UNI-T UPO3000E Series Digital Phosphor Oscilloscopes User Guide
- June 9, 2024
- UNI-T
Table of Contents
UNI-T UPO3000E Series Digital Phosphor Oscilloscopes User Guide
Copyright Information
Uni-Trend Technology (China) Co., Ltd. All rights reserved.
Trademark Information
UNI-T is the registered trademark of Uni-Trend Technology (China) Co., Ltd.
Document Version
MSO/UPO300020221130-V1.00
Statement
- UNI-T products are protected by patent rights in China and foreign countries, including issued and pending patents.
- UNI-T reserves the rights to any product specification and pricing changes.
UNI-T reserves all rights. Licensed software products are properties of Uni-Trend and its subsidiaries or suppliers, which are protected by national copyright laws and international treaty provisions. Information in this manual supersedes all previously published versions.
Safety Requirements
This section contains information and warnings that must be followed to keep the instrument operating under safety conditions. In addition, user should also follow the common safety procedures.
Safety Precautions
Warning
Please follow the following guidelines to avoid possible electric shock and risk to personal safety
Users must follow the following conventional safety precautions in operation, service and maintenance of this device. UNI-T will not be liable for any personal safety and property loss caused by the user’s failure to follow the following safety precautions. This device is designed for professional users and responsible organizations for measurement purposes.
Do not use this device in any way not specified by the manufacturer. This device is only for indoor use unless otherwise specified in the product manual.
Safety Statement
Warning
“Warning” indicates the presence of a hazard. It reminds users to pay attention to a certain operation process, operation method or similar. Personal injury or death may occur if the rules in the “Warning” statement are not properly executed or observed. Do not proceed to the next step until you fully understand and meet the conditions stated in the “Warning” statement.
Caution
“Caution” indicates the presence of a hazard. It reminds users to pay attention to a certain operation process, operation method or similar. Product damage or loss of important data may occur if the rules in the “Caution” statement are not properly executed or observed. Do not proceed to the next step until you fully understand and meet the conditions stated in the “Caution” statement
Note
“Note” indicates important information. It reminds users to pay attention to
procedures, methods and conditions, etc. The contents of the “Note” should be
highlighted if necessary.
Safety Sign
| Danger| It indicates possible danger of electric shock, which may cause
personal injury or death.
---|---|---
| Warning| It indicates that you should be careful to avoid personal injury
orproduct damage.
| Caution| It indicates possible danger, which may cause damage to this device
or other equipment if you fail to follow a certain procedure or condition. If
the “Caution” sign is present, all conditions must bemet before you proceed to
operation.
| Note| It indicates potential problems, which may cause failure of thisdevice
if you fail to follow a certain procedure or condition. If the
|
| “Note” sign is present, all conditions must be met before thisdevice will
function properly.
---|---|---
| AC| Alternating current of device. Please check the region’s voltagerange.
| DC| Direct current device. Please check the region’s voltage range.
| Grounding| Frame and chassis grounding terminal.
| Grounding| Protective grounding terminal.
| Grounding| Measure grounding terminal.
| OFF| Main power off.
| ON| Main power on.
| PowerSupply| Standby power supply: when the power switch is turned off,
thisdevice is not completely disconnected from the AC power supply.
CAT I| Secondary electrical circuit connected to wall sockets through
transformers or similar equipment, such as electronic instruments and
electronic equipment; electronic equipment with protective measures, and any
high-voltage and low-voltage circuits, such as the copier in the office
CAT II
| CATII: Primary electrical circuit of the electrical equipment connected to the indoor socket via the power cord, such as mobile tools, home appliances, etc. Household appliances, portable tools (e.g. electric drill), household sockets, sockets more than 10 meters away from CAT III circuit or sockets more than20 meters away from CAT IV circuit
CAT III
| Primary circuit of large equipment directly connected to the distribution
board and circuit between the distribution board and the socket (three-phase
distributor circuit includes a single commercial lighting circuit). Fixed
equipment, such as multi-phase motor and multi-phase fuse box; lighting
equipment and lines inside large buildings; machine tools and
powerdistribution boards at industrial sites (workshops)
CAT IV| Three-phase public power unit and outdoor power supply line equipment.
