hippomed AirOne Flex Battery Ultrasound Inhaler for Horses
Read and pay attention to these operating instructions before assembly and initial commissioning! Keep these operating instructions for future reference!
Foreword
Information about the manual
We reserve the right to make changes due to further technical development
compared to the data and illustrations mentioned in this manual. Reprints,
translations and reproductions in any form, including excerpts, require the
written consent of the manufacturer. Abbreviations, units, technical terms,
special designations or industry-standard terminology used in this manual are
explained in more detail in the “Appendix” section. This manual is part of the
scope of delivery.
- It must be kept close at hand and remains with the unit even if it is sold.
- This manual is not subject to a revision service. The respective current status can be obtained from the specialist trader or directly from the manufacturer.
- It has a modular structure and refers exclusively to the named product.
- These operating instructions describe the operation, use and maintenance of the unit.
The safety instructions are intended to prevent personal injury, damage to animals and the unit. All operators are required to read and observe these safety instructions at all times. The manufacturer accepts no liability or warranty for damage and malfunctions resulting from non-observance of the operating instructions!
Notes on the operating instructions
- Version/revision: 1.0
- Creation date: 2021-11-23
- Last change: 2022-04-07
Language of the original operating instructions: German
Representation of safety instructions
Safety instructions in the operating instructions
Danger!
The imminent danger which leads to serious physical injury or death.
Caution!
Risk of physical injury to people and animals and material damage if used incorrectly!
Important!
The potentially harmful situation that could damage the product or an object nearby
Safety
General
- Read the following instructions carefully before commissioning and using the unit.
- Anyone tasked with commissioning, operation or maintenance must have read and understood the operating instructions and in particular the chapter on safety.
- The unit must only be operated, maintained and repaired by trained and authorised personnel.
- Failure to comply with the individually listed points in the operating instructions can result in physical injury or material damage for which the manufacturer assumes no
liability.
General information regarding safety and warning instructions
Danger!
The imminent danger which leads to serious physical injury or death.
Lithium-ion batteries (Li-ion) pose a risk of explosion and fire. All safety
instructions must be observed in order to reduce the risk and to prevent
misuse.
- The unit contains a Li-ion battery.
- Fire hazard.
- The Li-ion battery cannot be replaced. Do not open the product!
- Only connect the unit to a supply network with a protective earth conductor.
- Only use the supplied mains adapter and cable to charge the unit/battery.
- Check the unit, mains adapter and charging cable for damage before use.
- Only operate the unit if it is undamaged.
- Do not operate the unit in an explosive atmosphere.
- Do not subject the unit to mechanical shocks, as this could damage the battery.
- Do not allow children to play with the unit.
- Protect the unit from heat (> +60 °C ), excessive sunlight and fire, as there is a risk of explosion.
- Store the unit at room temperature. Do not store the unit near heat sources or heaters.
- Keep the unit dry.
- Do not open faulty units. Faulty units may only be repaired by the manufacturer.
- Do not dispose of the unit with the Li-ion battery in household waste, fire or water. Observe local disposal regulations.
Caution!
Risk of physical injury to people and animals and material damage if used incorrectly!
-
Children and infirm individuals must only use the device under supervision.
-
Note the safety and maintenance instructions.
-
The unit must be used exclusively in accordance with intended use and in a technically safe condition! The operational safety of the unit is only guaranteed if it is used for
its intended purpose. Failure to comply with the individually listed points in the operating instructions can result in physical injury or material damage for which the manufacturer assumes no liability. -
Keep all safety and danger notices on/at the unit complete and always in a legible condition.
-
The maintenance intervals specified in these operating instructions must be observed.
-
Only use genuine spare parts.
Residual risks and foreseeable misuse
Risks for persons:
under the following circumstances, the safety of persons and material objects
can be endangered:
- Troubleshooting, adjustment, cleaning, repair and maintenance work contrary to the instructions in the operating instructions.
- If untrained personnel are working on electronic parts.
- Failure to comply with the safety instructions in the operating instructions.
- Unauthorised modifications to the unit.
- Use of non-original manufacturer spare parts
The following points summarise further measures aimed at protecting against personal and material risks:
- Mount the unit in a place that is well-ventilated and free from electromagnetic interference.
Risk for animals:
- The animals are not yet accustomed to using the inhaler (risk of panic reaction).
