MEDI-FLO MDE050500 Elastomeric Pump User Guide
- June 1, 2024
- MEDI-FLO
Table of Contents
- MEDI-FLO MDE050500 Elastomeric Pump
- Product Diagram
- Choosing the Right MEDI-FLO® Elastomeric Pump for Your Patient
- MEDI-FLO® Short Duration Filling Guide
- MEDI-FLO® Long Duration Filling Guide
- MEDI-FLO® Chemotherapy Filling Guide
- Preparing the MEDI-FLO® Elastomeric Pump
- Guidelines for Infusion
- Troubleshooting
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
MEDI-FLO MDE050500 Elastomeric Pump
Why Wolf-Pak® MEDI-FLO® Elastomeric Pumps?
MEDI-FLO® Elastomeric Pumps are proven to be safe and effective and are a
convenient alternative to electronic infusion pumps. Ideally suited for home
infusion, long-term care and outpatient chemotherapy treatments. MEDI-FLO®
pumps give the patient mobility and freedom to maintain an active lifestyle.
How Does It Work?
The MEDI-FLO® Elastomeric Pump delivers medication using a specially designed
multi-layered balloon-like reservoir.
It exerts mechanical pressure thereby administering the pump contents through
an orifice tube at a predetermined
flow rate. The rate is controlled by a flow restrictor at the end of the
tubing and by flow restrictive tubing (MEDI-FLO® Chemotherapy Pumps only).
When used according to manufacturer’s recommendations and instructions for
use, flow accuracy is within +/- 15% of the nominal (label) flow rate (at 99%
confidence level) when filled at nominal volume.
Nominal Flow Rate
Product Diagram
Choosing the Right MEDI-FLO® Elastomeric Pump for Your Patient
The MEDI-FLO® Elastomeric Pump is easy to fill and color-coded for quick and accurate device identification. A wide variety of sizes and flow rates are available along with an extensive library of drug stability data. The range of SKUs available offers dosing flexibility to administer various infusion therapies. Please refer to the Fill Volume and Delivery Time Tables on pages 4-7, as well as the Drug Stability Guide to help determine which product is best suited for the needed therapy.
Drug Stability
Drug Stability data is available on a wide range of medications. The Drug
Stability Guidelines for the administration of medications using the MEDI-FLO®
Elastomeric Pump were developed as a result of testing performed by
independent laboratories, review of various medical publications and
manufacturers’ product information, and available elastomeric infusion pump
drug stability data. The stability data relates to chemical stability of the
drugs tested, not to sterility.
The pharmacist dispensing the drug is responsible for ensuring proper preparation using validated aseptic techniques to prevent microbiological contamination. For practice and quality standards, refer to USP 797 Pharmaceutical Compounding – Sterile Preparations and USP 800 Hazardous Drugs – Handling in a Healthcare Setting.
Contact your sales representative at 1-800-335-9653, or refer to our website, www.medi-flo.com, for the most up to date drug stability information.
Fill Volumes and Delivery Times
Refer to the tables on the following pages to determine the appropriate pump
model based on the fill volume and desired delivery time. Residual volume
information is also included.
The MEDI-FLO® nominal flow rates are based on sodium choloride (0.9%, 31°
C/88° F) as reference. Use of 5% dextrose will result in 10% slower flow rate
or correspondingly 10% longer delivery time.
PLEASE NOTE
- Delivery times for partial or overfill volumes are approximate values.
- Filling the pump more than the nominal fill volume results in a faster flow rate.
- Filling the pump less than nominal fill volume results in a slower flow rate.
- Do not fill the pump less than the minimal or more than the maximum fill volume specified on the chart.
- It is recommended that the MEDI-FLO® pump be filled with diluent before adding the drug/medication.
MEDI-FLO® Short Duration Filling Guide
Item Code| MDE050500| MDE100500| MDE101000|
MDE102000| MDE200100| MDE200200| MDE250050| MDE250100
---|---|---|---|---|---|---|---|---
Nominal Fill Volume| 50 mL| 100 mL| 100 mL| 100 mL| 200 mL| 200 mL| 250
mL| 250 mL
Nominal Flow Rate| 50 mL/H| 50 mL/H| 100 mL/H| 200 mL/H| 100 mL/H| 200
mL/H| 50 mL/H| 100 mL/H
Min Fill Volume| 50 mL| 75 mL| 50 mL| 50 mL| 150 mL| 150 mL| 150 mL| 200
mL
Max Fill Volume| 60 mL| 110 mL| 120 mL| 100 mL| 300 mL| 300 mL| 275 mL|
300 mL
Residual Volume| 2 mL| 2 mL| 2 mL| 2 mL| 2 mL| 2 mL| 3 mL| 3 mL
Approximate Delivery Time (Hours:Min)
0:15| | | | 50 mL| | | |
---|---|---|---|---|---|---|---|---
0:30| | | 50 mL| 100 mL| | | |
0:45| | | 75 mL| | | 150 mL| |
1:00| 50 mL| | 100 mL| | | 200 mL| |
1:15| 60 mL| | 120 mL| | | 250 mL| |
1:25| | | | | | | |
1:30| | 75 mL| | | 160 mL| 300 mL| |
1:45| | | | | 180 mL| | |
1:55| | | | | | | |
2:00| | 100 mL| | | 200 mL| | | 200 mL
2:15| | 110 mL| | | 220 mL| | | 220 mL
2:30| | | | | 240 mL| | | 250 mL
2:45| | | | | | | | 275 mL
3:00| | | | | | | 150 mL| 300 mL
3:30| | | | | | | |
4:00| | | | | | | |
4:30| | | | | | | |
5:00| | | | | | | 250 mL|
5:30| | | | | | | 275 mL|
Item Code| MDE250125| MDE251750| MDE252500|
MDE250500| MDE271750| MDE401000| MDE402000| MDE502500
---|---|---|---|---|---|---|---|---
Nominal Fill Volume| 250 mL| 250 mL| 250 mL| 250 mL| 270 mL| 400 mL| 400
mL| 500 mL
Nominal Flow Rate| 125 mL/H| 175 mL/H| 250 mL/H| 500 mL/H| 175 mL/H| 100
mL/H| 200 mL/H| 250 mL/H
Min Fill Volume| 150 mL| 140 mL| 150 mL| 250 mL| 250 mL| 275 mL| 200 mL|
360 mL
Max Fill Volume| 300 mL| 300 mL| 300 mL| 300 mL| 340 mL| 550 mL| 500 mL|
550 mL
Residual Volume| 3 mL| 3 mL| 3 mL| 3 mL| 3 mL| 5 mL| 5 mL| 5 mL
Approximate Delivery Time (Hours:Min)
0:15| | | | | | | |
---|---|---|---|---|---|---|---|---
0:30| | | 150 mL| 250 mL| | | |
0:45| | 140 mL| 200 mL| 300 mL| | | |
1:00| | 175 mL| 250 mL| | | | 200 mL|
1:15| 150 mL| 220 mL| 300 mL| | | | 250 mL| 360 mL
1:25| | | | | 250 mL| | |
1:30| 180 mL| 250 mL| | | 270 mL| | 300 mL| 400 mL
1:45| 220 mL| 300 mL| | | 300 mL| | 350 mL| 475 mL
1:55| | | | | 320 mL| | |
2:00| 250 mL| | | | 340 mL| | 400 mL| 500 mL
2:15| 280 mL| | | | | | 450 mL| 550 mL
2:30| 300 mL| | | | | | 500 mL|
2:45| | | | | | 275 mL| |
3:00| | | | | | 300 mL| |
3:30| | | | | | 360 mL| |
4:00| | | | | | 400 mL| |
4:30| | | | | | 450 mL| |
5:00| | | | | | 500 mL| |
5:30| | | | | | 550 mL| |
MEDI-FLO® Long Duration Filling Guide
LONG DURATION | |
---|---|
Item Code | MDL100002 |
MDL270010 | MDL500020 |
Nominal Fill Volume | 100 mL |
Nominal Flow Rate | 2 mL/H |
Min Fill Volume | 75 mL |
Max Fill Volume | 110 mL |
Residual Volume | 2 mL |
Approximate Delivery Time
Hours:Min| Day(s)| | | | |
---|---|---|---|---|---|---
18:00| | | | | | 380 mL
20:00| | | | | 230 mL| 440 mL
25:00| 1| | | | 250 mL| 500 mL
27:00| 1| | | | 270 mL|
30:00| 1| | | | 300 mL|
36:00| 1.5| 72 mL| | | 330 mL|
40:00| 1.5| 80 mL| | | |
46:00| 1.5| 96 mL| | 240 mL| |
50:00| 2| 100 mL| | 250 mL| |
54:00| 2| 108 mL| | 270 mL| |
62:00| | | | | |
72:00| 3| | | 330 mL| |
75:00| 3| | | | |
80:00| 3| | 200 mL| | |
100:00| 4| | 220 mL| | |
125:00| 5| | 250 mL| | |
135:00| 6| | 270 mL| | |
144:00| 6| | 300 mL| | |
MEDI-FLO® Chemotherapy Filling Guide
CHEMOTHERAPY | |
---|---|
Item Code | MDC100002 |
MDC270050 | MDC300060 |
Nominal Fill Volume | 100 mL |
Nominal Flow Rate | 2 mL/H |
Min Fill Volume | 75 mL |
Max Fill Volume | 110 mL |
Residual Volume | 2 mL |
Approximate Delivery Time
Hours:Min| Day(s)| | | | |
---|---|---|---|---|---|---
18:00| | | | | |
20:00| | | 230 mL| | |
25:00| 1| | 250 mL| | |
27:00| 1| | 270 mL| | |
30:00| 1| | 300 mL| | |
36:00| 1.5| 72 mL| 330 mL| | |
40:00| 1.5| 80 mL| | | | 250 mL
46:00| 1.5| 96 mL| | | 240 mL| 280 mL
50:00| 2| 100 mL| | | 250 mL| 300 mL
54:00| 2| 108 mL| | | 270 mL| 320 mL
62:00| | | | | | 360 mL
72:00| 3| | | | 330 mL|
75:00| 3| | | | |
80:00| 3| | | 200 mL| |
100:00| 4| | | 220 mL| |
125:00| 5| | | 250 mL| |
135:00| 6| | | 270 mL| |
144:00| 6| | | 300 mL| |
Preparing the MEDI-FLO® Elastomeric Pump
Instructions for Filling
Use Aseptic Technique
- Unscrew the fill port cap.
- MEDI-FLO® Elastomeric Pump can be filled with a syringe or automated fluid dispensing device. Remove trapped air from the filling device and attach it securely to the fill port.
- Close the ON-OFF clamp and fill the MEDI-FLO® Elastomeric Pump with no more than the maximum recommended volume. When using a syringe to fill, push the plunger to dispense the fluid. Do not push the barrel onto the fill port as the syringe tip or fill port may break. Repeat as necessary.
- Remove filling device from the fill port. Screw on the fill port cap.
- Label with appropriate pharmaceutical and patient information.
PLEASE NOTE: Actual infusion time may vary due to the following:
- Filling the device less than the nominal volume generally results in slower flow rate.
- Filling the device more than the nominal volume generally results in faster flow rate.
Priming the Administration Tubing
Use Aseptic Technique
- Open the ON-OFF clamp.
- Loosen the patient end cap. Medication will start to flow and fill the tubing. When all air is expelled, tighten the patient end cap.
- Close the ON-OFF clamp.
Priming Technique for Drugs
For Drugs Prone to Precipitation
- Fill MEDI-FLO® Elastomeric Pump with 10 mL of diluent first.
- Using the above priming method, prime the tubing.
- Fill the remaining volume with diluent and medication. At completion, the diluent will fill the entire tubing, protecting it from precipitation, while the pump reservoir will contain medication.
Storage
The MEDI-FLO® Elastomeric pump should be stored under general warehouse
conditions at 68° F to 77° F (20° C to 25° C) and be protected from light
sources and heat, prior to filling with medication. Refer to drug
manufacturer’s requirements for storage after filling the pump with
medication.
Prior to starting infusion, the pump should be allowed to reach room
temperature.
The table below provides guidelines for the estimated amount of time it will
take for the pump to reach room temperature after refrigeration based on the
nominal fill volume. Please note the MEDI-FLO® pump should not be stored in a
freezer.
Nominal Fill Volume| Refrigerated Temperature| Estimated Time to
Reach Room Temp
---|---|---
50 mL – 100 mL| 35.6° to 46.4° F (+2° to +8° C)| 6 hours from refrigerator
50 mL – 100 mL| -0.4° F (-18° C)| 12 hours from refrigerator
100 mL+| 35.6° to 46.4° F (+2° to +8° C)| 12 hours from refrigerator
100 mL+| -0.4° F (-18° C)| 18 hours from refrigerator
Guidelines for Infusion
Starting Infusion
Use Aseptic Technique
- Allow the MEDI-FLO® Elastomeric Pump to warm to room temperature before use, especially when infusate has been refrigerated.
- Infusion should preferably be started 1 – 8 hours after filling.
- Verify that the ON/OFF clamp is closed.
- Clean patient access point site as directed by facility protocol. Attach the patient connector to the injection site.
- Flow restrictor should be taped to patient’s skin to maintain the appropriate temperature for accurate medication delivery.
- Begin infusion by opening the ON/OFF clamp, fluid will begin to flow immediately.
- Ensure the ON/OFF clamp remains in the ON/OPEN position and the tubing remains free of kinks.
- If kinks are observed in the tubing they can be released by rolling the tubing between fingers to restore shape of tubing and facilitate fluid flow.
During Infusion
Patient Guides for the MEDI-FLO® Elastomeric pump are available and should be
provided to the patient prior to infusion. Healthcare provider is responsible
for educating the patient on proper use.
Depending on the pump size, a change in the look and size of the pump may
appear fairly quickly. It may take longer to see this change if the length of
the infusion is more than 24 hours.
Full Pump
The inside of the pump will be filled with medication and will be round and
firm. The outer cover will be smooth and stretched over the ball.
During Infusion
During the infusion, the inside reservoir will shrink in size and the outer
cover will begin to form wrinkles.
Empty Pump
At the end of the infusion, the inner ball will become a tall cylinder and the
outer cover will collapse.
During Infusion (continued)
The filled pump should not be squeezed or played with. Applied pressure may
result in rupture or breakage and will result in increased flow rate.
During use, the pump may be placed in a carrying case, patient’s pocket or on
a table/bed next to the patient. Carrying cases are available for purchase
from Wolf Medical Supply.
Ensure the section of tubing from the filter to the patient connector is kept
underneath clothing and in contact with the patient’s body.
Depending on the therapy, the patient may be using the pump while sleeping. If
so:
- The pump should be placed on a bedside table or on top of the bed covers. If placed underneath the bed covers the pump may become too warm, and flow rate may increase.
- Do not place the pump on the floor or hang the pump on a bed post.
It is not recommended that the patient use the pump while bathing or swimming. Patient may shower with proper direction from facility or healthcare provider, if pump and access site are protected from water.
End of Infusion
Infusion is complete when the elastomeric membrane is no longer expanded.
Close the clamp and disconnect from access site. Flush access site and dispose
of pump and supplies as directed by facility protocol.
Cautions
The MEDI-FLO® Elastomeric Pump is designed for optimal performance,
effectiveness and safety as a single use device and not for reuse.
Performance, effectiveness and safety may be compromised if the device is
reused.
MEDI-FLO® is sterile and nonpyrogenic. The pump should not be used if sterile
pouch has been opened, damaged, or if either protector cap is not in place.
Do not exceed the maximum labeled fill volume of the pump as this may affect
the indicated flow rate accuracy.
Prevent the filter from getting wet. Do not get alcohol or detergents on the
filter as this may cause a leak.
The MEDI-FLO® Elastomeric pump is not intended for use for the infusion of
insulin, blood or blood products, TPN, lipids or fat emulsions.
Troubleshooting
If pump does not seem to be working properly, please make sure:
- The pump is at room temperature. If higher than room temperature, infusate will flow faster than the labeled rate.
- The ON/OFF clamp is in the open position and moves freely on the tubing.
- All clamps on the catheter are open and filter is not covered.
- There are no kinks in the pump tubing.
- Verify that the fill volume is within the guidelines provided for the device that is being used. Underfilling the pump will cause it to flow slower than the labeled rate. Overfilling the pump will cause it to flow faster than the labeled flow rate.
If you have checked all of the above and your pump still does not seem to be working, please contact Preferred Medical at 1-800-722-7865 for further instructions.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>