Equipment designed to “initial connection”, such as power distribution system
of power station, power instrument, front-end overload protection, and
anyoutdoor transmission line
| Certification| CE indicates a registered trademark of EU.
| Certification| UKCA indicates a registered trademark of UK.
| Certification| Conforms to UL STD 61010-1, 61010-2-030, Certified to CSA STD
C22.2 No. 61010-1, 61010-2-030.
| Waste| Do not place equipment and its accessories in the trash. Items must
be properly disposed of in accordance with local regulations.
|
EFUP
| This environment-friendly use period (EFUP) mark indicates that dangerous or
toxic substances will not leak or cause damage within this indicated time
period. The environment-friendly use period of this product is 40 years,
during which it can be used safely. Upon expiration of this period, it should
enter therecycling system.
---|---|---
Safety Requirements
Warning
Preparation before use| Please connect this device to AC power supply with the
power cable provided; The AC input voltage of the line reaches the rated value
of this device. See the product manual for specific rated value.The line
voltage switch of this device matches the line voltage; The line voltage of
the line fuse of this device is correct;It not used for measuring the main
circuit.
Check all terminal rated values| Please check all rated values and marking
instructions on the product to avoid fire and impact of excessive current.
Please consult the product manual fordetailed rated values before connection.
Use the power cord properly| You can only use the special power cord for the
instrument approved by the local and state standards. Please check whether the
insulation layer of the cord is damaged or the cord is exposed, and test
whether the cord is conductive. If the cord is damaged, please replace it
before using theinstrument.
Instrument Grounding| To avoid electric shock, the grounding conductor must be
connected to the ground. This product is grounded through the grounding
conductor of thepower supply. Please be sure to ground this product before it
is powered on.
AC power supply| Please use the AC power supply specified for this device.
Please use the power cord approved by your country and confirm that the
insulation layer isnot damaged.
Electrostatic prevention| This device may be damaged by static electricity, so
it should be tested in the anti-static area if possible. Before the power
cable is connected to this device, the internal and external conductors should
be grounded briefly to release static electricity. The protection grade of
this device is 4 kV forcontact discharge and 8 kV for air discharge.
Measurement accessories| Measurement accessories are of lower class, which are
definitely not applicable to main power supply measurement, CAT II, CAT III or
CAT IV circuit measurement .Probe subassemblies and accessories within the
range of IEC 61010-031 and current sensor within the range of IEC 61010-2-032
can meetits requirements.
Use the input / output port of this device properly| Please use the input /
output ports provided by this device in a properly manner. Do not load any
input signal at the output port of this device. Do not load any signal that
does not reach the rated value at the input port of this device. The probe or
other connection accessories should be effectively grounded to avoid product
damage or abnormal function. Please refer to theproduct manual for the rated
value of the input / output port of this device.
Power fuse| Please use power fuse of specified specification. If the fuse
needs to bereplaced, it must be replaced by maintenance personnel authorized
by UNI-T to replace the fuse that meets the specified specifications.
---|---
Disassembly and cleaning| There are no components available to operators
inside. Do not remove the protective cover.
Maintenance must be carried out by qualified personnel.
Service environment| This device should be used indoors in a clean and dry
environment with ambient temperature from 0 ℃ to 40 ℃.Do not use this device
in explosive, dusty or humid air.
Do not operate in humidenvironment| Do not use this device in a humid
environment to avoid the risk of internalshort circuit or electric shock.
Do not operate inflammable and explosive environment| Do not use this device
in a flammable and explosive environment to avoid product damage or personal
injury.
Caution
Abnormality| If this device may be faulty, please contact the authorized
maintenance personnel of UNI-T for testing. Any maintenance, adjustment or
partsreplacement must be done by the relevant personnel of UNI-T.
Cooling| Do not block the ventilation holes at the side and back of this
device;Do not allow any external objects to enter this device via ventilation
holes; Please ensure adequate ventilation, and leave a gap of at least 15 cm
on both sides, front and back of this device.
Safe transportation| Please transport this device safely to prevent it from
sliding, which maydamage the buttons, knobs or interfaces on the instrument
panel.
Proper ventilation| Poor ventilation will cause the device temperature to
rise, thus causing damage to this device. Please keep proper ventilation
during use, andregularly check the vents and fans.
Keep clean and dry| Please take actions to avoid dust or moisture in the air
affecting theperformance of this device. Please keep the product surface clean
and dry.
Note
Calibration| The recommended calibration period is one year. Calibration
should only becarried out by qualified personnel.
