equate UB546L Wrist Blood Pressure Monitor Instruction Manual

June 1, 2024
equate

equate LOGO Wrist Blood Pressure Monitor
6500 Series – Wireless
Instruction Manual
Original 1WMPD4004434C

Introduction

The Equate wrist blood pressure monitor is designed for ease of use and accuracy. This device will facilitate your daily blood pressure regimen.

  • We recommend that you read through this manual carefully before using the device for the first time.
  • This device is designed for use on adults, and is not intended for infants and children.
  • This device is designed for use to operate by yourself in the home healthcare environment to measure blood pressure and pulse rate of people for diagnosis.

Precautions

FCC Compliance Information
This device complies with Part 15 of FCC rules. Operation is subject to the following two conditions:

  1. This device may not cause harmful interference, and
  2. This device must accept any interference received, including interference that may cause undesired operations.
  • Precision components are used in the construction of this device. Extremes in temperature, humidity, direct sunlight, shock or dust should be avoided.
  • Clean the device and cuff with a dry, soft cloth or a cloth dampened with water and a neutral detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean the device or cuff.
  • Avoid tightly folding the cuff for long periods, as such treatment may shorten the life of the components.
  • The device and cuff are not water resistant. Prevent rain, sweat and water from soiling the device and cuff.
  • Measurements may be distorted if the device is used close to televisions, microwave ovens, cellular telephones, X-ray or other devices with strong electrical fields.
  • Wireless communication devices, such as home networking devices, mobile phones, cordless phones and their base stations, and walkie-talkies can affect this blood pressure monitor. Therefore, a minimum distance of 12 in/30 cm should be kept from such devices.
  • When reusing the device, confirm that the device is clean.
  • Used equipment, parts and batteries are not treated as ordinary household waste, and must be disposed of according to the applicable local regulations.
  • Do not modify the device. It may cause accidents or damage to the device.
  • To measure blood pressure, the arm must be squeezed by the cuff hard enough to temporarily stop blood flow through the artery. This may cause pain, numbness or a temporary red mark to the arm. This condition will appear especially when measurement is repeated successively. Any pain, numbness, or red marks will disappear with time.
  • Measuring blood pressure too frequently may cause harm due to blood flow interference.
  • Check that the operation of the device does not result in prolonged impairment of blood circulation, when using the device repeatedly.
  • Clinical testing has not been conducted on newborn infants and pregnant woman. Do not use on newborn infants or pregnant woman.
  • If you have had a mastectomy, please consult a doctor before using the device.
  • Do not let children use the device by themselves and do not leave the device within the reach of children, it may cause accidents or injury.
  • There are small parts that may cause a choking hazard if swallowed by mistake by infants.
  • Do not touch the batteries and the patient at the same time. That may result in electrical shock.
  • Use of accessories not detailed in this manual may compromise safety.
  • Should the battery short-circuit, it may become hot and potentially cause burns.
  • Allow the device to adapt to the surrounding environment before use (about one hour).
  • Do not inflate without wrapping the cuff around your wrist.
  • Do not apply the cuff on an arm in which another medical device is attached. The equipment may not function properly.
  • People who have a severe circulatory deficit in the arm must consult a doctor before using the device, to avoid medical problems.
  • Do not self-diagnose the measurement results and start treatment by yourself. Always consult your doctor for evaluation of the results and treatment.
  • Do not apply the cuff on a wrist with an unhealed wound.
  • Do not apply the cuff on a wrist receiving an intravenous drip or blood transfusion. It may cause injury.
  • Do not use the device where flammable gases such as anesthetic gases are present. It may cause an explosion.
  • Do not use the device in highly concentrated oxygen environments, such as a high-pressure oxygen chamber or an oxygen tent. It may cause a fire or explosion.
  • In the case of single components failure enclosure of near cuff may become hot and potentially cause malfunction (Max.44°C).

Parts Identification

equate UB546L Wrist Blood Pressure Monitor - Parts
Identification

Display

equate UB546L Wrist Blood Pressure Monitor - Display

Symbols

Symbols that appear on the display

Symbols Function/Meaning Recommended Action
Appears while measurement is in progress. It blinks when the pulse is
detected. Measurement is in progress.

