchantability, fitness for a particular purpose, satisfactory quality, or as
to the use of reasonable skill and care. In so far as the above warranties
cannot be disclaimed, Respiree limits the duration and remedies of the
warranties to the duration and to the remedies set forth above and as
permitted by law. See the licensing terms for the terms and conditions that
apply to and Software accompanying the Product.
Additionally, Respiree will not be liable for any incidental, indirect,
special or consequential loss, damage or expense arising from the use or loss
of any Products or Software. In no event shall Respiree’s liability arising
from any Product or Software (under contract, warranty, tort, strict liability
or otherwise) exceed the amount paid by purchaser for the Product or Software.
The above limitations do not preclude any liability that cannot legally be
disclaimed by contract.
Sales and End-user License Agreement
This document is a legal agreement between you (“purchaser”) and Respiree Pte
Ltd
(“Respiree”) for the purchase of this Product (“Product”) and a license in the
included or embedded Software (“Software”) except as otherwise expressly
agreed in a separate contract for the acquisition of this Product, the
following terms are the entire agreement between the parties regarding your
purchase of this Product. If you do not agree to the terms of this agreement,
promptly return the entire Product, including all accessories, in their
original packages, with your sales receipt to Respiree for a full refund.
Restrictions
- Copyright restrictions: The Software and the accompany written materials are copyrighted. Unauthorized copying of the Software, including Software that has been modified, merged or included with other software or the written materials is expressly forbidden. Purchaser may be held legally responsible for any copyright infringement that is caused or incurred by Purchaser’s failure to abide by the terms of the Agreement. Nothing in this License provides any rights beyond those provided by 17 U.S.C 117.
- Use Restrictions: Purchaser may physically transfer the Product from one location to another provided that the Software is not copied. Purchaser may not electronically transfer the Software from the Product to any other instrument. Purchaser may not disclose, publish, translate, release, distribute copies of, modify, adapt, translate, reverse engineer, decompile, disassemble or create derivates works based on the Software or the written materials or the hardware.
- Transfer Restrictions: In no event may Purchaser transfer, assign, rent, lease, sell or otherwise dispose of the Product of the Software on a temporary basis. Purchaser shall not assign or transfer this License, in whole or in part, by operation of law or otherwise without Respiree’s prior written consent, except that the Software and all of Purchaser’s rights hereunder shall transfer automatically to any party that legally acquires title to the Product with which this Software is included. Any attempt to assign any rights, duties or obligations arising hereunder other than as set forth in this paragraph shall be void.
Specifications
Display Ranges
Parameter | Display Ranges |
---|---|
Sp02 (Oxygen Saturation) | 0% to 100% |
PR (Pulse Rate) | 25 bpm to 240 bpm |
RR (Respiration Rate) | 4 rpm to 70 rpm |
Performance Specifications
Sp02 Accuracy
Condition| Range| Population| ARMS*
No Motion| 70% to 100%| Adults| 3%
Low Perfusion| 70% to 100%| Adults| 3%
Pulse Rate Accuracy
Condition| Range| Population| ARMS*
No Motion| 40 bpm to 200 bpm| Adults| 3 bpm
Low Perfusion| 40 bpm to 200 bpm| Adults| 3 bpm
Respiration Rate Accuracy
Condition| Range| Population| ARMS*
No Motion| 4 rpm to 40 rpm| Adults| 3 bpm
Motion| 4 rpm to 40 rpm| Adults| 3 bpm
- ARMS accuracy is a statistical calculation of the difference between the device measurements and the reference measurements. Approximately at least two-thirds of the device measurements fell within +/- ARMS of the reference measurements in a controlled study.
Sp02 Performance Specifications
Figure 1. Bload-Altman plot of the RS001 Cardio-respiratory Monitor Pulse Oximeter (ARMS 70-100%)
RR Performance Specifications
Figure 2. Bland-Altman plot of the Respiratory Rate measurement with respect to the respiration rate value on a simulator (RRef) Figure 3. Bland-Altman plot of the Respiratory Rate measurement with respect to the clinician-scored respiration rate value on hospitalized subjects
Medical Conditions
Medical Conditions from clinical study of hospitalized patients
Disease Type | Number |
---|---|
Chronic Obstructive Pulmonary Disease | 27 |
Asthma | 37 |
Pneumonia | 44 |
Type 1 and Type 2 Respiratory Failure | 10 |
Congestive Heart Failure | 7 |
Influenza | 3 |
Sepsis | 2 |
Bronchiectasis | 2 |
Pulmonary Embolism | 1 |
Pulmonary Odema | 1 |
Myocardial Infarction | 4 |
Urosepsis | 1 |
Cellulitis | 1 |
Battery Life
Continuous use-case battery life of 24-48 hours. Operates with a rechargeable battery. Not to be replaced.
Environment Conditions
- The recommended storage and environment conditions are as follows:
- Operating temperature: 15 – 35 degrees
- Storage temperature: 15 – 40 degrees
- Relative humidity: 15% – 95% (no condensing)
Transportation Conditions
- The recommended transportation conditions are as follows:
- Operating temperature: 15 – 35 degrees
- Storage temperature: 15 – 40 degrees
- Relative humidity: 15% – 95% (no condensing)
Physical Characteristics
- Dimensions: 43 mm x 43 mm x 13 mm
- Weight: 22.1 grams
Compliance
- ISO 80601-2-61: 2017
- IEC 60601-1-2: 2014
- AAMI/IEC 60601-1:2005 + AMD 1:2012
- FCC 47 CFR Part 15
- IEC 60601-1-6:2010, AMD1:2013 with IEC 62366-1:2015 and IEC 60601-1:200
- FCC ID: 2A3T2RS001-2
- The radiated output power of this device meets the limits of FCC radio frequency exposure limits.
For a Class B digital device or peripheral, the instructions furnished to the user shall include the following or similar statement, placed in a prominent location in the text of the manual: NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Symbols
The following symbols may appear on the product or product labeling:
References
Read User Manual Online (PDF format)