- The animals are unsupervised during use.
- The animals are not tethered correctly to the site.
Description
Description of the unit
Ultrasound Inhaler for Horses
The devices are designed to split (nebulise) inhalants by ultrasound into
components small enough for them to be optimally absorbed by the horse through
inhalation. The choice of inhalant is the operator’s responsibility.
Caution!
The unit must be used exclusively in accordance with intended use and in a
technically safe condition! The operational safety of the unit is only
guaranteed if it is used for its intended purpose. Failure to comply with the
individually listed points in the operating instructions can result in
physical injury or material damage for which the manufacturer assumes no
liability.
Scope of delivery
The AirOne Flex ultrasound inhaler is made up of the following components:
- Nebuliser and charger (1.1)
- Nebulising pot (fixed component of the nebuliser)
- Removable nebulising pot cover with connecting pieces (3.1)
- Breathing mask with connecting pieces (4.1), adjustable fresh air supply (4.2), rubber seal (4.3), adjustable head strap (4.4) and one inhalation (4.5) and two removable exhalation valves (4.6)
- Connecting pieces, breathing tube 170 cm (5.1)
- Wall bracket with 4 screws, 4 dowels and installation instructions
- Transport box with cover (not shown)
The accessories also include:
a measuring cup, 30 nebuliser cups, 1 litre of contact liquid.
Correct use
This unit must only be used to generate ultrasonic mist with the original
contact liquid. The use of nebuliser add-ons is the sole responsibility of the
operator. The regulations
set out in these operating instructions must be complied with. Any other use
of the unit, e.g. in any other area than those described, shall be deemed
incorrect use. The manufacturer accepts no liability for any resulting
physical and/or material damage.
Technical data
Item number/type designation: AirOne Flex battery-powered ultrasound nebuliser
- Power consumption: 700 mA (0.7 A)
- Nebuliser output: average: 6.7 ml/min, maximum: 15 ml/min
- Protection class: III
- Particle size: 0.47 – 6 µm
- Protection type: IP54
- Input voltage: AC 14.4 V 50 Hz
- Weight of base unit: 700 g
- Charger: 100 V-240 V AC 50-60 Hz
Ambient conditions
during operation: 0°C to 40°C and 700 to 1060 hPa
Preparation and commissioning
You have chosen a quality product from Hippomed! With the AirOne Flex you have
acquired an innovative inhaler that combines all the experience of the last 20
years on the
subject of ultrasound-based cold nebulisation. Of course, you can assume that
the proven quality features and characteristics of the AirOne series in terms
of ultra-fine particle size and optimal nebulisation have been retained in the
AirOne Flex. This means that not only are the bronchi reached, but also the
further branches down to the alveoli.
Preparation
Positioning the AirOne Flex
The unit can be wall mounted or used integrated with the mask directly on the
horse. We recommend the integrated use including mask directly on the horse.
Nebulising chamber and nebulizer cups
Grip the nebulizing chamber cover at the tube connection piece and pull it
upwards with a slight tilting movement. The locking mechanism releases and the
nebulizing chamber cover can be removed and placed to one side. Pour in 60 ml
of contact fluid into the nebulizing chamber using the measuring cup. Shake
well beforehand. Then place the nebulizer cup in the nebulizing chamber and
add the medium to be nebulized to the cup up to a maximum fill level of 50 ml.
Replace the nebulizing chamber cover and push it down firmly until it closes
flush with the upper edge of the housing. Important: the AirOne Flex delivers
its best performance with a MAXIMUM fill level of 50 ml in the nebulizer cup.
You should therefore avoid exceeding this amount. The duration of inhalation
is determined by the properties of the preparation to be nebulized. Please do
not nebulize any oil-containing substances. The generally recommended duration
for inhalation is around 15 – 20 minutes. For technical reasons, the
nebulization output diminishes towards the end of a treatment session. A
residual volume of around 5 ml may remain in the nebuliser cup. A
comparatively long nebulising session is needed to nebulise this remaining
amount. We therefore recommend ending the treatment session. Also, always make
sure that the white clamping ring in the attachment piece of the nebulising
chamber cover is located as far as possible at the front of the opening
towards the tube and that the slot in the clamping ring is facing upwards.
