To reach better result, please read this instruction before use.

Attention

Please read carefully attention before you use the product, and should stow for later query

Camera Blur

  1. Please use the sheath for every time of use. This will protect the camera lens, and avoid of foggy, and it can extend the lifetime of the device. There is only 50 pieces sheath in every package, please buy more from the original seller if you have used up of the sheath.
  2. Please do not use sharp, metal or liquid insert or touch signal port, in order to avoid short circuit and product damage.

Connection

  1. Please power off the machine before connection the cables.
  2. Please don’t let sharp thing, metal or liquid into or touch the cable connector, in order to circuit and product damage

Power

  1. Please use the power supply with the package.
  2. Please cut off the power supply if you don’t use it for a long time.
  3. Please don’t use the machine if the power supply damaged.
  4. please don’t compress line of power supply or signal connection line expose in gangway.
  5. Please don’t put the power supply line pass the walkway.
  6. Please don’t put the power supply in damp place

Working environment

  1. Please put the product in dry and ventilated place.
  2. Don’t put the product in super-cooling, super-hot, super-damp, direct-irradiation, dusty place.

Disassemble

  1. Please don’t disassemble any part of the product by yourself, or it will out of warranty.
Product features

Dear users,
Thanks for choosing our company’s product. The 15 inch monitor with oral camera system could freeze/capture images and save into U disk. It is good for doctors physician analysis, communication with Patients, so as patients have a better understanding of their condition, and improve work efficiency.
This unit support multimedia U disk player and video input, and wifi connect to computer (windows system only).
The monitor can take picturer for patient face with build-in face camera.
Face camera and Intraoral camera can make video with remote control.
(But cannot record sound).

Product Parts

1. Standard accessories

Wired unit:

  1. 15 inch LED monitor                        1 PC
  2. Oral camera Handpiece                      1 PC
  3. 8GB U disk (with software)                1 PC
  4. Oral camera sheath                             50 PCS
  5. Oral camera holder                             1 PC
  6. 12V power supply                               1 PC
  7. Remote control                                   1 PC
  8. Handpiece cable                                  1 PC
  9. Monitor adjusting handle                    1 PC
  10. Monitor arm for dental unit(Option)

Wireless unit(Option):

  1. Wireless launcher                                1 PC
  2. Wireless receiver                                 1 PC
  3. Charging base                                     1 PC
  4. Power adapter                                     1 PC
Functions

1. Monitor

Specifications:

Model

| MD-1500A
---|---
Size|

15 inch

Resolution

| 1024*768
Contrast|

500:1

Brightness

| 400(CD/M2)
Response time|

5ms

Display view:

  1. Upward angle of visibility
  2. Downward angle of visibility
  • We can adjust the brightness to brighter at “Menu”, When the monitor is in the downward angle of visibility.
  • We can adjust the brightness to darker at “Menu”, When the monitor is in the upward angle of visibility.

Front

  1. PC VGA input button

  2. Multimedia U disk player button

  3. Video input button

  4. Face camera button

  5. Oral camera button

  6. IR Receiver

  7. Power indicator

  8. Pause button(work under Multimedia U disk player mode)
    Capture (Work under face camera and oral camera mode)

  9. Mute button (work under Multimedia U disk player/PC/Video mode)
    Quad picture(work under oral camera mode)

  10. Volume down (work under Multimedia U disk player/PC/Video mode)
    Page down(work under oral camera mode)

  11. Volume up (work under Multimedia U disk player/PC/Video mode)
    Page up(work under oral camera mode)

  12. Power on/off

  13. Monitor adjusting handle

  14. Oral camera holder (the light will off when the camera put in the holder)

  15. Face camera (Adjustable focus, and Can be adjusted up and down 75 degrees

Back

a) Built-in 3W speaker

1. The plugs of the monitor

A: Pedal in                                F: Voice output
B: Multimedia U disk player     G: PC voice input
C: Video voice R-CH input      H: VGA input
D: Video voice L-CH input      I: DC12V power adapter input
E: Video input                          J: Oral camera input

2. U disk for saving pictures plug

 The U disk have to plug into the monitor when we saving the pictures.