Remain as still as possible.
| I.H.B. symbol appears when an irregular heartbeat is detected. It may appear when a very slight vibration like shivering or shaking is detected.| ––––––––––––
| Previous measurements stored in memory.| ––––––––––––
| Average data| ––––––––––––
| FULL BATTERY
The battery power indicator during measurement.| ––––––––––––
| LOW BATTERY
The battery power is low when it blinks.| Replace all batteries with new ones when the symbol blinks.
| User1 and User2| ––––––––––––
SYS| Systolic blood pressure in mmHg| ––––––––––––
DIA| Diastolic blood pressure in mmHg| ––––––––––––
PUL/min| Pulse per minute| ––––––––––––
AM| Time in the morning| ––––––––––––
PM| Time in the afternoon| ––––––––––––
| The device is connecting to the Bluetooth ® devices.| ––––––––––––
| Device internal error| Remove the Batteries and press the START/ STOP button, and then install the batteries again. If the error still appears, contact customer service.
| Unstable blood pressure due to movement during measurement.| Take another measurement. Remain still during measurement.
The systolic and diastolic values are within 10 mmHg of each other.| Apply the cuff correctly, and take another measurement.
| The pressure value did not increase during the inflation.
The cuff is not applied correctly.
| PUL DISPLAY ERROR The pulse is not detected correctly
| Pairing has not been performed correctly.| Remove and reinstall the batteries. Try pairing again.
| Pairing in progress.| ––––––––––––

Symbols printed on the device case

Symbols Function/Meaning
Standby and Turn the device on
Battery installation guide
**** Direct current
Type BF: Cuff is designed to provide spe- cial protection against

electrical shocks
SN| Serial number
| Refer to instruction manual/booklet
IP| International protection symbol
| Clock setting
| Clock adjustment and memory recall
| User selection
BT| Bluetooth® address
****| Used equipment, parts and batteries are not treated as ordinary household waste, and must be disposed of according to the applicable local regurations.

Using the Monitor

5.1. Installing / Changing The Batteries

  1. Remove the battery cover.
  2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that the polarities (+ and -) are correct. Use only LR03 or AAA batteries.
  3. Attach the battery cover.

CAUTIONS

  • Insert the batteries as shown in the battery compartment. If installed incorrectly, the device will not work.
  • When (LOW BATTERY mark) appears on the display, replace all batteries with new ones. Do not mix old and new batteries. It may shorten the battery life, or cause the device to malfunction.
  • (LOW BATTERY mark) does not appear when the batteries are drained.
  • The battery life varies with the ambient temperature and may be shorter at low temperatures.
  • Generally, wo new LR03 batteries will last approximately for three months when used twice for measurement each day.
  • Use the specified batteries only. The batteries provided with the device are for testing monitor performance and may have a limited life.
  • Remove the batteries if the device is not to be used for a long period of time. The batteries may leak and cause a malfunction.

5.2. Applying the Cuff

  1. Wrap the cuff around your wrist about 1 cm below your hand as shown in the figure below.
  2. Fasten the cuff tightly using the Velcro strip.

Note: For proper measurements, fasten the cuff tightly and measure on a bare wrist.
5.3. Measurements
During the measurement, it is normal for the cuff to feel very tight.
5.4. After Measurement

  • While readings are displayed, if you press the START button to turn the device off, new readings are stored in memory.
  • Remove the cuff and record your data.

Notes: The device is provided with an automatic power shut-off function which stores the current data in the memory and turns the device off automatically one minute after measurement. Allow at least three minutes between measurements on the same person.

Pairing with a Mobile Device

CAUTION

  • In the unlikely event that this monitor causes radio wave interference to a different wireless station, change the location where this monitor is used or stop use immediately.
  • The communication distance between this monitor and the mobile device is about 10 m. This distance is reduced by the conditions in the surrounding environment, so be sure to check that the distance is short enough for a connection to be made after measurement is complete.
  • Be sure to use in a location where visibility between the two devices that you want to connect is good. The connection distance is reduced by the structure of buildings or other obstructions. In particular, connection may be impossible when devices are used on either side of reinforced concrete.
  • Do not use Bluetooth® connection in the range of a wireless LAN or other wireless devices, near devices that emit radio waves such as microwaves, in locations where there are many obstructions, or in other locations where signal strength is weak. Doing so may result in frequent loss of connection, very slow communication speeds and errors.
  • In this case, switch off the power supply to the device that is not being used or use the monitor in a different location.
  • If the monitor does not connect normally when used near a wireless station or broadcast station, use the monitor in a different location.
  • Walmart cannot accept liability for any damages incurred due to impaired operation or data loss, etc that occur through the use of this product.
  • This product is not guaranteed to connect to all Bluetooth® compatible devices.