Introduction of MSO/UPO3000E Series
MSO/UPO3000E series digital phosphor oscilloscope includes ten models
Model | Analog Channel | Analog bandwidth | LA | AWG |
---|---|---|---|---|
UPO3354E | 4 | 350MHz | ○ | × |
UPO3352E | 2 | 350MHz | ○ | × |
UPO3504E | 4 | 500MHz | ○ | × |
UPO3502E | 2 | 500MHz | ○ | × |
MSO3354E | 4 | 350MHz | ● | × |
MSO3352E | 2 | 350MHz | ● | × |
MSO3504E | 4 | 500MHz | ● | × |
MSO3502E | 2 | 500MHz | ● | × |
MSO3354E-S | 4 | 350MHz | ● | ● |
MSO3504E-S | 4 | 500MHz | ● | ● |
●:Standard ○: Optional ×: Not support
Getting Started Guide
This chapter introduces on using the oscilloscope for the first time, the front and rear panels, the user interface, as well as the built-in help system.
General Inspection
It is recommended to inspect the instrument follow the steps below before using the MSO/UPO3000E series for the first time.
-
Check for Damages caused by Transport
If the packaging carton or the foam plastic cushions are severely damaged, please contact the UNI-T distributor of this product immediately. -
Check Attachment
Please check appendix for the list of accessories. If any of the accessories are missing or damaged, please contact UNI-T or local distributors of this product. -
Machine Inspection
If the instrument appears to be damaged, not working properly, or has failed the functionality test, please contact UNI-T or local distributors of this product.
If the equipment is damaged due to shipping, please keep the packaging and notify both the transportation department and UNI-T distributors, UNI-T will arrange maintenance or replacement.
Before Use
To perform a quick verification of the instrument’s normal operations, please follow the steps below:
-
Connect to the Power Supply
The power supply voltage range is from 100 VAC to 240 VAC, the frequency range is 50 Hz to 60 Hz. Connect the oscilloscope to the power supply line that came with the oscilloscope or any power supply line that meets the local country standards. If the power switch on the rear panel is not turned on, the indicator of power soft button on the left corner on the front panel will be extinguished and no function. If the power switch on the rear panel is turned on, the indicator of power soft button on the left corner on the front panel will be on red, press this button to boot-up the instrument. -
Boot-up Check
Press the soft power button and the light should change to green. The oscilloscope will show a boot animation, and then enter the normal interface. -
Connect Probe
Use probe’s BNC port connect to the channel 1 BNC port of the oscilloscope. Connect the probe to the “Compensating signal port” and the ground clip is connected to the “Ground terminal” below the compensating signal port as shown in Figure 1. The output of the compensating signal should be amplitude 3 Vpp, default frequency is 1 kHz
Figure 1 Compensating signal and Ground terminal
-
Function Check
Press the AUTO key, a 3 Vpp 1 kHz square wave should appear. Repeat step 3 to check all channels. -
Probe Compensation
When the probe is connected to any input channel for the first time, this step might be adjusted in order to match the probe and the input channel. Probes that are not compensated may lead to measurement errors or mistake. If need to adjust the probe compensation, please follow the following steps:
-
Set the attenuation coefficient in the probe menu and the switch on the probe 10x, and connect the probe to CH1. Make sure the probe’s connector is properly connected with the oscilloscope. Connect the probe’s main clip and ground clip to the oscilloscope’s compensating signal and ground terminal respectively. Open CH1 and press the AUTO button.
-
View the displayed waveform, as shown in Figure 2
Excessive Compensation Correct Compensation Insufficient Compensation
Figure 2 Compensating Calibration of Probe
-
If the displayed waveform is look like the above “Insufficient Compensation” or “Excessive Compensation”, use a non-metallic screwdriver to adjust the probe’s variable capacitance until the display matches the “Correct compensation” waveform.
Warning: To avoid electric shock when using the probe to measure high voltage, please ensure that the probe insulation is in good condition and avoid physical contact with any metallic part of the probe.