The following statements only apply to stationary use with the 175 cm tube: this prevents too much of the bubbling medium from travelling into the tube and accumulating there. Condensation may occur in the breathing tube during treatment. Occasional raising of the tube can cause a backflow of the condensation water into the nebuliser cup
Preparing the horse for inhalation
For safety reasons, the horse being treated must wear a halter and be tethered
on both sides with rubber ropes (www.kerbl.de, Art No. 3210370 – 3210372)
Stationary use
Check the correct fit of the rubber seal on the breathing mask. The arrow on
the rubber seal must point towards the front of the mask and down towards the
connection piece. Push the mask with the rubber ring over the horse’s nostrils
and fasten it securely with the head strap.
Mobile use
Ideally, you already adjust the length of the head strap before the first
inhalation so that it only needs to be attached directly to the mask with the
small hook-and-loop fastener.
After filling and preparing the nebuliser, the mask can be attached to the it:
-
slide the guide rail on the mask into the slot at the back of the nebuliser
-
If the mask is correctly seated in the rail, you can pull the fastening strips on both sides over the retaining pins on the nebuliser
-
Check the tight fit between the nebuliser and the mask and the correct fit of the grey seal on the mask
-
Take the connecting piece and turn it slightly clockwise onto the connection of the nebulising chamber cover, turn the other end of the connector slightly clockwise onto
the connection on the mask (it does not have to be turned as far as it will go, as there is no tension) -
Stand to the left of the horse
-
Slip the mask on from the neck over the nostrils
-
Hold the AirOne Flex with one hand and place the blue head strap over the neck with the other hand, then fasten the short hook-and-loop fastener to the mask.
The AirOne Flex has a special feature. Use can be interrupted for a short time
in the event of sideways movements of the head, depending on the filling
quantity and the
movement. The water level indicator lights up and goes out as soon as the head
returns to the starting position. Inhalation continues automatically. If the
LED does not go out automatically, please check the filling quantity of the
contact liquid.
Caution!
The fresh air supply must be fully open to avoid possible panic reactions by
the horse.
Commissioning and setting the fresh air supply
Switch the AirOne Flex on by pressing the ON/OFF button on the control panel.
The function is signalled by a green operation indicator lighting up.
Inhalation of the mist is best checked on the top of the nebulising chamber
cover.
Adjusting the fresh air supply on the breathing mask
Stepless closure of the fresh air supply on the breathing mask means that the
nebulisation quantity and thus the efficiency can be varied. However, this
must be done only very carefully and at the same time observing the horse’s
behaviour. The animal should show calm, even breaths. If the fresh air supply
to the breathing mask is closed completely too soon, the horse may panic. The
horse must first learn inhalation (breathing) and therefore the inhalation
times vary depending on the horse and its state of health. The inhalation time
period is 10 to max 20 minutes.
Meaning of the LED displays
Temperature LED lights up
If the “Temperature” indicator on the AirOne Flex’s control panel lights up, the unit has switched off due to overheating of the contact liquid. Allow the contact liquid to cool down for at least 10 minutes The warmer the contact medium, the sooner the unit switches off. The same also applies to extremely high outside temperatures, in which case you should use a cooled contact medium.
Water level LED lights up
When the “Water level indicator” lights up on the AirOne Flex’s control panel
(membrane keypad), there is normally too little contact liquid in the
nebulising chamber. Switch
off the unit, remove the nebulising chamber cover and nebuliser cup and
replace the contact liquid. After switching on, treatment with the AirOne Flex
can be continued. If
the control indicator continues to light up, this could also be due to
deposits on the sensors in the nebulising chamber. Carefully remove the
deposits with a soft brush and a little vinegar-based cleaner. If the unit
still does not work after this, there is probably a technical defect and you
must send it to the service centre for inspection.
Instructions for battery and charger
The battery in the AirOne Flex must be charged both before and after
inhalation. If the device is not used for a prolonged period of time, you
should charge the battery every two to three months to increase the life of
the battery pack and avoid deep discharge.
Caution!
To charge, first insert the plug of the charger supplied into the socket and
then the plug of the power adapter into the charging socket on the back of the
AirOne Flex.