6.1 Bluetooth® Transmission

Bluetooth® devices carry the Bluetooth® logo mark.
To connect with your mobile device – download and install Equate Heart Chart app.

Follow the instructions in the app to connect.
6.2 Cautions for Pairing

  • Only one device can be paired with this monitor at one time. If the mobile device cannot receive measurement data, try pairing again.
  • In case a 5th mobile device is registered, the monitor will delete the oldest mobile device.

6.3 Pairing Procedure

  1. Turn on Bluetooth® settings on your mobile device
  2. Press and hold the START button until “Pr” is displayed, and then release the button. The monitor will be in a state that can be found by the mobile device for about one minute.
  3. If “E 10” is displayed or pairing is failed, remove the batteries and try steps 1-3 again.
  4. Accept the pairing request on the Equate Heart Chart app.

6.4 Transmitting Temporarily Stored Data
In cases when the mobile device cannot receive measurement data, the measurement data is temporarily stored in the monitor memory. The data stored in the memory is transmitted the next time a connection is successfully made to the mobile device. A total of 90 sets of measurement data can be stored per user. When the amount of data exceeds 90, the oldest data is deleted and the new data is stored.
6.5 Time
This monitor has a built-in clock. The date and time that a measurement was taken is included in the measurement data. The built-in clock is automatically adjusted by syncing the clock of a moblie device. Sync of the time is done in the timing Bluetooth® symbol lights up, in the pairing process.

Measurements

7.1. Normal Measurement

  1. Wrap the cuff around your wrist. Sit comfortably with the cuff at the same level as your heart and relax.

  2. Press the SET button. Select a user from user and user .

  3. Adjust and keep the height of the cuff to the same level as your heart using the C.P.G. indicator.

  4. Press the START button. All of the display segments are displayed. Zero is displayed blinking briefly. Then the display changes, as indicated in the figure below, as the measurement begins. The cuff starts to inflate. It is normal for the cuff to feel very tight. The measurement starts automatically when inflation starts, and the (heart mark) blinks.
    Note: If you wish to stop inflation at any time, press the START button again.
    Note: If an appropriate pressure is not obtained, the device starts to inflate again automatically.

  5. When the measurement is complete, the systolic and diastolic pressure readings and pulse rate are displayed. The cuff exhausts the remaining air and deflates completely. In case the monitor connected to a mobile device, it’s possible to transfer the measurement data to the app, when the measurement is done.

  6. Press the START button again to turn off the power and store the measurement data.

Note: This device has an automatic power shut-off function.
Allow at least 3 minutes between measurements on the same person.

equate UB546L Wrist Blood Pressure Monitor - shut off
function

The C.P.G indicator:

equate UB546L Wrist Blood Pressure Monitor - CPG
indicator

7.2. Notes for Accurate Measurement

  • Sit comfortably on a chair with your feet on the floor and your back straight. Do not cross your legs.
  • Place your arm on a table with your palm facing upward and the cuff at the same level as your heart.
  • Relax for about five to ten minutes before measurement.
  • Remain still and keep quiet during measurement.
  • Do not smoke, exercise, or consume anything for at least 30 minutes beforehand.
  • This device bases its measurements on the heartbeat. If you have a very weak or irregular heartbeat, the device may have difficulty determining your blood pressure.
  • Should the device detect a condition that is abnormal, it will stop the measurement and display an error symbol. Refer to the section “4 Symbols” for the description of the symbols.
  • Try to measure your blood pressure at the same time every day.
  • The automatic blood pressure monitor’s performance may be affected by excessive temperature or humidity, or altitude.