Front Panel
Figure 3 Front Panel
Table 1 Front Panel
No. | Description | No. | Description |
---|---|---|---|
1 | Screen display area | 12 | Factory setting, LA, AWG, protocol decoding |
2 | Multipurpose knob | 13 | Trigger control (TRIGGER) |
3 | Recording waveform | 14 | Horizontal control (HORIZONTAL) |
4 | Jog dial knob | 15 | Analog channel input port |
5 | Function menu | 16 | Vertical control (VERTICAL) |
6 | Numeric keypad | 17 | Menu control soft key |
7 | Automatic control key | 18 | USB HOST interface |
8 | Run/Stop control key | 19 | Power supply soft key |
9 | Single trigger control key | 20 | Digital channel input port |
10 | Clear all control key | 21 | HOME menu |
11 | Compensating signal connector and ground terminal |
Rear Panel
Figure 4 Rear Panel
Table 2 Rear Panel
No. | Description | No. | Description |
---|---|---|---|
1 | EXT Trig port | 5 | LAN |
2 | AUX Out and arbitrary waveform signal output | 6 | AC Power Input Socket |
3 | VIDEO Out | 7 | Power Switch |
4 | USB Device port | 8 | Safety Lock |
User Interface
Figure 5 User Interface
Table 3 User Interface
No. | Description | No. | Description |
---|---|---|---|
1 | Trigger status identification | 9 | System time |
2 | Time base scale | 10 | USB DEVICE icon |
3 | Sampling rate/storage depth | 11 | CH4 state icon |
4 | Waveform indicator | 12 | CH3 state icon |
5 | Waveform trigger position | 13 | CH2 state icon |
6 | Horizontal displacement | 14 | CH1 state icon |
7 | Trigger status | 15 | HOME menu |
8 | Operation menu | 16 | Analog channel icon and waveform |
Touch Screen
MSO/UPO3000E series provides 8 inch super capacitive touch screen, multiple
point touch control and
gesture control. MSO/UPO3000E has easily operating system with flexible and
high sensitive touch screen
features for great waveform display and excellent user experience. Touch
control function includes tap,
squeeze, drag and drawing rectangle.
Hint : The menu displayed on the screen of the oscilloscope can all use the touch control function.
- Tap
- Squeeze
- Drag
- Drawing Rectangle
- Touch Shortcut Operating
Tap
Use one finger to slightly tap icon or word on the screen as shown in Figure 7. Tap can use for:
- Tap menu displayed on the screen and then to setup.
- Tap the function guide icon on the left corner of the screen to enable it.
- Tap to pop out numeric keypad to set parameter.
- Tap virtual keyboard to set label and file name.
- Tap message to pop out close button on the right corner to close it.
- Tap other window displayed on the screen to setup
Figure 7 Touch Gesture
Squeeze
Squeeze two fingers together or separate. Squeeze gesture can zoom out or zoom in the waveform. If the waveform need to zoom out, squeeze two finger together and then slide away; If the waveform need to zoom out, separate two fingers and then squeeze two fingers together as shown in Figure 8. Squeeze gesture can use for:
- Adjusting horizontal time base of waveform on the horizontal direction.
- Adjusting vertical scale of waveform on the vertical direction.
Figure 8 Squeeze Gesture
Drag
Use one finger to press and drag the selected item to the aimed position as
shown in Figure 9. Drag gesture
can use for:
- Drag waveform to change waveform displacement or offset.
- Drag window control to change window position (such as numeric keyboard).
- Drag cursor to change cursor position.
Figure 9 Drag Gesture
\
Drawing Rectangle
Open function guide and tap drawing rectangle icon to switch drawing rectangle mode. Drag finger on the screen to drawing rectangle as shown in Figure 10(a), 10(b). Move finger away from the screen, menu displayed on the screen, tap to select “Area A enable”, “Area B enable”, “Intersection”, “Non intersecting”, “Source”. Drag finger on the screen form bottom right up to top left to drawing the trigger area.
Figure 10 Drawing Rectangle Gesture
Select “Area A”:
- Drawing the area of trigger area A;
- Open trigger area A;
- Open “trigger area” menu.
Select “Area B”:
- Drawing the area of trigger area B;
- Open trigger area B;
- Open “trigger area” menu.
Hints: Press “drawing rectangle” icon to switch drawing rectangle and operating waveform mode. Press “drawing rectangle” icon, if icon displays , which presents the drawing rectangle mode is opened; press “drawing rectangle” **** icon, if icon displays , which presents the operating waveform mode is opened. Operating waveform mode of the oscilloscope is enabled by default.
Touch Shortcut Operating
Separate two fingers and meanwhile slide two fingers to one direction, this gesture can quickly turn on/off time measurement and voltage measurement cursor.
a. Time measurement cursor
- Slide down in vertical to directly turn on time measurement cursor as shown in Figure 11 (a);
- Slide up in vertical to directly turn off time measurement cursor;
b. Voltage measurement cursor
- Slide right in horizontal to directly turn on voltage measurement cursor as shown in Figure 11 (b);
- Slide left in horizontal to directly turn off voltage measurement cursor;
Note: Turn on/off cursor requires that slide distance should be >3div.
c. Quick Guide
When the oscilloscope is stop, touch and drag the T identifier to left or
right around quickly, or rotate the horizontal Position knob, a quick guide
icon will displayed at the top of the screen, as shown in Figure 12. The quick
guide icon can be closed when the oscilloscope is operating. In quick guide,
the oscilloscope will automatically adjust the delay on horizontal direction
at a constant speed according to the user-defined, it
is convenient for user to observe the waveform.