The AirOne Flex cannot be switched on during the charging process. The
charging time is approx. 2 to 3 hours, depending on the charge status of the
battery. The charger’s
LED changes its colour from permanent to flashing yellow-orange to green
(fully charged). If the LED flashes green after connecting to the charging
socket, please check
that the plug is inserted correctly as far as it will go, otherwise there is
no connection and the battery will not be charged. In any case, red means that
there is a technical fault and says nothing about the state of charge. If you
pull the plug out of the charging socket after charging and immediately plug
it in again, the charger will not show green (fully charged), but will
initially flash yellow-orange for a period of up to approx. 4 min. and then
return to green. This is a normal process while the charger checks the state
of charge.
See also the enclosed manufacturer’s operating instructions The charger can
only charge the battery in a temperature range of 0 °C to + 40°C. If you are
outside this temperature, this is signalled by a white flashing LED. With a
fully charged battery pack, you can provide inhalations for at least two
20-minute cycles, with a 10-minute break between cycles.
Caution!
Only use the original charger supplied with the AirOne Flex. Other chargers
will inevitably damage the AirOne Flex. Failure to do so automatically voids
the warranty. This special charger must be used exclusively with the AirOne
Flex and not with the previous-version AirOne battery. Similarly, the charger
for the AirOne battery must not be
connected to the AirOne Flex. Both have a different output voltage and the
units would be damaged. For safety reasons, the battery has an automatic cut-
off in the event of a possible overload. If such a case occurs, the AirOne
Flex can no longer be switched on. A reset is performed by briefly plugging
and unplugging the charging cable into/from the charging socket. The unit
should now work again by pressing the ON/OFF membrane button. If this is not
the case, there is a fault and the unit must be checked by a service centre.
Maintenance and care
Careful cleaning and disinfection
To ensure a sustained high level of treatment success, it is essential that
the AirOne Flex is carefully cleaned and disinfected after each use.
Commercially available washing up liquid is suitable for external cleaning.
For disinfection, please use the disinfectant included with the AirOne. Follow
the instructions for use on the disinfectant label. Remove the exhalation
valves on the breathing mask by turning them a quarter turn to the left
(bayonet catch) and remove the nebulising chamber cover on the nebuliser.
Carefully spray all components, namely the mask, valves and nebulising chamber
cover. Hang up the breathing tube and spray into the tube. Please observe the
information on exposure time on the disinfectant’s instructions for use. The
nebuliser cup can be used again after cleaning. However, it is recommended
that the cup is replaced after 3 treatments. After the exposure time, rinse
all parts with fresh water and leave to dry in a suitable place. Then
reassemble all the parts and store them in the transport box provided. The
AirOne Flex is thus perfectly prepared for the next use.
Important information about the contact liquid
Caution!
Long-term trouble-free operation is only guaranteed if the contact fluid
supplied is used. If other agents such as distilled or tap water are used as a
contact medium,
the oscillator unit, the heart of the unit, will be permanently damaged and
the unit will then have to be repaired and all warranty claims will be
invalidated. The contact
agent is produced in the laboratory with a conductivity specially adapted to
the AirOne Flex. If this is not or no longer correct, the water level
indicator lights up.
The contact liquid is only renewed when the water level indicator lights up or
there is visual contamination. If the AirOne Flex is not used for more than a
week, please
remove the contact fluid and only refill the AirOne Flex when it is used
again.
Further care instructions
Caution!
Do not carry out any descaling activities. These permanently damage the
surface at the bottom of the nebulising chamber (ultrasonic oscillator /
transducer) and it will need to be replaced. Never touch or clean the
oscillator with a hard object. Do not use acidic cleaning agents. For optimal
care of the nebulising chamber, take a soft brush, pour water without
additives into the nebulising chamber and use the brush to swirl up any
possible residue and then rinse it out. You can wipe the housing with a damp
cloth and then rub it dry.
Any questions?
We’ll be happy to advise you – simply contact us on:
T: +49 (0)8086 933 327
T: +49 (0)5204 92970
E: inhalatoren@kerbl.com
For chargeable or warranty repairs, please download the Hippomed repair
application and fill it out in full:
https://www.kerbl.com/de/service/downloads
Troubleshooting
Error | Remedy |
---|---|
Temperature LED lights up | Chapter 5.2.2.1 |
Water level LED lights up | Chapter 5.2.2.2 |
Horse is not inhaling properly | Check the inhalation valves for contamination |
Check the tube connection between the unit and mask Check the air supply setting on the mask
No inhalation mist visible| Check the content of the nebuliser cup
Unit not switched on (switch LED must light up green) Oscillator contaminated – see Chapter 6.3
Oscillator faulty – contact the service centre
Unit does not switch on| No battery charge – charge the battery
Switch faulty – contact the service centre
Battery not charging/no LED display on the charger| Check the power supply to
the charger
Battery not charging/LED display on the charger| See chapter 5.3 or the
charging socket is oxidised: wet the plug with contact spray, plug it in and
out several times and turn it in the socket at the same time.