The C.P.G.Indicator

8.1 The C.P.G. Indicator
The C.P.G. (Correct Position Guidance) indicator is the function to inform a difference between the height (wrist angle) of the blood pressure monitor and your cardiac height in the correct posture (Example: sitting posture, height of table and chair, etc.) during the measurement. The indicator can be used to get more stable measurement condition.

equate UB546L Wrist Blood Pressure Monitor - CPG indicator
2

The position of the device is checked before measurement. If the check shows a correct measurement position, the C.P.G indicator is lit Correct LEVEL(Blue). For Low or High measurement position, the indicator will be lit Low or High LEVEL(Orange)

equate UB546L Wrist Blood Pressure Monitor - Correct
LEVEL

8.2 The Indicator During Measurement And Recalling Memory
The C.P.G. indicator can be displayed at the beginning of measurement and is included in data stored in the memory. Adjust and keep the angle during measurement.

Recalling the Memory Data

The device automatically stores up to 90 x 2 blood pressure and pulse measurements in memory. Data stored in memory are assigned a data number in the order of the newest to the oldest. The oldest data displays as “ “. The symbol in the upper left corner of the display indicates that you are viewing previous data stored in memory.
9.1. Recalling Data

  1. Press the button.

  2. Select the user you want to review the memory for by pressing the button.
    Select a user from user or user .

  3. Press the button.
    The average of the last three measurements is displayed. (If no data, “0” is displayed. Press the or START button to turn device off.)

  4. Each time the button is pressed, the memory data is displayed as follows. Most recent data (ex. nex. no. 35). Three seconds after the data number displays, the measurement data is displayed.

    Last data (ex. no.1).
    Three seconds after the data number displays, the measurement data is displayed.

  5. After the last data is displayed, press the  button to return the average display of the last three measurements.

  6. Press the START button to turn the device off.
    After one minute of nonoperation the device will turn off automatically.equate UB546L Wrist Blood Pressure Monitor - Recalling Data
4

  7. Each time the button is pressed, the user is changed and the average of the last three measurements for that user is displayed.

Note: If this device stores the two measurements, the average of the two measurements is displayed. If this device stores only one measurement, this is displayed as the average.

9.2. Clearing Data

  1. Press the button to move to the memory display mode
  2. Select the user you want to delete with user button.
  3. Turn off the power by pressing START button.
  4. Press and hold button.
  5. Memory icon blinks and the data is deleted.
  6. The device power will be turned off automatically.

Setting Date and Time

Set the date and time prior to use.

  1. Press the button, the year starts blinking.
  2. Select the year using the  button. Press the button to set the current year and move to month / day selection. The date can be set anywhere between the years 2021 and 2059.
  3. Select the month using the button. Press the button to set the current month and move to day selection.
  4. Select the day using the  button. Press the button to set the current day and move to hour / minute selection.
  5. Select the hour using the  button. Press the button to set the current hour and move to minute selection.
  6. Select the minute using the button. Press the button to turn the device off.

equate UB546L Wrist Blood Pressure Monitor - Setting Date and
TimePressing the START button will turn the device off anytime.
Note: After three minutes of non-operation, the device will turn off automatically.

When the clock has not been set, the clock display indicates dashes as shown to the right.

When using the device for the first time, the clock is not adjusted. When the device is disconnected from the power supply, the set date and time will be erased.
When the set date and time is erased, please adjust again.
The clock of the monitor can sync to the clock of the mobile device during Bluetooth® communication.

What is an Irregular Heartbeat

The blood pressure monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies from the average of all heartbeats during the blood pressure measurement. It is important that you are relaxed, remain still and do not talk during measurements.
Note: We recommend contacting your physician if you see this indicator frequently.

Pressure Bar Indicator

The indicator monitors the progress of pressure during measurement.

equate UB546L Wrist Blood Pressure Monitor - Bar
Indicator

About Blood Pressure

What is Blood Pressure?
Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg). One’s natural blood pressure is represented by the fundamental pressure, which is measured first thing in the morning while one is still at rest and before eating.