Press the guide key on the front panel, it can adjust to play forward, stop or
backward. When in guiding, press Multipurpose knob can also stop the guide.
Operation Menu
Press any soft key to activate the corresponding menu; the following icon may be displayed on the menu.
This symbol represents that multipurpose knob on the front panel can be used
to select and adjust parameter.
This symbol represents that there are several options.
This symbol represents that there has next menu.
This symbol represents that the menu has two options.
This symbol represents that multipurpose knob and jog dial on the front panel
can This symbol represents that numeric keypad on the front panel or touch to
pop out numeric keyboard to input content.
This symbol represents that the current page can be flipped up or down.
Circle numbers represents the total pages of menu. Single page has no circle
display, two pages or more have circle display. Use key to turn page.
Remote Control
MSO/UPO3000E series digital phosphor oscilloscope communicates with the computer through USB interface and LAN interface to achieve remote control. Remote control is implemented based on the SCPI (Standard Commands for Programmable Instruments) command set.
MSO/UPO3000E series digital phosphor oscilloscope supports three remote control modes:
a. User-defined programming
b. Use PC software (instrument manager)
c. Web Control
Troubleshooting
-
When press the button, the oscilloscope is blank screen.
a. Check whether the power is correctly connected, power supply is normal or not.
b. Check whether the power switch is really turned on. The soft power switch button should light in green when the instrument is normal boot. And it should have relay sound.
c. If there is a relay sound, it indicates that the oscilloscope starts normally. Try the following steps: press the DEFAULT key, then press F1, if device returns to normal, it means backlight brightness is set too low.
d. After completing the above steps, restart the oscilloscope.
e. If the instrument still cannot work, please contact UNI-T. -
After signal acquisition, signal of waveform does not appear on the screen.
a. Check whether BNC port is connected properly.
b. Check whether the signal output channel has turned on.
c. Check whether the input signal channel has turned on.
d. Check whether has DC displacement in signal of information source.
e. Plug out the input signal, check whether the baseline is in the screen range (if not, please perform auto-calibration.)
f. If the instrument still cannot work, please contact UNI-T. -
The measured voltage amplitude value is 10 times larger or smaller than the actual value: Check the channel probe attenuation coefficient setting is consistent with the used probe attenuation rate.
-
There is a waveform display but not stable
a. Check the trigger source in the trigger menu and confirm that it matches the input channel of the actual signal.
b. Check the trigger type: general signal should use “Edge” trigger. The waveform can display stably only the proper trigger type is used.
c. Change the trigger coupling setting to high-frequency or low-frequency reject, in order to filter out the high-frequency or low-frequency noise that interferes with the trigger. -
No display after pressing Run/Stop button
a. Check whether the mode at the trigger panel (TRIGGER) is on “Normal” or “Single” and whether the trigger level exceeds the waveform range.
b. If it exceeds,set the trigger level to the middle or set the trigger mode to Auto.
c. Press AUTO button could automatically finish the above setting. -
Waveform refresh is very slow:
a. Check whether the acquisition method is average and the average times are large.
b. Check whether the storage depth is the maximum.
c. Check whether the trigger release time is large.
d. Check whether the trigger is normal and the current time base is slow.
e. All the preceding causes slow refreshing of waveforms. Users can restore the factory settings so that the waveforms can be refreshed normally.
Appendix Contact Us
If the use of this product has caused any inconvenience, if you in mainland China, you can contact UNI-T company directly.
Service support: 8am to 5.30pm (UTC+8), Monday to Friday or via email. Our email address is infosh@unitrend.com.cn
For product support outside mainland China, please contact your local UNI-T distributor or sales center.
Many UNI-T products have the option of extending the warranty and calibration period, please contact your local UNI-T dealer or sales center.
To obtain the address list of our service centers, please visit our official website http://www.unitrend.com
References
- UNI-T Voltage Meter, Multimeter, Oscilloscope | UNI-T-UNI-T Voltage Meter, Multimeter, Oscilloscope | UNI-T
- Trend Micro (US) | Industry-Leading Cyber Security Platform
- Default page
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>