Quality Aspirators 0483 Osseous Coagulum Trap Instruction Manual

June 1, 2024
Quality Aspirators

Quality Aspirators 0483 Osseous Coagulum Trap

  1. Any aspirator which fits into a standard high volume cut-off valve (11mm opening) will fit into the cap of the O.C.T.
  2. The O.C.T. will fit into any standard High Volume Cut-Off valve (11mm opening).Quality Aspirators 0483 Osseous Coagulum Trap 1
  3. Any aspirator which fits into a standard high volume cut-off valve (11mm opening) will fit into the cap of the O.C.T.
  4. The O.C.T. will fit into 1/4″ I.D. surgical tubing. Placing a small amount of lubricant inside the opening of the tubing before connecting to the O.C.T. will make this process much easier.

Quality Aspirators 0483 Osseous Coagulum Trap 2

Disassembled View

Quality Aspirators 0483 Osseous Coagulum Trap 3

Osseous Coagulum Trap Procedures and Sterilization Instructions

The Osseous Coagulum Trap offers an improved method of collecting osseous material. The O.C.T. is also excellent for retrieving root tips and precious metals. Osseous material is collected directly from the aspirator tip into a sterile filter.
The O.C.T. connects into a high volume cut-off valve or 1/4″ I.D. surgical tubing. Any surgical aspirator which fits into a high volume cut-off valve will fit into the O.C.T. cap.

PROCEDURE : When you reach the point in your procedure where you are ready to collect the osseous material, simply place the O.C.T. between the aspirator and cut-off valve or surgical tubing. As soon as you have collected the osseous material remove the O.C.T. trap. Failure to do so will cause the filter to become clogged or will dry out the osseous material because of the constant air flow through the trap. After harvesting the osseous material, dispose of the filter or keep the filter moist for cleaning and sterilization.

CLEANING

  • Filter basket should be kept moist to prevent material from drying to the mesh.
  • Rinse filter basket thoroughly and place in ultrasonic cleaner.(CAUTION: Do NOT place the metal cap or body in the utlrasonic)
  • Visually inspect filter mesh for tears or defects.

STERILIZATION

  • O.C.T. may be autoclaved, or chemiclaved.
  • Always sterilize filters and trap before use.

PRECAUTIONS
DO NOT put the metal O.C.T. cap or body in an ultrasonic cleaner. This will cause discoloration of the O.C.T.