AHA Classification Indicator

Each segment of the bar indicator corresponds to the AHA blood pressure classification.
:The indicator displays a segment, based on the current data, corresponding to the AHA classification.
AHA Classification Indicator

Troubleshooting

Problem Possible Reason Recommended Action
Nothing appears on the display, even when the power is turned on. Batteries
are drained. Replace all batteries with new ones.
Battery terminals are not in the correct position. Reinstall the batteries

with negative and positive terminals matching those indicated on the battery compartment.
The cuff does not inflate.| Battery voltage is too low. blinks. If the batteries are drained completely, the symbol does not appear.| Replace all batteries with new ones.
The device does not measure.
Readings are too high or too low.| The cuff is not applied properly.| Apply the cuff correctly.
You moved your wrist or body during the measurement.| Make sure you remain very still and quiet during the measurement.
The cuff position is not correct.| Sit comfortably and still. Raise your hand so that the cuff is level at the same level as your heart.
––––––––––––| If you have a weak or irregular heart beat, the device may have difficulty in determining your blood pressure.
Other| The value is different from that measured at a clinic or doctor’s office.| At a clinic or doctor’s office, apprehension may cause an elevated reading. Home measurement reduces the effects of outside influences on blood pressure readings & supplements the doctor’s readings.
––––––––––––| Remove the batteries. Place them back properly and take another measurement.

Note: If the actions described above do not solve the problem, contact customer service. Do not attempt to open or repair this product, as any attempt to do so will make your warranty invalid.

Maintenance

Do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, contact customer service.

Warranty

Product Warranty Term
Monitor 5 year
Cuff 2 year

Limited Warranty:
This warranty warrants to the first purchaser (“You”) that the product You purchased (the “Product”) will be free from defects in material, workmanship and design for the applicable Warranty Term stated above from the date You purchased the Product under normal use. This Limited Warranty is personal to You and is not transferable. If the Product is defective, then You return the Product in accordance with the procedure set forth below. This warranty obligation is limited to the repair or replacement, at our option, of the defective Product that has been returned by You within the warranty period. Such repair or replacement will be at no charge to You. The repaired or replacement Product is warranted here-under for the longer of the remainder of the original warranty period or 90 days from the date of shipment of the repaired or replacement Product.
To obtain a warranty service, please contact us at 1-888-726-9966 for return address, shipping and handling fee, and other instructions for processing warranty. Please ensure you have satisfactory proof of the date of Your purchase and a description of the defect. Returns will not be accepted unless a Return Material Authorization (RMA) Number has been issued from our Customer Service Representative.
This Limited Warranty does not cover, and we will not be liable for (i) any shipment damage, (ii) any damage or defect due to misuse, abuse, failure to use reasonable care, failure to follow written instructions enclosed with the Product, accident, subjecting the Product to any voltage other than the specified voltage, improper environmental conditions, or modification, alteration or repair by anyone other than persons authorized, or (iii) expendable or consumable components.
THIS LIMITED WARRANTY IS THE ONLY WARRANTY PROVIDED; THERE ARE NO OTHER EXPRESS WARRANTIES.
The above remedy of repair or replacement is your only and exclusive remedy. IN NO EVENT SHALL WE BE LIABLE FOR ANY DAMAGES, INCLUDING, WITHOUT LIMITATION, LOST PROFITS, LOST INFORMATION OR REPLACEMENT COSTS, ARISING OUT OF YOUR USE OF OR INABILITY TO USE THE PRODUCT, INCLUDING, WITHOUT LIMITATION, ANY SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF WE HAVE BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.
Some states do not allow the exclusion of incidental or consequential damages, so that the above exclusions may not apply to you. This warranty gives you specific legal rights, and you may also have other rights that may vary from state to state.
No distributor, dealer or other party is authorized to make any warranty or to modify this warranty, or to assume any liability with respect to its products.

Satisfaction Guaranteed

Satisfaction guaranteed – Or we’ll replace it or give you your money back. For questions or comments or to report an undesired reaction or side effect, please call 1-888-287-1915.
Walmart Inc.
702 S.W. 8th St.
Bentonville, AR 72716
www.walmart.com
1-888-287-1915
Equate wordmark and logos are trademarks of Walmart Apollo LLC.
The Bluetooth word mark and logos are registered trademarks of Bluetooth SIG, Inc.