Warranty conditions
Albert Kerbl GmbH grants you a 2-year warranty on the ultrasound base unit in
accordance with our general warranty conditions. These general warranty
conditions can be
found at www.kerbl.com/guarantee. We will
also be happy to send you the warranty conditions by post. Your statutory
warranty rights remain unaffected by our guarantee.
This warranty begins on the date of purchase. The warranty does not cover:
- Damage caused by natural wear and tear (wear parts such as the oscillator/transducer, battery), excessive strain and incorrect use (non-compliance with operating instructions).
- Damage caused by force majeure or sustained during transport.
- Damage to rubber parts, e.g. on access lines.
- Damage to the mask including its attachments (valves) and the tube.
Warranty repairs are only carried out by the service centre for DE/AT or an authorised contracted workshop! The manufacturer reserves the right to reject any warranty claims in the event of non-compliance. The unit must for this purpose be delivered to the specified service address, at the risk of the unit’s owner, complete, packaged and with sufficient postage, including the original invoice. Please provide as much information as possible on the nature of the problem. Costs incurred as a result of unjustified defect claims will be invoiced to the unit’s owner.
CE-/UKCA-conformity declaration
Albert KERBL GmbH hereby declares that the product/device described in these
instructions complies with the fundamental requirements and other relevant
stipulations and regulations. The CE-/UCKA-mark confirms compliance with the
Directives of the European Union or the relevant UK legislation.
Electrical waste
It is up to the operator to dispose of the device responsibly after its useful life in accordance with local regulations. Consult the relevant regulations for your country. The device must not be disposed of in household waste. Under the terms of the EU Directive on the disposal of old electrical and electronic equipment, the unit can be taken to local authority-run refuse sites or recycling plants free of charge, or returned to dealers who offer a returns service. Correct disposal protects the environment and prevents any possible harmful impacts on humans and the environment.
Spare parts and accessories
Item number | Article designation | Illustration |
---|---|---|
3211805 | Complete inhaler with warmblood mask and accessories in transport | |
box | ——– | |
3211804 | Inhaler with accessories without mask | **** |
………
3211804-0| Inhaler without accessories| 1
3211821| Wall bracket incl. 4 screws and 4 dowels with installation
instructions for AirOne| 6
3211826| Transport box with cover| 7
3211816| Complete warmblood mask with valves, seal and head straps| 4-4.6
3211817| Complete coldblood mask with valves, seal and head straps| 4-4.6
3211816-0| Warmblood mask without accessories| 4
3211832| Seal for warmblood mask| 4.3
3211830| Inhalation valve for breathing mask| 4.5
3211828| Coldblood mask without accessories| 4
3211859| Seal for coldblood mask| 4.3
3211831| Exhalation valve for breathing mask| 4.6
3211825| Head strap for breathing mask| 4.4
3211823| Breathing tube 170 length| 5.1
3211824| Connecting piece 25 cm| 5
3211839| Nebulising chamber cover| 3
3211841| Snap ring for nebulising chamber cover| 8
3211818| Nebuliser cup, pack of 30| 10
3211822| Measuring cup 250 ml| 9
3211819| Contact fluid 1 litre| 11
299310| Disinfectant 1 litre (Please ask for availability outside DE/ AT!)| 12
Albert Kerbl GmbH
Felizenzell 9
84428 Buchbach, Germany
Tel. +49 8086 933 – 100
Fax +49 8086 933 – 500
info@kerbl.de
www.kerbl.de
References
- Home - www.austria.at
- electronic.de
- Everything around animal breeding, husbandry and agricultural supplies - Albert Kerbl GmbH
- Page d'accueil - Kerbl France
- Home - Kerbl UK Ltd
- Guarantee - Albert Kerbl GmbH
- Everything around animal breeding, husbandry and agricultural supplies - Albert Kerbl GmbH
- Downloads - Albert Kerbl GmbH
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>