3. Wireless Receiver plug(Don’t insert U disk)

2. Oral camera

Specifications:

Model

| MD-910
---|---

Sensor

| 1/4 ” Sony CCD

TV standard

| PAL

Pixels

| 2.0 mega pixels

Light

| 6 white LEDs

Focus distance

| 3~50mm

Angle of view

| 105°

Cable length

| 2m

Cable port

| 6 pins

Oral camera functions:

  1. 6 Pcs white LED
  2. Freeze/Release
  3. Delete
  4. Page down
  5. Page up
  6. Quad
  7. Freeze/Release
3. Wireless launcher/Receiver (option)

Specifications:

Battery

| 350mA rechargeable Lithium battery
---|---

Channel

| 4 Channels

Frequency

| 5.8GHz

Wireless distance

| Within 8m

Wireless launcher:

  1. The fourth channel indicator

  2. The second channel indicator

  3. Battery capacity display
    (1) Red light means lack of battery
    (2) Green light means saturation

  4. Power on/off

  5. Charge hole

  6. Channel switch

  7. The first channel indicator

  8. The third channel indicator

Charge

  1. Launcher
  2. Sender inserted into handpiece
  3. Handset
  4. Power input
  5. Power adapter (DC 5V)

Wireless receiver:

  1. Channel switch
  2. The first channel indicator
  3. The second channel indicator
  4. The third channel indicator
  5. The fourth channel indicator

 The channel of launcher and receiver have to be same

4. Remote control
  1. Power on/off

  2. Video recording (work on face camera and oral camera)

  3. Mute

  4. Oral camera

  5. Face camera

  6. Zoom picture(zoom 1x to 4x)

  7. Page up

  8. Delete picturer

  9. Page down

  10. Freeze/capture

  11. Menu up

  12. Full/quad picture

  13. Volume down
    Menu up left

  14. Enter

  15. Volume up
    Menu down

  16. Menu

  17. Menu down

  18. Return

  19. Multimedia U disk

  20. Video input

  21. PC VGA input

Installation instructions

1. Install the handle to the monitor

2. Plug the U disk to the U disk saving plug

3. Put the monitor arm to 4 screws of the monitor (See the details as arm installation instructions)

4. Plug in the power adapter, connect the oral camera to monitor (for wired unit)

  1. Power adapter
  2. Camera cable

Handpiece combination to split schemes

1.Combination
Camera cable “ ” marked direction is front, connecting with camera and align with the front of the camera cable, the camera inserted front hole.

2.Split
When splitting camera and camera cable, please hold the non slip down and off.

5. Plug in the power adapter, connect the wireless oral camera (for wireless unit)

  1. Plug the Launcher into the camera handpiece
  2. Plug the receiver into monitor
  3. Power adapter

The channel of launcher and receiver have to be same

Take picture of the face

Monitor                             Remote control

  1. Press the “ ” button of the monitor or remote control to face camera.
  2. Adjust the camera to suitable focal distance.
  3. Press the “ ” button of the monitor or “ ” of remote control to capture.
Power off the light of oral camera

  1. Put the oral camera into holder, The button toward the outside as the picture.
    The camera light will off.

  2. Put the oral camera 10 minutes. The camera light will off.

How to determine the U disk available

There is a USB icon at the top right of screen after plug in the U disk. It means the U disk was recognized and works

There is a word “Save okay” flash at the bottom left of screen, when the picture save into U disk.

The picture can’t save if the U disk don’t plug in.

Video recording

Video recording button

  1. Adjust the monitor to oral camera or face camera mode.

  2. Press the video recording button “ ” to start.

  3. There is a word “REC” at the top right of screen.
    It means is recording.

  4. Press the video recording button “ ” again to end of the recording

  5. There is a word “Save okay” flash at the bottom left of screen, when the video save into U disk.

  6. The video recording will save into U disk automatically.

(The sound can’t be recorded)

Playback

  1. Adjust the monitor to oral camera mode.
  2. Press the Page up or Page down button of the oral camera or the remote control button “ ” “ “, find out the file which with “REC” word at the top right of screen.
    Press the video recording button of the remote control to play.
How to zoom the picture

The picture can be zoom 1x to 4x before freeze or after capture

  1. Zoom button

1. Adjust the monitor to oral camera mode.

2. Press the zoom button, the icons as below will be flash at the bottom left of screen.
XI>—1: Magnify 1 time.
XI>>–2: Magnify 2 times.
XI>>>-3: Magnify 3times.
XI>>>>4: Magnify 4 times.
Back to x1: Back to the original picture.

How to play the multimedia file of the U disk

1. Please plug in the U disk with videos, music or texts to Multimedia U disk player plug.
2. Press the “ ” of the monitor or the “ ” of the remote control to multimedia u disk player mode.
3. The windows as below will be display.

4. Choose photo, music, movie or text with “ ” “ ” button of the remote control.
Choose “C “, press “OK” to enter.

5. Choose the file press “OK” to play.

The formats are supported

Formats

| Size| Number of plays per second
---|---|---
Dat, mpg. mpeg|

MPG,MPEG

| Mpeg 1.2| 1920×1080@30P| 20Mbit/sec
Ts, trp, tp|

MPEG2-TS H264-S

| MPEG2.H.264| 1920×1080@30P| 20Mbit/sec
Vob|

MPEG2-PS

| MPEG2| 1920×1080@30P| 20Mbit/sec
Mp4|

AVI,MP4

| MPEG4
Divx 3.11,Divx 4.12.
Divx 5.x,Divx 6.
Xvid 1.00.
Xvid 1.01,Xvid 1.02.
Xvid 1.03,Xvid
1.10-beta 1/2
H.264| 1920×1080@30P| 20Mbit/sec
Rm,rmvb|

RMVB,RM

| Rm code RV8.RV9.
Rv10| 1280×720@30P| 10Mbit/sec
Mkv|

MKV

| H.264,MPEG1.2.4| 1920×1080@30P| 20Mbit/sec
Divx,avi|

AVI(1.0,2.0)
DMF0.1.2

| MPEG2.MPEG4
Divx 3.11,Divx
4.Divx 5,Divx 6,Xvid 1.00
Xvid 1.01.Xvid 1.02.
Xvid 1.03,Xvid
1.10-beta1/2
H.264| 1920×1080@30P| 20Mbit/sec
Asf| ASF
(Revision 01.20.03)| VC1| 1920×1080@30P| 20Mbit/sec


How to change the language of the monitor

1. Press the button “ ” of remote control to enter the menu.

2. Press the “ ” “ ” button of the remote control to choose “ ” icon.

3.Choose the first “language” and press “OK” button to enter.
Press the “ ” “ ” button of the remote control to change the language.
Press menu button “ ” to confirm and back

How to connect Wifi to computer

1. Your computer must be used with wireless WIFI card or use a wireless- enabled laptops. If not please buy a wireless WIFI receiver. If you need the internet, you need buy one more Wifi receiver.

2. Connect well the monitor and oral camera, and power on.
Plug in the wifi receiver and confirm it install well.

3. Find the Wifi name “Dental”.

4. Click “Dental” and enter the password “12345678” to connect. And check it connect successfully.

How to install the software

1. Please plug in the u disk which come with the package, and find the software “Intraoral Camera System”, And double click it to install.

2. Click “Next”.

3. Choose the folder which install.

4. Click “Next”.

6. Click “Finish”.

How to use the software

1. A icon will be established on the desktop, double click the icon.

2. Open the software, a window will prompt to enter patients information

3. Click the “Add” after entered patients information.

4. Click “Enter” or double click the serial number to enter the patients management

5. Click the “Capture” button at the bottom left to capture or video recording

6. The pictures will be save to “DentalWIFI” folder of D disk.

How to rename the WIFI name

We need use difference WIFI name when we use several units intraoral camera. Please click the “WIFI setting” to rename the name and password.

  1. WIFI name
  2. WIFI password
Simple troubleshooting

Please try to solve problems as the below ways. Please contact the seller or factory if can’t solve the problems after do these ways.

Items

| Problems| Solutions
---|---|---

1

| The host no responding after power on|

  1. Check the power adapter socket connection ok or not.
  2. Check the power adapter damaged or not.
  3. Contact the seller or factory.

2

| Can’t save the pictures|

  1. Check the U disk plug into U disk storage jack.
  2. Format the U disk, and check again.
  3. Contact the seller or factory.

3

| No image or blue screen of the monitor or computer|

  1. Check the cable connected ok between monitor and oral camera.
  2. Check the monitor under the right mode.
  3. Check the WIFI connect OK, and the software installed right.
  4. Restart power on.
  5. Please contact the seller or factory.

4

| The image is blur or 4 dirty on screen|

  1. Check the screen dirty or not.
  2. Check the lens dirty or not, and wipe it with cotton swab lightly.
  3. Please contact the seller or factory if can’t clean as above ways.
Monitor arm install (option)

Necessary parts:

Gasket
Bracket
Installed plate
Handpiece Holder
No.1,No.2,No.3,No.4,No5. bolts
Inter-hexagon spanner

Arm Structure
  1. Lock Nut
    (Adjust the angle or position, loosen this nut and then adjust).

  2. Can be adjusted angle up or down and also be rotated 10 degrees

  3. Monitor Installed plate
    (can rotate 10 degree)

  4. Endoscope Bracket

  5. Threading hole

  6. engine arm

Installation Instruction:

1. Install the bracket onto holder with the gasket, use No.1 bolts install bracket.
(The gasket is to avoid the slip of the bracket from the holder).

2. Connect the bracket and installed plate with No.2 bolts, the bolts could adjust tightness.
Use No.2 bolts install handpiece bar.(Note: Handpiece holder is adjustable according to the size of display.

3. Connect the plate to LCD use No.4 bolts and accessory, line go through the plate from rear covet. fix with No.3 bolts.

4. Install handpiece and docking station holder with No.5 bolts, connect the plug and turn on power.

         This product can bear the maximum weight is 4KG.