Technical Data

Type UB-6500BLEWM
Measurement method Oscillometric measurement
Measurement range Pressure:
Systolic pressure: 60 – 279 mmHg
Diastolic pressure: 40 – 200 mmHg
Pulse: 40 – 180 beats/minute
Measurement accuracy Pressure:
Pulse: ±5 %
Power Supply 2 x 1.5V alkaline batteries (LR03 or AAA)
Number of measurements Approx. 110 measurements, when AAA alkaline batteries

are used.
With pressure value 170 mmHg, room temperature 23 °C.
Classification| Internally powered ME equipment (by batteries) Continuous operation mode
Clinical test| According to ISO81060-2:2013
In the clinical validation study, K5 was used on 85 subjects for determination of diastolic blood pressure.
EMD| IEC 60601-1-2: 2014
Wireless Communication| Bluetooth:| Ver.5.1LE BLP
Frequency band:| 2402 MHz to 2480 MHz
Maximum RF output power:| < 10 dBm
Modulation:| GFSK
Supported Data:| Systolic Pressure, Diastolic Pressure, Pulse Rate
Communication distance:| About 10 m (This distance is reduced by the conditions in the surrounding environment)
Paired device:| 4 devices
Memory| 90 measurements per user
Operating conditions| +10 to +40 °C/15 to 85 %RH/800 to 1060 hPa
Transport/Storage conditions| – 20 to +60 °C/10 to 95 %RH/700 to 1060 hPa
Cuff Size| 5.3”- 8.5”(13.5 – 21.5cm)
Dimensions| Approx. 2.2”[W]×3.5”[H]×0.8”[D] (56[W]×88[H]×20[D]mm)
Weight| Approx. 3.9 oz (110 g), excluding the batteries
Ingress protection| Device: IP22
Applied part| Cuff: Type BF
Useful life| Device: 5 years (when used six times a day)

Note: Specifications are subject to change without prior notice. IP classification is the degree of protection provided by enclosures in accordance with IEC 60529. This device is protected against solid foreign objects of 12 mm diameter and greater such as a finger. Vertically dripping water shall have no harmful effect when the enclosure is tilted at an angle of 15° from its normal position.
FCC Caution
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
This transmitter must not be co-located or operated in conjunction with any other antenna or transmitter.
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency (RF) Exposure Guidelines in Supplement C to OET65. This equipment has very low levels of RF energy that it deemed to comply without maximum permissive exposure evaluation (MPE).
Note: This equipment has been tested and found to comply with the limits for Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. It this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver.
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

EMD Technical Data Battery-operated Blood Pressure Monitor
Medical Electrical Equipment needs special precautions regarding EMD and needs to be installed and put into service according to the EMD information provided in the following. Portable and mobile RF communication equipment (e.g. cell phones) can affect Medical Electrical Equipment. The use of accessories and cables other than those specified may result in increased emissions or decreased immunity of the unit.
Table 1 – EMISSION Limits

Phenomenon Compliance
Conducted and radiated RF EMISSION CISPR 11 Group 1, Class B

Table 2 – IMMUNITY TEST LEVELS : Enclosure Port

Phenomenon IMMUNITY TEST LEVELS
Electrostatic discharge IEC 61000-4-2 ±8 kV contact

±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF EM fields IEC 61000-4-3| 10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
Proximity fields from RF wireless communications equipment IEC 61000-4-3| See table 3
Rated power frequency magnetic fields IEC 61000-4-8| 30 A/m
50 Hz or 60 Hz

Table 3 – Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment

Test frequency (MHz)| Band (MHz)| Service| Modulation| Maximum power (W)| Distance (m)| IMMUNITY TEST LEVEL (V/m)
---|---|---|---|---|---|---
385| 380 – 390| TETRA 400| Pulse modulation 18 Hz| 1.8| 0.3| 27
450| 430 – 470| GMRS 460
FRS 460| FM
±5 kHz deviation
1 kHz sine| 2| 0.3| 28
710| 704 – 787| LTE Band 13, 17| Pulse modulation 217 Hz| 0.2| 0.3| 9
745
780
810| 800 – 960| GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5| Pulse modulation 18 Hz| 2| 0.3| 28
870
930
1720| 1700 – 1990| GSM 1800
CDMA 1900
GSM 1900 DECT
LTE Band 1, 3, 4, 25
UMTS| Pulse modulation 217 Hz| 2| 0.3| 28
1845
1970
2450| 2400 – 2570| Bluetooth ®
WLAN 802.11 b/g/n
RFID 2450
LTE Band 7| Pulse modulation 217 Hz| 2| 0.3| 28
5240| 5100 – 5800| WLAN 802.11 a/n| Pulse modulation 217 Hz| 0.2| 0.3| 9
5500
5785

equate LOGO

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals